MEATEST<br />
M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />
Zeroing function can be used e.g. to compensate the voltage drops at the cables, to suppress any remanent value<br />
of the signal etc.<br />
Overloading can damage the multimeter.<br />
Notes:<br />
* The current inputs of the multimeter are protected from current overloading by a fuse located in 140-41<br />
cable adapter. The fuse can be replaced after turning the cap of the fuse holder. The procedure is<br />
described in the adapter’s manual.<br />
* The multimeter can be used as a DC milivoltmeter with 20 mV, 200 mV, 2 V, 10 V ranges. The<br />
appropriate input terminals are PTHU and PTLU.<br />
* Maximum allowed voltage on the input terminals against ground terminal is 20 V.<br />
* Frequency measurement is possible up to 15 kHz. Input signal must be within 0.2 to 5 V range.<br />
squarewave or pulse shape of the input signal is expected.<br />
Operation manual v42 43