06.04.2014 Views

M140 Multifunction Calibrator / Tester - meatest.cz

M140 Multifunction Calibrator / Tester - meatest.cz

M140 Multifunction Calibrator / Tester - meatest.cz

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

Operation manual


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

Content<br />

Operation manual ...................................................................................................................... 1<br />

Basic Information ...................................................................................................................... 3<br />

Preparation for operation .......................................................................................................... 5<br />

Inspecting package contents, selecting the installation location .....................................................5<br />

Power-on ..............................................................................................................................................5<br />

Warm-up time .....................................................................................................................................5<br />

Replacement of fuse ............................................................................................................................6<br />

Safety precautions...............................................................................................................................6<br />

Description of controls .............................................................................................................. 7<br />

Front panel ..........................................................................................................................................7<br />

Rear panel .........................................................................................................................................13<br />

Control of the calibrator .......................................................................................................... 14<br />

Selection of function .........................................................................................................................14<br />

Setting the value of output signal ....................................................................................................14<br />

Setting relative deviation ..................................................................................................................16<br />

Change of value by factor of ten ......................................................................................................17<br />

Connection / disconnection of output terminals ............................................................................17<br />

Setting the frequency ........................................................................................................................18<br />

Generation of calibrated voltage .....................................................................................................19<br />

Generation of calibrated current ....................................................................................................21<br />

Generation of non-harmonic shapes ...............................................................................................22<br />

Simulation of resistance and capacitance .......................................................................................23<br />

Generation of electric power and energy ........................................................................................26<br />

Generation of frequency ..................................................................................................................31<br />

Simulation of temperature sensors..................................................................................................34<br />

Multimeter ................................................................................................................................ 38<br />

Basic menu ........................................................................................................................................38<br />

Function selection .............................................................................................................................39<br />

Setting the measurement range .......................................................................................................39<br />

Units of measurement .......................................................................................................................40<br />

Use of calculation formula ...............................................................................................................40<br />

Setting function parameters.............................................................................................................41<br />

Start of measurement .......................................................................................................................42<br />

Zero function .....................................................................................................................................42<br />

Simultaneous functions ....................................................................................................................44<br />

Operation manual v42 1


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

<strong>Tester</strong> ....................................................................................................................................... 46<br />

Basic menu ........................................................................................................................................ 46<br />

Execution of test program ............................................................................................................... 46<br />

Programming the test ...................................................................................................................... 47<br />

Setting the type of signals and the number of steps ....................................................................................... 48<br />

Setting the numeric values of the test ............................................................................................................ 48<br />

Setting the relays............................................................................................................................................ 49<br />

Setup menu .............................................................................................................................. 50<br />

Calibration mode ..................................................................................................................... 54<br />

Error messages ........................................................................................................................ 72<br />

Functional description of the calibrator ................................................................................ 74<br />

<strong>Calibrator</strong>’s maintenance ....................................................................................................... 80<br />

Verification test ....................................................................................................................... 82<br />

System control ......................................................................................................................... 90<br />

IEEE-488 bus properties ................................................................................................................. 90<br />

RS232 bus properties ....................................................................................................................... 90<br />

Command syntax ............................................................................................................................. 91<br />

Standard Status Data Structures .................................................................................................. 105<br />

Examples of use..................................................................................................................... 108<br />

Calibration of measurement instruments .................................................................................... 108<br />

Multimeters .................................................................................................................................................. 108<br />

Powermeters ................................................................................................................................................ 110<br />

Counters and oscilloscopes .......................................................................................................................... 112<br />

Thermometers .............................................................................................................................................. 112<br />

Measurement .................................................................................................................................. 113<br />

Voltage, current and frequency .................................................................................................................... 113<br />

Measurement of resistance or temperature using resistance temperature sensors ....................................... 114<br />

Measurement of temperature using thermocouples ..................................................................................... 115<br />

Strain gauge sensors for non-electrical values ............................................................................................. 115<br />

Testing of regulation and measurement sets and evaluation units ............................................ 117<br />

Use of Opt. 140-41 cable adapter ................................................................................................................ 117<br />

Use of Option 40/60 cable adapter............................................................................................................... 118<br />

Use of Option 70 .......................................................................................................................................... 118<br />

Examples of tests ......................................................................................................................................... 118<br />

Testing ......................................................................................................................................................... 120<br />

Specification .......................................................................................................................... 121<br />

Accessories............................................................................................................................. 128<br />

Operation Manual Supplement ............................................................................................ 129<br />

Change 1 ......................................................................................................................................... 129<br />

2 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

Basic Information<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> is a multifunction calibrator-tester, to be used primarily as a standard<br />

for calibration laboratories. It can be used for calibration of any measuring instrument which measures voltage,<br />

current, resistance, capacitance and frequency. It generates fixed non-harmonic signals to allow calibration<br />

of measuring instruments using signals with non-zero harmonic distortion. Frequency, amplitude and duty cycle<br />

of output signal can be adjusted. M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> is also suitable for basic calibration<br />

of oscilloscopes.<br />

The calibrator includes a function which simulates resistance and thermocouple temperature sensors<br />

and a built-in multimeter, which can be used simultaneously. Transducers of various types, regulators and<br />

sensing units can be therefore checked without the need for additional measuring instruments.<br />

Programmable functions of the calibrator, when used as a tester, include programming of a 10-step testing<br />

procedure, which completes automatically and displays a PASS/FAIL information in the end. This feature is<br />

linked to an independent relay output, which allows the control of other equipment.<br />

Basic features of the calibrator include: generation of calibrated DC and AC voltage in the range of 0<br />

µV to 1000 V, DC and AC current in the range of 0 µA to 20 A (50 µA to 1000 A when using a 50-turn coil).<br />

Maximum precision of the calibrator is 0.0035 % for DC voltage, 0.03 % for AC voltage, 0.013 % for DC<br />

current and 0.055 % for AC current. Maximum frequency range is 20 Hz to 50 kHz. The calibrator can generate<br />

periodic non-harmonic signal with defined duty cycle. This facilitates especially the checks of multimeters and<br />

their accuracy when measuring non-harmonic DC signals.<br />

The calibrator can also simulate a resistance or capacitance. Resistance range is 0 Ω to 50 MΩ;<br />

capacitance range is 1 nF to 50 µF, the accuracy suits the calibration of common multimeters. Basic accuracy<br />

of resistance ranges is 0.03 %. Basic accuracy of capacitance ranges is 0.5 %. The resistance can be used with<br />

AC signals up to 300 Hz to 1 kHz, depending on set-up value.<br />

Frequency ranges of the calibrator can generate a squarewave signal with definable and calibrated duty<br />

cycle and amplitude in the 1 mV to 10 V range and 0 to 10 kHz frequency range. Moreover, squarewave signal<br />

with very steep rising edge can be generated up to 20 MHz. Frequency ranges can be used to calibrate the<br />

corresponding frequency ranges of multimeters, as well as to calibrate the input sensitivity and time bases of<br />

oscilloscopes.<br />

Powermeter mode can be used to calibrate DC and AC single phase powermeters and energy meters.<br />

Voltage range is up to 240 V and current range is up to 10 A, power factor range is -1 to +1 and the resolution is<br />

1 % in the 40 Hz to 400 Hz frequency range. The voltage output can supply loads up to 30 mA, which allows the<br />

calibration of mechanical powermeters.<br />

Simulation of temperature sensors is yet another feature which can be used to calibrate thermometers<br />

and heat sensing units. The calibrator allows the simulation of all common Pt and Ni resistance sensors and R, S,<br />

B, J, T, E, K, N type thermocouples. Compensation of cold junction of thermocouple is achieved by entering<br />

the respective temperature using the calibrator’s keyboard. The accuracy of simulated temperature sensors<br />

depends on the value and type of sensor and ranges from 0.04 o C to 0.5 o C for resistance sensors and from<br />

0.4 o C to 4.3 o C for thermocouples.<br />

Internal multimeter with 20 mA, 20 mV, 200 mV and 10 V basic ranges and 0.01 % accuracy can be<br />

used to measure normalized signals coming from transducers, external thermocouples or resistance sensors or to<br />

measure pressure and force using strain gauge sensors.<br />

The calibrator includes many other features which facilitate easy use. For example relative deviation from set<br />

value of the output, currently displayed uncertainty of the output signal, calibration and testing procedures etc.<br />

The concept of calibrator control and indication of its status is based on flat luminiscent display, which provides<br />

all necessary information. The calibrator is controlled by opening menus on the display and selection from<br />

menus. Frequently used functions are assigned direct-control keys. The calibrator comes with standard GPIB bus<br />

and RS-232 serial line, which allow the calibrator to be controlled from a PC.<br />

The calibrator can easily fit within calibration systems featuring MBASE/WinQbase software support.<br />

Operation manual v42 3


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

ATTENTION !<br />

The calibrator generates life-threatening high voltage.<br />

The calibrator can only be used in line with this<br />

Manual.<br />

4 Operation manual v42


MEATEST<br />

Preparation for operation<br />

Inspecting package contents, selecting the installation location<br />

Basic package includes the following items:<br />

• <strong>Multifunction</strong> calibrator<br />

• Power cord<br />

• Spare fuse T4L250/T, T8L250/T<br />

• Operation manual.<br />

• Test report<br />

• Test cable 1000V/20 A 2 pcs<br />

• Cable adapter Option 40<br />

• Cable adapter Option 60<br />

• Cable adapter Option 70<br />

• RS 232 cable<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

The calibrator should be powered by 230/115 V – 50/60 Hz mains. It is a laboratory instrument whose<br />

parameters are guaranteed at 23±2 o C. Before powering on the instruments, place it on a level surface. Do not<br />

cover the vents at the bottom side and the fan opening at the rear panel.<br />

Power-on<br />

• Before connecting the calibrator to the mains, check the position of the mains voltage selector located at the<br />

rear panel.<br />

• Plug one end of the power cord into the connector located at the rear panel and connect the other end of the<br />

power cord into a wall outlet.<br />

• Switch on the mains switch located at the rear panel. Flat display is lit.<br />

• The calibrator performs internal hardware checks for 5 seconds.<br />

• After the tests conclude, the calibrator resets to its reference state, i.e. the following parameters are set:<br />

Function<br />

DC voltage<br />

Range<br />

20 V<br />

Set value<br />

10 V<br />

Output terminals OFF<br />

GPIB address of the calibrator is factory-preset to 2. This value is valid until the user changes it.<br />

Note. The calibrator resets to its reference status in case of power switching off and reconnection.<br />

Warm-up time<br />

The calibrator works after it is switched on and the initial checks complete. Specified parameters are only<br />

guaranteed after the instrument warms up for 60 minutes. During this period, the instrument cannot be calibrated.<br />

The display shows “cannot access the calibration” message if calibration is attempted during this period.<br />

Operation manual v42 5


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

Replacement of fuse<br />

The calibrator includes a fuse located in the mains connector at the rear panel. Replace the fuse as follows:<br />

• Switch off the calibrator<br />

• Remove the end of power cord from the mains connector at the rear panel.<br />

• Insert the blade of a flat screwdriver into the opening cut in the mains voltage selector and pull out the fuse<br />

holder.<br />

• Remove the fuse and replace it with new fuse of the same rating.<br />

Safety precautions<br />

The instrument has been designed in Safety Class I according to EN 61010-1. The design reflects the<br />

requirements of A2 amendment of the standard.<br />

Safety is ensured by the design and by the use of specific component types.<br />

The manufacturer is not liable for the damage caused by modification of the construction or replacement of parts<br />

with non-original ones.<br />

Safety symbols used on the equipment<br />

Warning, reference to the documentation<br />

Warning - risk of electric shock<br />

Danger - high voltage<br />

6 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

Description of controls<br />

Front panel<br />

The front panel of the calibrator includes a flat luminiscent display, control buttons and output terminals. The<br />

following picture shows the control part of the front panel.<br />

3 5<br />

1 2 4 6 7<br />

1 Display buttons<br />

There are five buttons below the display, whose meaning changes depending on the contents of the display.<br />

These buttons usually call-up the MENU, allow range change, step, logging of values etc.<br />

2 Cursor buttons<br />

Using these buttons, the cursor can be controlled within allowed limits on the display. The keyboard includes<br />

two buttons () which allow the cursor to be set to the required position at the display. The cursor can be<br />

moved to the left or right. These buttons are usually used to step through the options and to move from one<br />

option to another or between the menu levels. Numeric values can be set in some control modes as well. In these<br />

cases, the buttons marked (∧, ∨) allow the user to increase or decrease the number at the cursor button.<br />

The central button is used to confirm the selection (ENTER), or to SELECT from the menu.<br />

3 Potentiometer<br />

The potentiometer integrates several functions. By turning the knob to the left or right, the user can:<br />

• step through the options<br />

• enter numeric values<br />

Operation manual v42 7


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

The function of the potentiometer can usually be performed by the cursor buttons. The central button is used to<br />

confirm the selection (ENTER).<br />

4 Numeric keyboard<br />

The keyboard allows the entry of numeric values on the display. The central button is used to confirm the<br />

selection (ENTER). CANCEL button can be used to cancel the entry.<br />

5 Function buttons<br />

Function buttons can be used to call-up the functions of the calibrator directly. The following buttons are<br />

provided:<br />

function<br />

DC voltage<br />

AC voltage<br />

DC current<br />

AC current<br />

resistance / capacitance<br />

power / energy<br />

frequency<br />

internal multimeter<br />

simulation of temperature sensors<br />

button<br />

U / DC<br />

U / AC<br />

I / DC<br />

I / AC<br />

R – C<br />

P – E<br />

F<br />

METER<br />

T<br />

After the function mode is changed, the parameters of the respective function are restored. If the respective<br />

function was never used, the calibrator resets to its reference values. Reference values for individual functions<br />

are listed below.<br />

function value parameters<br />

DC voltage 10V - -<br />

AC voltage 10 V f = 1000 Hz<br />

DC current 100 mA - -<br />

AC current 100 mA f = 1000 Hz<br />

resistance<br />

100 kΩ<br />

capacitance 1 µF<br />

power 100 W f = 100 Hz *1<br />

energy<br />

frequency 1000 Hz U = 1 Vsym<br />

multimeter 10 V DC voltage<br />

simulation of temperature sensors 100 o C Pt 100/1.385, ITS90<br />

cold junction temperature of TC sensors 23 o C R<br />

*1<br />

U = 100 V, I = 1 A, PF(power factor) = 1 LA, active power is displayed in Watts<br />

8 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

6 Output / input terminals buttons<br />

OUTPUT button is used to connect the output signal of the calibrator to the output terminals. The connection is<br />

confirmed by red LED and a symbol at the display.<br />

METER button can be used to connect the input terminals to the internal multimeter. The connection is<br />

confirmed by green LED.<br />

7 Output / input terminals<br />

Output signal of the calibrator is connected to the output terminals. Current ranges are connected to +I / -I<br />

terminals, frequency output is connected to FREQ terminal. All other functions (voltage, resistance,<br />

capacitance) are connected to Hi / Lo terminals.<br />

GND terminal is connected to the chassis of the calibrator. It is connected to the ground terminal of the mains<br />

plug. Using the SETUP MENU of the calibrator, the output terminals of the calibrator can be grounded as well.<br />

Grounding is done internally by connecting Lo and GND terminals using a relay. This circuit design is suitable<br />

for most calibrations, when the object (multimeter) being calibrated is floating.<br />

AUXILIARY connector creates input of internal multimeter. It includes a limited range of output signals of the<br />

calibrator as well. The layout of individual pins and their meanings are listed in the following table.<br />

Auxiliary connector can be used with one of cable adapters Opt. 40, 60, 70, Opt. 140-41. <strong>Calibrator</strong> can<br />

recognize which type of adapter is connected and displays the information on front panel display.<br />

Operation manual v42 9


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

pin label signal Limitation<br />

1 0V5MER common terminal of multimetr power supply source<br />

2 GND ground (protection earth)<br />

3 SIMLI RC simulator output, current terminal Li Umax.= 10Vpp, Imax.=40mA<br />

4 SIMLU RC simulator output, voltage terminal Lu Umax.= 10Vpp, Imax.=40mA<br />

5 GND ground (protection earth)<br />

6 L common terminal of multimeter input<br />

7 -U low output terminal for DC voltage range<br />

8 -I low output terminal for DC current range<br />

9 NG2 sort function output, contact 2 of sort relay Umax.=50Vpp, Imax.=100 mA<br />

10 PTLI resistance temperature sensor input terminal Li Umax.= 10Vpp. R


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

8 Display<br />

1a 1b 1c 1d 1e 2a 3 2b 1g 1f<br />

The display is divided to three horizontal sections:<br />

1. OUTPUT section<br />

This section displays the set-up values of generated signals and the data related to the calibrator status. The<br />

section includes the following types of data:<br />

a) Information line<br />

• designation of display section: OUTPUT<br />

• error messages. The messages appear when an attempt is made to set up an invalid state of the<br />

calibrator, if analogue circuits of the calibrator are overloaded or if a communication error occurs<br />

when the calibrator is controlled using GPIB bus.<br />

• real date and time, if its display is set-up in the setup menu.<br />

b) Auxiliary data<br />

This line displays the total value of output signal if a non-zero relative deviation is set.<br />

c) Main data<br />

This line displays the main data of the output signal and the unit of measurement (using double size<br />

signs). The line also includes two symbols (▼▲) to define the actual position of the cursor during<br />

adjustment of the value. buttons can be used to move the cursor and ∧, ∨ buttons to change the<br />

value. (The value can be also changed using the potentiometer).<br />

d) Monitoring line<br />

This line displays the numbers entered using the numeric keyboard when the main data are set using the<br />

numeric keyboard. The information allows the entered information to be checked.<br />

e) Minor data<br />

There are two lines displaying the minor data of the output signal, especially:<br />

• set relative deviation from main set value in %<br />

• frequency (for DC voltage, current, power, energy functions)<br />

• set value of current, voltage or power factor (phase) for power, energy functions<br />

• value of R0 resistance and the type of resistance temperature sensor<br />

Operation manual v42 11


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

• cold junction temperature of TC sensors and the selected type of TC sensor<br />

• value of amplitude and shape type for frequency function<br />

f) Information section<br />

The information section located in the right part of the display displays additional information related to<br />

the selected function:<br />

• symbol of connected or disconnected output terminals.<br />

At the same time, a LED located above the OUTPUT button is lit.<br />

• information about remote/local control of the calibrator. If the calibrator is controlled remotely,<br />

REM is displayed. If the calibrator is controlled locally using the keyboard, LOCAL is displayed.<br />

• information about the use of 50-turn coil (COIL x50) at the current output of the calibrator, if this<br />

feature is turned on using the SETUP menu.<br />

• information about the type of connected cable adapter, if used<br />

• information about the grounding method of output terminals: GND I, GND U as set up using the<br />

setup menu.<br />

g) Information about the uncertainty of the output signal<br />

This section displays the maximum error of the main value of the output signal. The value is calculated<br />

using the main specification listed in the User’s Manual and it is displayed in %.<br />

2. INPUT section<br />

This section displays the values measured by the multimeter. The section includes following data:<br />

a) Main value of measured signal<br />

This line displays the measured value and the unit of measurement. If the input signals exceed the<br />

permitted range, OVERFLOW is displayed.<br />

b) Designation of selected function of the multimeter<br />

Symbolic display of selected function of the multimeter: V DC, mA DC, mV DC, R 4W, Freq, T TC, T<br />

RTD, SGS, ACAL.<br />

3. Display buttons section<br />

This line displays the symbolic descriptions which define the meaning of four related display buttons. The<br />

respective meanings are as follows:<br />

symbol button function note<br />

x 10<br />

increase set value 10 x<br />

: 10 decrease set value 10 x<br />

Shape selection of signal shape only for U, I, F functions<br />

+/- reversed polarity of output voltage and current only for DC U, DC I functions<br />

EXIT move up one level only for F, P-E functions<br />

Calib.<br />

SETUP<br />

enter the calibration menu<br />

enter the setup menu<br />

TC type selection of thermocouple sensor type only for T function<br />

RTD type selection of resistance temperature sensor type only for T function<br />

f enter the frequency of the signal only for U, I function<br />

MODE select the unit of measurement only for AC P-E function<br />

12 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

Rear panel<br />

The rear panel of the calibrator includes ventilation holes, power cord socket with fuse, mains voltage selector,<br />

mains switch, IEEE 488 connectors for connection to GPIB bus and type plate with serial number.<br />

5<br />

1<br />

2<br />

4 3<br />

1 air inlet - forced ventilation<br />

2 air outlet - forced ventilation<br />

3 GPIB, RS-232 connectors<br />

4 power cord socket with fuse, mains voltage selector, mains switch<br />

5 type plate<br />

Operation manual v42<br />

13


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

Control of the calibrator<br />

Selection of function<br />

After the power is switched on and the initial checks complete, the calibrator resets to its reference status, i.e. DC<br />

voltage output with set value of 10 V and output terminals disconnected. Internal multimeter is switched off. The<br />

status of the calibrator can be changed using the buttons located at the front panel in one of the following ways:<br />

1. Change of function by pressing one of direct function buttons<br />

After pressing one of the U, I, DC-AC, R-C, P-E, F, T, METER buttons, the calibrator switches to the desired<br />

function mode and resets to the reference or to the most recently used parameter setting.<br />

2. Connection /disconnection of output terminals<br />

After pressing the OUTPUT button, the output terminals of the calibrator are connected/disconnected.<br />

3. Connection /disconnection of multimeter<br />

After pressing the INPUT button, the multimeter starts measuring the value present at the input terminals,<br />

depending on the function mode of the multimeter. The measurement is only possible when any of Opt. 140-xx<br />

adapters is connected to the AUXILIARY connector.<br />

4. Entry to the setup menu<br />

After pressing the SETUP button, options of the SETUP MENU appear on the display and the display buttons<br />

allow the entry to the calibration mode (CALIB) or entry to the mode when the calibrator is used as a tester<br />

(TESTER). Previous function is restored by pressing of EXIT display button.<br />

Setting the value of output signal<br />

All function modes allow several methods of setting the main value of the output signal:<br />

Entry of the value using numeric keyboard<br />

• use the numeric keyboard to select the desired value. After the first digit is entered, symbols of unit<br />

of measurements are displayed above the display buttons. The monitor line displays the symbols<br />

[ _ _ _ _ _ _ _ _ ].<br />

• the same entry can be started by pressing the central cursor button<br />

• after the entry is complete (the value is displayed on the monitor line), press the display button<br />

below the desired unit of measurement (V, mV or µV in the example below)<br />

• the value is copied to the main display and the monitor line disappears.<br />

14 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

Entry of the value using cursor buttons<br />

• press , ∧ or ∨ button. The display now includes cursor marks which point to the active digit.<br />

• ∧ and ∨ buttons can be used to change the active digit. buttons can be used to change the<br />

position of the cursor marks<br />

• to get to the default screen, press EXIT button or keep pressing the center cursor button until there<br />

is no [ _ _ _ _ _ _ _ ] under any value. All values can be set using the buttons or the potentiometer.<br />

Entry of the value using the potentiometer<br />

• press the potentiometer knob. The display now includes cursor marks which point to the active<br />

digit<br />

• turn the knob to change the active digit<br />

• press the potentiometer knob to change to the mode which allows to change the value of the active<br />

digit. ← and → symbols are displayed above the active digit. Active digit can be changed by<br />

turning the knob.<br />

• turn the knob to change back to the mode which allows to change the position of the active digit.<br />

• to get to the default screen, keep pressing the center cursor button until there is no [ _ _ _ _ _ _ _ ]<br />

under any value, or press EXIT button. All values can be set using the buttons or the potentiometer.<br />

Reverse polarity<br />

In DC voltage and DC current modes, the polarity of the output value can be reversed by pressing +/- display<br />

button. „ - “ symbol appears in front of the main data value.<br />

Operation manual v42 15


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

MEATEST<br />

Setting relative deviation<br />

All function modes of the calibrator except frequency mode allow a relative deviation of output value from the<br />

main data to be set using a separate display. Relative deviation is displayed in the “minor data” section of the<br />

display and is designated with „ ∆%= 00.0000 % “ symbol. The relative deviation can be entered using one of<br />

the methods described above, e.g. using the numeric keyboard, cursor keys or the potentiometer.<br />

Setting relative deviation using numeric keyboard<br />

• keep pressing the center cursor button until [<br />

_ _ _ _ _ _ _ ] symbols appear under the relative<br />

deviation value in the “minor data” section of the display<br />

• enter the desired deviation and confirm the value by pressing “ % “ display button or by pressing<br />

ENTER on the numeric keyboard<br />

• the auxiliary line below the main data on the display displays the total value of output signal<br />

including the unit of measurement<br />

• the value of the signal at output terminals is: the value indicated by the main display + ∆ %.<br />

