Technical Information<br />
RS 232/RS 485 Communication Description<br />
5<br />
The controller can answers with the following response types:<br />
Tab. 8<br />
Type Length Value Description<br />
Logic 1 byte -<br />
Numeric 6 bytes -<br />
Alpha-numeric 10 bytes -<br />
ACK 1 byte (0x6)<br />
NACK 1 byte (0x15)<br />
Unknown Window 1 byte (0x32)<br />
Data Type Error 1 byte (0x33)<br />
Out of Range 1 byte (0x34)<br />
Win Disabled 1 byte (0x35)<br />
After a read instruction of a logic<br />
window<br />
After a read instruction of a<br />
numeric window<br />
After a read instruction of an<br />
alphanumeric window<br />
The command execution has been<br />
successfully completed<br />
The command execution has been<br />
failed<br />
The specified window in the<br />
command is not a valid window<br />
The data type specified in the<br />
command (Logic, Numeric or<br />
Alphanumeric) is not accorded with<br />
the specified Window<br />
The value expressed during a write<br />
command is out of the range value<br />
of the specified window<br />
The specified window is Read Only<br />
or temporarily disabled (some<br />
values cannot be changed while<br />
pump is running)<br />
NOTE<br />
Using the RS 485 interface, the message structure remains identical to the one<br />
used for the RS 232 interface, the only difference being that the value assigned to<br />
the ADDRESS <br />