Gebrauchsanleitung <strong>SDU</strong> <strong>1100</strong><br />

min of [W .3(z1),ITA.3(z1)] (mW)<br />

Associated z1 (cm)<br />

Acoustic zbp (cm)<br />

Parameter zsp (cm)<br />

deq(zsp) (cm)<br />

fc (MHz)<br />

Dim of Aaprt X (cm)<br />

Y (cm)<br />

PD (ì sec)<br />

PRF (kHz)<br />

Other pr@PIImax (MPa)<br />

Information deq@PIImax (cm)<br />

Focal FLx(cm) Length FLy(cm) IPA.3@MImax (W/cm 2 )<br />

Operating<br />

Control<br />

Conditions<br />

Control 1<br />

Control 2<br />

Control 3<br />

Control 4<br />

Transducer Model UB10R-065U Operating Mode Pulsed Doppler<br />

TIS TIB TIC<br />

Index Label MI scan non-scan non-<br />

Aaprt 1 Aaprt 1 scan<br />

Global Maximum Index Value < 1 < 1 1.3<br />

pr.3 (MPa)<br />

W0 (mW) 49.0<br />

min of [W .3(z1),ITA.3(z1)] (mW)<br />

Associated z1 (cm)<br />

Acoustic zbp (cm)<br />

Parameter zsp (cm) 4.0<br />

deq(zsp) (cm) 0.21<br />

fc (MHz) 5.2<br />

Dim of Aaprt X (cm) 1.6<br />

Y (cm) 0.8<br />

PD (ì sec)<br />

PRF (kHz)<br />

Other pr@PIImax (MPa)<br />

Information deq@PIImax (cm) 0.21<br />

Focal FLx(cm) Length FLy(cm) IPA.3@MImax (W/cm 2 )<br />

Operating<br />

Control<br />

Conditions<br />

Transducer Model UB10R-065U Operating Mode Color Flow inc. M mode<br />

Control 1 Focus 40mm<br />

Control 2<br />

Control 3<br />

Control 4<br />

Freq. Lowest<br />

TIS TIB TIC<br />

Index Label MI scan non-scan non-<br />

Aaprt 1 Aaprt 1 scan<br />

Global Maximum Index Value < 1 < 1 < 1 < 1<br />

pr.3 (MPa)<br />

W0 (mW)<br />

min of [W .3(z1),ITA.3(z1)] (mW)<br />

Associated z1 (cm)<br />

Acoustic zbp (cm)<br />

Parameter zsp (cm)<br />

deq(zsp) (cm)<br />

fc (MHz)<br />

Dim of Aaprt X (cm)<br />

Y (cm)<br />

PD (ì sec)<br />

PRF (kHz)<br />

Other pr@PIImax (MPa)<br />

Information deq@PIImax (cm)<br />

Focal FLx(cm) Length FLy(cm) IPA.3@MImax (W/cm 2 )<br />

Operating<br />

Control<br />

Conditions<br />

Control 1<br />

Control 2<br />

Control 3<br />

Control 4<br />

Acoustic uncertainties for acoustic quantities are as follows power : 30 % +15 % pressure : 2 % +0 % intensity : 32 % +19 % center frequency : 3 % +3 %<br />

D-18 Anhang D-Rev D.doc <strong>SDU</strong> <strong>1100</strong>/Rev.D/1006