Maximum relative deviation which can be entered is ± 30.000 %.<br />

The deviation can be positive or negative. If negative deviation is desired, press the display button labeled +/-. If<br />

positive deviation is then desired, press “ +/- “ button again. The polarity of the relative deviation can be<br />

reversed using the cursor buttons or the potentiometer as well.<br />

Setting relative deviation using cursor keys<br />

• keep pressing the center cursor button until [<br />

_ _ _ _ _ _ _ ] symbols appear under the relative<br />

deviation value<br />

• press , ∧ or ∨ button. The display now includes cursor marks which point to the active digit<br />

• ∧ and ∨ buttons can be used to change the active digit. buttons can be used to change the<br />

position of the cursor marks<br />

• to get to the default screen, keep pressing the center cursor button until there is no [ _ _ _ _ _ _ _ ]<br />

under any value, or press EXIT button. All values can be set using the buttons or the<br />

potentiometer..<br />

16<br />

Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

Setting relative deviation using potentiometer<br />

• keep pressing the center cursor button until [ _ _ _ _ _ _ _ ] symbols appear under the relative<br />

deviation value in the “minor data” section of the display<br />

• press the potentiometer knob. The display now includes cursor marks which point to the active<br />

digit. Turn the knob to change the value of the active digit<br />

• press the potentiometer knob to change to the mode which allows to change the position of the<br />

active digit. ← and → symbols are displayed above the active digit. The position of the active digit<br />

can be changed by turning the knob.<br />

• turn the knob to change back to the mode which allows to change the value of the active digit<br />

• to get to the default screen, keep pressing the center cursor button until there is no [ _ _ _ _ _ _ _ ]<br />

under any value, or press EXIT button. All values can be set using the buttons or the potentiometer.<br />

If a non-zero relative deviation is set, the main data can be changed as well. The value of the output signal is<br />

always recalculated. If a zero relative deviation is set, the “minor data” section is not displayed.<br />

Change of value by factor of ten<br />

All functions of the calibrator allow the increase of the output value by 10 or reduction of the output value by 10.<br />

Such operation is equivalent to the change of internal range only in U, I, P-E modes. If the change results in<br />

overflow or underflow of calibrator’s range, an error message appears:<br />

Value too large !<br />

Value too small !<br />

if the resulting value is too large<br />

if the resulting value is too small<br />

Range change<br />

• Press the display button labeled “ x10 “ if you want to increase the range, “ :10 “ to decrease the<br />

range.<br />

• The main value shown on the display is increased 10x (reduced 10x)<br />

P-E function changes the current, not voltage, when the range is changed.<br />

R-C function changes the set value 10x. The procedure, however, cannot be used to step the internal<br />

resistance/capacitance ranges, which are not decimal.<br />

T function also has other than decimal ranges and the change of set value 10x therefore does not correspond to<br />

internal range change. Internal range change in this case depends on the temperature sensitivity of the<br />

resistance/capacitance.<br />

Connection / disconnection of output terminals<br />

After switching on the output terminals are disconnected in all modes. Press the OUTPUT button to connect the<br />

output signal to the terminals. Red LED above the OUTPUT button is lit and the information field on the<br />

display shows the following symbol .<br />

Press the OUTPUT button again to disconnect the output terminals. Red LED goes off and the information field<br />

on the display shows the following symbol .<br />

During mode change, output terminals are always disconnected. Output terminals are disconnected also when<br />

changing between voltage and current ranges or when changing between AC and DC ranges is performed.<br />

Operation manual v42 17


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

MEATEST<br />

If voltage over 100 V is set in the voltage mode, special algorithm must be followed to connect the output<br />

terminals. The algorithm is described in the „Generation of calibrated voltage“ chapter of this Manual.<br />

Setting the frequency<br />

Frequency can only be selected in AC voltage (ACU) mode, AC current (ACI) mode , power (P-E) mode and<br />

frequency (f) mode. In each mode the frequency has a slightly different meaning and the frequency is therefore<br />

set in a different manner.<br />

AC voltage (ACU), AC current (ACI)<br />

Set value of frequency is included in the “minor data” section of the display in ACU, ACI, P-E modes.<br />

Frequency change<br />

• First select the AC voltage or AC current mode by pressing U (I), AC buttons or selecting the P-E<br />

mode using the display. Frequency value „f = xxx.xx Hz“ appears in the “minor data” section of<br />

the display. “ f “ symbol is displayed above one of the display buttons.<br />

• After “ f “ display button is pressed, [ _ _ _ _ _ _ _ ] symbols appear below the frequency value.<br />

Numeric keyboard can be used to enter the desired value. Press “ Hz “ or “ kHz “ to confirm the<br />

value. The value can be set using the buttons or the potentiometer.<br />

If too large or too small value is entered, the calibrator displays the maximum (minimum) value which is allowed<br />

for the selected function.<br />

Frequency (F)<br />

Set value of frequency is the main data on the display and the main parameter of the signal. Main data can be set<br />

by direct entry using the numeric keyboard, potentiometer or by changing the digit at the current cursor position.<br />

The setting procedure is described in the “Setting the value of output signal”.<br />

18<br />

Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

If frequency larger or smaller than the calibrator’s range is entered, the calibrator displays an error message:<br />

„Value is too large (small)“.<br />

Generation of calibrated voltage<br />

The multifunction calibrator provides calibrated DC and AC voltage. Output terminals for voltage ranges are<br />

labeled “ Hi “ and “ Lo “ at the front panel. Depending on the setting of the calibrator, voltage up to 1000 Vef<br />

can be present at the terminals .<br />

DC voltage range is 0 to 1000 V.<br />

AC voltage range is 100 µV V to 1000 V.<br />

Output voltage up to 20 V is available at AUXILIARY connector. It can be used only with cable adapter Opt.<br />

140-41.<br />

Control in the voltage mode<br />

• Press “U” button on the calibrator and then select AC or DC mode by pressing “DC-AC” button. The<br />

display shows the following data:<br />

* main data of set voltage<br />

* relative deviation<br />

* uncertainty of output voltage<br />

* frequency (when AC voltage is generated)<br />

* total value of output voltage when non-zero relative deviation is set<br />

• Set the desired value of voltage, including polarity when necessary, frequency and relative deviation. The<br />

signal is yet not connected to the output terminals. The information section of the display shows the symbol<br />

which informs about the disconnection of output terminals.<br />

• Press OUTPUT button.<br />

• Red LED is lit above the OUTPUT terminals to signal the connection of the signal to the output terminals;<br />

the information ion section of the display shows the symbol .<br />

• Calibrated voltage corresponding to set parameters is present at the output terminals.<br />

Operation manual v42<br />

19


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

Control sequence when output voltage over 100 V is selected<br />

When output voltage over 100 V is selected, the information section of the display shows the symbol which<br />

informs that a life-threatening voltage will be present at the output terminals. If the output terminals are currently<br />

connected, they will be disconnected when output voltage over 100 V is selected. OUTPUT button must be<br />

pressed to reconnect the output signal to the output terminals. After the OUTPUT button is pressed, an<br />

interrupted beep is sound, OUTPUT LED is lit and the information section of the display shows the symbol<br />

notifying the user about the connection of the dangerous output signal to the output terminals.<br />

Voltage, polarity, frequency, absolute and relative deviation can be set without the outputs being disconnected.<br />

The output terminals are automatically disconnected when changing between AC and DC ranges or when<br />

changing the function mode.<br />

Using AUTOCAL function<br />

To remove the effect of short-term drift and temperature dependency of small DC voltages, AUTOCAL function<br />

can be used. It can only be activated in the calibration mode. “Calibration mode” lists the respective procedure.<br />

Overloading of terminals<br />

If the output terminals are overloaded or short-circuited in the voltage mode, the calibrator disconnects the signal<br />

from the output terminals and reports „Overload U output“ error.<br />

ATTENTION DANGEROUS VOLTAGE<br />

Never touch the measurement circuit when voltage over 50 V is set and<br />

output terminals are connected!<br />

When working with voltages over 50 V, rules for work with dangerous<br />

voltage must be adhered to.<br />

ATTENTION DANGEROUS VOLTAGE<br />

When the calibrator is controlled remotely, it is not possible to disconnect<br />

the output voltage using the buttons located at the front panel!<br />

The calibrator must be first switched to local control mode by pressing the<br />

LOCAL button and then the output terminals can be disconnected or the<br />

mains switch must be switched off !<br />

20 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

Generation of calibrated current<br />

The multifunction calibrator provides calibrated DC and AC current. Output terminals for voltage ranges are<br />

labeled “ +I “ and “ -I “ at the front panel. The terminals can carry high current and are the only terminals to<br />

which the calibrated object can be connected. Depending on the setting of the calibrator, current up to 20 A ef can<br />

be driven by the terminals.<br />

DC current range is 0 to 20 A<br />

AC current range is 1µA to 20 A<br />

When 50-turn coil (option 140-50) is used, AC current range is 50µA to 1000 A. Output current up to 20 mA is<br />

available at AUXILIARY connector and it is accessible via cable adapter Opt. 41-41 only.<br />

Control in the current mode<br />

• Press “I” button on the calibrator and then select AC or DC mode by pressing “DC-AC” button. The display<br />

shows the following data:<br />

* main data of set current<br />

* relative deviation<br />

* uncertainty of output current<br />

* frequency (when AC current is generated)<br />

* total value of output current when non-zero absolute or relative deviation is set<br />

* time after which the output terminals will be disconnected when the output current over 10 A is<br />

selected.<br />

• Set the desired value of voltage, including polarity when necessary, frequency and relative deviation. The<br />

signal is yet not connected to the output terminals. The information section of the display shows the<br />

symbol which informs about the disconnection of output terminals.<br />

• Connect the load or short the output terminals labeled +I, -I.<br />

• Press OUTPUT button.<br />

• Red LED is lit above the OUTPUT terminals to signal the connection of the signal to the output terminals;<br />

the information section of the display shows the symbol .<br />

• Calibrated current corresponding to set parameters is driven by the output terminals.<br />

• If COILx50 function is activated (see below - Setup functions menu), the optional 50-turn coil must be<br />

connected to output terminals. The calibrator can be used to calibrate 50 µA to 1000 A ammeters. The<br />

calibrator generates AC and DC current within the range up to 20 A.<br />

CAUTION<br />

If GND terminal is connected to Lo, -I terminals, it is prohibited to connect<br />

external load to GND / Hi or GND / +I terminals. Such connection can<br />

damage the calibrator.<br />

Operation manual v42 21


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

Overloading the terminals<br />

When external circuit connected to current output terminals is disconnected or there is higher voltage at the load<br />

than permitted, the calibrator disconnects the output terminals and displays „Overload I output“ message. The<br />

same message can be displayed when 50-turn coil is used for AC current output at frequencies above 80 Hz. It<br />

depends on the set current and the type of ammeter connected.<br />

If the output terminals are disconnected due to time limitation of output current over 10 A, the calibrator displays<br />

„Current timeout !“ message.<br />

Generation of non-harmonic shapes<br />

The multifunction calibrator can generate non-harmonic periodic signals with predefined shape. To allow the<br />

setting of a non-harmonic output shape, the calibrator must be switched to AC U or AC I mode. In both cases, an<br />

indication of the type of output shape (Shape xxxxx) is displayed under the frequency value. Press the respective<br />

display button to change the shape of the output signal.<br />

The calibrator can generate the following shapes:<br />

• SINE harmonic<br />

• PWM POS squarewave - positive, with adjustable duty cycle<br />

• PWM SYM squarewave - symmetrical, with adjustable duty cycle<br />

• PWM NEG squarewave - negative, with adjustable duty cycle<br />

• RAMP A ramp, symmetrical positive<br />

• RAMP B ramp, symmetrical negative<br />

• TRIANGLE triangular, symmetrical<br />

• LIM SINE harmonic with amplitude limitation (truncated sin)<br />

Generation of non-harmonic signals has some limitations:<br />

• non-harmonic voltages can be generated in the 0.1 Hz to 1000 Hz frequency range<br />

• non-harmonic currents can be generated in the 0.1 Hz to 120 Hz frequency range<br />

• generation of these signals is limited to the voltage range up to 200 V and current range up to 2 A<br />

• non-harmonic signals cannot be generated in the P-E (power-energy) mode.<br />

Control in the non-harmonic mode<br />

• Select AC voltage or AC current mode. The main section of the display shows the following data:<br />

* main data of set current or voltage, unit of measurement<br />

* relative deviation<br />

* frequency<br />

* selected SHAPE of the output signal<br />

• Keep pressing SHAPE display button to select the desired shape of the output signal:<br />

The output terminals are automatically disconnected when changing the shape of the output signal or when<br />

changing the relative deviation ∆%, if a non-zero relative deviation is set.<br />

Displayed information<br />

When non-harmonic output shape is selected, the display shows additional information:<br />

• besides the main amplitude data, „pk“ index is displayed, notifying that the displayed main value is the peak<br />

value. Symbol which displays the shape of the output signal is displayed too.<br />

22 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

• below the main data, an information about the shape type „Shape xxxxx“ is displayed.<br />

• below the main data, calculated effective value of the output signal is displayed.<br />

• for squarewave signals, set value of duty cycle „PWM= xx %“ is displayed.<br />

Simulation of resistance and capacitance<br />

The multifunction calibrator can simulate an exact value of resistance or capacitance. The outputs of the<br />

simulator are connected to Hi – Lo terminals and to AUXILIARY connector (pins 20, 21, 22, 23). 4W resistance<br />

is accessible only via cable adapter Opt. 70.<br />

Only two-wire wire connection is available on the front panel terminals Hi-Lo. Both two-wire and four-wire<br />

connection is possible only through AUXILIARY connector. The terminals SIMHI - SIMLI are current<br />

terminals and SIMHU - SIMLU are voltage sensing terminals. Cable adapter Option 70 or cable adapter Option<br />

140-41 must be used for four-wire connection. Type of cable adapter currently connected to the AUXILIARY<br />

connector, is displayed on the display. If Option 70 is connected, label CA 140-70 is displayed in the right side.<br />

If Option 140-4141 is connected, label CA 140-41 is displayed.<br />

Cable adapter Option 70 can be used for four-wire connecting of simulated resistance only. In compare with<br />

direct two-wire wire connection through output terminals Hi – Lo, accuracy of resistance is better with Option 70, see<br />

Technical data. If Cable adapter Option 70 is connected to the AUXILIARY connector, only resistance mode<br />

and resistance temperature simulation mode can be chosen.<br />

The resolution of resistance and capacitance depends on the set value and corresponds to 0.01 % of set value.<br />

Minimum set value is 0.01 Ω<br />

Resistance simulation range is 0 Ω to 50 MΩ.<br />

Capacitance simulation range is 0.9 nF to 50 µF.<br />

Control in the resistance and capacitance mode<br />

• press R-C button on the calibrator. The display shows the set resistance.<br />

• If you want to simulate a capacitance, press R-C button again. The display shows the set capacitance.<br />

• The display shows the following data:<br />

* main data of set resistance (capacitance)<br />

* relative deviation of resistance (capacitance)<br />

* uncertainty of set resistance (capacitance)<br />

* total value of resistance (capacitance) if non-zero deviation is set<br />

Operation manual v42<br />

23


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

MEATEST<br />

• Set desired value of resistance (capacitance) or relative deviation. The value can be set using numeric<br />

keyboard, potentiometer or cursor buttons. Simulated resistance (capacitance) is not yet connected to the<br />

output terminals. The information section of the display shows the symbol which informs about the<br />

disconnection of output terminals.<br />

• Connect the object to be calibrated to the output terminals labeled Hi - Lo.<br />

• Press OUTPUT button.<br />

• Red LED is lit above the OUTPUT terminals to indicate the connection of simulated resistance<br />

(capacitance) to the output terminals. The information section of the display shows the symbol<br />

Simulated resistance (capacitance) is connected to output terminals.<br />

Setting relative deviation<br />

• Keep pressing the center cursor button until [ _ _ _ _ _ _ _ ] symbols appear under the relative deviation<br />

value (∆% = xx.xxxx %).<br />

• The value can be set using numeric keyboard, potentiometer or cursor buttons. Confirm the value by<br />

pressing “%” display button or by pressing ENTER.<br />

Limitations resulting from electronic simulation<br />

Electronic simulation of resistance and capacitance allows setting of a wide range of values with accuracy<br />

sufficient for calibration of common multimeters. Electronic simulation has the following limitations:<br />

• measurement current supplied by the multimeter to be calibrated must not exceed the value specified by the<br />

calibrator’s documentation. If the current is exceeded, the accuracy of simulated value is not guaranteed.<br />

• maximum peak voltage at Hi - Lo terminals supplied by the multimeter to be calibrated must not exceed<br />

specified limits. If the test voltage is exceeded, calibrator disconnects output terminals. Overload message is<br />

displayed on the display.<br />

24<br />

Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

Frequency dependence of resistance and capacitance<br />

Electronic simulator of resistance can be used with DC and AC test signal. Electronic simulator of capacitance<br />

can be used in AC range from 20 Hz to 1000 Hz.<br />

Operation manual v42 25


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

MEATEST<br />

Generation of electric power and energy<br />

The multifunction calibrator can generate exact value of electric power and energy. P-E function provides output<br />

voltage at Hi - Lo terminals and output current at +I - -I terminals. Lo and -I terminals are electrically connected.<br />

Power setting range:<br />

Voltage setting range:<br />

Current setting range:<br />

Power factor setting range:<br />

Frequency setting range:<br />

0.0 VA to 2400 VA<br />

0.2 V to 240 V<br />

0.01 A to 10 A<br />

-1 to +1 (phase –90 to +90 °)<br />

DC, 40 Hz to 400 Hz<br />

Control in power generation mode<br />

• Press “P-E” button on the calibrator and then select AC or DC mode by pressing “DC-AC” button. The<br />

display shows set power value.<br />

• The display shows the following data::<br />

* main value of set power in selected unit of measurement VA, W, VAr<br />

* power factor value PF in negative polarity LA or positive polarity LE or the phase shift between the<br />

voltage and current in °.<br />

* frequency, if AC power is selected<br />

* voltage at Hi - Lo terminals<br />

* current through +I - -I terminals<br />

* uncertainty of set power<br />

• Set desired value of power using numeric keyboard, potentiometer or cursor buttons. Output power is not<br />

yet connected to the output terminals. The information section of the display shows the symbol which<br />

informs about the disconnection of output terminals..<br />

• Connect the instrument to be calibrated to Hi - Lo and +I - -I terminals or short the +I - -I terminals.<br />

• Press OUTPUT button.<br />

• Red LED is lit above the OUTPUT terminals to indicate the connection of simulated electrical power to the<br />

output terminals; the information section of the display shows the symbol<br />

Desired power is connected to output terminals.<br />

Display modes<br />

26<br />

Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

The calibrator can display AC power in one of three ways:<br />

• apparent power in VA<br />

• active power in W<br />

• reactive power in VAr<br />

Keep pressing MODE display button to change the function mode. Along with mode change, the power display<br />

(depending on set power factor) and unit of measurement change as well. If DC power is generated, it is always<br />

displayed in Watts.<br />

The calibrator can display the phase relation of output voltage and current as power factor (–1 to +1) or as phase<br />

shift in degrees (0 to 360 °). SETUP setup menu is used to change the method of displaying the phase relation.<br />

Power setting modes<br />

The calibrator allows several ways of setting the value of generated power.<br />

1. Setting the main power value<br />

• The main value can be changed using numeric keyboard, changing the digit at the cursor position<br />

after selecting the cursor position with buttons, by range change using „x10“, „:10“ display<br />

buttons, or using the potentiometer.<br />

• Output power is changed by changing the value of output current.<br />

Operation manual v42 27


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

2. Setting the voltage<br />

• The main power value can be changed by changing the voltage.<br />

• Select P-E mode and then keep pressing the center cursor button until [ _ _ _ _ _ _ _ ] symbols<br />

appear under the voltage (U = xxx.xxxx V).<br />

• The value can be set using numeric keyboard and confirmed by pressing µV, mV, V display button.<br />

The value can be set using cursor buttons or potentiometer as well.<br />

• Main power value is recalculated using new set voltage and existing setting of current and power<br />

factor.<br />

3. Setting the current<br />

• The main power value can be changed by changing the current..<br />

• Select P-E mode and then keep pressing the center cursor button until [ _ _ _ _ _ _ _ ] symbols<br />

appear under the current (I = xx.xxxx A).<br />

• The value can be set using numeric keyboard and confirmed by pressing µA, mA, A display button.<br />

The value can be set using cursor buttons or potentiometer as well..<br />

• Main power value is recalculated using new set current and existing setting of voltage and power<br />

factor.<br />

28 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

4. Setting the power factor (AC power only)<br />

• If W or Var is indicated, the main power value can be changed by changing the power factor.<br />

Change of power factor does not change the output apparent power.<br />

• Select P-E mode and then keep pressing the center cursor button until [ _ _ _ _ _ _ _ ] symbols<br />

appear under the power factor symbols (PF = x.xxx LA (LE) or Phase = xxx.x).<br />

• The value can be set using numeric keyboard and confirmed by pressing LA/LE (°) button or by<br />

pressing ENTER.<br />

• Main power value is recalculated using new set power factor and existing setting of current and<br />

voltage. The calculation is only made if active or reactive power is displayed.<br />

In the power generation mode, relative deviation cannot be set.<br />

If the power factor is set to define the phase relation of voltage and current, confirmation of entered<br />

value by pressing LA button means positive phase, LE means negative phase.<br />

Setting the energy<br />

Keep pressing the P-E button to switch to the energy generation mode. Auxiliary data display area<br />

shows the time in seconds and the energy delivered to output terminals after pressing the OUTPUT<br />

button given the existing setting of voltage, current, frequency and power factor. Time setting range is<br />

1.1 s to 1999 s.<br />

Operation manual v42 29


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

The energy value can be set in two ways:<br />

Direct setting of energy<br />

• Select the energy mode and then keep pressing the center cursor button until [ _ _ _ _ _ _ _ ] symbols<br />

appear under the time value (E = xxx.xxxx) supplemented with indication of the set mode.<br />

• The value can be set using numeric keyboard, cursor buttons or potentiometer and confirmed by pressing<br />

Ws/kWs/MWs, VAs/kVAS/MVAs or VArs/kVArs/MVArs display buttons depending on the set mode. The<br />

value can also be confirmed by pressing ENTER.<br />

• Time value is recalculated using new set energy.<br />

Setting the time<br />

• Select the energy mode and then keep pressing the center cursor button until [ _ _ _ _ _ _ _ ] symbols<br />

appear under the time value (t = xxx.x s).<br />

• The value can be set using numeric keyboard, cursor buttons or potentiometer and confirmed by pressing<br />

“s” display button depending on the set mode. The value can also be confirmed by pressing ENTER.<br />

• Energy value is recalculated using new set time.<br />

Grounding of the calibrator and the instrument to be calibrated in the P-E mode<br />

When calibrating the power and energy meters with separate voltage and current circuits, it is advisable<br />

to select GND U ON and GND I ON (both grounding methods on) on the M-140 calibrator. This<br />

setting will ground both the current and voltage output of the calibrator.<br />

If the instrument to be calibrated has electrically connected and not grounded current and voltage<br />

inputs, GND U ON and GND I OFF should be selected on the M-140 calibrator.<br />

If Lo and -I terminals on the calibrator are connected AND the same terminals are connected at the<br />

instrument to be calibrated, resulting voltage drop at the current cables can damage the relay which<br />

interconnects Lo and -I terminals with GND terminal in the calibrator.<br />

“Operating examples“ chapter provides more information concerning correct connection of powermeters and<br />

energy meters to the calibrator.<br />

30 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

Uncertainty calculation of set power<br />

Uncertainty of set power displayed on the Accuracy line of the display is calculated according to the following<br />

formula:<br />

for active power d P = √ ( dU 2 + dI 2 + dPF 2 + 0.03 2 ) [%]<br />

for reactive power d P = √ ( dU 2 + dI 2 + dPF* 2 + 0.03 2 ) [%]<br />

for apparent power d P = √ ( dU 2 + dI 2 + 0.03 2 ) [%]<br />

where dP is the uncertainty of set power [%]<br />

dU is the uncertainty of set voltage [%]<br />

dI is the uncertainty of set current [%]<br />

dPF is the uncertainty of set PF (cosϕ) [%]<br />

dPF* is the uncertainty of set sinϕ [%]<br />

Generation of frequency<br />

The multifunction calibrator can generate several different voltage shapes with exact frequency, amplitude and<br />

duty cycle. The output signal is present at BNC coaxial connector FREQ located at the front panel. The signal is<br />

not present at any other output terminal.<br />

There are two frequency generation modes. The first mode (PWM) allows the generation of squarewave output<br />

signal with calibrated amplitude, frequency and duty cycle. Frequency range is up to 10 kHz. The second mode<br />

(HF) also provides squarewave output signal with very steep rising edge, typically less than 3 ns.<br />

