Technical Information<br />
RS 232/RS 485 Communication Description<br />
5<br />
Communication Format<br />
• Physical level: both RS232 e RS485<br />
• 8 data bit<br />
• no parity<br />
• 1 stop bit<br />
• baud rate: 600/1200/2400/4800/9600 programmable<br />
Communication Protocol<br />
The communication protocol is a MASTER/SLAVE type where:<br />
• Host = MASTER<br />
• Pump = SLAVE<br />
The communication is performed in the following way:<br />
1. the host (MASTER) send a MESSAGE + CRC to the controller<br />
(SLAVE);<br />
2. the controller answer with an ANSWER + CRC to the host.<br />
The MESSAGE is a string with the following format:<br />
++++++<br />
Where:<br />
NOTE<br />
When a data is indicated between two quotes (‘...’) it means that the indicated<br />
data is the corresponding ASCII character.<br />
• (Start of transmission) = 0x02<br />
• (Unit address) = 0x80 (for RS 232)<br />
• (Unit address) = 0x80 + device number (0 to 31)<br />
(for RS 485)<br />
<strong>HS452</strong>, <strong>HS652</strong> <strong>Rotary</strong> <strong>Vane</strong> <strong>Pumps</strong> User Manual / 87-900-978-01 (L) 107/126