<strong>CMG</strong>-<strong>40T</strong>-1<br />
Signal meter<br />
The upper section of the HCU contains a simple voltmeter for monitoring<br />
various signals from the instrument.<br />
• To monitor the low-gain outputs, switch the dial to V, N/S or E/W LOW<br />
VEL according to the component you want to monitor.<br />
• To monitor the high-gain outputs (on a <strong>40T</strong>-1 with that option), switch<br />
the dial to V, N/S or E/W HIGH VEL.<br />
• To monitor the mass position outputs, switch the dial to V, N/S or E/W<br />
MASS POS. Whilst you are adjusting mass position offsets, you should<br />
also switch the instrument out of broadband mode by switching the<br />
rightmost CENTRING SELECT switch to 1 SEC VEL, or by holding<br />
down the CENTRE button on a breakout box.<br />
• You can set the range of the meter with the RANGE switch. When<br />
switched to 10 V, the meter ranges from –10 to + 10 V (as marked.)<br />
When switched to 1 V, the range is –1 to +1 V.<br />
8 Issue A