25.09.2015 Views

Sherman Plate 4.5 mm Sherman Platte 4.5 mm Plaque de Sherman ...

Sherman Plate 4.5 mm Sherman Platte 4.5 mm Plaque de Sherman ...

Sherman Plate 4.5 mm Sherman Platte 4.5 mm Plaque de Sherman ...

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Section of reconstruction plates for<br />

operation on pelvis<br />

<strong>Sherman</strong> <strong>Plate</strong> <strong>4.5</strong> <strong>mm</strong><br />

<strong>Sherman</strong> <strong>Platte</strong> <strong>4.5</strong> <strong>mm</strong><br />

<strong>Plaque</strong> <strong>de</strong> <strong>Sherman</strong> <strong>4.5</strong> <strong>mm</strong><br />

Indication : Used principally in the reconstruction of the pelvis. The<br />

<strong>de</strong>sign of these plates allows for three dimensional contouring. Due<br />

to their reduced stiffness, they should only be used for tension-band<br />

fixation or adaption.<br />

<strong>Plate</strong> <strong>de</strong>tails : Reconstruction plates <strong>4.5</strong> <strong>mm</strong> are manufactured<br />

from relatively soft material. Design and softness of the material<br />

enable three-dimensional contouring. Do not bend more than 15º<br />

between holes.<br />

<strong>Plate</strong> thickness : 2.5 <strong>mm</strong><br />

<strong>Plate</strong> width : 11.0 <strong>mm</strong><br />

Hole spacing : 13.0 <strong>mm</strong><br />

Screws to be used : Fixation with diam <strong>4.5</strong> <strong>mm</strong> cortical screws and<br />

diam. 6.5 <strong>mm</strong> cancellous screws.<br />

Material<br />

F.138 - ISO 5832/1.<br />

: Certified Stainless Steel according to ASTM<br />

Ref. Number<br />

Stainless Steel<br />

Holes<br />

Lengthening<br />

L1 (<strong>mm</strong>)<br />

Length<br />

L (<strong>mm</strong>)<br />

10313001003<br />

10313001004<br />

10313001005<br />

10313001006<br />

10313001007<br />

10313001008<br />

10313001009<br />

10313001010<br />

10313001011<br />

4<br />

5<br />

6<br />

7<br />

8<br />

9<br />

10<br />

11<br />

12<br />

21<br />

21<br />

21<br />

21<br />

21<br />

21<br />

21<br />

21<br />

21<br />

58<br />

71<br />

84<br />

97<br />

110<br />

123<br />

136<br />

149<br />

162<br />

D-13


Section of reconstruction plates for<br />

operation on pelvis<br />

Reconstruction <strong>Plate</strong> <strong>4.5</strong> <strong>mm</strong><br />

Rekonstrutionplatte <strong>4.5</strong> <strong>mm</strong><br />

<strong>Plaque</strong> <strong>de</strong> reconstruction <strong>4.5</strong> <strong>mm</strong><br />

Indication : Used principally in the reconstruction of the pelvis. The<br />

<strong>de</strong>sign of these plates allows for three dimensional contouring. Due<br />

to their reduced stiffness, they should only be used for tension-band<br />

fixation or adaption.<br />

<strong>Plate</strong> <strong>de</strong>tails : Reconstruction plates <strong>4.5</strong> <strong>mm</strong> are manufactured<br />

from relatively soft material. Design and softness of the material<br />

enable three-dimensional contouring. Do not bend more than 15º<br />

between holes.<br />

<strong>Plate</strong> thickness : 2.7 <strong>mm</strong><br />

<strong>Plate</strong> width : 12.0 <strong>mm</strong><br />

Hole spacing : 16.0 <strong>mm</strong><br />

Screws to be used : Fixation with diam <strong>4.5</strong> <strong>mm</strong> cortical screws and<br />

