FIELD SERVICEREPLACEMENT OF THE THERMOCOUPLE ASSEMBLY (PN 18913)1. Remove the bottom front panel and the upper and lower rear panels of the furnace. (Note the position ofthe vent louvers on the rear panels.)2. Note and record the color and location of the wires on the thermocouple terminals located on the MainPC Board (Figure 5).3. Remove the thermocouple wires from the thermocouple terminals on the Main PC Board (Figure 5).4. Remove the clamp which secures the thermocouple to the ceramic terminal block located in the upper rearof the furnace (Figure 4).5. Remove the heating plate wire which crosses over the thermocouple and bend it out of the way just enoughto permit sliding the thermocouple out of the rear of the heating chamber (Figure 4).6. Remove the insulating filler that covers the holes in the bottom of the upper section (Figure 4). Straightenthe thermocouple wires where they were bent at a 90° angle.7. Withdraw the thermocouple down into the lower rear of the furnace through the holes in the upper andlower sections. Discard thermocouple assembly.8. Feed the new thermocouple from the lower rear of the furnace up through the holes in the upper and lowersections and into the upper rear of the furnace.12345679. Bend the new thermocouple wires at a 90° angle approximately 3-3/8 inches (8.6cm) from the exposedtip of the thermocouple. Be sure that there are three ceramic insulating sleeves between the thermocoupletip and the bend. Insert the new thermocouple into the hole at the rear of the heating chamber.10. Replace the heating plate wire that was removed in Step 3. Check that there is a ceramic insulating sleevecovering the thermocouple wires where they cross the heating plate wire. Replace the clamp that holds thethermocouple wires to the ceramic terminal block. Replace the insulating filler that covers the holes in thebottom of the upper section.11. Connect the thermocouple wires to the thermocouple terminals on the Main PC Board. Observe the colorcoding you noted in Step 2.12. Replace the front panel and the upper and lower rear panels. (Be sure the vent louvers on the rear panelsprotrude out and face down.)25

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!