82<br />

<strong>Pressure</strong> transmitters<br />

Programmable digital display<br />

Type series AP<br />

with 1 or 2 limit value switches for PT100, PT1000<br />

and voltage and current signals<br />

AP ...<br />

Various routines for setting the following parameters<br />

are integrated into the microprocessor-controlled<br />

digital display:<br />

· Measuring range (start and end point)<br />

· Display range (start and end point)<br />

· Setting of decimal point<br />

· 2 limit values (relays) and their hysteresis<br />

· Relay dropout or pickup delay<br />

· Scanning of minimum and maximum<br />

measured value<br />

· Rounding up or down of last digit<br />

· Averaging<br />

Technical data<br />

Input signals Freely selectable by<br />

setting jumpers.<br />

See Product Summary<br />

Housing front 48 x 96 mm (DIN)<br />

Dimension display On APT: °C<br />

On APV: none<br />

Display<br />

3 1/2 digit,<br />

LED 12.5 mm, red,<br />

automatic “–” sign<br />

Display range See Product Summary<br />

Other ranges<br />

adjustable<br />

Decimal point Programmable<br />

Measuring rate 2.5 measurements/<br />

second<br />

Keyboard lockable with jumper (prevents<br />

input of commands)<br />

Switching outputs 2 x NO/NC contacts<br />

programmable<br />

Switching capacity 2 x 230 V, 5 A AC<br />

of output relay<br />

Supply voltage 230 V, 50–60 Hz, 3 VA<br />

Degree of protection IP 60, DIN 40 050<br />

(front)<br />

Working temperature –10 to +50°C<br />

Connection method Lift terminals<br />

Front panel cut-out H x W: 44.5 x 90.5 mm<br />

Max. installation 115 mm<br />

depth<br />

All routines and parameters can be set with keys on the front. The switching status of the relays is displayed<br />

by LEDs. Setting is buffered. If the supply voltage is interrupted, the set parameters are<br />

retained. For powering transmitters, an electrically isolated 24 VDC power supply (max. 30 mA) is<br />

available.<br />

Product Summary<br />

Type Input signals Display range Suitable Stages<br />

(programmable) (programmable) for<br />

APV 600 0–1 V DC –1999 <strong>Pressure</strong> and 1 switching point<br />

APV 630 0–10 V DC to temperature 2 switching points<br />

0–20 mA DC +1999 transmitters<br />

APT 600 Pt 100 / Pt 1000 –150°C to Temperature 1 switching point<br />

APT 650 +199.9°C sensors 2 switching points<br />

–200°C to Pt 100 /<br />

+800°C Pt 1000<br />

Dimensions