19.09.2012 Views

Alternative applications of ndd spirette

Alternative applications of ndd spirette

Alternative applications of ndd spirette

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Frequently asked questions about EasyOne Pro<br />

Sales Issues<br />

1. Why do I not need to calibrate the EasyOne Pro?<br />

o <strong>ndd</strong> TrueFlow to measure flow is proven long term stable. In fact there is<br />

no calibration button on any <strong>of</strong> our many thousand equipment in use for<br />

over a decade now.<br />

� Over 40’000 EasyOne equipment throughout the world are since<br />

2000 in use and frequently checked for proper functioning . For<br />

more information on references in your area see:<br />

http://www.<strong>ndd</strong>.ch/AboutUs/references.aspx<br />

� For publications on the subject long term stability refer to:<br />

http://www.<strong>ndd</strong>.ch/Downloads/scientific_docs.aspx screen for “no<br />

calibration”<br />

� The technology is described in:<br />

http://www.<strong>ndd</strong>.ch/UserData/Download_00210_00.pdf<br />

o <strong>ndd</strong> MolMass to measure He dilution for lung volume is applied physics<br />

(molar mass difference in direct proportion to He concentration) and<br />

cannot / must not be calibrated. For more explanation refer to<br />

http://www.<strong>ndd</strong>.ch/AboutUs/Technology/molmass.aspx<br />

o The CO sensor in fact is automatically calibrated frequently in the<br />

background with the patient diffusion gas itself.<br />

o For all QC assurance procedures you may <strong>of</strong> course follow the internally<br />

provided procedures with Levy Jennings plots and other QC diagnostics:<br />

� Bio Cal check using standard persons from your crew<br />

� Syringe procedures variable or single flow<br />

� Simulator cal checks<br />

2. Why do other systems need to calibrate daily?<br />

Flow<br />

All other systems which do not apply <strong>ndd</strong> patented TrueFlow technology use<br />

mechanical sensors; placed within and subjected to the sample breath<br />

stream.<br />

All flow measurement systems based on pressure drop (orifices, Fleisch<br />

pneumotachograph etc.) and all systems based on turbines and heated wires<br />

are subject to:<br />

o A transducer / amplifier converting the measured secondary effect<br />

(pressure drop, turbine speed, heat loss into an electric signal have<br />

naturally<br />

� Linearity issues<br />

� Stability issues<br />

o Any mechanical sensor in contact with the breath suffers contamination<br />

and hence a drift to be considered and cleaned away frequently.


� Some contamination like patient discharge may be filtered out, but<br />

this causes resistance and low flow threshold problems<br />

� Other effects like condensation <strong>of</strong> existing humidity are being<br />

formed regardless <strong>of</strong> filters and affect the measurement in a similar<br />

manner as contamination<br />

o Contamination builds up<br />

� And needs to be adjusted for<br />

� Needs to be cleaned away frequently and then re adjusted<br />

Gas composition<br />

All other systems which do not apply <strong>ndd</strong> patented Molar Mass technology<br />

apply some classical gas component measuring system.<br />

These gas sensors need to be calibrated using reference gases:<br />

o A basic calibration can be achieved uisng 2 reference gases<br />

o For higher quality 3 reference gases are required in order to take into<br />

account the non-linearity <strong>of</strong> the sensor.<br />

o hence involving complex calibration methods<br />

o and relying on the accuracy <strong>of</strong> locally supplied gases !!!<br />

Why is <strong>ndd</strong> calibration free?<br />

<strong>ndd</strong> applies the <strong>ndd</strong> patented TrueFlow and Molar Mass measurement which<br />

is uniquely different:<br />

Flow by <strong>ndd</strong> TrueFlow<br />

As implemented with EasyOne and meanwhile proven by evidence <strong>ndd</strong> True<br />

flow is “life time stable”.<br />

L td<br />

− tu<br />

F = ⋅<br />

2cosϕ<br />

t ⋅t<br />

d<br />

u<br />

A simple time measurement (<strong>of</strong> course calibration free, quarz based)<br />

measures the transit-times <strong>of</strong> ultrasonic pulses traveling up- and downstream<br />

the gas flow:<br />

o up-stream pulses are faster (i.e. smaller transit-time)<br />

o down-stream pulses are slower (i.e. larger transit-times)<br />

Using the above formula flow speed can be computed independently <strong>of</strong> any<br />

variables (temperature, humidity, ambient pressure, gas density etc.).<br />

Using flow speed the volume can be computed easily.<br />

Since there is no mechanical sensor in the gas stream, there is no<br />

maintenance, no drift, no effect <strong>of</strong> contamination.<br />

Gas composition by <strong>ndd</strong> Molar Mass<br />

The Molar Mass measurement determines the mean molar mass <strong>of</strong> the gas<br />

