Technical Information<br />

RS 232/RS 485 Communication Description<br />

5<br />

Command: READ PUMP STATUS<br />

Source: PC<br />

Destination: Pump (with address = 3)<br />

02 83 32 30 35 30 03 38 37<br />

STX ADDR WINDOW RD ETX CRC<br />

Source: Pump (with address = 3 in stop status)<br />

Destination: PC<br />

02 83 32 30 35 30 30 30 30 30 30 03 38 37<br />

STX ADDR WINDOW DATA (STATUS) ETX CRC<br />

Windows meaning<br />

Tab. 9<br />

WIN TYPE DESCRIPTION<br />

000 L START/STOP<br />

(1=START; 0=STOP)<br />

001 L LOW SPEED<br />

[0=OFF / 1=ON]<br />

008 L REMOTE/SERIAL CONFIGURATION<br />

(1=REMOTE; 0=SERIAL)<br />

102 N SET POINT VALUE (Hz)<br />

105 N SET POINT HYSTERESIS<br />

(%) [0-100]<br />

108 N BAUD RATE<br />

(0-4) [600-1200-2400-4200-9600]<br />

117 N LOW SPEED ADJUST (Hz)<br />

120 N ROTATIONAL FREQUENCY SETTING<br />

[Hz] (High speed adjust)<br />