Site Preparation<br />

___________________________________________________________________ Exhaust System<br />

1.6 Exhaust System<br />

It is your responsibility as the user to provide an adequate exhaust system.<br />

Most of what is introduced into the <strong>TRACE</strong> <strong>MS</strong> will eventually be<br />

exhausted from the rotary pump, along with a small amount of oil vapor that<br />

these pumps characteristically emit. Therefore, the pump should be<br />

connected to a fume exhaust system.<br />

The rotary pump connects to the rear of the <strong>TRACE</strong> <strong>MS</strong> at bench level by a<br />

reinforced vacuum hose. A connector suitable for 13 mm (1/2") i.d. tubing is<br />

fitted to the pump exhaust.<br />

Consult local regulations for the proper method of exhausting the fumes<br />

from your system.<br />

To carry out the venting, a user-supplied fume cupboard or oil mist filter is<br />

recommended.<br />

Caution. Do not use an oil mist filter with CI instruments.<br />

Note. Instrument damage resulting from incorrectly plumbed exhaust lines<br />

is not covered under warranty.<br />

_____________ <strong>Finnigan</strong> <strong>TRACE</strong> <strong>MS</strong> Preinstallation Requirements Guide _____________ 1-13