PWM mode<br />

Frequency range:<br />

Voltage range:<br />

Signal shapes:<br />

HF mode<br />

Frequency range:<br />

Voltage range:<br />

Signal shapes:<br />

0.1 Hz to 100 kHz<br />

1 mV to 10 Vpp<br />

squarewave, negative PWM NEG – symmetrical PWM SYM – positive<br />

PWM POS<br />

0.1 Hz to 20 MHz<br />

5 V pk-pk 0, -10, -20, -30 dB<br />

symmetrical squarewave<br />

PWM mode can be used to calibrate the input sensitivity of oscilloscopes at frequencies up to 10 kHz. HF mode<br />

can be used to calibrate the time base of oscilloscopes.<br />

To switch between the modes, keep pressing “F” direct mode button. The display includes the symbols for<br />

currently selected mode (PWM or HF).<br />

Control in the frequency mode<br />

• Press F direct mode button. The calibrator switches to PWM mode. If HF mode is desired, press F button<br />

once more. The main data on the display is the frequency.<br />

• The display shows the following data:<br />

* set frequency<br />

* relative deviation of frequency<br />

* signal amplitude (PWM mode ) or attenuation (HF mode)<br />

* duty cycle (PWM mode only)<br />

* signal shape: PWM NEG / POS / SYM (PWM mode only)<br />

Operation manual v42 31


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

• Set the frequency using numeric keyboard, cursor buttons or potentiometer. Output signal is not yet<br />

connected to the output terminals. The information section of the display shows the symbol<br />

which informs about the disconnection of output terminals...<br />

• Connect the object to be calibrated to FREQ terminal.<br />

• Press OUTPUT button.<br />

• Red LED is lit above the OUTPUT terminals to indicate the connection of signal to the output connector.<br />

• Output signal with set frequency is present at the output connector.<br />

Note<br />

• “FREQ” connector must not be overloaded. In 100 mV to 10V voltage range, maximum load is 5 mA.<br />

In other voltage ranges, maximum load is 0.1mA. If the output is overloaded, the set value is not<br />

guaranteed.<br />

• The output is short-circuit proof.<br />

• The outer casing of the connector is electrically connected to the chassis of the calibrator.<br />

Setting relative deviation<br />

• Keep pressing the center cursor button until [ _ _ _ _ _ _ _ ] symbols appear under the relative deviation<br />

value ∆% = xx.xxxx %.<br />

• The value can be set using numeric keyboard, potentiometer or cursor buttons. Confirm the value by<br />

pressing “%” display button or by pressing ENTER.<br />

Setting the amplitude<br />

Signal amplitude in Volts can only be set in PWM mode.<br />

32 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

• Select the frequency mode and keep pressing the center cursor button until [ _ _ _ _ _ _ _ ] symbols appear<br />

under the amplitude value (U = x.xxx V).<br />

• Set the value using numeric keyboard and confirm by pressing “V” display button or by pressing ENTER.<br />

Setting the attenuation<br />

Signal attenuation in dB can only be set in HF mode. Attenuation can be set in steps of (0, -10, -20, -30) dB.<br />

• Select the frequency mode and keep pressing the center cursor button until [ _ _ _ _ _ _ _ ] symbols appear<br />

under the attenuation value (a = x.xxx dB).<br />

• Set the value using numeric keyboard and confirm by pressing dB display button or by pressing ENTER. If<br />

other than permitted value is set, the closest permitted value is used.<br />

Operation manual v42 33


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

Setting the duty cycle<br />

Duty cycle can only be set in PWM mode.<br />

• Select the frequency mode and keep pressing the center cursor button until [ _ _ _ _ _ _ _ ] symbols appear<br />

under the duty cycle value (PWM = xx %).<br />

• Set the value using numeric keyboard, cursor buttons or potentiometer and confirm by pressing % display<br />

button or by pressing ENTER.<br />

Setting the signal shape<br />

Signal shape can only be set in PWM mode.<br />

• Keep pressing SHAPE display button to select desired signal shape NEG – negative, SYM – symmetrical,<br />

POS – positive.<br />

• Output signal of desired shape is connected to the output connector.<br />

Simulation of temperature sensors<br />

The multifunction calibrator can simulate resistance temperature sensors and thermocouples. When resistance<br />

temperature sensors are simulated, a simulated resistance corresponding to set temperature, sensor type and<br />

temperature scale is connected to Hi - Lo terminals. When thermocouples are simulated, a simulated voltage<br />

corresponding to set temperature, sensor type and temperature of cold end of thermocouple is connected to Hi -<br />

Lo terminals.<br />

Simulated values of temperature sensors are also available at the AUXILIARY connector. Thermocouple voltage<br />

is available at +U, -U terminals. Four-wire connection of resistance temperature sensors is provided by current<br />

terminals PTLI, PTHI and voltage terminals PTLU, PTHU. 140-41 cable adapter is recommended.<br />

Temperature setting range:<br />

Sensor types:<br />

Temperature scale:<br />

-250 to +1820 o C depending on simulated sensor type<br />

resistance temperature sensor Pt 1.385, Pt 1.392, Ni<br />

thermocouple K, N, R, S, B, J, T, E<br />

ITS 90, PTS 68 for resistance temperature sensors and thermocouples<br />

Switching between resistance temperature sensors and thermocouples<br />

• Press T button on the calibrator. The main value on the display is set temperature. The calibrator simulates a<br />

resistance temperature sensor.<br />

• Press T button on the calibrator once again. The calibrator simulates a thermocouple.<br />

Setting the temperature<br />

• Press T button on the calibrator. The main value on the display is set temperature.<br />

• The display shows the following data:<br />

* main data of temperature in o C or K<br />

* sensor type thermocouples: K, N, R, S, B, J, T, E<br />

34 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

resistance temperature sensors: Pt 1.385, Pt 1.392, Ni<br />

*<br />

*<br />

resistance at 0 o C labeled R0 (resistance temperature sensors only)<br />

cold junction temperature of thermocouple sensors labeled RJ<br />

* set value of relate deviation in %, labeled ∆T = xxxx.x °C (K)<br />

the information section shows:<br />

* temperature scale type<br />

* uncertainty of simulated temperature value of selected temperature sensor type<br />

• Set the main value of temperature using numeric keyboard, cursor buttons or potentiometer. Output<br />

terminals are disconnected, the information section of the display shows the symbol which shows that<br />

output terminals are disconnected.<br />

• Connect the object to be calibrated to Hi - Lo terminals.<br />

• Press OUTPUT button.<br />

• Red LED is lit above the OUTPUT terminal to indicate that the output signal is connected to output<br />

terminals. The display shows the symbol of connected output terminals.<br />

Note<br />

• Load of output terminals is limited similarly to corresponding voltage or current ranges.<br />

• Output signals provided at Hi - Lo terminals and AUXILIARY connector are short-circuit proof.<br />

Switching between temperature sensor types<br />

• Keep pressing “TC type” or “RTD type” display button to select desired sensor type.<br />

• If resistance temperature sensors are selected, each press of the button selects Pt1.385, Pt1.392 or Ni<br />

resistance thermometer. The display shows current setting as Pt385 / Pt392 / Ni.<br />

• If thermocouples are selected, each press of the button selects K, N, R, S, B, J, T, E types. The display<br />

shows current setting as TC TYPE x, where x is the type of the thermocouple<br />

Operation manual v42 35


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

Entry of R0 coefficient for resistance temperature sensors<br />

For resistance temperature sensors, resistance at 0 o C labeled R0 can be set. The range is 20 Ω to 2kΩ for all<br />

types of resistance temperature sensors.<br />

• Select the resistance temperature sensor mode and keep pressing the center cursor button until<br />

[ _ _ _ _ _ _ _ ] symbols appear under the R0 coefficient value (R0 = xxxx Ω).<br />

• Set the value using numeric keyboard and confirm by pressing “ Ω” or “kΩ ” display button or<br />

by pressing ENTER.<br />

Note<br />

After the calibrator is switched on or before the first change of the coefficient, R0 is set to 100 Ω. This setting<br />

corresponds to Pt 100 resistance thermal sensor.<br />

Entry of cold junction temperature<br />

For thermocouples, the temperature of cold junction can be entered. The entry is performed by setting the RJ<br />

field in the auxiliary data section of the display.<br />

• Select the thermocouple mode and keep pressing the center cursor button until [ _ _ _ _ _ _ _ ]<br />

symbols appear under the (RJ = xxxx.x o C) value, if o C unit of measurement is used, or under<br />

(RJ = xxxx.x K) value, if K unit of measurement is used.<br />

• Set the value using numeric keyboard.<br />

• Confirm the value by pressing o C or K display button or by pressing ENTER.<br />

36 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

Automatic compensation of cold junction temperature<br />

Automatic cold junction TC sensors compensation can be performed, when cable adapter Option 140-01 is used<br />

for simulating. Ambient temperature measured by in cable adapter mounted Pt1000 sensor is taken as<br />

temperature of cold junction. This automatic compensation is performed always, when measuring of ambient<br />

temperature is activated on display (push button INPUT ON, green led lights). When temperature measuring is<br />

not activated or cable adapter Option 140-01 is not connected, manual compensation only is available. Set value<br />

of RJ on the display to the appropriate ambient temperature to compensate manually influence of cold junction.<br />

Use of AUTOCAL function<br />

To remove the effects of short-term drift and thermal dependency of the simulation, AUTOCAL function can be<br />

used. It can only be activated in the calibration mode. Procedure is following:<br />

• Use a display button to enter the calibration menu. Enter the calibration code and confirm by pressing<br />

ENTER.<br />

• Use the cursor buttons or potentiometer to select the AUTOCAL function from the calibration menu. After<br />

the function is activated, only one option - OFFSET ACAL - is provided. Press SELECT display button to<br />

confirm the option.<br />

• Proceed according to the instructions provided on the display. Automatic calibration takes ca 8-10 minutes<br />

and prompts the user to short Hi-Lo circuits and then to disconnect them.<br />

• After the calibration, the calibrator remains in the calibration mode. Press EXIT display button to return to<br />

normal display.<br />

Do not connect anything to any terminals during the automatic calibration, with the exception of the prompt to<br />

short the Hi-Lo terminals. The procedure is described in “Calibration mode” chapter.<br />

Operation manual v42 37


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

Multimeter<br />

The calibrator includes a built-in multimeter which can measure many electrical and non-electrical values.<br />

Besides DC voltage and current, it can measure frequency, temperature and when external strain gauge sensors is<br />

connected, even other non-electrical values can be measured. The multimeter can only be connected through<br />

AUXILIARY connector. „Description of controls“ chapter lists the pins of the connector.<br />

The manufacturer recommends 140-41 cable adapter or Option 40 cable terminal to connect the multimeter. 140-<br />

41 cable adapter includes a fuse which protects the multimeter from current overload.<br />

The manufacturer does not recommend connection of multimeter inputs to<br />

calibrator outputs.<br />

Such connection can result in presence of high voltage at multimeter inputs,<br />

which can damage the multimeter.<br />

Basic menu<br />

• Multimeter setting menu can be opened by pressing the METER direct control button. After the button is<br />

pressed, SETUP METER menu opens, which allows to set-up the multimeter. Press EXIT display button to<br />

return to previous menu.<br />

• Press ∧ or ∨ cursor button to browse the menu options. Active option is always inverted.<br />

• The parameters of the active option can be changed if the respective function is allowed to be changed. Use<br />

UP, DOWN, CLR, NEXT and display buttons to change the parameters of the active option.<br />

• Multimeter parameters can also be set using the potentiometer. The potentiometer can be used either to<br />

browse the menu options or to change the value of the active option. Press the knob to switch between these<br />

modes.<br />

• Keep pressing EXIT display button to return to normal display when you are finished setting up the<br />

parameters.<br />

38 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

When the multimeter is being set-up, measurement cannot be initiated. Measurement can only be initiated after<br />

the menu is left using the INPUT button.<br />

Function selection<br />

• Press METER direct control button and select FUNCTION option from the menu using cursor buttons.<br />

• Press UP, DOWN display buttons or use the potentiometer to select one of the following functions:<br />

* basic DC voltage range VDC 10 V<br />

* DC current mA DC 20 mA<br />

* small DC voltage range mV DC up to 2 V<br />

* four-wire resistance measuring range up to 2 kΩ<br />

* frequency up to 15 kHz<br />

* temperature measurement using thermocouples T TC<br />

* temperature measurement using resistance temperature sensors T RTD<br />

* measurement of strain gauges (pressure, force) SGS<br />

• By selection of the measurement function, some related options are automatically changed as well<br />

(measuring range, unit of measurement).<br />

• Press EXIT display button the close the function selection menu and to return to the basic display.<br />

Setting the measurement range<br />

• Press METER direct control button and select RANGE option from the menu using cursor buttons.<br />

• Press UP, DOWN display buttons or use the potentiometer to select the measurement range for selected<br />

function. Ranges of individual functions can be changed:<br />

* small DC voltages mV DC ranges: 20, 200, 2000 mV<br />

* temperature measurement using thermocouples T TC, types: K, N, R, S, B, J, T, E<br />

* temperature measurement using RTD sensors, types: Pt 1.385, Pt 1.392<br />

* setting of sensitivity of strain gauge sensors (pressure, force) SGS<br />

• After setting the range, use ∧ or ∨ cursor buttons to switch to the previous or next menu option. The same<br />

can be achieved using the potentiometer after pressing the knob. Press EXIT display button the close the<br />

range selection menu and to return to the basic display.<br />

Operation manual v42 39


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

Units of measurement<br />

Unit of measurement can be set for each measurement function. The unit of measurement is displayed along with<br />

the measured value on the display. The description of the unit of measurement can consist of up to 4 characters.<br />

• Press METER direct control button and select UNIT option from the menu using cursor buttons.<br />

• Press the knob of the potentiometer to display ∧ and ∨ symbols above the active character.<br />

• Press UP, DOWN display buttons or use the potentiometer to select desired character at the active position.<br />

• Press NEXT display button to move onto next character.<br />

• Press the knob of the potentiometer to complete the setting.<br />

• After the unit of measurement is selected, use ∧ or ∨ cursor buttons to move onto the next menu option. The<br />

same can be achieved using the potentiometer after pressing the knob. Press EXIT display button the close<br />

the range selection menu and to return to the basic display.<br />

If you want to reset the unit of measurement to original setting, press CLR display button. Original settings are<br />

listed below:<br />

* basic DC voltage range VDC V<br />

* DC current mA DC 20 mA mA<br />

* small DC voltages mV DC up to 100 mV mV<br />

* four-wire resistance measurement Ω<br />

* frequency up to 15 kHz Hz<br />

* temperature measurement using<br />

thermocouples °C<br />

* temperature measurement using resistive<br />

temperature sensors °C<br />

* measurement of strain gauge sensors mV/V<br />

Use of calculation formula<br />

Each function of the multimeter can display recalculated measured value. The value shown on the display is<br />

always calculated according to the formula:<br />

40 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

where<br />

Y = A0 + A1*X + A2 * X 2<br />

X is the value measured by the multimeter<br />

Y is the value shown on the display<br />

Default setting of the coefficients is A0 = 0, A1 = 1, A2 = 0. This means that directly measured value is<br />

displayed on the display. If any other recalculation is required, enter the desired parameters. Parameter A0 must<br />

be of the same dimension as is in row UNIT.<br />

• Press METER direct control button and select A0 (A1, A2) option from the menu using cursor buttons.<br />

• Enter new value using numeric keyboard.<br />

• Confirm the value by pressing ENTER.<br />

• After setting the coefficients, use ∧ or ∨ cursor buttons to switch to the previous or next menu option. The<br />

same can be achieved using the potentiometer after pressing the knob. Press EXIT display button the close<br />

the range selection menu and to return to the basic display.<br />

If you want to reset a coefficient to original setting, press CLR display button.<br />

Each measurement function of the multimeter has its own independent set of coefficients.<br />

Setting function parameters<br />

Some measurement functions have additional parameters. When measuring the temperature using RTD sensors,<br />

R0 (resistance at 0 °C) can be set, when measuring the temperature using TC sensors, the temperature of cold<br />

junction can be set and when measuring using SGS sensors, supply voltage of the bridge can be set.<br />

Original settings are listed below:<br />

T RTD<br />

R0 = 100 Ω<br />

T TC RJ =23 °C<br />

SGS<br />

Voltage = 5 V<br />

The parameters can be changed as follows:<br />

• Press METER direct control button and select a function, then use cursor buttons to select the line which<br />

includes the function parameter.<br />

• Enter new value using numeric keyboard.<br />

• Confirm the value by pressing ENTER.<br />

• After setting the coefficients, use ∧ or ∨ cursor buttons to switch to the previous or next menu option. The<br />

same can be achieved using the potentiometer after pressing the knob. Press EXIT display button the close<br />

the range selection menu and to return to the basic display.<br />

Operation manual v42 41


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

Start of measurement<br />

To start a measurement:<br />

• Press INPUT direct control button in the basic state of the calibrator.<br />

• INPUT field on the display shows the measured value. The measurement is indicated by a green LED above<br />

the INPUT button.<br />

• Press INPUT button again to stop the measurement. The LED goes off and the input connectors are<br />

disconnected.<br />

The multimeter does not display uncertainty of measuring. If the input range is exceeded, OVERFLOW message<br />

is displayed.<br />

140-41 Cable adapter must be used to connect the signal.<br />

Zero function<br />

The multimeter has a zeroing function. It can be activated after the measurement was activated by pressing<br />

INPUT direct control button. Zeroing is performed by pressing METER direct control button. After the button is<br />

pressed, most recent value is stored and subtracted from all ongoing measured values. A “Zero” message appears<br />

on the right side of the display, indicating that this function is active. Press METER a second time to deactivate<br />

the Zeroing function.<br />

Multimeter’s menu can be entered only after the measurement is terminated by pressing INPUT button.<br />

42 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

Zeroing function can be used e.g. to compensate the voltage drops at the cables, to suppress any remanent value<br />

of the signal etc.<br />

Overloading can damage the multimeter.<br />

Notes:<br />

* The current inputs of the multimeter are protected from current overloading by a fuse located in 140-41<br />

cable adapter. The fuse can be replaced after turning the cap of the fuse holder. The procedure is<br />

described in the adapter’s manual.<br />

* The multimeter can be used as a DC milivoltmeter with 20 mV, 200 mV, 2 V, 10 V ranges. The<br />

appropriate input terminals are PTHU and PTLU.<br />

* Maximum allowed voltage on the input terminals against ground terminal is 20 V.<br />

* Frequency measurement is possible up to 15 kHz. Input signal must be within 0.2 to 5 V range.<br />

squarewave or pulse shape of the input signal is expected.<br />

Operation manual v42 43


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

Simultaneous functions<br />

The multifunction calibrator allows simultaneous generation of calibrated signal along with measurement of<br />

another signal using the built-in multimeter. To use both parts of the calibrator simultaneously, cable adapters<br />

supplied by the manufacturer are necessary.<br />

When using the calibrator, the following states are possible; each state has its limitation, listed in the table.<br />

Method of use Limitation of output signals Limitation of multimeter’s<br />

function<br />

Method of connection<br />

1 No cable adapter is used. <strong>Calibrator</strong>’s output signals are not<br />

limited in any way and can be fully<br />

used.<br />

<strong>Calibrator</strong>’s output signals are<br />

available only at the terminals<br />

located at the front panel.<br />

The multimeter cannot be<br />

used.<br />

If the multimeter is activated,<br />

FAIL error message appears.<br />

2 The calibrator is used with<br />

140-01 cable adapter.<br />

3 The calibrator is used with<br />

140-41 designed for<br />

simultaneous measurement.<br />

OUTPUT 140-41 function is<br />

set to AUX<br />

4 The calibrator is used with<br />

140-41 designed for<br />

simultaneous measurement.<br />

OUTPUT 140-41 function is<br />

set to PANEL.<br />

5 The calibrator is used with<br />

Option 40 Canon – 2 banana<br />

cable end.<br />

6 The calibrator is used with<br />

Option 60 Canon – 4 banana<br />

cable end..<br />

<strong>Calibrator</strong>’s output signals are not<br />

limited in any way and can be fully<br />

used.<br />

<strong>Calibrator</strong>’s output signals are<br />

available only at the terminals<br />

located at the cable adapter.<br />

<strong>Calibrator</strong>’s output signals are<br />

available only at the terminals<br />

located at the cable adapter; ranges<br />

are limited to:<br />

- DC voltage up to 20 V<br />

- DC current up to 20 mA<br />

- four-wire resistance<br />

measurement.<br />

<strong>Calibrator</strong>’s output signals are only<br />

available at the terminals located at<br />

the front panel in full ranges and<br />

can be fully used.<br />

<strong>Calibrator</strong>’s output signals are not<br />

available at the terminals of the<br />

cable adapter.<br />

<strong>Calibrator</strong>’s output signals are not<br />

limited in any way and can be fully<br />

used.<br />

<strong>Calibrator</strong>’s output signals are only<br />

available on the terminals located<br />

at the front panel.<br />

<strong>Calibrator</strong>’s output signals are not<br />

limited in any way and can be fully<br />

used.<br />

<strong>Calibrator</strong>’s output signals are only<br />

available at the terminals located at<br />

the front panel.<br />

The multimeter can only be<br />

used to measure the external<br />

temperature using a Pt100<br />

sensor built into the adapter.<br />

The multimeter can be used<br />

without limitations and in full<br />

ranges.<br />

SGS sensors can be<br />

connected.<br />

The multimeter can be used<br />

without limitations and in full<br />

ranges.<br />

SGS sensors cannot be<br />

connected<br />

The multimeter can be used in<br />

the following ranges:<br />

- DC voltage up to 12 V<br />

- DC voltage up to 25 mA<br />

- frequency up to 15 kHz.<br />

The multimeter can only be<br />

used in the following ranges:<br />

- temperature using TC<br />

sensors<br />

- temperature using RTD<br />

sensors<br />

- resistance up to 2 kOhm<br />

SGS sensors cannot be<br />

connected.<br />

140-01 cable adapter<br />

connected onto calibrator’s<br />

terminals.<br />

140-41 cable adapter<br />

connected onto calibrator’s<br />

terminals.<br />

OUTPUT 140-41 AUX<br />

140-41 cable adapter<br />

connected onto calibrator’s<br />

terminals.<br />

OUTPUT 140-41 PANEL<br />

Option 40 Cable, connected<br />

onto calibrator’s<br />

AUXILIARY connector.<br />

Option 60 cable end ,<br />

connected onto calibrator’s<br />

AUXILIARY connector<br />

7 The calibrator is used with<br />

Option 70 Canon – 4<br />

terminals.<br />

Following output signals can be set<br />

only:<br />

- resistance in four-wire<br />

connection<br />

– resistance temperature sensor<br />

simulation in four-wire connection<br />

Multimeter cannot be used.<br />

If the multimeter is activated,<br />

FAIL error message appears.<br />

Cable adapter Option 70,<br />

installed on AUXILIARY<br />

connector.<br />

Output signals are available on<br />

adapter Option 70 only.<br />

44 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

If current is drawn from +I and –I output terminals located at 140-41 cable adapter, +I and –I output terminals<br />

located at calibrator’s front panel must not be connected simultaneously.<br />

“Operating examples“ chapter provides examples of correct connection during simultaneous measurements.<br />

Operation manual v42 45


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

<strong>Tester</strong><br />

The calibrator includes an application SW, which facilitates automated testing of regulators and transducers.<br />

<strong>Tester</strong> function combines the use of the calibrator as the source of precision signal with automatic change of<br />

output signal, while the response of the unit being tested is measured by the calibrator’s built-in multimeter. The<br />

calibrator is capable of subsequent processing of measured values, providing PASS/FAIL indication.<br />

<strong>Tester</strong> function comprises of performance of a programmed sequence of up to 10 steps. For each step, the type<br />

and value of output signal and the type and tolerance of input signal can be defined. If the input signal is not<br />

within the tolerance limit, the respective step of the test program is considered failed. The time between<br />

individual steps can be set as well.<br />

Basic menu<br />

• Press SETUP display button to enter the <strong>Tester</strong> mode. After the button is pressed, the bottom line of the<br />

display shows TESTER option. Press the respective display button to open the menu of test programs.<br />

• The menu includes a sequence number of the test program, date of creation and program name.<br />

• Use the cursor buttons or the knob of the potentiometer to select the desired test program. Press EXECUTE<br />

display button to execute the program or SELECT to edit the program.<br />

Execution of test program<br />

After EXECUTE display button is pressed to execute the test program, the calibrator displays a screen with set<br />

parameters of output signal and set configuration of the multimeter. The calibrator displays PROGRAM<br />

NUMBER and STEP information above the main output data when the test program runs. PROGRAM<br />

NUMBER xx, where xx can be 00 to 09, designates the test program which currently runs. STEP (max. 10)<br />

designates current step of the program.<br />

46 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

After the test program concludes, the calibrator displays a table showing the results. The table includes the<br />

sequence number of each step, set output value of the calibrator, the value measured by the multimeter and the<br />

result of the respective test step (PASS/FAIL).<br />

Total number of failed steps is displayed above the table (xx ERRORS) along with the designation of performed<br />

test.<br />

After reviewing the results of the test, press EXIT display button to return to the previous level. The test can be<br />

interrupted by pressing CANCEL button.<br />

Programming the test<br />

Test procedures can be programmed after you select a procedure in the basic <strong>Tester</strong> menu and press SELECT<br />

display button. The calibrator displays a programming table.<br />

Two basic test parameters can be set:<br />

• definition of the type of input and output signals and the number of steps<br />

