Verify Shorts and Opens<br />

Substitute a known good short and open of the same connector type and sex as<br />

the short and open in question. If the devices are not from one of the standard<br />

calibration kits, refer to the <strong>HP</strong> <strong>8753E</strong> <strong>Network</strong> A~~~lyzer User’s <strong>Guide</strong> for<br />

hfomation on how to use the ~~~~~~~~~~~~ fMction* Set aside the short<br />

and open that are causing the problem.<br />

1. Perform an Sll l-port calibration using the good short and open. Then press<br />

zpQ&g,c,,<br />

. . . . .._. . ;ii.....:L.:!. i I-- ~~~~~~~~<br />

-._- ..__.._... - . . . . .../..<br />

to view the devices h S&h chart format.<br />

2. Come& the good shod to poti 1. Press c-1 ~~~~~~~~~ ad<br />

turn the front panel knob to enter enough electrical delay so that the trace<br />

appears as a dot at the left side of the circle. (See Figure 9-2a, left.)<br />

Replace the good short with the questionable short at port 1. The trace of<br />

the questionable short should appear very similar to the known good short.<br />

3. CoMect *e good open to port 1. mess lscale, ~.~~~~~~~~ and<br />

turn the front panel knob to enter enough electrical delay so that the trace<br />

appears as a dot at the right side of the circle. (See Figure 9-2b, right.)<br />

Replace the good open with the questionable open at port 1. The trace of the<br />

questionable open should appear very similar to the known good open.<br />

g-6 Accessories Troubleshooting