What to do if... <strong>pH</strong> <strong>7110</strong>10 What to do if...10.1 <strong>pH</strong>More information and instructions on cleaning and exchange ofsensors are given in the documentation of your sensor.10.1.1 No stable measured valueCause• Junction of the electrode contaminated• Glass membrane of the electrodecontaminatedRemedy• Clean the junction• Clean the glass membraneCause• <strong>pH</strong> value of the test sample notstable• Temperature of the test samplenot stableRemedy• Measure with air excluded ifnecessary• Adjust temperature if necessary10.1.2 Error message CalErrorCause• The values determined for zeropoint and slope of the electrodeare outside the allowed limits.• RecalibrateRemedy• Junction contaminated • Clean the junction• <strong>pH</strong> electrode broken • Replace the <strong>pH</strong> electrode38 ba75925e01 07/2011