SECTION 3, continued<br />

Table 2 Formazin Standard Preparation<br />

Standard Step 1 Step 2 Step 3<br />

20 NTU<br />

200 NTU<br />

1000 NTU<br />

4000 NTU<br />

7500 NTU<br />

Ampule<br />

Add 100 mL of dilution water to a<br />

clean 200-mL class A volumetric<br />

flask<br />

Add50mLofdilutionwatertoa<br />

clean 100-mL class A volumetric<br />

flask<br />

Add50mLofdilutionwatertoa<br />

clean 100-mL class A volumetric<br />

flask<br />

Transfer approximately 30 mL of well<br />

mixed 4000-NTU Formazin stock<br />

solution to a clean sample cell. No<br />

dilution is required.<br />

Use in the sealed ampule as<br />

provided. DO NOT open the vial or<br />

use its contents as dilution stock.<br />

* Class A volumetric pipets can be used in place of a TenSette Pipet.<br />

With a TenSette ® * Pipet, add 1.00<br />

mL of well-mixed 4000-NTU<br />

Formazin stock solution to the<br />

200-mL flask.<br />

With a TenSette* Pipet, add 5.00 mL<br />

of well-mixed 4000-NTU Formazin<br />

stock solution to the 100-mL flask.<br />

With a TenSette* Pipet, add 25.00<br />

mL of well-mixed 4000-NTU<br />

Formazin stock solution to the<br />

100-mL flask.<br />

Note: Install each cell with the orientation mark aligned with the cell holder reference mark<br />

when using matched vials.<br />

38<br />

Dilute to the mark with dilution water.<br />

Stopper and mix.<br />

Dilute to the mark with dilution water.<br />

Stopper and mix.<br />

Dilute to the mark with dilution water.<br />

Stopper and mix.