composition (i.e. the percentage weighted sum <strong>of</strong> all molar mass values <strong>of</strong><br />

the gas components).<br />

4 ⎛ tu<br />

⋅ td<br />

M = κ ⋅ R ⋅T<br />

⋅ 2 ⎜<br />

L ⎝ tu<br />

+ t<br />

d<br />

⎞<br />

⎟<br />

⎠<br />

2<br />

Molar Mass can therefore be used to measure changes in gas compositions.<br />

The molar mass measurement must not be calibrated and it is completely<br />

linear over its measurement range. We therefore do not need to calibrate with


the exception <strong>of</strong> CO, and this is done automatically and in a new way as<br />

explained below.<br />

EasyOne Pro measures:<br />

Flow and main-stream Molar Mass at the very same spot (same sensors) near<br />

patient.<br />

Helium by MolMass and CO by NDIR in a side stream arrangement.<br />

To have a precise time-alignment <strong>of</strong> all measurements (Flow, He, CO) the<br />

delay <strong>of</strong> each signal is automatically determined using a patented technology.<br />

The automatic delay correction is also calibration free and is performed for<br />

each test.<br />

He; Helium by Molar Mass<br />

We perform a 2 point calibration:<br />

1. Air being zero He<br />

2. Test gas being the working range maximum (concentration <strong>of</strong> the bottle<br />

as indicated by supplier)<br />

The absolute concentration is for <strong>ndd</strong> technology relatively unimportant since<br />

the relative Molar Mass measurement contains all the information needed to<br />

do accurate DLCO<br />

The absolute concentration is IMPORTANT for all simulator based<br />

measurements since the absolute concentrations <strong>of</strong> the reference gas bottles<br />

are compared to the absolute concentration <strong>of</strong> the DLCO test gas.<br />

The Molar Mass calibration is automatic an never requires any external gases.<br />

CO, Carbon Monoxide<br />

We perform an fully automatic 3 point calibration<br />

3. Air being the zero CO<br />

4. test gas being the working range max<br />

5. mid range: The exact gas mixture is determined by Molar mass and then<br />

used to calibrate the CO sensor at the mid point. Using this technology<br />

(patent pending) the non-linearity <strong>of</strong> the CO sensor can automatically be<br />

determined for each test.<br />

All calibration steps are fully automatic and with no need for any external gas.<br />

Ambient variables input<br />

No variables must be entered since all are either measured or irrelevant<br />

For the side-stream systems the humidity difference between inspiratory and<br />

expiratory measurements is equalized by conditioning <strong>of</strong> the sample to<br />

environmental level through Nafion® (parmapure) technology.<br />

3. How long does it take to set up EasyOne Pro “from scratch” and then<br />

every time I want to use it?<br />

o The first time the intuitive process (the proverbial 1,2,3) may take up to<br />

10 minutes, following the instructions in the manual<br />

� Connect the gas , the patient tube and adjust the regulator pressure<br />

� Connect the mains power (100-240VAC) and switch on<br />

� Run through the set-up routine


o Any subsequent times, including in different locations, it just takes 2-3<br />

minutes to be ready for measurements, no warm up, no checks, no<br />

maintenance and <strong>of</strong> course no calibration.<br />

4. How long does it take to measure one patient (LFT: spirometry and<br />

DLCO)?<br />

o It typically takes under 19 minutes for one patient: explanations,<br />

spirometry (BD) and DLCO in practical application with EasyOne Pro; on<br />

remote locations and similar in home site (reference by Respiratory Testing Services,<br />

Fremantle, WA, Australia; May 2009; timed on 2 locations with over 30 patients)<br />

o Important EasyOne Pro features for time saving:<br />

� Hand held sensor assembly, fast, easy and lip tight docking to<br />

patient, no positioning manipulation <strong>of</strong> support arms needed<br />

� Touch screen display, all operations and operator / patient<br />

interactions on the same screen<br />

� Customer designed report format and print out from main<br />

equipment (no PC required)<br />

� On board s<strong>of</strong>tware for data processing etc. (free export)<br />