• numeric values for each step<br />

Operation manual v42 47


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

Setting the type of signals and the number of steps<br />

Type of output signal can be set using OUTPUT display button. Press the button repeatedly to select the signal<br />

which will be generated at the output terminals after the test program is executed. The following options are<br />

available:<br />

VDC – VAC – ADC – AAC – WDC – WAC – R – C – T RTD – T TC – F PWM – F HF<br />

Selected type of output signal is displayed in the header of the programming table, “Output” column. The values<br />

in the table change when the type of the output signal is changed.<br />

Type of measured signal (input) can be set using INPUT display button. Press the button repeatedly to select<br />

the signal which will be measured at the input terminals by the built-in multimeter after the test program is<br />

executed. The following options are available:<br />

V DC – mA DC – mV DC – R 4W – Freq – T TC – T RTD<br />

Explanation of symbols:<br />

V DC DC voltage measurement in 12 V range<br />

mA DC DC current measurement in 25 mA range<br />

mV DC DC voltage measurement in 20 mV to 2 V range(automatic range switching)<br />

R 4W four-wire measurement of resistance<br />

Freq measurement of frequency up to 15 kHz<br />

T TC measurement of temperature using thermocouple<br />

T RTD measurement of temperature using resistance temperature sensor<br />

If T TC or T RTD functions are used, the parameters set for the multimeter mode apply. The settings include the<br />

type of unit of measurement (K – °C), temperature scale (ITS68 – PTS90), sensor type, temperature of cold<br />

junction of thermocouple (RJ) and resistance at 0 °C.<br />

Selected type of input signal is displayed in the header of the programming table, “Input” column. The values in<br />

the table change when the type of the input signal is changed.<br />

Number of test steps can be set by pressing STEPS display button in the range 1 to 10. Set number of steps is<br />

displayed in the header of the programming table, “Steps” column. The number of the rows in the table changes<br />

when the number of steps is changed. You can select the lines in the table by pressing ∨, ∧ buttons or by turning<br />

the knob of the potentiometer.<br />

Setting the numeric values of the test<br />

After the type of output signal is selected, numeric values of individual steps can be set. The following<br />

parameters can be set for each step:<br />

1. Value of OUTPUT signal. The range valid for the respective output signal can be used.<br />

If DC voltage or DC current are programmed, only the voltage or current can be set, not the frequency. The<br />

calibrator uses the frequency which has been previously set for the respective output function.<br />

If PWM or HF frequency output is selected, only the frequency can be set. The calibrator uses the output<br />

voltage and duty cycle which has been previously set for the respective output function.<br />

If power output (W AC, W DC) is selected, only W can be selected. The calibrator uses constant voltage,<br />

power factor (phase) and frequency which have been previously set in power generation mode. Change of<br />

power set in the programming table is achieved by change of output current during the test.<br />

2. Lower tolerance limit of input value (LOW). The limit can be set within the selected range of the<br />

multimeter.<br />

3. Upper tolerance limit of input value (HIGH). The limit can be set within the selected range of the<br />

multimeter.<br />

4. Duration of each program step in seconds (TIME). Setting range is 0.5 to 100 s.<br />

48 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

Programming of values is done as follows:<br />

• Select the value to be changed by cursor buttons or the knob of the potentiometer.<br />

• Enter the numeric value using the units of measurement displayed in the header of the table.<br />

• Confirm the entry by pressing ENTER. New entry is copied to the respective position.<br />

• This way, all numeric entries in the table can be changed.<br />

• Press EXIT display button to return to the previous level when the programming is complete.<br />

Note:<br />

When the test program is executed, all steps are performed.<br />

Within one test program, only one type of input signal and output signal can be used. It is impossible to switch to<br />

other function of the calibrator in the middle of a program.<br />

Setting the relays<br />

SWITCH POLARITY and SWITCH ACTIVITY items of SETUP MENU are used to set-up the relay. The<br />

following table lists the states of the relay depending on the values of SWITCH POLARITY, SWITCH<br />

ACTIVITY and the test result.<br />

SETUP MENU RELAY BEFORE TEST TEST RESULT RELYA AFTER TEST<br />

Switch polarity OFF disconnected FAIL disconnected<br />

Switch activity PASS PASS connected<br />

Switch polarity ON connected FAIL connected<br />

Switch activity PASS PASS disconnected<br />

Switch polarity OFF disconnected FAIL connected<br />

Switch activity FAIL PASS disconnected<br />

Switch polarity ON connected FAIL disconnected<br />

Switch activity FAIL PASS connected<br />

Operation manual v42 49


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

Setup menu<br />

The multifunction calibrator allows many other, less frequently used parameters to be set. Setup menu is used to<br />

set these parameters. Setup menu is opened by pressing SETUP display button. If output terminals are<br />

connected, they will be disconnected and the following display appears:<br />

Use ∧ or ∨ cursor button or the knob of the potentiometer to browse the menu options. Active option is always<br />

inverted and when changed, the descriptions of display buttons change as well. Display buttons show how the<br />

respective parameter can be set. Each parameter can be changed after the knob of the potentiometer is pressed.<br />

Press EXIT display button twice to save the parameters when the setting is completed. New settings are retained<br />

when the calibrator is switched off. Setup menu offers the following options:<br />

1. Coil x50 .... xx ON/OFF<br />

This parameter can be set on when 50-turn current coil is going to be used for clamp ammeter calibration. The<br />

coil multiplies the output current. OFF is set by the manufacturer.<br />

2. GND U .... xx ON/OFF<br />

This parameter connects Lo terminal to GND. In practice this means that Lo terminal is grounded. By pressing<br />

the display buttons, the terminal can be grounded or ungrounded. ON is set by the manufacturer, output terminal<br />

is grounded.<br />

3. GND I .... xx ON/OFF<br />

This parameter connects -I to GND. In practice this means that -I terminal is grounded. By pressing the display<br />

buttons, the terminal can be grounded or ungrounded. OFF is set by the manufacturer, output terminals are not<br />

grounded.<br />

It is recommended to ground only the voltage channel GND U ON, GND I OFF, for all ranges except the<br />

generation of power or energy. If the meter to be calibrated has Lo terminal grounded, it is recommended to<br />

unground both outputs of the calibrator, GND U OFF, GND I OFF to exclude ground loops.<br />

Note<br />

50 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

If neither the calibrator’s output, nor the meter’s inputs are grounded, signal/noise ratio can arise at the<br />

calibrator’s output.<br />

4. Temp.scale .... xx ITS90/PTS68<br />

This parameter allows the temperature scale for platinum resistance temperature sensors to be selected. Pressing<br />

the display buttons allows to switch between ITS90 and PTS68 temperature scales. ITS90 is set by the<br />

manufacturer.<br />

5. Temp.unit .... xx<br />

o C/K<br />

This parameter allows the temperature unit for simulation of temperature sensors to be selected. Pressing the<br />

display buttons allows to switch between °C and K. °C is set by the manufacturer.<br />

6. Phase.unit .... xx<br />

o /cos<br />

This parameter allows the unit of phase shift between the voltage and current output to be selected in the<br />

power/energy generation mode. Pressing the display buttons allows to switch between ° and COS (degrees and<br />

power factor).<br />

7. Output 140-41 .... xx AUX/PANEL<br />

This parameter allows the output terminals to be selected. AUX means that the output signals are present only at<br />

the cable adapter’s terminals, PANEL means that the output signals are present only at the front panel.<br />

8. Meter average .... xx UP/DOWN<br />

Displays the number of measurements taken before the average value of build-in multimeter is displayed<br />

(integration constant). UP, DOWN display buttons can be used to select a value in the range of 1 to 20. The<br />

higher the value, the longer the calibrator needs to measure a value, but displayed value is more stable. If 20 is<br />

set, one measurement takes approximately 2.5 s.<br />

9. Interface .... xx GPIB/RS232<br />

Displays the type of interface used to control the calibrator from a PC. By pressing GPIB or RS232 buttons, the<br />

respective type can be selected. The calibrator can be remotely controlled only using the selected interface.<br />

10. GPIB address .... xx UP/DOWN<br />

Displays the calibrator’s address at the GPIB bus. UP, DOWN display buttons can be used to select any valid<br />

GBIP address in the range of 00 to 30. The address 02 is set by the manufacturer.<br />

11. RS232 baud rate .... xx UP/DOWN<br />

Indicates the communication speed of RS232 bus. UP/DOWN display buttons can be used to select 150, 300,<br />

600, 1200, 2400, 4800, 9600, 19200. Perfect communication with the PC requires equal values set at the PC and<br />

the calibrator.<br />

12. Handshake .... xx OFF/Xon-Xoff<br />

Indicates the communication handshake. Display buttons can be used to select OFF or Xon/Xoff. Perfect<br />

communication with the PC requires equal values set at the PC and the calibrator.<br />

Operation manual v42 51


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

13. Keyb.beep .... xx ON/OFF<br />

This parameter allows the acoustic indication of pressed buttons to be switched off or on. ON and OFF display<br />

buttons can be used to switch the indication off or on. ON is set by the manufacturer.<br />

This parameter does not control the acoustic indication of output voltages over 100 V and identification of<br />

errors.<br />

14. Keyb.volume .... xx UP/DOWN<br />

This parameter allows the volume of acoustic indication to be set. UP and DOWN display buttons allow to set<br />

the value in the range of 00 to 15. The bigger the value, the louder sound. This parameter controls the volume of<br />

keyboard beep (if switched on), indication of output voltages over 100 V and identification of errors when<br />

controlling the calibrator.<br />

15. Brightness .... xx UP/DOWN<br />

This parameter sets the contrast of the display. UP and DOWN display buttons allow to set the value in the range<br />

of 00 to 15.<br />

16. Rotary change .... xx ON/OFF<br />

This parameter controls the range of functions of the potentiometer. If ON is set, the potentiometer can move the<br />

cursor both to the left and right (← and → symbols) and up and down (∧ and ∨ symbols). If OFF is set, the<br />

potentiometer can move the cursor up and down only (∧ and ∨ symbols).<br />

17. Switch polarity .... xx ON/OFF<br />

This parameter controls the function of the built-in relay. If ON is set, the relay is closes before the test. If OFF is<br />

set, the relay is opens before the test.<br />

18. Switch activity .... xx PASS/FAIL<br />

This parameter controls the function of the built-in relay. If PASS is set, the relay is active (changes its status) if<br />

the test results in PASS status. If FAIL is set, the relay is active (changes its status) if the test results in FAIL<br />

status.<br />

19. Cal.code .... 00000<br />

Entry of calibration code. Calibration code is a five-digit number, which must be entered to access the calibration<br />

mode. If the calibration code is set to “00000”, this information is displayed in the Setup menu. Calibration code<br />

can be changed. New calibration code can be directly entered using numeric keyboard and confirmed by pressing<br />

ENTER. If non-zero calibration code is set, correct calibration code must be entered to access the calibration<br />

mode. Non-zero calibration code is not displayed further on the display.<br />

52 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

The purpose of the calibration code is to prevent unauthorized users from changing the calibration of the<br />

instrument.<br />

Note<br />

It is advisable to write down actual calibration code if changed. If you forget the calibration code, you have to<br />

send the calibrator to the manufacturer.<br />

20. Cal.date .... xx.yyyy<br />

Displays the date of last calibration of the calibrator (month/year). The parameter cannot be changed, as it is<br />

automatically recorded when leaving the calibration mode.<br />

21. Serial No .... xxxxxx<br />

Displays the serial number of the calibrator. The parameter cannot be changed.<br />

22. Time .... xx:yy<br />

Displays real time. The parameter can be changed using HOUR UP, HOUR DO, MIN UP, MIN DO. display<br />

buttons.<br />

23. Date .... xx.yy.zzzz<br />

Displays real time. The parameter can be changed using DAY UP, MONTH UP, YEAR UP, YEAR DO display<br />

buttons.<br />

24. Time on display .... xx ON/OFF<br />

If set to ON, time and date are displayed in the upper part of the display. If OFF is set, time and date are not<br />

displayed. ON is set by the manufacturer.<br />

Operation manual v42 53


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

Calibration mode<br />

The multifunction calibrator includes a calibration procedure, which allows calibration of the calibrator. Zero<br />

point and slope of the characteristics of individual generation and measurement ranges are set during the<br />

calibration in predefined order. The calibration can only be controlled using the buttons and menu on the<br />

calibrator.<br />

The calibration mode includes AUTOCAL autocalibration function. This function can automatically correct the<br />

short-term instability of zero point for voltage ranges up to 20 V. This function is not a part of the complete<br />

calibration procedure, but it can be used to reduce the time and temperature dependent drift of zero point when<br />

calibrating using small DC voltages and when simulating the thermocouples.<br />

Calibration principles<br />

The calibrator can be calibrated:<br />

• completely, i.e. all functions are calibrated in all recommended points<br />

• partially, i.e. only selected functions are calibrated in all recommended points<br />

• partially, i.e. only selected functions are calibrated in selected points<br />

Complete calibration consists of all partial calibrations performed in the order defined by the calibration menu. If<br />

an item of the calibration menu, e.g. “VOLTAGE DC” is selected, it is not necessary to calibrate all ranges<br />

defined by the calibration algorithm. If new calibration of all ranges is not possible (e.g. the required standard is<br />

not available), old calibration data can be confirmed, i.e. current step of the calibration can be skipped.<br />

Calibration interruption can be performed in any point of the calibration<br />

procedure. However this particular calibration influences parameters of the<br />

calibrator.<br />

Accuracy of the calibrator is guaranteed when full calibration was done.<br />

DC voltage calibration is performed by setting the zero and slope of the scale in all ranges and in both signal<br />

polarities (+ and -) (except the 1000 V range, where zero correction is not necessary).<br />

AC voltage calibration is performed by setting the zero and slope of the scale in all ranges at 1000 Hz (except<br />

the 1000 V range, where the calibration is performed at 500 Hz).<br />

DC current calibration is performed by setting the zero and slope of the scale in all ranges and in both signal<br />

polarities (+ and -).<br />

AC current calibration is performed by setting the zero and slope of the scale in all ranges at 1000 Hz (except<br />

the 20 A, where the calibration is performed at 120 Hz).<br />

AC and DC power calibration is performed by calibration of DC and AC current. The phase relationship of<br />

voltage and current (power factor) is defined digitally and cannot be calibrated. It can only be checked. The<br />

respective check is not included in the calibration procedure.<br />

Resistance calibration is performed by setting the zero and slope of the scale in all ranges. Slope of each range is<br />

calibrated twice. In practice, 9 internal range resistances (50 Ω to 10 MΩ) must be measured and the measured<br />

value entered into the calibrator; then the zero point of the scale is adjusted for each range. Calibration of<br />

resistance ranges must be performed through AUXILIARY connector. Cable adapter Option 70 must be<br />

connected to the auxiliary output.<br />

Capacitance calibration is performed by setting the zero and slope of the scale in all ranges. Slope of each range<br />

is calibrated twice. In practice, 9 internal capacitance ranges (1 nF to 10 µF) must be measured and the measured<br />

value entered into the calibrator; then zero point of the scale is adjusted for each range.<br />

54 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

Frequency function calibration is performed by calibration of DC voltage range and zero point check. DC<br />

voltage levels are checked, between which the calibrator switches when generating the signal. The accuracy of<br />

PWM of the squarewave output signal depends on the frequency accuracy and cannot be calibrated, it can only<br />

be checked. The respective check is not included in the calibration procedure.<br />

Multimeter is calibrated on voltage ranges 2 V and 10 VDC voltage ranges, on current range 20 mADC and on<br />

resistance range 2000 Ω. For calibration cable adapters Option 40 (voltage, current) and Option 60 (resistance)<br />

must be used.<br />

Access to the calibration procedure<br />

Calibration code is required to access the calibration procedure.<br />

• Press SETUP to open the setup menu.<br />

• Press CALIB display button.<br />

• Concerning cable adapters, access to the calibration mode is permitted without connected any of cable<br />

adapters or with connected cable adapter Option 40, Option 60 and Option 70. If cable adapter Option 140-<br />

01 or Option 140-41is connected to the AUXILIARY connector, access to calibration mode is not permitted.<br />

<strong>Calibrator</strong> will display message:<br />

Err 9<br />

Bad cable adapter !<br />

• If an attempt is made to access the calibration procedure within 60 minutes after the calibrator was switched<br />

on, the calibrator does not open the respective menu and displays the following message instead:<br />

Err 21<br />

Time warm up !<br />

xx minutes remain<br />

• If the calibrator is already on for at least 60 minutes, it requests the entry of the calibration code after CAL.<br />

MODE display button is pressed.<br />

• Enter the correct calibration code using numeric keyboard and press ENTER.<br />

• If incorrect calibration code is entered, an error message appears on the display for approximately 3<br />

seconds:<br />

Err 20<br />

Bad calib. code!<br />

Operation manual v42 55


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

• If correct calibration code is entered, calibration menu appears:<br />

• Use ∧ and ∨ cursor buttons to move the cursor through the list:<br />

1. VOLTAGE DC All DC voltage ranges calibration<br />

2. VOLTAGE AC All AC voltage ranges calibration<br />

3. CURRENT DC All DC current ranges calibration<br />

4. CURRENT AC All AC current ranges calibration<br />

5. POWER DC DC power calibration<br />

6. POWER AC AC power calibration<br />

7. RESISTANCE Resistance calibration<br />

8. CAPACITANCE Capacitance calibration<br />

9. FREQUENCY Amplitude in frequency mode calibration<br />

10. ANALOG INPUT Multimeter calibration<br />

11. AUTOCAL DC voltage offset correction<br />

Selection of calibration type<br />

After the calibration menu is displayed, one of partial calibrations can be selected. Use ∧ and ∨ cursor buttons to<br />

move the cursor through the list. Having selected the required function to be calibrated, press SELECT display<br />

button. The following data are shown (the following example is valid for VOLTAGE DC range):<br />

The table lists recommended calibration points. Having selected the required calibration point using SELECT<br />

display button, the following data are shown.<br />

56 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

Display buttons have the following meaning:<br />

WRITE<br />

SKIP<br />

EXIT<br />

new calibration value is entered into the memory, old value is irreversibly lost<br />

current calibration step is skipped, old value is retained in the memory<br />

current calibration is terminated. After this button is pressed, the calibration memory<br />

hold all data (old or new entered) and the calibrator returns to the calibration<br />

menu. It is not necessary to calibrate all ranges; calibration of only selected ranges<br />

is possible by skipping the ranges which do not need to be calibrated.<br />

Moreover, the display shows the range which is being calibrated (RANGE), and the value to be set at the<br />

external standard multimeter (VALUE).<br />

Setting the new calibration data<br />

Use ∧, ∨, cursor buttons to set such main data on the display, when the output signal measured by external<br />

standard multimeter reaches the required calibration point. When the standard output value is reached, press<br />

“WRITE” to write new calibration value to the calibration memory. If you press “SKIP” button, the calibrator<br />

ignores the new value and old value is retained. After you press “WRITE” (or “SKIP”), the calibrator moves on<br />

to the next calibration point.<br />

The procedure is repeated for all calibration points of the selected function. If you press “EXIT” button before<br />

completing the calibration, the calibrator returns to the calibration menu.<br />

Termination of calibration<br />

The calibration can be terminated in the following cases:<br />

• complete calibration has been performed, new calibration data have been entered, the program has returned<br />

to the calibration menu,<br />

• calibration of selected function has been performed, new calibration data have been entered, the program<br />

has returned to the calibration menu,<br />

• calibration of selected range(s) of selected function has been performed, new calibration data have been<br />

entered, the program has returned to the calibration menu,<br />

• the calibration has been started but no calibration data have been entered, the program has returned to the<br />

calibration menu after “EXIT” display button has been pressed,<br />

Operation manual v42 57


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

Press “EXIT” display button to terminate the calibration. After the button is pressed, the calibration date is saved<br />

internally and the calibrator returns to the state it was in before the calibration has been started.<br />

Calibration points<br />

Each function of the calibrator has assigned fixed calibration points which have to be set during the calibration.<br />

For VOLTAGE DC, VOLTAGE AC, CURRENT DC, CURRENT AC, POWER AC, POWER DC, F functions,<br />

the signal value is set using the keyboard. For R-C functions and partially also during the calibration of the<br />

multimeter, calibration data of the range resistances must be entered. T function does not require any calibration,<br />

as the output voltage or resistance is based on arithmetic interpolation using standard tables of temperature<br />

sensor values.<br />

The calibrator needs no calibration of the following parameters:<br />

• frequency<br />

• phase relationship (power factor) of output voltage and frequency in AC power/energy generation<br />

mode<br />

• DC and AC power (except DC and AC current in this mode)<br />

58 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

VOLTAGE DC function<br />

nominal value [V] set limits [V] range [V] note<br />

0.0 2 u 20 m zero calibration<br />

0.0 2 u -20 m zero calibration<br />

19 m 4 u 20 m slope calibration<br />

-19 m 4 u -20 m slope calibration<br />

0.0 2 u 200 m zero calibration<br />

0.0 2 u -200m zero calibration<br />

190 m 6 u 200 m slope calibration<br />

-190 m 6 u -200m slope calibration<br />

0.0 5 u 2 zero calibration<br />

0.0 5 u -2 zero calibration<br />

1.9 12 u 2 slope calibration<br />

-1.9 12 u -2 slope calibration<br />

0.0 20 u 20 zero calibration<br />

0.0 20 u -20 zero calibration<br />

19 100 u 20 slope calibration<br />

-19 100 u -20 slope calibration<br />

0.0 200 u 200 zero calibration<br />

0.0 200 u -200 zero calibration<br />

190 600 u 200 slope calibration<br />

-190 600 u -200 slope calibration<br />

1000 20 m 1000 slope calibration<br />

-1000 20 m -1000 slope calibration<br />

Table DC VOLTAGE<br />

VOLTAGE AC function<br />

nominal value [V] set limits [V] range [V] recommended frequency[Hz]<br />

1.9 m 5 u 20 m 1000<br />

19 m 10 u 20 m 1000<br />

19 m 15 u 200 m 1000<br />

190 m 40 u 200 m 1000<br />

190 m 30 u 2 1000<br />

1.9 100 u 2 1000<br />

1.9 200 u 20 1000<br />

19 1 m 20 1000<br />

19 5 m 200 1000<br />

190 10 m 200 1000<br />

190 50 m 1000 1000<br />

750 50 m 1000 500<br />

Table AC VOLTAGE<br />

Other then recommended frequencies can be used for calibration. Specification of the calibrator is valid when<br />

recommended frequency is used.<br />

Operation manual v42 59


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

CURRENT DC function<br />

nominal value [A] set limits [A] range [A] note<br />

0.0 3 n 200 u zero calibration<br />

0.0 3 n -200 u zero calibration<br />

190 u 5 n 200 u slope calibration<br />

190 u 5 n -200 u slope calibration<br />

0.0 20 n 2 m zero calibration<br />

0.0 20 n -2 m zero calibration<br />

1.9 m 50 n 2 m slope calibration<br />

1.9 m 50 n -2 m slope calibration<br />

0.0 100 n 20 m zero calibration<br />

0.0 100 n -20 m zero calibration<br />

19 m 200 n 20 m slope calibration<br />

-19m 200 n -20 m slope calibration<br />

0.0 1 u 200 m zero calibration<br />

0.0 1 u -200 m zero calibration<br />

190 m 2 u 200 m slope calibration<br />

-190 m 2 u -200 m slope calibration<br />

0.0 20 u 2 zero calibration<br />

0.0 20 u -2 zero calibration<br />

1.9 50 u 2 slope calibration<br />

-1.9 50 u -2 slope calibration<br />

0.0 300 u 20 zero calibration<br />

0.0 300 u -20 zero calibration<br />

10 600 u 20 slope calibration<br />

-10 600 u -20 slope calibration<br />

Table DC CURRENT<br />

CURRENT AC function<br />

nominal value [A] set limits [A] range [A] recommended frequency[Hz]<br />

19 u 5 n 200 u 1000<br />

190 u 50 n 200 u 1000<br />

190u 40 n 2 m 1000<br />

1.9 m 200 n 2 m 1000<br />

1.9 m 200 n 20 m 1000<br />

19 m 2 u 20 m 1000<br />

19 m 2 u 200 m 1000<br />

190 m 20 u 200 m 1000<br />

190 m 20 u 2 500<br />

1.9 200 u 2 500<br />

1.9 1 m 20 120<br />

10 3 m 20 120<br />

Table AC CURRENT<br />

Other then recommended frequency can be used for calibration. Specification of the calibrator is valid when<br />

recommended frequency is used.<br />

60 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

POWER DC function<br />

DC current calibration<br />

nominal value [A] set limits [A] range [A] note<br />

0.0 400 n 20 m zero calibration<br />

0.0 400 n -20 m zero calibration<br />

19 m 2 u 20 m slope calibration<br />

-19m 2 u -20 m slope calibration<br />

0.0 2 u 200 m zero calibration<br />

0.0 2 u -200 m zero calibration<br />

190 m 20 u 200 m slope calibration<br />

-190 m 20 u -200 m slope calibration<br />

0.0 50 u 2 zero calibration<br />

0.0 50 u -2 zero calibration<br />

1.9 200 u 2 slope calibration<br />

-1.9 200 u -2 slope calibration<br />

0.0 200 u 10 zero calibration<br />

0.0 200 u -10 zero calibration<br />

10 1 m 10 slope calibration<br />

-10 1 m -10 slope calibration<br />

Table DC POWER<br />

POWER AC function<br />

AC current calibration<br />

nominal value [A] set limits [A] range [A] recommended frequency[Hz]<br />

1.9 m 400 n 20 m 120<br />

19 m 2 u 20 m 120<br />

19 m 2 u 200 m 120<br />

190 m 20 u 200 m 120<br />

190 m 20 u 2 120<br />

1.9 200 u 2 120<br />

1.9 200 u 10 120<br />

10 1 m 10 120<br />

Table AC POWER<br />

Other then recommended frequency can be used for calibration. Specification of the calibrator is valid when<br />

recommended frequency is used.<br />

Operation manual v42 61


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

RESISTANCE function<br />

nominal value [Ω] set limits [Ω] range [Ω]<br />

Table R<br />

10 0.01 10 – 50<br />

0 0.005 10 – 50<br />

20 0.005 10 – 50<br />

50 0.005 50 - 100<br />

20 0.005 50 - 100<br />

100 0.01 50 - 100<br />

200 0.01 100 – 400<br />

100 0.01 100 – 400<br />

400 0.03 100 – 400<br />

1 k 0.02 400 – 2 k<br />

400 0.02 400 – 2 k<br />

2 k 0.06 400 – 2 k<br />

5 k 0.1 2 k – 10 k<br />

2 k 0.1 2 k – 10 k<br />

10 k 0.3 2 k – 10 k<br />

20 k 1 10 k – 40 k<br />

10 k 1 10 k – 40 k<br />

40 k 3 10 k – 40 k<br />

100 k 3 40 k – 200 k<br />

40 k 3 40 k – 200 k<br />

200 k 10 40 k – 200 k<br />

500 k 20 200 k – 1 M<br />

200 k 20 200 k – 1M<br />

1 M 100 200 k – 1M<br />

2 M 200 1M – 4 M<br />

1 M 200 1M – 4 M<br />

4 M 1 k 1M – 4 M<br />

10 M 2 k 4 M – 20 M<br />

4 M 2 k 4 M – 20 M<br />

20 M 10 k 4 M – 20 M<br />

20 M 20 k 20M – 50 M<br />

50 M 50 k 20M – 50 M<br />

For resistance mode calibration, cable adapter Option 70 or Option 140-41 must be used. Connect standard<br />

multimeter to the four output terminals on the currently connected cable adapter.<br />

