SECTION 3, continued<br />
Table 2 Formazin Standard Preparation<br />
Standard Step 1 Step 2 Step 3<br />
20 NTU<br />
200 NTU<br />
1000 NTU<br />
4000 NTU<br />
7500 NTU<br />
Ampule<br />
Add 100 mL of dilution water to a<br />
clean 200-mL class A volumetric<br />
flask<br />
Add50mLofdilutionwatertoa<br />
clean 100-mL class A volumetric<br />
flask<br />
Add50mLofdilutionwatertoa<br />
clean 100-mL class A volumetric<br />
flask<br />
Transfer approximately 30 mL of well<br />
mixed 4000-NTU Formazin stock<br />
solution to a clean sample cell. No<br />
dilution is required.<br />
Use in the sealed ampule as<br />
provided. DO NOT open the vial or<br />
use its contents as dilution stock.<br />
* Class A volumetric pipets can be used in place of a TenSette Pipet.<br />
With a TenSette ® * Pipet, add 1.00<br />
mL of well-mixed 4000-NTU<br />
Formazin stock solution to the<br />
200-mL flask.<br />
With a TenSette* Pipet, add 5.00 mL<br />
of well-mixed 4000-NTU Formazin<br />
stock solution to the 100-mL flask.<br />
With a TenSette* Pipet, add 25.00<br />
mL of well-mixed 4000-NTU<br />
Formazin stock solution to the<br />
100-mL flask.<br />
Note: Install each cell with the orientation mark aligned with the cell holder reference mark<br />
when using matched vials.<br />
38<br />
Dilute to the mark with dilution water.<br />
Stopper and mix.<br />
Dilute to the mark with dilution water.<br />
Stopper and mix.<br />
Dilute to the mark with dilution water.<br />
Stopper and mix.