� State <strong>of</strong> the art design: latest technology, latest hardware<br />

generations, windows based dedicated built-in PC<br />

5. What are the running costs for EasyOne Pro?<br />

o You will need a new <strong>spirette</strong> and barriette for hygienic reasons for each patient,<br />

for prices refer to the local source, usually well under USD 5. These are the ONLY<br />

consumable required.<br />

o The gas (0.3%CO; 10%He; 18-25% O2; N2) consumption is ca. 5 liters per<br />

inhalation, so typically 10-15 liters per patient<br />

o The power consumption is so low that you may easily use an uninterrupted power<br />

supply (UPS) for “total independence”: bed side, occupational sites etc.<br />

6. Why should I prefer <strong>ndd</strong> EasyOne Pro over other brands / models?<br />

o <strong>ndd</strong> is the company that has set the new standard in spirometry:<br />

� true flow (ultrasonic transit time measurement)<br />

� <strong>spirette</strong> the new hygienic consumable that allows calibration free<br />

(no sensor) use <strong>of</strong> TrueFlow based instruments.<br />

� References are published from Platino to BOLD, from USA to<br />

Australia<br />

o EasyOne Pro is the first and only instrument based on the new ATS/ERS<br />

standard for DLCO measurement<br />

o And developed by using the new ATS ERS approved simulator<br />

o EasyOne Pro features the same <strong>ndd</strong> benefits as does the word wide<br />

successful and best sold spirometer EasyOne:<br />

� No calibration<br />

� Ready to use at the touch <strong>of</strong> a button<br />

� Test QC monitored (and logged) for ATS / ERS compliance<br />

� User prompts to achieve best trials<br />

� Maintenance free, no service parts required,<br />

� but with a new and convincing trouble shooting concept; assuring<br />

help when and wherever needed and backing up the <strong>ndd</strong> promises.<br />

� It is the brand / instrument preferred by opinion leaders.


• By design <strong>of</strong> its patented technology TrueFlow and<br />

MolMass currently only available at <strong>ndd</strong> and a few selected<br />

OEM Partners.<br />

o You will always need only one gas, the DLCO diffusion gas (CO/He/Air),<br />

nothing else.<br />

7. Do you follow the ERS ATS 2005 guidelines<br />

o Yes, EasyOnePro is the first (and so far only) equipment that has been<br />

developed AFTER these guidelines were published. Therefore we match<br />

them better than anyone else.<br />

See:<br />

http://www.<strong>ndd</strong>.ch/Downloads/scientific_docs.aspx<br />

o Also we are conform with the Hans Rudolf Simulator developed by Crapo /<br />

Jensen. Published results: best out <strong>of</strong> 6 major brands.<br />

Technical Issues:<br />

8. What size <strong>of</strong> gas bottle should I order?<br />

o The size is a consideration between the following key aspects depending<br />

on your own preferences:<br />

• # <strong>of</strong> patients you want to measure per bottle<br />

• Usually the gas has a certified stability for 2-3 years<br />

• Ease <strong>of</strong> portability required (down to for example 5l for<br />

bedside)<br />

o Attention the filling pressure may be higher with big<br />

bottles (250atm with 20 and higher l) and lower with<br />

small bottles (150 atm. at 5l)<br />

9. Uninterrupted Power Supply (UPS) for EasyOne Pro<br />

o If the power supply has interruption or large Voltage Fluctuations a UPS<br />

units is necessary.<br />

o For specific calculation see:<br />

http://www.apc.com/tools/ups_selector/index.cfm<br />

Note the following specifications:<br />

� Power consumption EasyOne Pro: 80VA (Volt-Ampere, similar to<br />

Watt)<br />

� Use only a slow fuse <strong>of</strong> 3 Ampere.<br />

� The UPS need to supply 160Wh (Energy for a 2 hour power<br />

interruption)<br />

� For USP usually the Rating is indicated in Watt or VA this value<br />

need to be at least 80 VA<br />

� Depending <strong>of</strong> the lengths <strong>of</strong> the power interruption the backup time<br />

value is important. UPS provider indicate this value usually in<br />

Minutes (Example: 700VA � Back uptime 30 min with 50% load)


That means for EasyOne Pro (80VA) the backup time is around 2<br />

hours.<br />

o UPS Units can be bought in most electronic shops.<br />

10. Why are there He and CO curves during tidal breathing?<br />

o The MolarMass Signal is influenced by the MM <strong>of</strong> the Inspiration gas (MM=28.95<br />

g mol -1 ) and the MM <strong>of</strong> the expiration (MM=29.78 g mol -1 ). (Green Line) During<br />

tidal breathing. T his has nothing to do with Helium. ( No He gas apply during tidal<br />

breathing)<br />

o The purple line (CO) during tidal breathing is measuring the blank value which<br />

will be deducted in the exhalation.<br />

The scientists believe that during this tidal breathing phase both the He (Green Line)<br />

and the CO (Brown Line) should be flat. Yes but we use it to measure and calibrate the<br />

molar mass signal<br />

11. Does EasyOne Pro give the “same results” as the instrument(s) I<br />

am using currently, do the results match?<br />

o The issue <strong>of</strong> quality <strong>of</strong> DLCO results is a very complex one. A published<br />

