51<br />

The remote control plug P2 can be used for the following functions:<br />

- Connection of outer safety switch opens when a malfunction apears.<br />

- Remote control from starting or stopping of the TMP motor (start and stop in front panel).<br />

- We advise to screen the remote control cable.<br />

See figures 6/2 and 6/3 - P2 plug connection.<br />

Interface output J6 : see figure 6/4 at the end of the manual.<br />

Note: If an electromagnetic valve is utilized on the same vacuum system as the turbomolecular pump then<br />

the valve must have a shielded cable. The shielded cable must properly grounded. This precaution must be<br />

followed to prevent interference with the TMP frequency converter.<br />

3.4 - STARTING<br />

I<br />

For a new TMP, or after two months, we advise the user that switch 110nl1 the TMP at atmospheric pressure<br />

for about 5 minutes with converter (it is necessary for a repartition of grease in ball bearings).<br />