Technical Information<br />

RS 232/RS 485 Communication Description<br />

5<br />

Operational limits:<br />

Tab. 10<br />

Input Voltage (V)<br />

Controller status<br />

< 85 Power fail<br />

85 – 140 Operative<br />

140 – 170 Power fail<br />

170 – 270 Operative<br />

> 270 Power fail<br />

Status Led<br />

Tab. 11<br />

Led status<br />

Off<br />

Green flashing<br />

Green<br />

Red<br />

Slow red flashing (2 Hz)<br />

Quick red flashing (10 Hz)<br />

Orange + Off<br />

Controller status<br />

Stop<br />

Ramp – Autotuning<br />

Normal Operation<br />

Fail<br />

Oil level fail<br />

Oil level sensor fail<br />

Reset + selftest<br />

NOTE<br />

When you switch the controller on an electronic self-test is performed, during<br />

such phase you’ll see the led Orange for 1 sec. And turned off for two other<br />

seconds. Immediately after the first phase the controller perform a specific test<br />

on the pump rotor in order to be sure that it isn’t locked.<br />

<strong>HS452</strong>, <strong>HS652</strong> <strong>Rotary</strong> <strong>Vane</strong> <strong>Pumps</strong> User Manual / 87-900-978-01 (L) 113/126