62 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

CAPACITANCE function<br />

nominal value [F] set limits [F] range [F]<br />

1 n 5 p 900 p – 2.5 n<br />

900 p 5 p 900 p – 2.5 n<br />

2.5 n 5 p 900 p – 2.5 n<br />

5 n 5 p 5 n - 10 n<br />

2.5 n 5 p 5 n - 10 n<br />

10 n 10 p 5 n - 10 n<br />

20 n 10 p 10 n - 50 n<br />

10 n 10 p 10 n - 50 n<br />

50 n 50 p 10 n - 50 n<br />

100 n 50 p 50 n - 250 n<br />

50 n 50 p 50 n - 250 n<br />

250 n 250 p 50 n - 250 n<br />

500 n 250 p 250 n – 1 u<br />

250 n 250 p 250 n – 1 u<br />

1 u 1 n 250 n – 1 u<br />

2 u 2 n 1 u – 2.5 u<br />

1 u 2 n 1 u – 2.5 u<br />

2.5 u 5 n 1 u – 2.5 u<br />

5 u 5 n 2.5 u – 10 u<br />

2.5 u 5 n 2.5 u – 10 u<br />

10 u 20 n 2.5 u – 10 u<br />

20 u 20 n 10 u – 50 u<br />

10 u 20 n 10 u – 50 u<br />

50 u 200 n 10 u – 50 u<br />

Table C<br />

FREQUENCY function<br />

amplitude calibration<br />

nominal value [V] set limits [V] range [V] note<br />

Table F<br />

0.0 10 u -- zero calibration<br />

10 1 m 10 slope calibration<br />

ANALOG INPUT multimeter function<br />

voltage, current, resistance calibration<br />

nominal value [-] set limits [-] range [-] note<br />

0 V 50 u 10 V zero calibration<br />

10 V 200 u 10 V slope calibration<br />

0 mA 50 nA 20 mA zero calibration<br />

19 mA 500 nA 20 mA slope calibration<br />

0 Ω 5 mΩ 200 Ω zero calibration<br />

100 Ω 5 mΩ 200 Ω slope calibration<br />

0 Ω 50 mΩ 2000 Ω zero calibration<br />

1000 Ω 50 mΩ 2 kΩ slope calibration<br />

0 mV 2 u 20 mV zero calibration<br />

19 mV 2 u 20 mV slope calibration<br />

0 mV 7 u 200 mV zero calibration<br />

190 mV 7 u 200 mV slope calibration<br />

0 mV 50 u 2 V zero calibration<br />

1.9 V 50 u 2 V slope calibration<br />

Table Multimetr<br />

For calibration of ranges 2, 10 V and 20 mA cable adapter Option 40 must be used. For calibration of resistance<br />

ranges cable adapter Option 60 is necessary.<br />

Operation manual v42 63


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

Full calibration procedure<br />

Following pages describe procedure of the full calibration.<br />

Required instruments:<br />

Following instruments are required for calibration:<br />

• 81/2 digit multimeter type Agilent HP3458A or Fluke 8508A or other type with accuracy 0.001 % on DC<br />

voltage<br />

• Resistance shunt 10 mΩ, 100 mΩ Burster 1280, or other type with accuracy 0.01%<br />

• Resistance standard 100 Ω, 1000 Ω Burster 1228, 1229, or other type with accuracy 0.005%<br />

• RLC meter BM 595, HP 4263A, HP4278A, ESI 2150, or other with accuracy 0.1 %<br />

• Counter HP 53181A, HO 53130, BM 642 or other with accuracy 0,001 %<br />

• Powermeter 0.02- 0.05 %<br />

HP8903A Distortion analyzer and scope with bandwidth min. 20 MHz are recommended for THD measuring of<br />

AC signals.<br />

Calibration procedure<br />

1. Connect the calibrator and the multimeter to the mains and let them switched on for at least three hours<br />

in a laboratory at 23±1 o C.<br />

2. Press SETUP display button to call up the setup menu and then CALIB display button to call up the<br />

calibration menu.<br />

3. Enter the calibration code and press ENTER (default calibration code is “00000”).<br />

4. DC voltage ranges calibration<br />

a) Connect the voltage input terminals of the multimeter to the Hi - Lo output terminals of the<br />

calibrator.<br />

b) Select VOLTAGE DC from the calibration menu and confirm by pressing SELECT button.<br />

Switch M-140 output terminals ON.<br />

c) Follow the instructions provided on the calibrator’s display and the DCU table to adjust the<br />

calibrator’s output in the calibration points.<br />

d) To adjust the calibrator’s output in the calibration points, press SELECT button and use ,<br />

∨, ∧ cursor buttons or potentiometer to adjust output voltage. Confirm correctly set value by<br />

pressing WRITE display button. If you want to skip the calibration point whose calibration you<br />

have already entered, press SKIP display button.<br />

e) Switch output terminals OFF<br />

5. AC voltage ranges calibration<br />

a) Select VOLTAGE AC from the calibration menu and confirm by pressing SELECT button.<br />

Switch M-140 output terminals ON.<br />

b) Follow the instructions provided on the calibrator’s display and the ACU table to adjust the<br />

calibrator’s output in the calibration points.<br />

64 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

c) To adjust the calibrator’s output in the calibration points, press SELECT button and use ,<br />

∨, ∧ cursor buttons or potentiometer to adjust the output voltage. Confirm correctly set value<br />

by pressing WRITE display button. If you want to skip the calibration point whose calibration<br />

you have already entered, press SKIP display button.<br />

d) Switch output terminals OFF. Disconnect multimeter and calibrator.<br />

6. DC current ranges calibration<br />

a) Connect current input terminals of the multimeter to the +I - -I output terminals of the<br />

calibrator. Select CURRENT DC from the calibration menu.<br />

b) Select DC current measurement range on external multimeter. Switch output terminals ON<br />

c) Follow the instructions provided on the calibrator’s display and the DCI table to adjust the<br />

calibrator’s output in the calibration points.<br />

d) To adjust the calibrator’s output in the calibration points, press SELECT button and use ,<br />

∨, ∧ cursor buttons or potentiometer to adjust the output current. Confirm correctly set value<br />

by pressing WRITE display button. If you want to skip the calibration point whose calibration<br />

you have already entered, press SKIP display button.<br />

e) Resistance shunt should be used on 2A, 20 A ranges, if standard multimeter does not cover this<br />

range.<br />

7. AC current ranges calibration<br />

a) Select CURRENT AC from the calibration menu. Set the same function on external<br />

multimeter.<br />

b) Follow the instructions provided on the calibrator’s display and the ACI table to adjust the<br />

calibrator’s output in the calibration points.<br />

c) To adjust the calibrator’s output in the calibration points, press SELECT button and use ,<br />

∨, ∧ cursor buttons or potentiometer to adjust the output current. Confirm correctly set value<br />

by pressing WRITE display button. If you want to skip the calibration point whose calibration<br />

you have already entered, press SKIP display button.<br />

d) Resistance shunt should be used on 2A, 20 A ranges, if standard multimeter does not cover this<br />

range.<br />

8. DC power ranges calibration<br />

a) DC power ranges calibration is performed by calibrating 20 mA, 200 mA, 2 A, 10 A DC<br />

current ranges. There is no need to calibrate voltage ranges.<br />

b) Select POWER DC from the calibration menu.<br />

c) Connect a standard ammeter to calibrator’s output terminals +I - -I.<br />

ATTENTION<br />

<strong>Calibrator</strong>’s Lo and -I terminals are electrically connected.<br />

Operation manual v42 65


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

d) Follow the instructions provided on the calibrator’s display and the POWER DC table to adjust<br />

the calibrator’s output in the calibration points.<br />

e) To adjust the calibrator’s output in the calibration points, press SELECT button and use ,<br />

∨, ∧ cursor buttons or potentiometer to adjust the output voltage. Confirm correctly set value<br />

by pressing WRITE display button.<br />

f) Shunt should be used on 2A, 10 A ranges.<br />

9. AC power ranges calibration<br />

a) DC power ranges calibration is performed by calibrating 20 mA, 200 mA, 2 A, 10 A AC<br />

current ranges. There is no need to calibrate the voltage ranges.<br />

b) Select POWER DC from the calibration menu.<br />

c) Connect a standard ammeter to calibrator’s output terminals +I - -I<br />

d) Follow the instructions provided on the calibrator’s display and the POWER AC table to adjust<br />

the calibrator’s output in the calibration points.<br />

e) To adjust the calibrator’s output in the calibration points, press SELECT button and use ,<br />

∨, ∧ cursor buttons or potentiometer to adjust the output voltage. Confirm correctly set value<br />

by pressing WRITE display button..<br />

f) Shunt should be used in 2A, 20 A ranges.<br />

10. Resistance ranges calibration<br />

Resistance ranges calibration is performed by both calibration methods, i.e. setting the checked value<br />

and measuring this value with the standard instrument, as well as setting the nominal value of the<br />

calibrator’s measurement range on the external multimeter by adjusting the calibrator’s output. If the<br />

entry of measured value to the calibrator is required, VALUE METER is displayed at the respective<br />

calibration point. The value on the calibrator’s display has the same format as the nominal value of the<br />

respective calibration point on the display. If, on the other hand, the user is requested to adjust the<br />

calibrator’s output to set the nominal value of the calibrator’s measurement range on the external<br />

multimeter, VALUE is displayed at the respective calibration point. The value on the calibrator’s<br />

display has general format.<br />

a) The calibrator have to be calibrated with attached adapter Option 70, in four-wire connection.<br />

Connect adapter to the AUXILIARY connector. Select RESISTANCE from the calibration<br />

menu<br />

b) Select four-wire resistance measurement function on the external multimeter. Adjust the zero<br />

value of the external multimeter including the cables, if necessary.<br />

c) Connect all four wires of adapter Option 70 to the input terminals of external multimeter. Use<br />

four wire technique. Connect correct terminals and wires.<br />

d) Measure the resistance in the recommended calibration points using the multimeter and write<br />

the values into the calibrator. Use numeric keyboard, potentiometer or cursor buttons to enter<br />

the value.<br />

e) Perform the adjustment for all calibration points as usual. Press WRITE display button to<br />

confirm correct values.<br />

f) Disconnect adapter Option 70 from AUXILIARY connector.<br />

66 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

11. Capacitance ranges calibration<br />

Capacitance ranges calibration is performed by both calibration methods, i.e. setting the checked value<br />

and measuring this value with the standard instrument, as well as setting the nominal value of the<br />

calibrator’s measurement range on the external multimeter by adjusting the calibrator’s output. If the<br />

entry of measured value to the calibrator is required, VALUE METER is displayed at the respective<br />

calibration point. The value on the calibrator’s display has the same format as the nominal value of the<br />

respective calibration point. If, on the other hand, the user is requested to adjust the calibrator’s output<br />

to set the nominal value of the calibrator’s measurement range on the external multimeter, VALUE is<br />

displayed at the respective calibration point. The value on the calibrator’s display has general format.<br />

a) Select CAPACITANCE from the calibration menu. Select GND U OFF and GND I OFF from<br />

the setup menu.<br />

b) Adjust the values of shorted and open terminals of the external RLC meter, set the frequency<br />

used during the measurement (1000 Hz for 1 nF to 10 nF range, 100 Hz for 10 nF to 10 uF<br />

ranges). Connect Hi, Hu terminals to calibrator’s Lo output and Li, Lu terminals to calibrator’s<br />

Hi output.<br />

c) Measure the capacitance of internal range capacitors using the RLC meter and write the values<br />

into the calibrator. The procedure is the same as when calibrating the resistance ranges.<br />

d) Perform the adjustment for all calibration points as usual. Press WRITE display button to<br />

confirm correct values.<br />

Operation manual v42 67


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

11. Amplitude calibration in frequency mode<br />

a) Select FREQUENCY in the calibration menu.<br />

b) Connect the external multimeter with DC voltage measurement range selected to calibrator’s<br />

Hi - Lo output terminals.<br />

c) Follow the instructions provided on the calibrator’s display and the F table to adjust the<br />

calibrator’s output in the calibration points. To adjust the calibrator’s output in the calibration<br />

points, use , ∨, ∧ cursor buttons to adjust the output value. Confirm correctly set value by<br />

pressing WRITE display button.<br />

d) When calibrating the amplitude of the frequency generation function, DC voltage<br />

corresponding to the amplitude of output squarewave signal is set.<br />

12. Built-in multimeter calibration<br />

Built-in multimeter calibration consists of the calibration of 20 mV, 200 mV, 2V and 10 V voltage<br />

ranges, one 20 mA current range and 200 Ω and 2 kΩ resistance ranges.<br />

Cable adapters Option 40 – two-wire adapter, and Option 60 – four-wire adapter are necessary to<br />

connect the multimeter.<br />

Due to the fact that the precision of the built-in multimeter is comparable to the precision of the<br />

calibrator in resistance, small voltage and current ranges, external calibrator of higher precision class<br />

should be used to calibrate the built-in multimeter. If external calibrator is not available, the calibrator<br />

can be used to calibrate the built-in multimeter, but it has to be taken into account the fact that the<br />

precision of the calibration can be insufficient. An 81/2 digit multimeter is recommended for<br />

calibration.<br />

a) 10 V DC voltage range calibration<br />

• Connect adapter Option 40 to the AUXILIARY connector on the front panel.<br />

• Select ANALOG INPUT from the calibration menu. Press SELECT display button to select<br />

the first calibration point, 0 mV.<br />

• Make short on adapter Option 40.<br />

• After the value stabilizes, press WRITE display button to write new calibration value.<br />

• Connect input voltage terminals of external multimeter to Hi – Lo output terminals of the<br />

calibrator. You can use test wires, which are part of basic delivery. Connect Option 40 H input<br />

wire of adapter Option 40 to the Hi output or the calibrator and L input wire of adapter Option<br />

40 to the Lo output on the calibrator. If necessary, adjust the zero point of the external<br />

multimeter before connecting the cables (internal and external multimeters are in parallel).<br />

• Press SELECT display button to select 10 V calibration point. Red light is lighting, all<br />

terminals on calibrator are automatically ON.<br />

• “ Output = xx.xxxxxx V “ is displayed in the bottom part of the display, showing the set value<br />

of output voltage. This value can be changed using the potentiometer or cursor button.<br />

• Adjust the calibrator’s output voltage so that the external multimeter shows 10.0000 V<br />

• After the value stabilizes, press WRITE display button to write new calibration value. If<br />

calibration in this point is not required, press SKIP. Terminals of multimeter will switch OFF.<br />

• Disconnect both the external and built-in multimeter.<br />

68 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

b) 20 mA DC current range calibration<br />

• Leave input terminals of Option 70 adapter OPEN.<br />

• Make short on calibrator output terminals +I - -I.<br />

• Press SELECT display button to select the first calibration point, 0 mA.<br />

• After the value stabilizes, press WRITE display button to write new calibration zero value.<br />

• Disconnect short on +I - -I terminals. Connect L input wire of adapter Option 40 to the -I<br />

output on the calibrator, H input wire of adapter to the minus current input on standard<br />

multimeter and +I output terminal of the calibrator to the positive current input of standard<br />

multimeter. <strong>Calibrator</strong> current output, input of the internal multimeter and current input of<br />

standard multimeter are connected in series. If necessary, adjust the zero point of the external<br />

multimeter before calibrating current range.<br />

• Press SELECT display button to select calibration point, 19 mA. Red LED is lit above<br />

OUTPUT button.<br />

Note:<br />

It can occur, that calibrator beeps and display error message. In this case push the<br />

button OUTPUT ON to switch calibrator output manually.<br />

• Adjust the calibrator’s output current so that the external multimeter shows 19.0000 mA. After<br />

the value stabilizes, press WRITE display button to write new calibration value. Terminals of<br />

multimeter will switch OFF.<br />

• Disconnect both the external and built-in multimeter. Disconnect cable adapter Option 40 from<br />

the AUXILIARY connector.<br />

c) 100 Ω resistance range calibration<br />

Calibration of both ranges 100 and 1000 Ω consists of following steps:<br />

• 0 Ω calibration<br />

• resistance range of multimeter calibration<br />

For calibration of resistance ranges external standard 100 Ω and 1000 Ω are required.<br />

Connect to the AUXILIARY connector cable adapter Option 60.<br />

0 Ω calibration<br />

• Make short of all four Option 60 input wires.<br />

• Press SELECT display button to select the first calibration point, 0 Ω.<br />

• After the value stabilizes, press WRITE display button to write new calibration value.<br />

100 Ω calibration<br />

• Connect resistance standard 100 Ω to the input terminals. Use four-wire technique.<br />

• Press SELECT display button to select the second calibration point, 100 Ω.<br />

• After the value stabilizes, write calibration value of resistance standard to the lower line and<br />

confirm by pressing WRITE.<br />

d) 1000 Ω resistance range calibration<br />

• Make short of all four Option 60 input wires.<br />

• Press SELECT display button to select the first calibration point, 0 Ω.<br />

• After the value stabilizes, press WRITE display button to write new calibration value.<br />

1000 Ω calibration<br />

• Connect resistance standard 1000 Ω to the input terminals. Use four-wire technique.<br />

• Press SELECT display button to select the second calibration point, 1000 Ω.<br />

Operation manual v42 69


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

• After the value stabilizes, write calibration value of resistance standard to the lower line and<br />

confirm by pressing WRITE.<br />

e) 20 mV voltage range calibration<br />

• Make short of test wires of Option 60 adapter.<br />

• Press SELECT display button to select the first calibration point of 20 mV range, 0 mV.<br />

• After the value stabilizes, press WRITE display button to write new calibration value.<br />

• Connect external standard multimeter set to the mVDC function to the Hi – Lo output<br />

terminals. Use test wires from basic delivery. Connect HU test wire of cable adapter to the Hi<br />

terminal on the calibrator and LU test wire of cable adapter to the Lo terminal on the<br />

calibrator. Both build-in meter and external multimeter are in parallel. If necessary, adjust zero<br />

of standard multimeter before calibration.<br />

• Press SELECT display button to select 19 mV calibration point. Red LED is lit above<br />

OUTPUT button.<br />

• Set such output voltage on the calibrator, that standard multimeter indicates 19.000 mV<br />

• After the value stabilizes, press WRITE display button to write new calibration value.<br />

Terminals of the calibrator will switch OFF automatically.<br />

• Disconnect input wires of Option 60.<br />

f) 200 mV voltage range calibration<br />

• Make short of test wires of Option 60 adapter.<br />

• Press SELECT display button to select the first calibration point of 200 mV range, 0 mV.<br />

• After the value stabilizes, press WRITE display button to write new calibration value.<br />

• Connect external standard multimeter set to the mVDC function to the Hi – Lo output<br />

terminals. Use test wires from basic delivery. Connect HU test wire of cable adapter to the Hi<br />

terminal on the calibrator and LU test wire of cable adapter to the Lo terminal on the<br />

calibrator. Both build-in meter and external multimeter are in parallel. If necessary, adjust zero<br />

of standard multimeter before calibration.<br />

• Press SELECT display button to select 190 mV calibration point. Red LED is lit above<br />

OUTPUT button.<br />

• Set such output voltage on the calibrator, that standard multimeter indicates 190.00 mV<br />

• After the value stabilizes, press WRITE display button to write new calibration value.<br />

Terminals of the calibrator will switch OFF automatically.<br />

• Disconnect input wires of Option 60.<br />

g) 2000 mV voltage range calibration<br />

• Make short of test wires of Option 60 adapter.<br />

• Press SELECT display button to select the first calibration point of 2000 mV range, 0 mV.<br />

• After the value stabilizes, press WRITE display button to write new calibration value.<br />

• Connect external standard multimeter set to the mVDC function to the Hi – Lo output<br />

terminals. Use test wires from basic delivery. Connect HU test wire of cable adapter to the Hi<br />

terminal on the calibrator and LU test wire of cable adapter to the Lo terminal on the<br />

calibrator. Both build-in meter and external multimeter are in parallel. If necessary, adjust zero<br />

of standard multimeter before calibration.<br />

• Press SELECT display button to select 1900 mV calibration point. Red LED is lit above<br />

OUTPUT button.<br />

70 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

• Set such output voltage on the calibrator, that standard multimeter indicates 1900.0 mV<br />

• After the value stabilizes, press WRITE display button to write new calibration value.<br />

Terminals of the calibrator will switch OFF automatically.<br />

• Disconnect input wires of Option 60.<br />

i) Exit the calibration mode.<br />

AUTOCAL function<br />

To remove the effects of short-term drift and thermal dependency of the simulation, AUTOCAL function can be<br />

used. It can only be activated in the calibration mode.<br />

Use of AUTOCAL function influences the zero value of DC voltage generated by the calibrator. Use of the<br />

function is recommended only after the calibrator reaches the operating temperature.<br />

Procedure:<br />

• Use a display button to enter the calibration mode. Enter the calibration code and confirm by pressing<br />

ENTER.<br />

• Use the cursor buttons or potentiometer to select the AUTOCAL function from the calibration menu. After<br />

the function is activated, only one option - OFFSET ACAL - is provided. Press SELECT display button to<br />

confirm the option.<br />

• The calibration includes two steps. During the first step, Hi-Lo terminals must be shorted. Use a short cable<br />

to connect the terminals. During the second step, the calibrator requests the disconnection of Hi-Lo<br />

terminals. Follow the instructions provided on the display:<br />

- when requested to do so, short calibrator’s Hi-Lo terminals. Press NEXT display button.<br />

- the calibrator performs internal measurement for ca 30 seconds. During this period, progress<br />

information is displayed in the bottom part of the display.<br />

- after the measurement completes, the calibrator requests the disconnection of Hi-Lo terminals.<br />

Disconnect the cable and confirm by pressing NEXT display button.<br />

- the calibrator performs internal measurement of voltage ranges for ca 8 minutes. During this period,<br />

progress information is displayed in the bottom part of the display.<br />

- after the measurement completes, the calibrator returns to the calibration mode.<br />

- press EXIT display button to return to normal state.<br />

• Do not connect any terminals during the calibration procedure, with the exception of requested short of Hi-<br />

