Site Preparation<br />

Gas Supplies______________________________________________________________________<br />

1.7 Gas Supplies<br />

Follow the guidelines in this section to ensure that the gases used by your<br />

<strong>TRACE</strong> <strong>MS</strong> system meet the required specifications. It is important to note<br />

that the purity of the gas supply (particularly the carrier gas) can affect the<br />

sensitivity of your <strong>TRACE</strong> <strong>MS</strong>.<br />

Carrier Gas<br />

• The carrier gas should be research grade helium (99.9999% purity).<br />

• Ensure that water and oxygen impurities are less than 1 ppm.<br />

• Oxygen, water and hydrocarbon traps are recommended in the supply<br />

line close to the <strong>TRACE</strong> GC 2000.<br />

• The carrier gas connects to the rear of the <strong>TRACE</strong> GC 2000 (top<br />

center). The maximum input pressure is 70 psig; pressure regulation<br />

should be delivered via a double-stage high purity regulator.<br />

CI Reagent Gas<br />

• The reagent gas should be research grade (99.9999% purity). Methane,<br />

ammonia and isobutane may be used depending on the application.<br />

• Ensure that water and oxygen impurities are less than 1 ppm.<br />

• The gas line connector is located at the rear of the <strong>TRACE</strong> <strong>MS</strong> at bench<br />

level. If a pressurized bottle is used, a suitable regulator should be<br />

supplied with the bottle. The normal operating pressure is 5 psig.<br />

Gas Connectors<br />

• All gas lines must be in place and leak-checked prior to the installation<br />

of your <strong>TRACE</strong> <strong>MS</strong> system.<br />

• Ensure all tubing and connectors are new to avoid contamination.<br />

• Sharing gas lines from other instruments is acceptable, providing their<br />

operation does not affect the pressure or flow available to your <strong>TRACE</strong><br />

<strong>MS</strong> system.<br />

Table 1-6 on page 1-15 provides information on the types of tubing suitable<br />

for different gases.<br />

1-14 ____________ <strong>Finnigan</strong> <strong>TRACE</strong> <strong>MS</strong> Preinstallation Requirements Guide_____________