Figure 3. Loading volume influences band width; 5-20% gradient. While the 0.1 and 0.25 ml<br />

samples gave identical peak shapes, the 0.5 and 0.75 ml sample volumes clearly lead to a broadening<br />

of the band and the resultant loss of resolution. Thus, for the SW41 rotor tubes, sample volumes in<br />

5-20% gradients must be limited to 0.25 ml and fairly short runs if high resolution is required. These<br />

gradients were fractionated using the hand crank (pre-motorized).<br />

A final limitation of these gradients will become evident when examining the 5-45% profiles<br />

below. The range of S values that can be separated on a given gradient is limited here because the<br />

low viscosity and density of the 20% end point does not retard the sedimentation of large particles.<br />

In Figure 1, for example, particles with an S value of 250-750S would have been separated adequately<br />

but 1000S particles would have pelleted in this run.<br />

14