SECTION 5, continued<br />

5.2 Flow-Cell Kits (Low Pressure)<br />

The Low Pressure Flow-Cell systems (manual or automated kits) use an<br />

innovative sample cell design* with a baffled inlet and dual outlets that minimize<br />

accumulation of entrapped air bubbles and heavy solid particles in the cell (see<br />

Figure 8). The glass cell is threaded on both ends to accommodate plastic end<br />

caps. The cell has an approximate volume of 22 mL when assembled. The parts<br />

disassemble easily for thorough cleaning.<br />

Figure 8 Low Pressure Flow Cell Assembly<br />

* U.S. patent D358,448; other patents pending.<br />

Sample is introduced into the top of the cell (see Figure 9). A baffle deflects<br />

the incoming sample to the side wall of the cell minimizing turbulence in the<br />

light path.<br />

Sample is discharged from top and bottom outlets. The top outlet collects and<br />

expels air bubbles and particles that tend to float. The conical-shaped bottom<br />

outlet collects settleable solids; water discharged from the bottom outlet carries<br />

the settled solids out of the cell. This novel, dual-outlet design eliminates dead<br />

volume in the cell to provide rapid, thorough flushing of the cell from one sample<br />

to the next.<br />