<strong>CMG</strong>-<strong>40T</strong>-1<br />

Again, this method produces inferior bonding to the instrument, and moisture<br />

becomes trapped between the hard surfaces. We recommend the use of fine dry<br />

sand (right) contained in a box if necessary, which can also insulate the<br />

instrument against convection currents and temperature changes. Sand has the<br />

further advantage of being very easy to install, requiring no preparation.<br />

Finally, many pit installations have a large space around the seismometer,<br />

covered with a wooden roof. Large air-filled cavities are susceptible to currents<br />

which produce lower-frequency vibrations, and sharp edges and corners can<br />

give rise to turbulence. We recommend that a wooden box is placed around the<br />

sensor to protect it from these currents. Once in the box, the emplacement may<br />

be backfilled with fresh turf to insulate it from vibrations at the surface, or<br />

simply roofed as before.<br />

By following these guidelines, you will ensure that your seismic installation is<br />

ready to produce the highest quality data.<br />