Getting StartedA - ground levelB - positive input of the supply voltage +24 to 35 V; this input is connected inparallel to the centre pin of the coaxial TNC power connector on the samepanelC - Larmor output signal, TTL compatible square voltage signal at LarmorfrequencyD - TTL compatible input: high level or open for operation in the southernoperating hemisphere, low level or connected to pin A for operation in thenorthern hemisphere. This input is effective only if the Hemisphere Controlswitch is in the Remote (R) position.1-10CS-3 Manual - part # 762701 Revision 2.0