ERS article deals with the aspects addressing this subject and is<br />

recommended reading for anyone concerned with quality issues:<br />

http://www.<strong>ndd</strong>.ch/UserData/Download_01471_00.pdf<br />

o EasyOne Pro is the first and so far only instrument that has been<br />

developed based on the ATS/ERS accepted standards <strong>of</strong> the Hans<br />

Rudolph (Crapo/Jensen) simulator (designed around 2005)<br />

o According to statements by the lead author <strong>of</strong> the above mentioned<br />

article EasyOne Pro is as close to the standards as he has seen it possible.<br />

And there is no calibration ! No “wrong” manipulation.<br />

o Another poster deals with the current situation <strong>of</strong> leading brands:<br />

http://www.<strong>ndd</strong>.ch/UserData/Download_00210_00.pdf<br />

o Your <strong>ndd</strong> dealer has a Hans Rudolph simulator available to make your<br />

evaluation easier and base it on the only internationally accepted standard<br />

available today.<br />

12. How does single breath DLCO compare to multi breath?<br />

o Single breath DLCO is the only DLCO method standardized by “ATS/ERS:<br />

Task Force; Standardization <strong>of</strong> Lung Function Testing” and as such the<br />

guide line: http://www.<strong>ndd</strong>.ch/UserData/Download_01384_00.pdf All<br />

considerations are dealt with in this publication.<br />

13. How does single breath TLC compare to other methods?<br />

o The relationship and circumstance are described in the publication linked<br />

here http://www.<strong>ndd</strong>.ch/UserData/Download_01468_01.pdf<br />

o If the TLC is "adjusted" with the Punjab equation, then RV would need to<br />

be recalculated. Since TLC = RV + VC, and using FVC as the VC then RV =<br />

TLC - FVC. This is simple, and the increase in RV will be equal to the<br />

increase in TLC due to the Punjab equation. So if the "Pre Punjab" TLC is


say 4.5 and FVC is 3.5 it follows that the RV is 1.0 (TLC - FVC) and it follows<br />

that RV/TLC ratio is 1.0/4.5 or 0.22. If the Punjab equation increases the<br />

TLC from 4.5 to say 5.0, FVC stays the same so the new RV = 5.0 - 3.5 =<br />

1.5, and the new RV/TLC ratio is 1.5/5.0 or 3.0. But keep in mind that the<br />

new RV/TLC ratio is "NOT" the old RV (1.0) divided by the "NEW" TLC (5.0),<br />

which would equal 0.20.<br />

14. What is UPG, the 3rd available mode in EasyOne Pro?<br />

o Ultrasound Pneumology: is a new <strong>ndd</strong> developed tidal breathing<br />

measurement combining the CO2 concentration measurement with flow<br />

(TrueFlow and MolMass at the same time and point <strong>of</strong> measurement).<br />

o The resulting CO2 / flow - volume capnography curve shows changes <strong>of</strong><br />

slopes and in the different phases.<br />

o The changes in appearance (shape) are characteristic for certain lung<br />

diseases and their severities<br />

o Several publications regarding Vd (Fowler), COPD classification:<br />

http://www.<strong>ndd</strong>.ch/UserData/Download_00158_00.pdf etc. have been<br />

published.<br />

o The optionally available UPG mode allows measurements for R&D<br />

purposes in a standardized hardware setting with a research SW..<br />

15. Displayed Device Messages<br />

Message 1 – Check Spirette insertion<br />

Message is shown in the following cases:<br />

- Spirette not correctly inserted<br />

- Use <strong>of</strong> old or defective Spirette<br />

- First measurement on different altitude as last<br />

measurement (e.g. first measurement in<br />

Mexico City after measurement in Switzerland)<br />

Message 2 – Save New Values<br />

How to avoid message:<br />

- Correctly insert Spirette (completely and with<br />

correct position, see position marker)<br />

See also chapter “inserting the <strong>spirette</strong>” in the<br />

operator’s manual.<br />

- Use new <strong>ndd</strong> Spirette<br />

- If this message is shown repeatedly, another<br />

message with a question to save values is<br />

shown, press “Yes” and repeat measurement.<br />

Message is shown in the case, where the<br />

“Check Spirette Insertion” message was<br />

shown several times. This means that the<br />

signal amplitude has changed. If Spirette<br />

is correct inserted press “Yes” to save<br />

the new values. This might be necessary<br />

on large altitude changes between<br />

measurements.


