Site Preparation<br />
Gas Supplies______________________________________________________________________<br />
1.7 Gas Supplies<br />
Follow the guidelines in this section to ensure that the gases used by your<br />
<strong>TRACE</strong> <strong>MS</strong> system meet the required specifications. It is important to note<br />
that the purity of the gas supply (particularly the carrier gas) can affect the<br />
sensitivity of your <strong>TRACE</strong> <strong>MS</strong>.<br />
Carrier Gas<br />
• The carrier gas should be research grade helium (99.9999% purity).<br />
• Ensure that water and oxygen impurities are less than 1 ppm.<br />
• Oxygen, water and hydrocarbon traps are recommended in the supply<br />
line close to the <strong>TRACE</strong> GC 2000.<br />
• The carrier gas connects to the rear of the <strong>TRACE</strong> GC 2000 (top<br />
center). The maximum input pressure is 70 psig; pressure regulation<br />
should be delivered via a double-stage high purity regulator.<br />
CI Reagent Gas<br />
• The reagent gas should be research grade (99.9999% purity). Methane,<br />
ammonia and isobutane may be used depending on the application.<br />
• Ensure that water and oxygen impurities are less than 1 ppm.<br />
• The gas line connector is located at the rear of the <strong>TRACE</strong> <strong>MS</strong> at bench<br />
level. If a pressurized bottle is used, a suitable regulator should be<br />
supplied with the bottle. The normal operating pressure is 5 psig.<br />
Gas Connectors<br />
• All gas lines must be in place and leak-checked prior to the installation<br />
of your <strong>TRACE</strong> <strong>MS</strong> system.<br />
• Ensure all tubing and connectors are new to avoid contamination.<br />
• Sharing gas lines from other instruments is acceptable, providing their<br />
operation does not affect the pressure or flow available to your <strong>TRACE</strong><br />
<strong>MS</strong> system.<br />
Table 1-6 on page 1-15 provides information on the types of tubing suitable<br />
for different gases.<br />
1-14 ____________ <strong>Finnigan</strong> <strong>TRACE</strong> <strong>MS</strong> Preinstallation Requirements Guide_____________