<strong>HP</strong> <strong>8753E</strong> System Verification<br />

Equipment Required<br />

Calibration Kit, 7-mm ................................................ <strong>HP</strong> 85031B<br />

Verification Kit, 7-mm ............................................... .<strong>HP</strong> 85029B<br />

Test Port Extension Cable Set, 7-mm ................................ <strong>HP</strong> 11857D<br />

Printer ........................................... <strong>HP</strong> ThinkJet/DeskJet/LaserJet<br />

Additional Equipment Required for Option 075 <strong>Analyzer</strong>s<br />

Minimum Loss Pad (2), 50 Q to 75 Q . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . <strong>HP</strong> 11852B<br />

Adapter (2), APC-7 to Type-N (m) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . <strong>HP</strong> 11525A<br />

Anulgzer warmup time: 1 hour<br />

This system veriEcation consists of three separate procedures:<br />

1. Initialization<br />

2. Measurement Calibration<br />

3. Device Verification<br />

2-9 System Verification and<br />

Performance Tests