HS452, HS652 Rotary Vane Pumps - Agilent Technologies
HS452, HS652 Rotary Vane Pumps - Agilent Technologies
HS452, HS652 Rotary Vane Pumps - Agilent Technologies
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Technical Information<br />
Connections to the Inlet and Exhaust Flanges<br />
5<br />
NOTE<br />
Application of an oil trap filter is necessary to avoid pollution of the surrounding<br />
atmosphere by the oil present in the exhaust duct during pump operation.<br />
CAUTION!<br />
Never block the pump exhaust line. This would cause overpressure in the casing<br />
with the risk of breaking the pumping block envelope.<br />
<strong>HS452</strong>, <strong>HS652</strong> <strong>Rotary</strong> <strong>Vane</strong> <strong>Pumps</strong> User Manual / 87-900-978-01 (L) 89/126