COUNTER-BALANCING THE VERTICAL MOVEMENT<br />

When accessories, including the photography unit, are fitted to the main body, the vertical<br />

counter-balance movement may need to be adjusted. To correct this, auxiliary springs must be<br />

fitted.<br />

Major Combinations of Accessories and Necessary Auxiliary Springs<br />

Each auxiliary spring consists of 2 identical springs.<br />

Do not use different springs in a set.<br />

COUNTER-BALANCE PROCEDURE<br />

1 Turn the control lever clockwise and raise the base to the top position, remove the center<br />

screw and take off the cover.<br />

22<br />

ASSEMBLY PROCEDURE<br />

Accessories<br />

Necessary auxiliary spring<br />

Without<br />

tonometer<br />

With<br />

tonometer<br />

TV relay lens TL-55 + SONY DXC-33 (DXC-390)<br />

Auxiliary<br />

TV attachment TL-56 + Nikon Microsystem<br />

None spring<br />

Digital camera unit DC-1<br />

SO-AS0<br />

Auxiliary<br />

TV relay lens TL-54 + JVC KY-F70<br />

spring<br />

SO-AS0<br />

Beam splitter + Observation tube Auxiliary Auxiliary<br />

Camera attachment SR-53 + NIKON mount + FUJI FINEPIX spring spring<br />

S2 Pro<br />

SO-AS0 SO-AS1<br />

Auxiliary<br />

Beam splitter + TV relay lens (1/2C) + JVC KY-F70<br />

spring<br />

SO-AS2<br />