Lo terminals.<br />

Operation manual v42 71


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

Error messages<br />

If an error occurs during the calibrator’s operation or control, error message is displayed on the display. Errors<br />

can be caused by:<br />

• incorrect control using the front panel, i.e. attempts to force a prohibited mode, e.g. setting an out-of-range<br />

value, overloading of output terminals etc.,<br />

• fault of the calibrator, e.g. internal communication error during the communication between individual<br />

functional blocks,<br />

• incorrect control using the GPIB or RS-232 bus.<br />

Below you can see a sample error message which appears when too large value is attempted to set up. All error<br />

messages are displayed in the center of the display.<br />

The following table lists all error messages, their meaning and simple troubleshooting.<br />

72 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

No<br />

error<br />

label description troubleshooting<br />

01 Overload 2V ! 2V range overloaded Output current is too high. Increase load resistance.<br />

02 Overload 20V ! 20 V range overloaded Output current is too high. Increase load resistance.<br />

03 Overload 200V ! 200, 1000 V ranges<br />

overloaded<br />

Output current is too high. Increase load resistance.<br />

04 Overload I output ! Current output overloaded Voltage on the load is too high. Decrease load resistance.<br />

05 High temperature ! Too high internal<br />

temperature<br />

Output stages are overloaded. Do not use ranges 200V, 1000V or<br />

20 A for at least 10 minutes. Check if the ventilation holes are<br />

free.<br />

06 Overload RC ! RC simulator overloaded Test current is too high. Use lower range of tested Ohmmeter.<br />

07 FBK error ! Internal error Turn off the calibrator and turn on it again.<br />

08 OUTPUT must be<br />

in OFF state !<br />

Cable adapter tried to be<br />

exchanged while output<br />

terminals was ON<br />

10 Interface error ! GPIB communication error Wrong data format on GPIB.<br />

Switch off output terminals with button OUTPUT, change the<br />

adapter, and switch output terminals on.<br />

11 Bad command ! Bad command of GPIB Not known command on GPIB.<br />

12 Bad<br />

communication !<br />

13 Over range ! Overcrossing the range via<br />

GPIB<br />

GPIB communication error Listener not connected to the GPIB. Check correct connection of<br />

GPIB cable.<br />

Value out of range was set via GPIB. Set correct value.<br />

20 Bad calib. code ! Bad calibration code Wrong calibration code was entered, calibration cannot start.<br />

Enter correct calibration code.<br />

21 Time warm up ! Attempt to start calibration<br />

before warm up<br />

24 Cable adapter must<br />

be off !<br />

Adapter isn’t allowed for<br />

autocalibration<br />

25 Use cable adapter ! Attempt to start calibration<br />

without cable adapter.<br />

30 Internal RxD<br />

timeout !<br />

31 Internal<br />

communication !<br />

37 <strong>Calibrator</strong> is not<br />

ready !<br />

40 Value too large ! Maximum value is out of<br />

limit<br />

41 Value too small ! Minimum value is out of<br />

limit<br />

42 Deviation too large<br />

!<br />

Attempt to start calibration before 60 minutes warm up period.<br />

Let the calibrator turned on for at least 60 minutes.<br />

Use another cable adapter or perform autocalibration procedure<br />

without cable adapter.<br />

Calibration of resistance ranges can be performed with adapter<br />

Option 70. Calibration of internal multimeter can be performed<br />

with cable adapters Option 40 and Option 60.<br />

Internal error Internal error of the calibrator. Turn the calibrator off and after 5<br />

s turn on. If the error will appear again, contact manufacturer.<br />

Internal error Internal error of the calibrator. Turn the calibrator off and after 5<br />

s turn on. If the error will appear again, contact manufacturer.<br />

Internal error Internal error of the calibrator. Turn the calibrator off and after 5<br />

s turn on. If the error will appear again, contact manufacturer.<br />

Deviation is too high<br />

44 Unable +/- ! Change of polarity is not<br />

allowed<br />

45 Unable – polarity ! Negative polarity is not<br />

allowed.<br />

46 Unable DC/AC ! DC/AC conversion is not<br />

possible<br />

47 Current timeout<br />

!<br />

48 Not allowed on<br />

AUX output !<br />

Time limit for current over<br />

10 A exceeded<br />

Function cannot be used on<br />

output AUXILIARY<br />

Attempt to set value over possible range. Set correct value.<br />

Attempt to set value under possible range. Set correct value.<br />

Set deviation is out of limit –30% to +30%. Set correct value.<br />

Attempt to change polarity, where it is not allowed.. Concerned<br />

modes F, P-E, R-C, ACV, ACI.<br />

Attempt to set negative polarity, where it is not allowed..<br />

Concerned modes F, P-E, R-C, ACV, ACI.<br />

Attempt to change parameter AC/DC where it is nonsense or<br />

where it is not allowed.<br />

Long-term loading current terminals with output current over A.<br />

Do not use this setting in combination with connected cable<br />

adapter.<br />

Operation manual v42 73


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

Functional description of the calibrator<br />

Basic blocks<br />

Basic functional blocks are:<br />

• front panel keyboard<br />

• LCD display<br />

• output terminals<br />

• output voltage amplifier 200 V<br />

• output voltage amplifier 20 A<br />

• main board<br />

• voltage amplifier 2 V<br />

• voltage amplifier 20 V<br />

• DC reference voltage with DAC<br />

• generator<br />

• feedback circuits<br />

• phase shift circuits<br />

• current ranges generator<br />

• multimeter<br />

• power line transformer<br />

• power supply board<br />

• interface GPIB and RS232<br />

74 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

Operation manual v42<br />

75


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

2, 20 V DC voltage ranges<br />

The functional scheme is shown in the picture below:<br />

There is a DC reference voltage source integrated within the 22bit measurement converter. Its output is fed to the<br />

output stage for 2 and 20 V range. Output voltage present at Hi and Lo terminals is sensed by sensing wires.<br />

Feedback eliminates the influence of amplifier’s output impedance and of the resistance of wires within the<br />

calibrator.<br />

200 V DC voltage range<br />

The block diagram functional scheme is shown in the picture below.<br />

The scheme is similar to 2 and 20 V ranges. A 240 V power amplifier with electronic fuses is connected to the<br />

output of 20 V amplifier.<br />

2 to 200 V AC voltage ranges<br />

The functional scheme is shown in the picture below:<br />

76 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

The calibrator’s built-in generator generates a sine wave with voltage-controlled amplitude. The frequency is<br />

derived from microprocessor control circuit’s crystal oscillator. The signal is fed to 20 V or 200 V amplifier and<br />

then to output terminals. Feedback circuits sense the voltage present at the output terminals, normalize its value<br />

and detects it. This results in a signal corresponding to average value of the output voltage. This signal is further<br />

filtered and compared to the set value of output voltage. The error value controls the amplitude of the generator’s<br />

output.<br />

20mV and 200 mV voltage ranges<br />

20 mV and 200 mV voltage ranges are derived from 2 and 20 V voltage ranges.<br />

The output of the amplifier is fed to an inverting attentuator with 1:100 nominal division ratio. The signal is then<br />

led to the output terminals, sensed via local feedback. This connection allows to load the calibrator output with<br />

output current of several mA without losing accuracy.<br />

1000 V AC, DC voltage range<br />

The highest voltage range of the calibrator uses 200 V amplifier. It is connected to a pair of transformers with ca<br />

1:6 transformation ratio.<br />

In 1000 V AC mode, the output of 200 V amplifier is transformed and led to the output terminals. The output<br />

voltage is sensed, rectified and compared to reference DC voltage provided by the DA converter. error value<br />

controls the amplitude of the generator’s output so that there is correct voltage at the output terminals.<br />

In 1000 V DC mode, 12 kHz signal is transformed, rectified, filtered and led to the output terminals. The output<br />

voltage is sensed equally to 1000 V AC mode.<br />

Operation manual v42 77


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

MEATEST<br />

Current converter<br />

The current converter and current amplifier form a separate design block based on a transconductivity converter<br />

with 10 -5 S nominal conversion ratio.<br />

A six-range switched current amplifier connects to the output of the current converter.<br />

input<br />

power stage<br />

Its input is fitted with overload protection and a separate circuit which measures the shift between output voltage<br />

and current in 2 and 20 A current ranges. The output stage is push-pull in B class.<br />

Resistance and capacitance simulator<br />

Resistance and capacitance are simulated using electronic simulator.<br />

DA convertor<br />

power stage<br />

The Hi and Lo terminals are calibrator’s output terminals. The stage including U1 operating amplifier converts<br />

voltage to current. U2 is a separation amplifier. R1 to R5 and C1 to C3 are range impedance. DA converter has 0<br />

to +1 and 0 to -1 conversion ration and allows the simulation of resistance and capacitance values different from<br />

the range impedances. The output stage increases the permitted current load of the output.<br />

ATTENTION<br />

78<br />

The simulator’s output voltage (Hi - Lo terminals) is limited to 8 Vpk.<br />

Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

Frequency synthesizer<br />

Frequency synthesis circuitry allows fine setting of frequency in the whole calibrator’s range. Frequency<br />

synthesis uses DSP circuits with basic frequency 20 MHz.<br />

Power and energy mode<br />

In the power/energy mode, the calibrator supplies the set voltage to Hi-Lo terminals and the set current through<br />

+I – -I terminals. Both signals have defined phase shift according to the set power factor, if AC mode is set.<br />

Output voltage is obtained from 200 V AC voltage range. Output current is obtained from 2 or 20 A current<br />

range. The output current generator is independent of the output voltage generator and synchronized by the<br />

microprocessor’s crystal oscillator. There is a phase shift between the signals of the output current generator and<br />

output voltage generator according to the set power factor. The phase of output current and voltage is compared<br />

by a phase comparator and compared to a reference value representing the set power factor; the output drives a<br />

voltage-controlled phase shift element, which controls the phase of output current.<br />

Operation manual v42 79


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

<strong>Calibrator</strong>’s maintenance<br />

The multifunction calibrator is electronic instrument with microprocessor control. All blocks which are heavily<br />

loaded during the operation are cooled by a fan.<br />

Rules for correct operation<br />

Especially the following rules should be adhered to guarantee correct operation of the calibrator:<br />

• The calibrator can only be switched on and off by pressing the mains switch located at the rear panel.<br />

• Do not connect the calibrator to other voltage than set by the voltage selector.<br />

• Do not block the vent openings located at the rear panel and bottom panel.<br />

• The calibrator must not be operated in dusty environment. It was designed to be used in a laboratory.<br />

• No liquid or small objects can be permitted to enter the calibrator through the vent openings..<br />

• Do not switch the calibrator outside its operating temperature range.<br />

• Connect the instruments to be calibrated to proper output terminals. There is no way of protecting the<br />

calibrator from the damage caused by some improper connections.<br />

• Do not damage the output terminals by plugging in “bananas” thicker than the terminals were designed<br />

for.<br />

• Whenever possible, use the setup menu to ground Lo output terminal (GND U ON setup function).<br />

• Do not overload the power stages by leaving the calibrator switched on with the load connected for a long<br />

time, especially on 20 A current range and 200V and 1000 V voltage ranges.<br />

• If the instruments to be calibrated are not connected to calibrator’s output terminals using original cables,<br />

ensure that cables suitable for the calibration voltage and current are used. Maximum output voltage can<br />

reach 1000 V AC and the maximum output current can reach 20 A AC.<br />

Regular maintenance<br />

The calibrator does not require any special maintenance of electrical or mechanical parts. If is gets dirty, the case<br />

and the display can be cleaned by a wool rag moistened with alcohol.<br />

The calibrator should be calibrated in the recommended 12-month intervals. A calibration center should perform<br />

the calibration.<br />

What to do in case of failure<br />

If an obvious failure occurs during the operation (e.g. the display is not lit, the fan is not turning), the calibrator<br />

must be switched off immediately. First, check the fuse located in the power cord receptacle. Procedure is<br />

following:<br />

• Remove the end of power cord from the mains connector at the rear panel.<br />

• Insert the blade of a flat screwdriver into the opening cut in the mains voltage selector and pry out the fuse<br />

holder.<br />

• Remove the fuse. Replace it with new fuse of the same rating if the fuse was broken.<br />

80 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

• Replace the fuse holder, reconnect the power cord and switch on the calibrator. If the problem persists,<br />

contact the manufacturer.<br />

If an obvious fault is evidenced, e.g. a measurement range or an operating mode is not functional, the user cannot<br />

correct the fault. Contact the manufacturer.<br />

Hidden faults can cause different symptoms and be caused by different causes. Usually, they cause instability of<br />

some parameter. Hidden defects can be caused by unacceptable distortion, degraded insulation etc. In this case<br />

contact the manufacturer.<br />

Sometimes it seems that the calibrator has hidden defect, when the rules for correct operation are not adhered to.<br />

In this case, the fault is caused by the operator. Most frequent cases of false “hidden defects”:<br />

• mains voltage out of tolerance limits or unstable<br />

• wrong grounding of the measurement circuit (bad connection of the ground terminal of the mains outlet, or<br />

several ground connection when grounding loops are formed)<br />

• proximity to sources of intensive influence, whose products are spread through the mains or propagated by<br />

the electromagnetic field<br />

• strong electrostatic or electromagnetic field which can cause major instability during calibration using higher<br />

impedance.<br />

Operation manual v42 81


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

Verification test<br />

Procedure recommended for verifying parameters of the calibrator is described in this chapter. During tests it is<br />

not necessary access to the interior of the instrument.<br />

Required equipment<br />

Following instruments are required for performance verification test:<br />

• 81/2 digit multimeter type Agilent HP3458A or Fluke 8508A or other type with accuracy 0.001 % on DC<br />

voltage<br />

• resistance shunt 10 mΩ, 100 mΩ Burster 1280, or other type with accuracy 0.01%<br />

• RLC meter BM 595, HP 4263A, HP4278A, ESI 2150, or other with accuracy 0.1 %<br />

• counter HP 53181A, HO 53130, BM 642 or other with accuracy 0,001 %<br />

• powermeter 0.02- 0.05 % like Zimmer LMG95, Yokogawa<br />

• resistance standard 100 Ohm, 1000 Ohm with accuracy 0.005%<br />

HP8903A Distortion analyzer and scope with bandwidth min. 20 MHz are recommended for THD measuring of<br />

AC signals.<br />

Configuration of the calibrator<br />

<strong>Calibrator</strong> should be tested directly from the front panel terminals and without use of terminal adapter 140-01 or<br />

140-41. For build-in multimeter testing cable adapters Option 40 and Option 60 are recommended. To suppress<br />

influence of noise or interference with power line frequency in measuring circuit it is recommended following<br />

setting of the calibrator (in SETUP MENU):<br />

• Coil x50 OFF<br />

• GND U ON (in capacitance test OFF)<br />

• GND I ON (in capacitance test OFF)<br />

Note:<br />

It is recommended to ground only the voltage channel GND U ON, GND I OFF, for all ranges<br />

except the generation of power or energy. If the meter to be calibrated has Lo terminal<br />

grounded, it is recommended to un-ground both outputs of the calibrator, GND U OFF, GND I<br />

OFF to exclude ground loops.<br />

• Meter average 05<br />

If it is grounded neither calibrator nor standard meter, higher level can occur on the output<br />

terminals.<br />

In general, when calibrator is connected to the standard meter, ground loops can be arise<br />

through power line connection. Ground loops can result apparently worse noise, short term<br />

stability or non-harmonic distortion of output signal. If necessary use toroidal chokes to<br />

suppress this products.<br />

In all other items in SETUP MENU the setting of parameters don’t influence accuracy of the calibrator.<br />

Use sin waveform of the output signal on all AC tests.<br />

Performance verification may be performed after warm-up period i.e. 1 hour after switching on. <strong>Calibrator</strong> have<br />

to be in temperature stabilize condition at minimum 8 hours before performance verification test is started.<br />

82 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

Basic steps of the performance verification test<br />

Verification procedure consists of following steps:<br />

• 20 V DC voltage test with linearity check<br />

• DC voltage internal ranges 20 mV, 200 mV, 2 V, 240 V, 1000 V test<br />

• 20 V AC voltage test with linearity check<br />

• AC voltage internal ranges 20 mV, 200 mV, 2 V, 20V, 240 V, 1000 V test<br />

• 200 mA DC current test with linearity check<br />

• DC current internal ranges 200 uA, 2 mA, 20 mA test<br />

• AC current internal ranges 200 uA, 2 mA, 20 mA, 200 mA test<br />

• AC/DC high current ranges 2 A, 20 A test<br />

• AC/DC power 480VA 2400VA (AC power with PF 0, +0.5, -0.5) test<br />

• Resistance test in points 10, 100, 1k, 10k, 100k, 1M, 10M, 50M Ohm on DC<br />

• Capacitance test in points 1n, 10n, 100n, 1u, 10u, 50u F on frequency 1000 Hz<br />

• Frequency nominal value 1 kHz test<br />

• Multimeter internal ranges 20 mVDC, 200 mVDC, 2 VDC, 10VDC, 25mADC, 200 Ohm, 2 kOhm,<br />

frequency 1 kHz test<br />

• Distortion checking of AC voltage, range 20 V.<br />

Procedure<br />

Following part describes procedure of performance verification test. Recommended measuring points are the<br />

same as the points in table of limits (see tables bellow).<br />

1. Connect the calibrator to the mains and let them switched on for at least one hour in a laboratory at 23±1 o C.<br />

2. Perform ACAL procedure (see chapter Calibration mode).<br />

3. Connect voltage input of the standard multimeter to the voltage output terminals of the calibrator. Set<br />

appropriate parameters on the standard multimeter to achieve its best accuracy.<br />

4. Perform 20 VDC linearity, DC voltage, 20 VAC linearity, AC voltage tests according to the tables I, II, III,<br />

IV. Deviations should not exceed specified limits.<br />

5. Connect current input of standard multimeter to the current output terminals of the calibrator. Set<br />

appropriate parameters on the standard multimeter to meet its best accuracy.<br />

6. Perform 200 mADC linearity, DC current, AC current tests according to the tables V, VI, VII. Deviations<br />

should not exceed specified limits.<br />

7. Connect current output terminals of the calibrator to the current terminals of resistance shunt 100 mOhm.<br />

Connect voltage input of standard multimeter to the voltage terminals of the resistance shunt. Set range 100<br />

(200)mV on standard multimeter.<br />

8. Perform AC/DC high current test on range 2 A according to the table VIII. Deviation should not exceed<br />

specified limit.<br />

9. Connect current output terminals of the calibrator to the current terminals of resistance shunt 10 mOhm.<br />

Connect voltage input of standard multimeter to the voltage terminals of the resistance shunt. Set range 100<br />

(200)mV on standard multimeter.<br />

Operation manual v42 83


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

10. Perform AC/DC high current test on range 20 A according to the table VIII. Deviation should not exceed<br />

specified limit.<br />

11. Connect standard powermeter to the appropriate output and current terminals of the calibrator.<br />

12. Perform AC/DC power test according to the table IX. Deviations should not exceed specified limits.<br />

13. Connect voltage terminals of the calibrator to the Hi/Lo and Sense H/L terminals respectively of the<br />

standard multimeter. Set resistance mode on multimeter and calibrator as well. Use four-wire technique for<br />

connection of multimeter for testing on nominal values bellow 10 kOhm. Use autozero function of standard<br />

multimeter to eliminate thermoelectric voltages and residual resistances of multimeter and cables before<br />

measuring.<br />

14. Connect adapter Option 70 to the AUXILIARY connector. Perform resistance test according to the table X.<br />

Deviations should not exceed specified limits.<br />

15. Connect voltage terminals of the calibrator to the RCL meter. Switch GND U and GND I OFF in SETUP<br />

MENU (In common RCL meters measuring circuit must not be grounded). Connect source terminal of RCL<br />

meter to the Lo output terminal of the calibrator.<br />

16. Perform capacitance test according to the table XI. Deviations should not exceed specified limits.<br />

17. Connect output voltage terminals of the calibrator to the counter. Set output voltage 1 VAC, frequency 1<br />

kHz.<br />

18. Perform frequency test according to the table XII. Deviation should not exceed specified limit.<br />

19. Connect Option 40 cable adapter to the auxiliary connector on the front panel. Connect Lo banana of the<br />

adapter to the Lo output terminal on the calibrator. Connect Hi banana of the adapter to the Hi output<br />

terminal on the calibrator. Set appropriate value of frequency and DC voltage range 20V on the calibrator<br />

(frequency and 10 V DC range on build in meter) according to the table XIII.<br />

20. Perform multimeter test in points FREQUENCY 1000Hz and 10 VDCV. Deviations should not exceed<br />

limits in the tables bellow.<br />

Note :<br />

Because accuracy of calibrator output is not accurate enough in all points, use external<br />

standard counter (frequency) or multimeter (DC voltage) connected in parallel to the output<br />

terminals to obtain exact value on the output.<br />

21. Connect Option 40 cable adapter to the +I, -I output terminals of the calibrator. Set appropriate value of<br />

output DC current.<br />

22. Perform multimeter test on range 25 mA DC, i.e. point 19 mA DC (see table of limits) according to the table<br />

XIII. Deviation should not exceed specified limit.<br />

23. Connect Option 60 cable adapter to the auxiliary connector on the front panel. Set in SETUP MENU of<br />

build-in meter function DC voltage, range 2V. Connect together Hu and Lu terminals (banana) of the<br />

adapter to make short and use ZERO function of the meter to reset zero point of the meter.<br />

24. Connect terminal Hu of the adapter with output terminal Hi on calibrator. Connect terminal Lu of the<br />

adapter with output terminal Li on the calibrator.<br />

25. Set appropriate values of DC voltage on the calibrator according to the table XIII and perform test of meter<br />

DC voltage ranges 20mV, 200mV and 2V. Deviation should not exceed specified limit.<br />

Note :<br />

Because accuracy of calibrator output is not accurate enough in all points, use external<br />

standard multimeter connected in parallel to the output terminals to obtain exact value of DC<br />

voltage on the output terminals.<br />

26. Disconnect Option 60 from output terminals Hi, Li. Set resistance function in SETUP MENU of build-in<br />

meter. Make four-wire short on the Option 60 banana terminals. Use ZERO function of the build-in meter to<br />

exclude parameters of the adapter. Connect standard resistor 100 Ohm to the adapter. Use four-terminal<br />

technique.<br />

27. Perform multimeter test on resistance range 200 Ohm according to the table XIII. Deviation should not<br />

exceed specified limit.<br />

28. Use the same procedure to test resistance range 2000 Ohm.<br />

84 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

29. Disconnect cable adapter Option 60 and connect distortion meter to the voltage output terminals of the<br />

calibrator. Set output voltage 10 VAC, frequency 1000 Hz and sin waveform.<br />

30. Check harmonic distortion of output signal. It should not exceed 0.05%.<br />

If calibrator is out of limits is in some points of this test, appropriate function and range should be recalibrated. It<br />

is not necessary to recalibrate all functions, but only this one, which does not meet specification. See chapter<br />