Message 3 - Device initialization failed! Retry or call your local service agent<br />

This message is shown if “No” was<br />

pressed in above showed “Save New<br />

Values” dialog. Press “Yes” to save the<br />

new signal amplitude values.<br />

Other Reported Messages<br />

volumen de muestreo en posición maxima: Appeared with the DLCO calibration with syringe.<br />

Sampling volume at maximum position.<br />

This message is shown if:<br />

- The expired volume was too small (expired volume should be > 1liter)<br />

- If it was not exhaled immediately after the DLCO valve has open after the breath hold.<br />

- If no stable gas concentration plateau during exhalation after breath hold was detected.<br />

E.g. too low expiration flow.<br />

- Do you have a screenshot <strong>of</strong> that test?<br />

Fallo en CO calib: in the biological control.<br />

Test Failure: CO-Calib? More details required and exact error message.<br />

Does this happen several times? Was the gas connected? Is everything correctly<br />

mounted and the system is tight?<br />

Is it only the case in the biological control? Could you send the database or a<br />

picture <strong>of</strong> the maneuver?<br />

16. Absence or faulty CO/MM Curve (Human Test)<br />

Different Error Messages occur when the Activate Button is pressed to late.<br />

� The DLCO Valve unit does not close


For correct timing proceed in the following way:<br />

Advice Patient for tidal breathing:<br />

Advice Patient to exhale<br />

Press Activate while patient exhale


Attention!<br />

Press Activate at the early Start <strong>of</strong> last expiration well before<br />

gas inhalation. The Instrument automatically detects end <strong>of</strong><br />

exhale and opens the gas demand valve (audible sound)<br />

17. Which breath hold time definition do you apply?<br />

We apply the Jones Meade breath hold definition as specified in the ERS ATS, you<br />

can select between 8 and 11 seconds hold time.<br />

18. Why are we not getting a predicted for the VA and the KCO?<br />

the predicted reference used is Miller – and also why is it not reporting the breathhold<br />

time?<br />

The VA parameter has only a predicted in the Knudson reference. Therefore also<br />

the KCO (which is DLCO / VA) will have a predicted in the Knudson reference.<br />

19. About breath hold time (BHT)<br />

The breath hold time (BHT) is only shown for each trial. There is no definition for a<br />

summary value <strong>of</strong> the BHT. Therefore in the summary report the BHT is not shown.<br />

In the DLCO report you have the table with values for each trial. There you should<br />

see all BHTs. You can add this table also to the summary report if you wish.


20. Mark DLCO attempts as unacceptable.<br />

Double click first at position (1.), then click on the scroll bar (2.) and then press<br />

the button (3.).<br />

21. Show date <strong>of</strong> test in patient list<br />

In order to increase the performance the last test date is not shown by default.<br />

The current logo seems not very intuitive ... The screenshot below shows you how<br />

to show the dates:<br />

22. How does the EasyOne Pro know that a VI target <strong>of</strong> 85% has been<br />

achieved?<br />

The Inspiratory Target Volume in the EasyOne Pro is set at 85% <strong>of</strong> the patient’s VC<br />

max value which we know from the patient’s previous tests (DLCO, FVL, or SVC) or<br />

in the case <strong>of</strong> a new patient where spirometry is not performed prior to the DLCO<br />

Test then we use the predicted VC, FVC value for that patient.<br />

23. How does the operator know when a patient has performed a<br />

Valsalva Manoeuvre or Muller Manoeuvre as there does not seem<br />

to be any mouth pressure measurement as there is with other<br />

DLCO systems?<br />

The EasyOne Pro does make a mouth pressure measurement during the DLCO Test<br />

which you can select at the end <strong>of</strong> the DLCO Test and you will see the pressure<br />

gradient and the pressure value displayed in Mbar and also the Flow plot and<br />

display in LPS.


24. Sample and Washout Volumes<br />

We “let it all out”, analyze the curve and pick the respective segments, including<br />

discard volume. That is (in fact quite <strong>of</strong>ten) more than 1 liter and goes against the<br />

traditional understanding. But it is correct.<br />

See the following extract from the ERS ATS guide lines 2005<br />

25. Restriction for EasyOne ProS<strong>of</strong>twareupdate<br />

Please do not use MAC for unzipping S<strong>of</strong>tware update package.<br />

26. Any new question?<br />

For any other question or to get references please do not hesitate to contact Juerg<br />

Anderauer: Anderauer@<strong>ndd</strong>.ch or mobile +41 79 238 6804.<br />

Sepember 2010<br />

Juerg Anderauer

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!