04.01.2014 Views

CMG-40T-1 - Güralp Systems Ltd

CMG-40T-1 - Güralp Systems Ltd

CMG-40T-1 - Güralp Systems Ltd

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>CMG</strong>-<strong>40T</strong>-1<br />

Handheld Control Unit's Display Select knob. If using a HCU, also<br />

check that the Centring Select knob is set to OFF.<br />

2. If the vertical component needs adjusting, remove the cap on the lid<br />

which protects the Verticalpot with a flat-bladed screwdriver (provided).<br />

3. Insert the screwdriver through the opening, and engage the pot. An LED<br />

flashlight may be useful for locating the head.<br />

4. Turn the pot either way until the offset readout is as close to 0 V as<br />

possible.<br />

5. Repeat steps 2 – 5 for the north/south and east/west components.<br />

10 Issue A