� Refer to Chapter 7, “Source Troubleshooting,” if you cannot perform the<br />

adjustment.<br />

How to Set Up the Fractional-N Spur Avoidance and<br />

FM Sideband Adjustment<br />

1. Press B ~~~:~~~,~~~~~~’ (where X is the sequence number).<br />

2. Remove the upper-rear comer bumpers and the top cover, using a TORX<br />

screwdriver.<br />

3. Follow the directions on the analyzer display and make alI of the API<br />

adjustments.<br />

Sequence Contents<br />

Sequence for the High/Low Band Transition Adjustment<br />

--Sequence HB~sets the hi-band to low-band switch point.-<br />

PRESET<br />

SYSTEM<br />

SERVICE MENU<br />

ANALOG BUS ON<br />

STARTllIWu<br />

STOP 21 M/u<br />

MEAS<br />

ANALOG IN 22 xl (Al2 GND)<br />

DISPLAY<br />

DATA>MEM<br />

DATA-MEM<br />

MEAS<br />

ANALOG IN 23 xl (VCO TUNE)<br />

MKR11IWl.l<br />

SCALELREF .l xl<br />

3-64 Adjustments and Correction Constants