Alarm and HOLD Messages<br />

Message Released by Cause<br />

Alarm Sensocheck Polarization / Cable<br />

(22 mA) Error messages Flow (CONTROL input)<br />

Alarm<br />

contact<br />

opens<br />

Flow (current input)<br />

ERR A / ERR B: conductance > 250,000 µS<br />

ERR A / ERR B: conductivity > 1,000 µS/cm<br />

HOLD HOLD HOLD via menu or input<br />

(Last/Fix) CONF <strong>Configuration</strong><br />

CAL<br />

Calibration<br />

SERVICE Service<br />

To generate a message via the CONTROL input:<br />

Open the “<strong>Configuration</strong>“ menu to activate the message:<br />

Power supply<br />

12...24 V AC/DC<br />

13<br />

12<br />

CONTROL<br />

input<br />

Stratos Pro A4...<br />

To generate a message via the current input:<br />

Open the “<strong>Configuration</strong>“ menu to activate the message:<br />

0(4) ... 20 mA:<br />

5<br />

6<br />

Current<br />

input<br />

Stratos Pro A4...<br />

30