Disinfection temperature setting (degrees centigrade)<br />
Thermal disinfection is defined by holding the chamber at a given minimum temperature for a given<br />
period of time. In the UK the minimum temperature is 80 Degrees Centigrade which has to be held<br />
for a period of 60 seconds.<br />
Regional thermal disinfection standards may vary around the world; adjustments made to this<br />
parameter define what temperature the chamber must achieve before the one minute cycle will<br />
start. Default setting is 82 Degrees Centigrade which provides a safety margin ensuring that the 80<br />
Degrees centigrade is met.<br />
Should the chamber temperature drop below the number set in this parameter; a 60 second timer<br />
(Thermal disinfection time) will reset to zero starting again once the temperature has reached the<br />
minimum temperature set within this parameter.<br />
Default setting: 82 Degrees Centigrade<br />
<strong>Mini</strong>mum setting: 75 Degrees Centigrade<br />
Maximum setting: 90 Degrees Centigrade<br />
Initial fill (seconds)<br />
The initial cold-wash time in seconds. Default setting 30 seconds. Extending this time may produce<br />
better results from the wash but should only be considered were the water pressure is sufficient to<br />
refill the water tank in line with the flow rate of the pump. If adjusting this parameter test and ensure<br />
that the water pump does not run dry.<br />
Default setting: 40 Seconds - <strong>Mini</strong><br />
Default setting: 30 Seconds - <strong>Midi</strong><br />
Default setting: 20 Seconds – <strong>Optima</strong> 2<br />
<strong>Mini</strong>mum setting: 5 Seconds<br />
Maximum setting: 45 Seconds<br />