Technical data of Transparent Thermostats <strong>CT</strong> <strong>53</strong>, <strong>CT</strong> <strong>53</strong> HT, <strong>CT</strong> <strong>53</strong> <strong>TT</strong>, <strong>CT</strong> 54, <strong>CT</strong> 54 HT<br />
Data transfer:<br />
Interface: bi-directional serial acc. to EIA RS232C<br />
Data format: selectable<br />
Baud rate: 1200, 2400, 4800 or 9600<br />
Parity: no / odd / even<br />
Length: 7 data bits<br />
Stop bit: 1<br />
Protocol: <strong>CT</strong>S/RTS: (on or off)<br />
Mains power connection:<br />
230 V~, 50 Hz; on request 115 V~, 60 Hz<br />
Total power consumption: at 230 V, [VA]: 2070; at 115 V, [VA]: 1070<br />
Measurements: Total, incl. heater, cooling coil and line cord connection:<br />
(W x D x H) [cm]: 38 x 26 x 34<br />
Housing above the cover plate of the baths:<br />
(W x D x H) [cm]: 12 x 15.5 x 14.5<br />
Weight: Approx. [kg]: 5<br />
Environment:<br />
Ambient temperature: +10 ... +40 °C (for storage and transportation)<br />
Air humidity acc. DIN EN 61 010, part 1:<br />
Max. rel. humidity 80 % for temperatures up to +31 °C,<br />
linear decrease down to 50 % rel. humidity at a temperature of +40 °C<br />
Safety measures:<br />
Variable excess temperature protection, adjustable from [°C] :+20 .. +160<br />
Low liquid level protection: fixed<br />
according to DIN 12 876 safety class III<br />
Protection class: IP 20 acc. EN 60 529<br />
Power supply:<br />
acc. to class 1, VDE 0106 T1<br />
not for use in explosive atmosphere