5 Technical Information<br />

RS 232/RS 485 Communication Description<br />

RS 232/RS 485 Communication Description<br />

Both the RS 232 and the RS 485 interfaces are available on the<br />

connector J2<br />

J2 – Serial Port<br />

Tab. 6<br />

PIN N.<br />

SIGNAL NAME<br />

1 + 5 V out<br />

2 TX (RS232)<br />

3 RX (RS232)<br />

4 Spare<br />

5 GND<br />

6 A+ (RS485)<br />

7 Spare<br />

8 B- (RS485)<br />

9 Spare<br />

The communication protocol is the same (see the structure below),<br />

but only the RS 485 manages the address field. Therefore to enable<br />

the RS 485 is necessary to select the type of communication as well<br />

as the device address by means of the T-plus software.<br />

NOTE<br />

The RS-485 is a 2-wires (gnd optional) half-duplex communication link.<br />

106/126 <strong>HS452</strong>, <strong>HS652</strong> <strong>Rotary</strong> <strong>Vane</strong> <strong>Pumps</strong> User Manual / 87-900-978-01 (L)