diam. 6.5 <strong>mm</strong> cancellous screws.<br />

Material<br />

: Certified Stainless Steel according to ASTM<br />

F.138 - ISO 5832/1. Certified Titanium alloy according to ASTM F.136<br />

- ISO 5832/3.<br />

Ref. Number<br />

Stainless Steel<br />

Ref. Number<br />

Titanium<br />

Holes<br />

Length<br />

<strong>mm</strong><br />

10311011001<br />

10311011002<br />

10311011003<br />

10311011004<br />

10311011005<br />

10311011006<br />

10311011007<br />

10311011008<br />

10311011009<br />

10311011010<br />

10311011011<br />

10311011012<br />

10311011013<br />

10311011014<br />

10311011015<br />

10311011016<br />

10311011017<br />

10311012001<br />

10311012002<br />

10311012003<br />

10311012004<br />

10311012005<br />

10311012006<br />

10311012007<br />

10311012008<br />

10311012009<br />

10311012010<br />

10311012011<br />

10311012012<br />

10311012013<br />

10311012014<br />

10311012015<br />

10311012016<br />

10311012017<br />

4<br />

5<br />

6<br />

7<br />

8<br />

9<br />

10<br />

11<br />

12<br />

13<br />

14<br />

15<br />

16<br />

17<br />

18<br />

19<br />

20<br />

61<br />

77<br />

93<br />

109<br />

125<br />

141<br />

157<br />

173<br />

189<br />

205<br />

221<br />

237<br />

253<br />

269<br />

285<br />

301<br />

317<br />

D-14


Section of reconstruction plates for<br />

operation on pelvis<br />

Reconstruction <strong>Plate</strong> <strong>4.5</strong> <strong>mm</strong> curved<br />

Rekonstrutionplatte <strong>4.5</strong> <strong>mm</strong> gebogen<br />

<strong>Plaque</strong> <strong>de</strong> reconstruction <strong>4.5</strong> <strong>mm</strong><br />

pour le bassin<br />

Indication : Used principally in the reconstruction of the pelvis. It is<br />

slightly thicker than the straight pelvic reconstruction plate and prevent<br />

for the pelvic crest.<br />

<strong>Plate</strong> <strong>de</strong>tails : This plate has been especially <strong>de</strong>signed for<br />

use on the pelvis. Material is thicker than straight reconstruction plate.<br />

<strong>Plate</strong> thickness : 2.7 <strong>mm</strong><br />

<strong>Plate</strong> width : 12.0 <strong>mm</strong><br />

Hole spacing : 16.0 <strong>mm</strong><br />

Screws to be used : Fixation with diam <strong>4.5</strong> <strong>mm</strong> small cortical screws.<br />

Material<br />

: Certified Stainless Steel according to ASTM<br />

F.138 - ISO 5832/1. Certified Titanium alloy according to ASTM F.136<br />

- ISO 5832/3.<br />

Ref. Number<br />

Stainless Steel<br />

Ref. Number<br />

Titanium<br />

Holes<br />

Length<br />

<strong>mm</strong><br />

10311061003<br />

10311061005<br />

10311061007<br />

10311061009<br />

10311061011<br />

10311061013<br />

10311061015<br />

10311062003<br />

10311062005<br />

10311062007<br />

10311062009<br />

10311062011<br />

10311062013<br />

10311062015<br />

6<br />

8<br />

10<br />

12<br />

14<br />

16<br />

18<br />

94<br />

126<br />

158<br />

190<br />

222<br />

254<br />

286<br />

D-15


Section of reconstruction plates for operation<br />

on pelvic, acetabular, clavicle and calcaneal<br />

Reconstruction <strong>Plate</strong> 3.5 <strong>mm</strong><br />

Rekonstrutionplatte 3.5 <strong>mm</strong><br />

<strong>Plaque</strong> <strong>de</strong> reconstruction 3.5 <strong>mm</strong><br />