Calibration mode, where recalibration procedure is described.<br />

Operation manual v42 85


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

Tables of limits<br />

20 V DC Basic range with linearity test<br />

Function Range Value (V) Frequency (Hz) Deviation<br />

allowed(%value)<br />

V-DC 20.0 V 2.0 0.008<br />

V-DC 20.0 V 4.0 0.006<br />

V-DC 20.0 V 6.0 0.005<br />

V-DC 20.0 V 8.0 0.004<br />

V-DC 20.0 V 10.0 0.004<br />

V-DC 20.0 V 12.0 0.004<br />

V-DC 20.0 V 14.0 0.004<br />

V-DC 20.0 V 16.0 0.004<br />

V-DC 20.0 V 18.0 0.004<br />

V-DC 20.0 V 19.0 0.004<br />

V-DC 20.0 V -2.0 0.008<br />

V-DC 20.0 V -4.0 0.006<br />

V-DC 20.0 V -6.0 0.005<br />

V-DC 20.0 V -8.0 0.004<br />

V-DC 20.0 V -10.0 0.004<br />

V-DC 20.0 V -12.0 0.004<br />

V-DC 20.0 V -14.0 0.004<br />

V-DC 20.0 V -16.0 0.004<br />

V-DC 20.0 V -18.0 0.004<br />

V-DC 20.0 V -19.0 0.004<br />

Table I<br />

DC voltage test<br />

Function Range Value (V) Frequency (Hz) Deviation<br />

allowed(%value)<br />

V-DC 2.0 V 1.9 0.004<br />

V-DC 2.0 V -1.9 0.004<br />

V-DC 240.0 V 190.0 0.004<br />

V-DC 240.0 V 240.0 0.003<br />

V-DC 240.0 V -190.0 0.004<br />

V-DC 240.0 V -240.0 0.003<br />

V-DC 1000.0 V 1000.0 0.010<br />

V-DC 1000.0 V -1000.0 0.010<br />

Table II<br />

20 V AC Basic range with linearity test<br />

Function Range Value (V) Frequency (Hz) Deviation<br />

allowed(%value)<br />

V-AC 20.0 V 2.0 1000 0.075<br />

V-AC 20.0 V 4.0 1000 0.050<br />

V-AC 20.0 V 6.0 1000 0.042<br />

V-AC 20.0 V 8.0 1000 0.037<br />

V-AC 20.0 V 10.0 1000 0.035<br />

V-AC 20.0 V 12.0 1000 0.033<br />

V-AC 20.0 V 14.0 1000 0.032<br />

V-AC 20.0 V 16.0 1000 0.031<br />

V-AC 20.0 V 18.0 1000 0.031<br />

V-AC 20.0 V 19.0 1000 0.030<br />

Table III<br />

86 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

AC voltage test<br />

Function Range Value (V) Frequency (Hz) Deviation<br />

allowed(%value)<br />

V-AC 20 mV 0.019 1000 0.358<br />

V-AC 200 mV 0.19 1000 0.142<br />

V-AC 2.0 V 1.9 1000 0.030<br />

V-AC 20.0 V 19.0 50 0.030<br />

V-AC 20.0 V 19.0 120 0.030<br />

V-AC 20.0 V 19.0 10000 0.030<br />

V-AC 20.0 V 19.0 20000 0.082<br />

V-AC 20.0 V 19.0 50000 0.082<br />

V-AC 240.0 V 190.0 1000 0.036<br />

V-AC 1000 V 750.0 120 0.057<br />

Table IV<br />

200 mA DC Basic range with linearity test<br />

Function Range Value (A) Frequency (Hz) Deviation<br />

allowed(%value)<br />

A-DC 200.0 mA 0.02 0.040<br />

A-DC 200.0 mA 0.04 0.025<br />

A-DC 200.0 mA 0.06 0.020<br />

A-DC 200.0 mA 0.08 0.018<br />

A-DC 200.0 mA 0.10 0.016<br />

A-DC 200.0 mA 0.12 0.015<br />

A-DC 200.0 mA 0.14 0.014<br />

A-DC 200.0 mA 0.16 0.014<br />

A-DC 200.0 mA 0.18 0.013<br />

A-DC 200.0 mA 0.19 0.013<br />

A-DC 200.0 mA -0.02 0.040<br />

A-DC 200.0 mA -0.04 0.025<br />

A-DC 200.0 mA -0.06 0.020<br />

A-DC 200.0 mA -0.08 0.018<br />

A-DC 200.0 mA -0.10 0.016<br />

A-DC 200.0 mA -0.12 0.015<br />

A-DC 200.0 mA -0.14 0.014<br />

A-DC 200.0 mA -0.16 0.014<br />

A-DC 200.0 mA -0.18 0.013<br />

A-DC 200.0 mA -0.19 0.013<br />

Table V<br />

DC current test<br />

Function Range Value (A) Frequency (Hz) Deviation<br />

allowed(%value)<br />

A-DC 200.0 uA 0.00019 0.061<br />

A-DC 200.0 uA -0.00019 0.061<br />

A-DC 2.0 mA 0.0019 0.025<br />

A-DC 2.0 mA -0.0019 0.025<br />

A-DC 20.0 mA 0.019 0.013<br />

A-DC 20.0 mA -0.019 0.013<br />

Table VI<br />

Operation manual v42 87


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

AC current test<br />

Function Range Value (A) Frequency (Hz) Deviation<br />

allowed(%value)<br />

A-AC 200.0 uA 0.00019 60 0.161<br />

A-AC 2.0 mA 0.0019 60 0.081<br />

A-AC 20.0 mA 0.019 60 0.055<br />

A-AC 20.0 mA 0.019 120 0.055<br />

A-AC 20.0 mA 0.019 1000 0.055<br />

A-AC 20.0 mA 0.019 1000 0.055<br />

A-AC 200.0 mA 0.19 60 0.055<br />

Table VII<br />

AC/DC high current test<br />

Function Range Value (A) Frequency (Hz) Deviation<br />

allowed(%value)<br />

A-DC 2.0 A 1.0 0.025<br />

A-DC 2.0 A -1.0 0.025<br />

A-AC 2.0 A 1.0 60 0.060<br />

A-DC 10.0 A 10.0 0.040<br />

A-DC 10.0 A -10.0 0.040<br />

A-AC 10.0 A 10.0 60 0.160<br />

Table VIII<br />

AC/DC power test<br />

Function Range Value (VA) Frequency (Hz) Deviation<br />

allowed(%value)<br />

P 1 480W 480 60 0.07<br />

P 0,5LA 480W 240 60 0.46<br />

P 0,5LE 480W 240 60 0.46<br />

P 1 2400W 2400 60 0.14<br />

P 0,5LA 2400W 1200 60 0.77<br />

P 0,5LE 2400W 1200 60 0.77<br />

Table IX<br />

Voltage = 240V<br />

Resistance test<br />

Function Range Value (Ohm) Frequency (Hz) Deviation<br />

allowed(%value)<br />

O-4W 100.0 Ohm 10.0 DC 0.130<br />

O-4W 100.0 Ohm 100.0 DC 0.015<br />

O-4W 1000 Ohm 1000.0 DC 0.015<br />

O-4W 10k Ohm 10000.0000 DC 0.015<br />

O-4W 100k Ohm 100000.0000 DC 0.015<br />

O-4W 1M Ohm 1.000000e+6 DC 0.050<br />

O-4W 10M Ohm 1.000000e+7 DC 0.200<br />

O-4W 50M Ohm 5.000000e+7 DC 0.500<br />

Table X<br />

Capacitance test<br />

Function Range Value (F) Frequency (Hz) Deviation<br />

allowed(%value)<br />

CAP 1 nF 1.00000e-09 1000 2.000<br />

CAP 10 nF 1.000000e-8 1000 0.500<br />

CAP 100 nF 1.000000e-7 1000 0.500<br />

CAP 1 uF 1.000000e-6 500 0.500<br />

CAP 10 uF 1.000000e-5 300 1.500<br />

CAP 50 uF 5.000000e-5 300 2.000<br />

Table XI<br />

88 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

Frequency test<br />

Function Range Value (Hz) Frequency (Hz) Deviation<br />

allowed(%value)<br />

FREQ 1 kHz 1000.0 0.005<br />

Table XII<br />

Multimeter test<br />

Function Range Value (Hz) Frequency (Hz) Deviation<br />

allowed(%value)<br />

FREQ 1 kHz 1000.0 Hz 0.005<br />

V-DC 10 V 10.0 V 0.013<br />

V-DC 10 V -10.0 V 0.013<br />

A-DC 25 mA 0.019 A 0.017<br />

A-DC 25 mA -0.019 A 0.017<br />

V-DC 20 mV 0.019 V 0.057<br />

V-DC 20 mV -0.019 V 0.057<br />

V-DC 200 mV 0.19 V 0.024<br />

V-DC 200 mV -0.19 V 0.024<br />

V-DC 2 V 1.9 V 0.020<br />

V-DC 2 V -1.9 V 0.020<br />

O-4W 200 Ohm 100.0 Ohm 0.030<br />

O-4W 2 kOhm 1000.0 Ohm 0.021<br />

Table XIII<br />

Operation manual v42 89


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

MEATEST<br />

System control<br />

The calibrator includes standardized IEEE-488 bus and RS232 serial line. System connectors are located at the<br />

rear panel. For the remote control to work properly, bus parameters must be set in the system menu. For IEEE-<br />

488 bus, address is important (0 to 30 setting range). For RS232 bus, communication speed can be set (150 to<br />

19200 Bd) and software handshake XON/XOFF can be set. The calibrator can be only controlled by one<br />

interface at a time. It is therefore necessary to select one of the interfaces (GPIB/RS232) using the system menu.<br />

IEEE-488 bus properties<br />

The instrument performs the following functions based on GPIB bus commands:<br />

SH1, AH1, T5, L3, RL1, DC1, SR1<br />

The instrument also recognizes the following general commands:<br />

DCL<br />

SDC<br />

EOI<br />

GTL<br />

LLO<br />

SPD<br />

SPE<br />

Device Clear<br />

Selected Device Clear<br />

End or Identify Message Terminator<br />

Go To Local<br />

Local Lock Out<br />

Serial Poll Disable<br />

Serial Poll Enable<br />

RS232 bus properties<br />

To transfer the data using RS232 bus, 8N1 data format is used, i.e. each data word includes 8 bits, no parity and<br />

one stop bit. The communication speed can be set using the system menu. Available values: 150, 300, 600, 1200,<br />

2400, 4800, 9600 and 19200 Bd. Software handshake (communication control) XON/XOFF can be set to control<br />

the transfer of the data through the bus.<br />

RS-232 connector layout<br />

9-pin connector D-SUB FEMALE<br />

Pin Label Direction Note<br />

2<br />

3<br />

5<br />

TXD output transmitter<br />

RXD input receiver<br />

GND - ground<br />

Cable between the calibrator and PC (configuration 1:1)<br />

PC<br />

D-Sub 1 D-Sub 2 M-140<br />

Receiver<br />

Transmitter<br />

Ground<br />

2<br />

3<br />

5<br />

2<br />

3<br />

5<br />

Transmitter<br />

Receiver<br />

Ground<br />

90<br />

Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

Command syntax<br />

The commands described in this chapter can be issued through both buses (IEEE-488 and RS232).<br />

All commands listed in this chapter are explained in two columns:<br />

KEYWORD and PARAMETERS.<br />

KEYWORD column includes the name of the command. Each command includes one or more keywords. If a<br />

keyword is in brackets ( [ ] ), it is not mandatory. Non-mandatory commands are used only to achieve<br />

compatibility with language standard SCPI.<br />

Capitals designate the abbreviated form of the commands; extended form is written in lowercase.<br />

Command parameters are in brackets (); each parameter is separated using a comma. Parameters in brackets (<br />

[ ] ) are not mandatory. Line ( | ) means “or” and is used to separate several alternative parameters.<br />

Semicolon ‘;’ is used to separate more commands written on one line.<br />

E.g. VOLT 2.5 ; OUTP ON<br />

Note:<br />

Each command must end in or . Both codes can be used at the same time. The calibrator<br />

performs all commands written on one line of the program after it receives , or code. Without<br />

this code, the program line is ignored.<br />

Description of abbreviations<br />

= Decimal Numeric Program Data, this format is used to express decimal number with or without the<br />

exponent.<br />

=<br />

Character Program Data. Usually, it represents a group of alternative character parameters. E.g.<br />

{ON | OFF | 0 | 1}.<br />

? = A flag indicating a request for the value of the parameter specified by the command. No other<br />

parameter than the question mark can be used.<br />

(?) = A flag indicating a request for the parameter specified by the command. This command permits a<br />

value to be set as well as requested.<br />

=<br />

=<br />

carriage return. ASCII code 13. This code executes the program line.<br />

line feed. ASCII code 10. This code executes the program line.<br />

OUTPut subsystem<br />

This subsystem allows to control the output terminals of <strong>M140</strong> calibrator, to activate the four-wire output or to<br />

switch the calibrator to x50 current coil (option 130-50, 140-50).<br />

Keyword<br />

Parameters<br />

OUTPut<br />

[:STATe] (?) { ON | OFF | 0 | 1 }<br />

: ISELection (?) { HIGHi | HI50turn }<br />

OUTP [:STAT] (?) { ON | OFF | 0 | 1 }<br />

This command activates or deactivates the output of <strong>M140</strong> calibrator.<br />

• ON or 1 - activates the output<br />

• OFF or 0 - deactivates the output<br />

If query is sent, <strong>M140</strong> returns ON if the output is active or OFF if it is inactive<br />

Operation manual v42 91


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

Example:<br />

OUTP 1 - activates the output<br />

OUTP ? - the calibrator returns ON or OFF<br />

OUTP :ISEL (?) { HIGH | HI50 }<br />

This command activates or deactivates the 1000A current range (using a 50-turn coil).<br />

• HIGH - deactivates the 50-turn coil<br />

• HI50 - activates the 50-turn coil (up to 1000A range)<br />

If query is sent, <strong>M140</strong> returns HIGH if the 50-turn coil is deactivated or HI50 if it is activated.<br />

Example:<br />

OUTP :ISEL HI50 activates the 50-turn coil<br />

OUTP :ISEL ? the calibrator returns HIGH or HI50<br />

SOURce subsystem<br />

This subsystem allows to control the individual functions of <strong>M140</strong> calibrator.<br />

Keyword<br />

Parameters<br />

[SOURce]<br />

: FUNCtion<br />

[: SHAPe] (?) { DC | SINusoid | PULPositive | PULSymmetrical |<br />

PULNegative | RMPA | RMPB | TRIangle | LIMSinusoid |<br />

PWMPositive | PWMSymmetrical | PWMNegative | .<br />

SQUare }<br />

: VOLTage<br />

[: LEVEl]<br />

[: IMMediate]<br />

[: AMPLitude] (?) <br />

: CURRent<br />

[: LEVEl]<br />

[: IMMediate]<br />

[: AMPLitude] (?) <br />

: RESistance<br />

[: LEVEl]<br />

[: IMMediate]<br />

[: AMPLitude] (?) <br />

: CAPacitance<br />

[: LEVEl]<br />

[: IMMediate]<br />

[: AMPLitude] (?) <br />

: POWEr<br />

[: LEVEl]<br />

[: IMMediate]<br />

[: AMPLitude] (?) <br />

: PHASe<br />

: UNITS (?) { DEG | COS }<br />

[: ADJust] (?) <br />

: VOLTage<br />

[: LEVEl]<br />

[: IMMediate]<br />

[: AMPLitude] (?) <br />

: CURRent<br />

[: LEVEl]<br />

[: IMMediate]<br />

[: AMPLitude] (?) <br />

92 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

: EARTh<br />

: VOLTage (?) { ON | OFF | 0 | 1 }<br />

: CURRent (?) { ON | OFF | 0 | 1 }<br />

: AUXiliary (?) { ON | OFF | 0 | 1 }<br />

: ADAPter (?)<br />

: FREQuency<br />

[: CW ] (?) <br />

: DUTY (?) <br />

: VOLT (?) <br />

: ATTE (?) <br />

: TEMPerature<br />

: UNITs (?) { C | CEL | K }<br />

: SCALe (?) { TS68 | TS90 }<br />

: THERmocouple<br />

[: LEVEl]<br />

[: IMMediate]<br />

[: AMPLitude] (?) <br />

: RJUNction (?) <br />

: TYPE (?) { B | E | J | K | N | R | S | T }<br />

: PRT<br />

[: LEVEl]<br />

[: IMMediate]<br />

[: AMPLitude] (?) <br />

: TYPE (?) { PT385 | PT392 | NI }<br />

: NRESistance (?) <br />

[SOUR] :FUNC [:SHAP] (?) { DC | SIN | PULP | PULS | PULN | RMPA | RMPB |<br />

TRI | LIMS | PWMP | PWMS | PWMN | SQU }<br />

This command sets the shape of the output signal. At the same time, the respective function must be set. E.g. for<br />

:VOLT or :CURR function, FUNC DC, FUNC SIN, or other signal shape must be set. Some functions (:RES,<br />

:CAP) do not require any other settings.<br />

• DC sets a DC output signal for voltage, current or power modes.<br />

• SINusoid sets AC output signal for voltage, current or power modes.<br />

• PULPositive sets AC squarewave signal for voltage or current modes. The amplitude and duty cycle can be<br />

set. The rectangles are positive, i.e. switching between 0 and +amplitude.<br />

• PULSymmetrical sets AC squarewave signal for voltage or current modes. The amplitude and duty cycle<br />

can be set. The rectangles are symmetrical, i.e. switching between -amplitude and +amplitude.<br />

• PULNegative sets AC squarewave signal for voltage or current modes. The amplitude and duty cycle can<br />

be set. The rectangles are negative, i.e. switching between –amplitude and 0.<br />

• RMPA sets AC output signal - ramp shape - for voltage or current modes. The amplitude can be set. -The<br />

output switches between -amplitude a +amplitude.<br />

• RMPB sets AC output signal - ramp shape - for voltage or current modes. The amplitude can be set. -The<br />

output switches between -amplitude a +amplitude.<br />

• TRIangel sets AC output signal - triangular shape - for voltage or current modes. The amplitude can be set.<br />

-The output switches between -amplitude a +amplitude.<br />

• LIMSinusoid sets AC output signal - sinus shape with limitation of amplitude- for voltage or current<br />

modes. The amplitude can be set. -The output switches between -amplitude a +amplitude.<br />

• PWMPositive sets POS type output signal - digital pulse width modulation.<br />

• PWMSymmetrical sets SYM type output signal - digital pulse width modulation.<br />

• PWMNegative sets NEG type output signal - digital pulse width modulation.<br />

• SQUare sets HSO type output signal - digital pulse output.<br />

Operation manual v42 93


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

If query is sent, <strong>M140</strong> returns a string containing { DC | SIN | PULP | PULS | PULN | RMPA | RMPB | TRI |<br />

LIMS | PWMP | PWMS | PWMN | SQU } depending on the current setting. If impedance or temperature sensor<br />

simulation is set, NONE is returned.<br />

[SOUR] :VOLT [:LEVE] [:IMM] [:AMPL] (?) <br />

This command activates the generation of DC or AC voltage (depending on the DC or SIN parameter of the<br />

FUNC command).<br />

<br />

Represents the value of DC or AC voltage expressed in Volts. Negative value is accepted for DC voltage.<br />

“Technical Data” chapter lists the acceptable ranges.<br />

If query is sent, <strong>M140</strong> returns the set value of current using standard exponential format. Example: 20.547mV is<br />

returned as -2.054700e-002. Positive numbers are not preceded with “+” sign.<br />

[SOUR] :CURR [:LEVE] [:IMM] [:AMPL] (?) <br />

This command activates the generation of DC or AC current (depending on the DC or SIN parameter of the<br />

FUNC command).<br />

<br />

Represents the value of DC or AC current expressed in Amperes. Negative value is accepted for DC current.<br />

“Technical Data” chapter lists the acceptable ranges.<br />

If query is sent, <strong>M140</strong> returns the set value of current using standard exponential format. Example: 20.547mA is<br />

returned as -2.054700e-002. Positive numbers are not preceded with “+” sign.<br />

[SOUR] :RES [:LEVE] [:IMM] [:AMPL] (?) <br />

This command activates the generation of a resistance.<br />

<br />

Represents the value of the resistance in Ohms. “Technical Data” chapter lists the acceptable ranges.<br />

If query is sent, <strong>M140</strong> returns the set value of resistance using standard exponential format. Example: 20.5Ω is<br />

returned as 2.050000e+001.<br />

[SOUR] :CAP [:LEVE] [:IMM] [:AMPL] (?) <br />

This command activates the generation of a capacitance.<br />

<br />

Represents the value of the capacitance in Ohms. “Technical Data” chapter lists the acceptable ranges.<br />

If query is sent, <strong>M140</strong> returns the set value of capacitance using standard exponential format. Example: 20.5nF is<br />

returned as 2.050000e-008.<br />

[SOUR] :POWE :PHAS :UNIT (?) { DEG | COS }<br />

This command sets the method used to specify the phase shift between the output voltage and current.<br />

• DEG activates a mode where all entries are done using the angle, „°“ in the range of 0.0 – 360.0°<br />

• COS activates a mode where all entries are done using the power factor in the range of 1.000 to –1.000,<br />

LAG or LEAD ( LAG = 0-180°, LEAD = 180-360° )<br />

The unit of measurement remains valid even after the calibrator is switched off and back on.<br />

If query is sent, <strong>M140</strong> returns the set unit of measurement { DEG | COS }.<br />

Example:<br />

PHAS :UNIT DEG - sets the angle as the unit of measurement<br />

PHAS :UNIT ? - the calibrator returns DEG<br />

94 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

[SOUR] :POWE :PHAS (?) [ , { LEAD | LAG } ]<br />

This command sets the phase shift between the output voltage and current. At the same time, the power<br />

generation mode is set.<br />

<br />

Represents the phase shift between the output voltage and current in degrees (DEG setting), or specified as a<br />

value of the power factor (COS setting). “Technical Data” chapter lists the acceptable ranges.<br />

,{LEAD|LAG}<br />

This parameter is entered only when specifying the value of the power factor. If the value is omitted, LAG is<br />

used.<br />

If query is sent, <strong>M140</strong> returns the set value of the phase shift between the output voltage and current using<br />

standard exponential format. Example: 156.3 ° is returned as 1.563000e+002.<br />

Example:<br />

Example:<br />

POWE :PHAS 250.2 - activates the power generation mode and sets 250.2° phase shift<br />

between the output voltage and current<br />

POWE :PHAS ? - the calibrator returns 2.502000e+002<br />

POWE :PHAS 0.554 ,LAG - activates the power generation mode and sets 0.554 LAG<br />

power factor<br />

POWE :PHAS ? - the calibrator returns 5.540000e-001,LAG<br />

[SOUR] :POWE :VOLT [:LEVE] [:IMM] [:AMPL] (?) <br />

This command sets the amplitude of the voltage for power generation. At the same time, the power generation<br />

mode is set.<br />

<br />

Represents the voltage, expressed in Volts, used to generate the power. “Technical Data” chapter lists the<br />

acceptable ranges.<br />

If query is sent, <strong>M140</strong> returns the set value of the voltage using standard exponential format. Example: 100.3V is<br />

returned as 1.003000e+002.<br />

[SOUR] :POWE :CURR [:LEVE] [:IMM] [:AMPL] (?) <br />

This command sets the amplitude of the current for power generation. At the same time, the power generation<br />

mode is set..<br />

<br />

Represents the current, expressed in Amperes, used to generate the power. “Technical Data” chapter lists the<br />

acceptable ranges.<br />

If query is sent, <strong>M140</strong> returns the set value of the current using standard exponential format. Example: 1.3A is<br />

returned as 1.300000e+000.<br />

[SOUR] :EART :VOLT (?) { ON | OFF | 0 | 1 }<br />

This command connects or disconnects the voltage Lo terminal to/from GND terminal.<br />

• ON or 1 grounds the voltage output<br />

• OFF or 0 ungrounds the voltage output<br />

If query is sent, <strong>M140</strong> returns ON when the output is grounded or OFF when ungrounded.<br />

Example:<br />

EART : VOLT 1 - grounds the voltage output terminals<br />

EART : VOLT ? - the calibrator returns ON or OFF<br />

[SOUR] :EART :CURR (?) { ON | OFF | 0 | 1 }<br />

Operation manual v42 95


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

This command connects or disconnects the current Lo terminal to/from GND terminal.<br />

• ON or 1 grounds the current output<br />

• OFF or 0 ungrounds the current output<br />

If query is sent, <strong>M140</strong> returns ON if the current terminals are gronded or OFF if ungrounded.<br />

Example:<br />

EART : CURR 1 - grounds the current output terminals<br />

EART : CURR ? - the calibrator returns ON or OFF<br />

[SOUR] :AUX (?) { ON | OFF | 0 | 1 }<br />

This command connects or disconnects the output signals to AUXILIARY connector.<br />

• ON or 1 - output signals are connected to AUXILIARY connector, front panel terminals are disconnected.<br />

• OFF or 0 - output signals are connected to front panel terminals, AUXILIARY connector is disconnected.<br />

If query is sent, <strong>M140</strong> returns ON when the signals are connected to AUXILIARY connector or OFF when the<br />

signals are connected front panel terminals.<br />

Example:<br />

AUX 1 - output signals are connected to AUXILIARY connector<br />

AUX ? the calibrator returns ON or OFF<br />

[SOUR] :AUX :ADAP (?)<br />

This command can be used to find out what type of cable adapter is connected to AUXILIARY connector.<br />

If query is sent, <strong>M140</strong> returns the type of cable adapter connected to AUXILIARY connector{ NONE |<br />

CA14001 | CA14041 | CA14040 | CA14060 | CA5 | CA6 | CA7 }.<br />

Example: AUX :ADAP ? - the calibrator returns NONE, when no cable adapter is connected,<br />

or the type of cable adapter connected to AUXILIARY connector<br />

[SOUR] :FREQ [:CW] (?) <br />

This command sets the frequency being generated.<br />

Examples:<br />

AC voltage frequency :<br />

FUNC :SIN ; :VOLT ; :FREQ <br />

AC current frequency :<br />

FUNC :SIN ; :CURR ; :FREQ <br />

Digital frequency :<br />

FUNC :SQU ; :FREQ <br />

<br />

Represents the frequency in Hz. . “Technical Data” chapter lists the acceptable ranges which depend on the<br />

selected function mode.<br />

If query is sent, <strong>M140</strong> returns the set value of the current using standard exponential format. Example: 20.5kHz<br />

is returned as 2.050000e+004.<br />

[SOUR] :FREQ :DUTY (?) <br />

This command sets the duty cycle of PULP, PULS, PULN, PWMP, PWMS, PWMN function modes.<br />

