Maximum cycle time (minutes)<br />
Parameter for controlling the time in minutes of the steam disinfection time out. If the heater takes<br />
longer than the set time within this parameter to reach the disinfection temperature, the cycle is<br />
aborted and a fault code is displayed<br />
Default setting: 10 minutes<br />
<strong>Mini</strong>mum setting: 02 minutes<br />
Maximum setting: 15 minutes<br />
Thermal disinfection time (seconds)<br />
Thermal disinfection is defined by holding the chamber at a given minimum temperature for a given<br />
period of time. In the UK the minimum temperature is 80 Degrees Centigrade which has to be held<br />
for a period of 60 seconds.<br />
Regional thermal disinfection standards may vary around the world; adjustments made to this<br />
parameter define what temperature the chamber must achieve before the one minute cycle will<br />
start. Default setting is 75 seconds which provides a safety margin ensuring that the 60 seconds is<br />
met.<br />
Default setting: 75 seconds<br />
<strong>Mini</strong>mum setting: 10 seconds<br />
Maximum setting: 120 seconds<br />
Blockage retries<br />
Parameter used to control the amount of times the machine will attempt to clear a blockage:<br />
Default setting: 1 attempt<br />
<strong>Mini</strong>mum setting: 1 attempt<br />
Maximum setting: 10 attempts<br />
Copyright © 2017 DDC Dolphin Ltd. Page 16 of 40 April 17 SGPC – <strong>V25.1</strong>