Operation (5-Lead, 5L, LT and XG) Operation (XG)<br />

NOTES: For optimal results, DO NOT place the airway adapter between the endotracheal<br />

tube and the elbow, as this may allow patient secretions to block the adapter windows.<br />

Position the airway adapter with it’s windows in a vertical and NOT a horizontal position.<br />

This helps keep patient secretions from “pooling” on the windows.<br />

5. Check that the connections have been made correctly by verifying a proper CO 2<br />

waveform (capnogram) on the monitor display.<br />

NOTE: The CO 2 waveform is displayed by choosing Auto Display<br />

CO 2 as Waveform 2 or 3 and the Resp. Source as CO 2 /Off<br />

or CO 2/ECG in the Resp menu.<br />

6. The sensor cable should face away from the patient. To secure the sensor cable safely<br />

out of the way, attach the sensor cable holding clips to the airway tubing, then connect<br />

the sensor cable to the clips.<br />

1.6.12.5 Use of Neonatal Airway Adapter<br />

The neonatal airway adapter should be used when monitoring patients with endotracheal<br />

tube diameters less than or equal to 4.0 mm.<br />

1. Verify the windows are clean and dry. Clean or replace the adapter if necessary.<br />

2. Snap the airway adapter into the Capnostat® sensor. Align the arrow on the bottom of<br />

the airway adapter with the arrow on the bottom of the Capnostat®. Press the sensor<br />

and airway adapter together until they “click”.<br />

3. If necessary, perform an adapter calibration. See “Adapter Calibration” on page 1-<br />

135.<br />

4. Place the Capnostat® / airway adapter assembly between the elbow and the ventilator<br />

circuit wye.<br />

To<br />

Ventilator<br />

CAPNOSTAT<br />

CO Sensor<br />

2<br />

Wye<br />

FIGURE 1-73 Neonatal Airway Adapter<br />

R<br />

To<br />

Patient<br />

Neonatal Airway Adapter<br />

NOTES: For optimal results, DO NOT place the airway adapter between the endotracheal<br />

tube and the elbow, as this may allow patient secretions to block the adapter windows.<br />

Position the airway adapter with it’s windows in a vertical and NOT a horizontal position.<br />

This helps keep patient secretions from “pooling” on the windows.<br />

<strong>Passport</strong> 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 137