COUNTER-BALANCING THE VERTICAL MOVEMENT<br />
When accessories, including the photography unit, are fitted to the main body, the vertical<br />
counter-balance movement may need to be adjusted. To correct this, auxiliary springs must be<br />
fitted.<br />
Major Combinations of Accessories and Necessary Auxiliary Springs<br />
Each auxiliary spring consists of 2 identical springs.<br />
Do not use different springs in a set.<br />
COUNTER-BALANCE PROCEDURE<br />
1 Turn the control lever clockwise and raise the base to the top position, remove the center<br />
screw and take off the cover.<br />
22<br />
ASSEMBLY PROCEDURE<br />
Accessories<br />
Necessary auxiliary spring<br />
Without<br />
tonometer<br />
With<br />
tonometer<br />
TV relay lens TL-55 + SONY DXC-33 (DXC-390)<br />
Auxiliary<br />
TV attachment TL-56 + Nikon Microsystem<br />
None spring<br />
Digital camera unit DC-1<br />
SO-AS0<br />
Auxiliary<br />
TV relay lens TL-54 + JVC KY-F70<br />
spring<br />
SO-AS0<br />
Beam splitter + Observation tube Auxiliary Auxiliary<br />
Camera attachment SR-53 + NIKON mount + FUJI FINEPIX spring spring<br />
S2 Pro<br />
SO-AS0 SO-AS1<br />
Auxiliary<br />
Beam splitter + TV relay lens (1/2C) + JVC KY-F70<br />
spring<br />
SO-AS2<br />
Cover