<strong>CMG</strong>-<strong>6T</strong><br />

5 Specifications<br />

Velocity output bandwidth<br />

Velocity output sensitivity<br />

Peak output<br />

Lowest spurious resonance<br />

Linearity<br />

Cross-axis rejection<br />

10s - 100 Hz (standard),<br />

30s - 100 Hz (optional) or<br />

1s - 100 Hz (model <strong>CMG</strong>-<strong>6T</strong>-1)<br />

2 x 1,200 V/ms -1 (standard),<br />

2 x 10,000 V/ms -1 (optional)<br />

2 x 2,000 V/ms -1 (optional) or<br />

2 x 1,000 V/ms -1 (optional)<br />

±10V (20V peak-to-peak)<br />

450 Hz<br />

> 90 dB<br />

> 65 dB<br />

Electronics self-noise -172 dB (relative to 1 m 2 s -4 Hz -1 )<br />

Operating temperature<br />

-40 to +75°C<br />

Temperature sensitivity < 0.6 V per 10 °C<br />

Mass recentring range<br />

Construction<br />

Case diameter<br />

Case height (with handle)<br />

Weight<br />

Power supply<br />

Optional low-power version<br />

Calibration controls<br />

Offset nulling<br />

Optional remote nulling<br />

Optional accessories<br />

±3° from horizontal<br />

Hard anodised aluminium case<br />

with 'O'-ring seals throughout<br />

154 mm<br />

207 mm<br />

2.49 kg<br />

10 - 36 VDC<br />

5 VDC supply, ±4.5 V output<br />

common signal and enable lines<br />

exposed on sensor connector<br />

potentiometers, adjustable<br />

through case<br />

microprocessor controlled motor<br />

drives<br />

Break-out box<br />

Hand-held control unit<br />

30 Issue B