A4, A5, A6 Samplers and A7 Pulse Generator<br />

Replacement<br />

1. Check the connector pins on the motherboard before reinstallation.<br />

2. Reverse the order of the removal procedure.<br />

Note � When reconnecting semirigid cables, it is recommended that<br />

the connections be torqued to 10 in-lb.<br />

� Be sure to route W8 and W9 as shown. No excess wire should<br />

be hanging in the All and Al4 board slots. Routing the wires<br />

in this manner will reduce noise and crosstalk.<br />