Operation (5-Lead, 5L, LT and XG) Controls, Indicators and Connectors (XG)<br />

Front Panel (Continued)<br />

SET UP CENTRAL ECG IBP NIBP ALARMS TRENDS<br />

BEEP<br />

SELECT ADMIT LEAD ZERO 1 START<br />

VOLUME DISPLAY VOLUME<br />

EXIT DISCHARGE SIZE ZERO 2 INTERVAL<br />

LIMITS<br />

HOLD 3 SEC. TO CLEAR<br />

PRINT<br />

SILENCE FREEZE CO2 PUMP DEFLATE<br />

HOLD 3 SEC. TO STOP<br />

15<br />

14 13<br />

13. START (NIBP)<br />

A single action key which initiates an NIBP measurement. This function is not available if a<br />

measurement is in progress.<br />

14. INTERVAL (NIBP)<br />

A repeat action key which selects the interval setting for NIBP measurements. Values wrap<br />

around at the lowest/highest choices. Choices are: Off, Continuous, 1, 2.5, 5, 10, 15, 20,<br />

30, 60, and 120 minutes. Five seconds after the selection is made and the key is released,<br />

the new interval takes effect.<br />

15. DEFLATE (NIBP)<br />

A single action key which stops any NIBP measurement in progress, including any timed<br />

measurement sequence, and deflates the cuff. When a timed measurement is stopped an<br />

“NIBP: DEFLATE” message followed by an “NIBP: IDLE” message displays in the NIBP<br />

message window until the timer mode is restarted by either pressing START or changing the<br />

interval.<br />

16. VOLUME (Alarms and Alarm LED)<br />

A repeat action key which adjusts the alarm volume in 5 steps. There is no off position for this<br />

tone. The tone wraps to the minimum volume once the maximum is reached. The red alarm<br />

bell LED illuminates when an alarm condition exists. See “Alarms” on page 1-149 for<br />

detailed information on alarm situations.<br />

17. LIMITS (Alarms)<br />

A single action key which provides access to the alarm menu to view and select alarm<br />

values. Press SELECT to change the alarm values. Press the LIMITS key again to access<br />

additional alarm parameters (when the Gas Module is installed). To return the display to the<br />

main screen, press the EXIT key.<br />