Operating the CS-3 in the FieldIt is important to mention that the disturbing effect is reduced, if the directionof the spurious AC field is perpendicular to the sensor optical axis. It hasbeen experimentally determined that in this case the tested magnetometeroperated satisfactory in presence of the disturbing field with magnitude3500nT peak at the frequency 3kHz.The operating range of the tumble angle, in the plane perpendicular to thedirection of the spurious field, was reduced from 10°-85° (without the ACfield) to 25°-65° (with the AC field).2-10CS-3 Manual - part # 762701 Revision 2.0