Indication : Used principally in the treatment of pelvic, clavicle and<br />

calcaneal fractures. The <strong>de</strong>sign of these plates allows for threedimentional<br />

contouring. Due to their reduced stiffness, they should<br />

be used only for cerclage or adaptations.<br />

<strong>Plate</strong> <strong>de</strong>tails : Reconstruction plates 3.5 <strong>mm</strong> are manufactured<br />

from relatively soft material. Design and softness of the material<br />

enable three-dimensional contouring. Do not bend more than 15º<br />

between holes.<br />

<strong>Plate</strong> thickness : 2.7 <strong>mm</strong><br />

<strong>Plate</strong> width : 10.0 <strong>mm</strong><br />

Hole spacing : 12.0 <strong>mm</strong><br />

Screws to be used : Fixation with diam 3.5 <strong>mm</strong> cortical screws and<br />

diam. 4.0 <strong>mm</strong> small cancellous screws.<br />

Material<br />

: Certified Stainless Steel according to ASTM<br />

F.138 - ISO 5832/1. Certified Titanium alloy according to ASTM F.136<br />

- ISO 5832/3.<br />

Ref. Number<br />

Stainless Steel<br />

Ref. Number<br />

Titanium<br />

Holes<br />

Length<br />

<strong>mm</strong><br />

10311021001<br />

10311021002<br />

10311021003<br />

10311021004<br />

10311021005<br />

10311021006<br />

10311021007<br />

10311021008<br />

10311021009<br />

10311021010<br />

10311021011<br />

10311021012<br />

10311021013<br />

10311021014<br />

10311021015<br />

10311021016<br />

10311021017<br />

10311022001<br />

10311022002<br />

10311022003<br />

10311022004<br />

10311022005<br />

10311022006<br />

10311022007<br />

10311022008<br />

10311022009<br />

10311022010<br />

10311022011<br />

10311022012<br />

10311022013<br />

10311022014<br />

10311022015<br />

10311022016<br />

10311022017<br />

4<br />

5<br />

6<br />

7<br />

8<br />

9<br />

10<br />

11<br />

12<br />

13<br />

14<br />

15<br />

16<br />

17<br />

18<br />

19<br />

20<br />

46<br />

58<br />

70<br />

82<br />

94<br />

106<br />

118<br />

130<br />

142<br />

154<br />

166<br />

178<br />

190<br />

202<br />

214<br />

226<br />

238<br />

D-16


Section of reconstruction plates for<br />

operation on pelvis<br />

Reconstruction <strong>Plate</strong> 3.5 <strong>mm</strong> curved<br />

Rekonstrutionplatte 3.5 <strong>mm</strong> gebogen<br />

<strong>Plaque</strong> <strong>de</strong> reconstruction 3.5 <strong>mm</strong><br />

pour le bassin<br />

Indication : Used principally in the reconstruction of the pelvis. It is<br />

slightly thicker than the straight pelvic reconstruction plate and prevent<br />

for the pelvic crest.<br />

<strong>Plate</strong> <strong>de</strong>tails : This plate has been especially <strong>de</strong>signed for<br />

use on the pelvis. Material is thicker than straight reconstruction plate.<br />

<strong>Plate</strong> thickness : 2.7 <strong>mm</strong><br />

<strong>Plate</strong> width : 10.0 <strong>mm</strong><br />

Hole spacing : 12.0 <strong>mm</strong><br />

Screws to be used : Fixation with diam 3.5 <strong>mm</strong> small cortical screws.<br />

Material<br />

: Certified Stainless Steel according to ASTM<br />

F.138 - ISO 5832/1. Certified Titanium alloy according to ASTM F.136<br />

- ISO 5832/3.<br />

Ref. Number<br />

Stainless Steel<br />

Ref. Number<br />

Titanium<br />

Holes<br />

Length<br />

<strong>mm</strong><br />

10311041003<br />

10311041005<br />

10311041007<br />

10311041009<br />

10311041011<br />

10311041013<br />

10311041015<br />

10311042003<br />

10311042005<br />

10311042007<br />

10311042009<br />

10311042011<br />

10311042013<br />

10311042015<br />

6<br />

8<br />

10<br />

12<br />

14<br />

16<br />

18<br />

70<br />

94<br />

118<br />

142<br />

166<br />

190<br />

214<br />

D-17


Section of reconstruction plates for<br />

operation on madibular<br />

Reconstruction <strong>Plate</strong> 2.7 <strong>mm</strong><br />

Rekonstrutionplatte 2.7 <strong>mm</strong><br />

<strong>Plaque</strong> <strong>de</strong> reconstruction 2.7 <strong>mm</strong><br />

