Maximum cycle time (minutes)<br />

Parameter for controlling the time in minutes of the steam disinfection time out. If the heater takes<br />

longer than the set time within this parameter to reach the disinfection temperature, the cycle is<br />

aborted and a fault code is displayed<br />

Default setting: 10 minutes<br />

<strong>Mini</strong>mum setting: 02 minutes<br />

Maximum setting: 15 minutes<br />

Thermal disinfection time (seconds)<br />

Thermal disinfection is defined by holding the chamber at a given minimum temperature for a given<br />

period of time. In the UK the minimum temperature is 80 Degrees Centigrade which has to be held<br />

for a period of 60 seconds.<br />

Regional thermal disinfection standards may vary around the world; adjustments made to this<br />

parameter define what temperature the chamber must achieve before the one minute cycle will<br />

start. Default setting is 75 seconds which provides a safety margin ensuring that the 60 seconds is<br />

met.<br />

Default setting: 75 seconds<br />

<strong>Mini</strong>mum setting: 10 seconds<br />

Maximum setting: 120 seconds<br />

Blockage retries<br />

Parameter used to control the amount of times the machine will attempt to clear a blockage:<br />

Default setting: 1 attempt<br />

<strong>Mini</strong>mum setting: 1 attempt<br />

Maximum setting: 10 attempts<br />