Indication : Indicate primarlly in co<strong>mm</strong>inuted fractures of smaller<br />

diameter bones specially used for mandibular reconstruction surgery.<br />

The <strong>de</strong>sign of these plates allows for three diamensional contouring.<br />

<strong>Plate</strong> <strong>de</strong>tails : Reconstruction plates 2.7 <strong>mm</strong> are manufactured<br />

from relatively soft material. Design and softness of the material<br />

enable three-dimensional contouring.<br />

<strong>Plate</strong> thickness : 2.7 <strong>mm</strong><br />

<strong>Plate</strong> width : 8.0 <strong>mm</strong><br />

Hole spacing : 8.0 <strong>mm</strong><br />

Screws to be used : Fixation with diam 2.7 <strong>mm</strong> mini cortical screws.<br />

Material<br />

: Certified Stainless Steel according to ASTM<br />

F.138 - ISO 5832/1. Certified Titanium alloy according to ASTM F.136<br />

- ISO 5832/3.<br />

Ref. Number<br />

Stainless Steel<br />

Ref. Number<br />

Titanium<br />

Holes<br />

Length<br />

<strong>mm</strong><br />

10311031001<br />

10311031002<br />

10311031003<br />

10311031004<br />

10311031005<br />

10311031006<br />

10311031007<br />

10311031008<br />

10311031009<br />

10311031010<br />

10311031011<br />

10311031012<br />

10311031013<br />

10311031014<br />

10311031015<br />

10311031016<br />

10311031017<br />

10311031019<br />

10311031021<br />

10311032001<br />

10311032002<br />

10311032003<br />

10311032004<br />

10311032005<br />

10311032006<br />

10311032007<br />

10311032008<br />

10311032009<br />

10311032010<br />

10311032011<br />

10311032012<br />

10311032013<br />

10311032014<br />

10311032015<br />

10311032016<br />

10311032017<br />

10311032019<br />

10311032021<br />

4<br />

5<br />

6<br />

7<br />

8<br />

9<br />

10<br />

11<br />

12<br />

13<br />

14<br />

15<br />

16<br />

17<br />

18<br />

19<br />

20<br />

22<br />

24<br />

32<br />

40<br />

48<br />

56<br />

64<br />

72<br />

80<br />

88<br />

96<br />

104<br />

112<br />

120<br />

128<br />

136<br />

144<br />

152<br />

160<br />

176<br />

192<br />

D-18


Section of reconstruction plates for<br />

operation on distal humerus<br />

Y - Reconstruction <strong>Plate</strong> 3.5 <strong>mm</strong><br />

Y - Rekonstrutionplatte 3.5 <strong>mm</strong><br />

<strong>Plaque</strong> <strong>de</strong> reconstruction<br />

en Y 3.5 <strong>mm</strong><br />

Indication : Used to treat co<strong>mm</strong>inuted fractures, particularly of the<br />

distal humerus. The <strong>de</strong>sign of these plates allows for three-dimensional<br />

contouring.<br />

<strong>Plate</strong> <strong>de</strong>tails : Y reconstruction plates are manufactured from<br />

relatively soft material. Design and softness of the material enable<br />

three-dimensional contouring. If necessary, the plate can be shortened<br />

by breaking or cutting. Adjustable in three-dimension with bending<br />

pliers.<br />

<strong>Plate</strong> thickness : 1.5 <strong>mm</strong><br />

<strong>Plate</strong> width : 9.0 <strong>mm</strong><br />

Hole spacing : 12.0 <strong>mm</strong><br />

Screws to be used : Fixation with diam 3.5 <strong>mm</strong> small cortical screws<br />

and 4.0 <strong>mm</strong> small cancellous screws.<br />

Material<br />

: Certified Stainless Steel according to ASTM<br />

F.138 - ISO 5832/1. Certified Titanium alloy according to ASTM F.136<br />

- ISO 5832/3.<br />

Ref. Number<br />

Stainless Steel<br />

Ref. Number<br />

Titanium<br />

Holes<br />

Shaft<br />

Total<br />

Holes<br />

10311051001 10311052001 4 4x4x4<br />

D-19

Hooray! Your file is uploaded and ready to be published.