Examples:<br />

To set symmetrical squarewave voltage signal, 10V with defined duty cycle:<br />

VOLT 10.0; FUNC :PULS ; FREQ :DUTY <br />

To set digital frequency POS signal with defined duty cycle :<br />

FUNC :PWMP ; FREQ :DUTY <br />

96 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

<br />

Represents the duty cycle in percent. 0 to 100% can be set.<br />

If query is sent, <strong>M140</strong> returns the set value of the duty cycle using standard exponential format. Example: 25% is<br />

returned as 2.50000e+001.<br />

[SOUR] :FREQ :VOLT (?) <br />

This command sets the amplitude of PWMP, PWMS, PWMN signals.<br />

Examples:<br />

To set the amplitude of PWMP digital frequency signal:<br />

FUNC :PWMP ; FREQ :VOLT <br />

<br />

Represents the voltage in Volts. 0.000 to 10.000 V can be set.<br />

If query is sent, <strong>M140</strong> returns the set value of the amplitude using standard exponential format. Example: 2.05V<br />

is returned as 2.05000e+000.<br />

[SOUR] :FREQ :ATTE (?) <br />

This command sets the attenuation of digital SQU frequency signal.<br />

Examples:<br />

To set the attenuation of high frequency signal:<br />

FUNC :SQU ; FREQ :ATTE <br />

<br />

Represents the attenuation in decibels. 0 to -30 dB can be set in 10 dB increments.<br />

If query is sent, <strong>M140</strong> returns the set value of the attenuation using standard exponential format. Example: -20<br />

dB is returned as -2.00000e+001.<br />

[SOUR] :TEMP :UNIT (?) { C | CEL | K }<br />

This command sets the unit of measurement of temperature.<br />

• C or CEL sets „Centigrade“<br />

• K sets „Kelvin“<br />

The unit of measurement remains valid after the calibrator is switched off.<br />

If query is sent, <strong>M140</strong> returns the set unit of measurement{ C | K }.<br />

[SOUR] :TEMP :SCAL (?) { TS68 | TS90 }<br />

This command sets one of temperature scales. The setting influences the simulation of resistance temperature<br />

sensors and thermocouples.<br />

• TS68 sets IPTS-68 temperature scale<br />

• TS90 sets ITS-90 temperature scale<br />

The temperature scale remains valid after the calibrator is switched off.<br />

If query is sent, <strong>M140</strong> returns the set temperature scale { TS68 | TS90 }.<br />

[SOUR] :TEMP :THER [:LEVE] [:IMM] [:AMPL] (?) <br />

This command activates the simulation of thermocouples (DC voltage generation).<br />

Operation manual v42 97


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

<br />

Represents the temperature expressed in the units set by the ‘UNIT’ command. “Technical Data” chapter lists the<br />

acceptable ranges.<br />

If query is sent, <strong>M140</strong> returns the set temperature using standard exponential format. Example: 20.5°C is<br />

returned as 2.050000e+001.<br />

[SOUR] :TEMP :THER :RJUN (?) <br />

This command sets the temperature of cold end of thermocouple.<br />

<br />

Represents the temperature expressed in the units set by the ‘UNIT’ command. “Technical Data” chapter lists the<br />

acceptable ranges.<br />

Example: to set the temperature of cold end of thermocouple to 25°C:<br />

:TEMP :THER :RJUN 25 <br />

If query is sent, <strong>M140</strong> returns the set temperature using standard exponential format. Example: 20.5°C is<br />

returned as 2.050000e+001.<br />

[SOUR] :TEMP :THER :TYPE (?) { B | E | J | K | N | R | S | T }<br />

This command sets the type of thermocouple to be simulated.<br />

Example: to activate the simulation of S type thermocouple at 350°C:<br />

:TEMP :THER 350; :TEMP :THER :TYPE S <br />

If query is sent, <strong>M140</strong> returns the set type thermocouple { B | E | J | K | N | R | S | T }.<br />

[SOUR] :TEMP :PRT [:LEVE] [:IMM] [:AMPL] (?) <br />

This command activates the simulation of resistance temperature sensors (resistance generation).<br />

<br />

Represents the temperature expressed in the units set by the ‘UNIT’ command. “Technical Data” chapter lists the<br />

acceptable ranges.<br />

If query is sent, <strong>M140</strong> returns the set temperature using standard exponential format. Example: 20.5°C is<br />

returned as 2.050000e+001.<br />

[SOUR] :TEMP :PRT :TYPE (?) { PT385 | PT392 | NI }<br />

This command sets the type of resistance temperature sensor to be simulated.<br />

Example: to activate the simulation of platinum resistance temperature sensor at 350°C approximated using<br />

PT385 table (Europe):<br />

:TEMP :PRT 350; :TEMP :PRT :TYPE PT385 <br />

If query is sent, <strong>M140</strong> returns the set type of approximation table { PT385 | PT392 | NI }.<br />

[SOUR] :TEMP :PRT :NRESistance (?) <br />

This command sets the nominal resistance of the resistance temperature sensor at 0°C. 10Ω to 2kΩ can be set.<br />

<br />

Represents the nominal resistance in Ohms.<br />

If query is sent, <strong>M140</strong> returns the set value of nominal resistance using standard exponential format. Example:<br />

20.5Ω is returned as 2.050000e+001.<br />

98 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

MEASure subsystem<br />

This subsystem allows to control the internal multimeter of <strong>M140</strong> calibrator. It sets the multimeter’s functions<br />

and reads measured values.<br />

Keyword<br />

Parameters<br />

MEASure<br />

?<br />

: CONFigure<br />

: VOLTage<br />

: CURRent<br />

: MVOLTage<br />

: RESistance<br />

: FREQuency<br />

: TEMPerature<br />

: RTD<br />

: TYPE (?) { PT385 | PT392 }<br />

: NRESistance (?) <br />

: THERmocouple<br />

: TYPE (?) { B | E | J | K | N | R | S | T }<br />

: RJUNction (?) <br />

: SGS<br />

: VOLTage (?) <br />

: OFF<br />

MEAS ?<br />

This command returns the measured value.<br />

Example:20.5Ω is returned as 2.050000e+001.<br />

MEAS :CONF ?<br />

<strong>M140</strong> returns the set measurement mode { VOLT | CURR | MVOLT | RES | FREQ | TEMPerature:RTD |<br />

TEMPerature:THERmocouple | SGS | OFF }.<br />

MEAS :CONF :VOLT<br />

This command selects the function VOLT (range 0 .. 12V) of internal multimeter and switches it on.<br />

Example: to set the multimeter to DC voltage 10V measurement mode:<br />

MEAS :CONF :VOLT <br />

MEAS :CONF :CURR<br />

This command selects the function CURR (range 0 .. 25mA) of internal multimeter and switches it on.<br />

Operation manual v42 99


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

Example: to set the multimeter to DC current 20mA measurement mode:<br />

MEAS :CONF :CURR <br />

MEAS :CONF :MVOLT<br />

This command selects the function MVOLT (range 0 .. 2000mV) of internal multimeter and switches it on.<br />

Example: to set the multimeter to DC voltage 2000mV measurement mode:<br />

MEAS :CONF :MVOLT <br />

MEAS :CONF :RES<br />

This command selects the function RES (range 0 .. 2000Ω) of internal multimeter and switches it on.<br />

Example: to set the multimeter to resistance measurement mode:<br />

MEAS :CONF :RES <br />

MEAS :CONF :FREQ<br />

This command selects the function FREQ of internal multimeter and switches it on.<br />

Example: to set the multimeter to frequency measurement mode:<br />

MEAS :CONF :FREQ <br />

MEAS :CONF :TEMP :RTD :TYPE (?) { PT385 | PT392 }<br />

This command selects the function RTD of internal multimeter, sets the type of resistance temperature sensor<br />

and switches the multimeter on.<br />

Example: to set the multimeter to resistance temperature sensor type PT385 measurement mode:<br />

MEAS :CONF :TEMP : RTD :TYPE PT385 <br />

If query is sent, <strong>M140</strong> returns the set type of approximation table { PT385 | PT392 }.<br />

MEAS :CONF :TEMP :RTD :NRESistance (?) <br />

This command selects the function RTD of internal multimeter, sets the nominal resistance of the resistance<br />

temperature sensor to 0°C and switches the multimeter on. 10Ω to 2kΩ can be set.<br />

<br />

Represents the nominal resistance in Ohms.<br />

Example: to set the multimeter to resistance temperature sensor Pt100 measurement mode:<br />

MEAS :CONF :TEMP : RTD :NRES 100 <br />

If query is sent, <strong>M140</strong> returns the set value of nominal resistance using standard exponential format. Example:<br />

20.5Ω is returned as 2.050000e+001.<br />

MEAS :CONF :TEMP :THER :TYPE (?) { B | E | J | K | N | R | S | T }<br />

This command selects the function TC of internal multimeter, sets the type of thermocouple and switches the<br />

multimeter on.<br />

Example: to set the multimeter to thermocouple R measurement mode:<br />

MEAS :CONF :TEMP : THER :TYPE R <br />

100 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

If query is sent, <strong>M140</strong> returns the set type of thermocouple { B | E | J | K | N | R | S | T }<br />

MEAS :CONF :TEMP :THER :RJUNction (?) <br />

This command selects the function TC of internal multimeter, sets the temperature of cold junction of selected<br />

thermocouple and switches the multimeter on.<br />

<br />

Represents the temperature expressed in the unit set by the ‘UNIT’ command. In “Technical Data” chapter there<br />

is list of acceptable ranges.<br />

Example: to set the temperature of cold end of thermocouple to 25°C:<br />

MEAS :CONF :TEMP :THER :RJUN 25 <br />

If query is sent, <strong>M140</strong> returns the set temperature using standard exponential format. Example: 20.5°C is<br />

returned as 2.050000e+001.<br />

MEAS :CONF :SGS :VOLT (?) <br />

This command selects the function SGS of internal multimeter, sets the supply voltage of sensor and switches<br />

the multimeter on.<br />

<br />

Represents the supply voltage in V. 0V to 20V can be set.<br />

Example: to set the multimeter to SGS measurement mode and power supply of sensor to 10V:<br />

:MEAS :CONF :SGS :VOLT 10 <br />

If query is sent, <strong>M140</strong> returns the supply voltage of SGS using standard exponential format. Example: 15V is<br />

returned as 1.500000e+001.<br />

MEAS :CONF :OFF<br />

This command switches the multimeter off.<br />

TESTer subsystem<br />

This subsystem allows to control features of <strong>M140</strong> calibrator when used as tester. It starts selected test procedure<br />

and reads its result.<br />

Keyword<br />

Parameters<br />

TESTer<br />

: RUN <br />

: RESUlt ?<br />

TEST :RUN <br />

This command starts selected procedure. Number 1 to 10 can be set.<br />

<br />

Represents number of procedure.<br />

Example: to run test procedure 3 :<br />

TEST :RUN 3 <br />

Operation manual v42 101


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

TEST :RESU ?<br />

<strong>M140</strong> returns the result of selected procedure { PASS | FAIL | RUN }.<br />

String „RUN“ is returned if selected procedure isn’t finished.<br />

I/D (instrument identification)<br />

*IDN?<br />

This command returns the identification of the manufacturer, model, serial number and firmware revision.<br />

The reply is formatted as follows:<br />

MEATEST,M-140,412341,4.6<br />

Operation complete<br />

*OPC <br />

This command sets the OPC bit in the ESR (Event Status Register) when all pending operations are complete.<br />

Operation complete?<br />

*OPC? <br />

This command returns “1” to the output queue after all pending operations are complete.<br />

Operation complete?<br />

*OPC? <br />

This command returns “1” to the output queue after all pending operations are complete.<br />

Wait-to-Continue command<br />

*WAI <br />

Prevents the instrument from executing any further commands or queries until all previous remote commands<br />

have been executed.<br />

Reset<br />

*RST <br />

This command resets the calibrator to its initial status.<br />

Test operation<br />

*TST? <br />

This command executes an internal self-test. Return the self-test result (“0” for pass or “1” for fail).<br />

102 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

Status byte reading (IEEE488 only)<br />

*STB? <br />

This query returns the Status Byte Register including the MSS bit.<br />

Service Request Enable setting (IEEE488 only)<br />

*SRE <br />

This command sets condition of the Service Request Enable register. Since bit 6 is not used, the maximum entry<br />

is 191.<br />

Service Request Enable reading (IEEE488 only)<br />

*SRE? <br />

This query returns the Service Request Enable Register.<br />

Event Status Register reading (IEEE488 only)<br />

*ESR? <br />

This query returns the contents of the Event Status Register and clears the register.<br />

Event Status Enable setting (IEEE488 only)<br />

*ESE <br />

This command programs the Event Status Enable register bits. If one or more of the enabled events of the Event<br />

Status Enable register is set, the ESB of Status Byte Register is set too.<br />

Event Status Enable reading (IEEE488 only)<br />

*ESE? <br />

This query returns the Event Status Enable register.<br />

Clear status (IEEE488 only)<br />

*CLS <br />

This command clears the Event Status Register and the Status Byte Register except the MAV bit and output<br />

queue.<br />

Remote control<br />

*REM <br />

This command activates the remote control. When the calibrator is controlled by GPIB bus, it goes to the remote<br />

control mode automatically. When remote control is active, the calibrator ignores all controls from the front<br />

panel, except LOCAL button.<br />

Operation manual v42 103


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

Local control<br />

*LOC <br />

This command activates the local control (using front panel buttons). When the calibrator is controlled by GPIB<br />

bus, it goes to the local control mode automatically.<br />

Local control lock<br />

*LLO <br />

This command locks out the local control; the calibrator cannot be returned to local control by pressing LOCAL<br />

button. Return to local control can only be performed by a command sent through the bus, or by switching the<br />

calibrator off and on.<br />

Local control unlock<br />

*UNL <br />

This command cancels the „*LLO“ command. Unlocked calibrator can be returned to local control mode by<br />

pressing LOCAL button.<br />

104 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

Standard Status Data Structures<br />

All status registers are defined by the IEEE-488.2 standard. The programmer has access to Status Register,<br />

Enable Register and Output Queue in the <strong>M140</strong> <strong>Calibrator</strong>.<br />

Status data structure of <strong>M140</strong> <strong>Calibrator</strong> contains following registers:<br />

STB – Status Byte Register<br />

SRE – Service Request Enable Register<br />

ESR – Event Status Register<br />

ESE – Event Status Enable Register<br />

Output Queue<br />

STB Status Byte Register<br />

The summary messages from the Event Status Register and Output Queue are used to set or clear the appropriate<br />

bits (B4 and B5) of the Status Byte Register. These bits do not latch, and their states (0 or 1) are solely<br />

dependent on the summary messages (0 or 1). For example, if the Standard Event Register is read, its register<br />

will clear. As a result, its summary message will reset to 0, which in turn will clear the ESB bit in the Status Byte<br />

Register.<br />

Bit configuration of Status Byte Register :<br />

Operation manual v42 105


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

RQS<br />

MSS<br />

Request Service, bit 6. The RQS bit is set to 1 whenever bits ESB or MAV change from 0 to 1 and are<br />

enabled (1) in the SRE. When RQS is 1, the <strong>M140</strong> asserts the SRQ control line on the IEEE-488<br />

interface. You can do a serial poll to read this bit to see if the <strong>M140</strong> is the source of an SRQ.<br />

Master Summary Status, bit 6. The MSS bit is set to 1 whenever bits ESB or MAV are 1 and enabled<br />

(1) in the SRE. This bit can be read using the *STB? command in serial remote control in place of doing<br />

a serial poll.<br />

ESB Event Summary Bit, bit 5. The ESB bit is set to 1 when one or more enabled ESR bits are set to 1.<br />

Conversely, the ESB bit is set to 0 when no enabled ESR bits are set to 1.<br />

MAV<br />

Message Available, bit 4. The MAV bit is set to 1 whenever data is available in the <strong>M140</strong> IEEE488<br />

Output Queue. This message is used to synchronize information exchange with the controller. The<br />

controller can, for example, send a query command to the <strong>M140</strong> and then wait for MAV to become<br />

TRUE. The IEC 625-1 bus is available for other use while an application program is waiting for a<br />

device to respond. If an application program begins a read operation of the Output Queue without first<br />

checking for MAV, all system bus activity is held up until the <strong>M140</strong> responds.<br />

SRE Service Request Enable Register<br />

The Service Request Enable Register is an 8-bit register that enables corresponding summary messages in the<br />

Status Byte Register. Thus the application programmer can select reason for a device (<strong>M140</strong>) to issue a service<br />

request by altering the contents of the SRE. The Service Request Enable Register is read with the *SRE?<br />

common query. The response message to this query represents the sum of the binary-weighted values of the<br />

SRE. The value of unused bit 6 shall always be zero. The Service Request Enable Register is written with the<br />

*SRE common command followed by an integer value (0 – 191). Sending the *SRE common command<br />

followed by a zero clears the SRE. A cleared register does not allow status information to generate a rsv local<br />

message and thus, no service request are issued. The Service Request Enable Register is cleared upon power-on.<br />

ESR Event Status Register<br />

The Event Status Register is a two-byte register in which the higher eight bits are always 0, and the lower eight<br />

bits represent various conditions of the <strong>M140</strong> calibrator. The ESR is cleared when the power is turned on, and<br />

every time it is read.<br />

Bit configuration of Event Status Register :<br />

PON<br />

URQ<br />

CME<br />

EXE<br />

DDE<br />

QYE<br />

OPC<br />

Power On, bit 7. This event bit indicates that an off-to-on transition has occurred in the device’s power<br />

supply.<br />

User Request, bit 6. This event bit indicates disconnecting or connecting of any cable adapter to the<br />

auxiliary connector on the front panel.<br />

Command Error, bit 5. This event bit indicates that an incorrectly formed command or query has been<br />

detected by the <strong>M140</strong>.<br />

Execution Error, bit 4. This event bit indicates that the received command cannot be executed, owing to<br />

the device state or the command parameter being out of bounds.<br />

Device Dependent Error, bit 3. This event bit indicates that an error has occurred which is neither a<br />

Command Error, a Query Error, nor an Execution Error. A Device-specific Error is any executed device<br />

operation that did not properly complete due to some condition, such as overload.<br />

Query Error, bit 2. This event bit indicates that either:<br />

1. an attempt is being made to read data from the Output Queue when no output is either present or<br />

pending<br />

2. or data in the Output Queue has been lost<br />

Operation Complete, bit 0. This event bit is generated in response to the *OPC command. It indicates<br />

that the device has completed all selected pending operations.<br />

106 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

ESE Event Status Enable Register<br />

The Event Status Enable Register allows one or more events in the Event Status Register to be reflected in the<br />

ESB summary-message bit. This register is defined for 8 bits, each corresponding to the bits in the Event Status<br />

Register. The Event Status Enable Register is read with the common query *ESE?. Data is returned as a binaryweighted<br />

value. The Event Status Enable Register is written to by the common command, *ESE. Sending the<br />

*ESE common command followed by a zero clears the ESE. The Event Status Enable Register is cleared upon<br />

power-on.<br />

Output Queue<br />

The Output Queue stores response messages until they are read. The availability of output is summarized by the<br />

MAV (message available) summary message. The MAV summary message is used to synchronize information<br />

exchange with the controller. The Output Queue shall be a first in, first out (FIFO) Queue. The Output Queue is<br />

cleared upon power-on.<br />

Operation manual v42 107


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

Examples of use<br />

Calibration of measurement instruments<br />

The calibrator can be used for direct calibration of various instruments which measure electrical values. Opt.<br />

140-01 cable adapter is recommended. The cable adapter includes a temperature sensor which allows the<br />

measurement of external temperature. Measured value can be displayed by pressing INPUT button.<br />

Multimeters<br />

The calibrator can be used for calibration of digital and analogue multimeters (DCV, ACV, DCI, ACI,<br />

resistance, capacitance, temperature, frequency and duty cycle.<br />

Voltage ranges<br />

Thanks to low output impedance and high output current, the calibrator can be used for calibration of analogue<br />

voltmeters and milivoltmeters having low input impedance. Voltage output is connected to Hi/Lo terminals. The<br />

calibrator does not allow the four-wire connection of the instrument to be calibrated.<br />

It is not recommended to connect non-standard load to the voltage output. The calibrator is designed to be used<br />

for calibration of voltmeters. Output terminals should be loaded with high and real impedance. Although the<br />

output is fitted with fast electronic and microprocessor protection, high capacitance or inductance loads can lead<br />

to oscillations of output amplifiers and result in damage.<br />

The instrument to be calibrated can be connected either directly to the front panel terminals, or through Opt.140-<br />

01 cable adapter. If L terminal of the instrument to be calibrated is not grounded, then calibrator’s Lo terminal<br />

should be grounded (GND U ON, see „Setup menu“ chapter).<br />

Connection of a multimeter to be calibrated (voltage range) to calibrator’s output terminals<br />

Truncated waveform with defined distortion can be used for testing of multimeters<br />

108 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

Connection of a multimeter to be calibrated (voltage range) to the terminals of 140-01 cable adapter<br />

Current ranges<br />

All DC and AC current ranges are connected to calibrator’s +I/-I terminals.<br />

If current is drawn from calibrator’s +I and –I output terminals, +I and –I output terminals located at<br />

140-41 cable adapter must not be connected simultaneously!<br />

When using the current output under heavy load (10 to 20 A), the runtime is limited to 0 to 60 s. The runtime<br />

depends on the set current and it is controlled by the microprocessor. The user cannot extend the runtime; if<br />

longer runtime is required, the output terminals must be unloaded, some time must elapse (for example1 min.)<br />

and the load can then be connected again.<br />

When feeding 2 to 20 A current to the output terminals, the output voltage must not exceed approx. 1.5 Vef. If<br />

the current induces higher voltage on the load, the calibrator disconnects the output terminals and displays an<br />

error message.<br />

When ammeters are being calibrated using currents over 1 A, it is important to connect the terminals properly,<br />

paying attention both to the calibrator’s output terminals and the instrument’s input terminals. Excessive contact<br />

resistance can heat up the terminals and cause calibration errors. Excessive and unstable contact resistance has<br />

non-linear characteristic and can distort the output AC current.<br />

It is not recommended to connect non-standard load to the current output. The calibrator is designed to be used<br />

for calibration of ammeters. Output terminals should be loaded with low and real impedance. Although the<br />

output is fitted with fast electronic and microprocessor protection, high capacitance or inductance can lead to<br />

oscillations of output amplifiers and result in damage.<br />

The instrument to be calibrated can be connected either directly to the front panel terminals, or to Opt.140-01<br />

cable adapter. If L terminal of the instrument to be calibrated is not grounded, then calibrator’s -U (-I) terminal<br />

should be grounded (GND U ON, GND I ON, see „Setup menu“ chapter).<br />

Operation manual v42 109


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

Connection of a multimeter to be calibrated (current range) to calibrator’s output terminals<br />

Optional current coil can extend the calibrator’s current range to 1000 A. The coil can be used for calibration of<br />

both DC and AC ammeters. The clamps of the ammeter must be positioned in angle 90 o to the coil. When using<br />

the current coil, no steel or other magnetic objects must be present in the vicinity (50 cm) of the current coil, as<br />

they would deform the magnetic field and cause big calibration error.<br />

Connection of current coil:<br />

Current coil<br />

130-50, 140-50<br />

Powermeters<br />

The calibrator can be used for calibration of DC and AC digital and analogue powermeters and energy meters.<br />

The following limitations apply: voltage 0.2 V to 240 V<br />

current 2 mA to 10 A<br />

frequency DC, 40 Hz to 400 Hz<br />

time<br />

1.1 s to 1999 s<br />

phase shift -180 to +180°<br />

The display can show active, reactive and apparent power and energy values. The phase shift between the output<br />

voltage and current can be set and displayed as well, either as the power factor (PF, cos ϕ) or as the phase shift in<br />

degrees.<br />

In the P-E mode, the permitted load of the voltage and current outputs is the same as in the voltage and current<br />

modes. If the output is overloaded, the calibrator disconnects the output terminals and displays an error message.<br />

For calibration of tong powermeters, current coil can be used. Power factor meters can be calibrated as well.<br />

110 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

Basic connection of a powermeter to the calibrator<br />

HU LU HI LI<br />

Possible methods of grounding of the measurement circuit<br />

To prevent the creation of ground loops and to ensure the guaranteed precision of the calibrator, the powermeter<br />

to be calibrated as well as the calibrator must be properly grounded.<br />

Depending on the nature of the measurement instrument to be calibrated, the following situations can occur:<br />

Device under test GND U GND I<br />

U input floating<br />

I input floating<br />

U and I inputs galvanically<br />

disconnected, not grounded<br />

U input floating<br />

I input floating<br />

galvanically connected terminals L<br />

both U and I inputs, not grounded<br />

voltage input grounded<br />

current input floating<br />

voltage input floating<br />

current input grounded<br />

both voltage and current inputs<br />

grounded<br />

ON<br />

ON<br />

(OFF)<br />

OFF<br />

ON<br />

OFF<br />

ON<br />

OFF<br />

(ON)<br />

ON<br />

OFF<br />

OFF<br />

ATTENTION<br />

The voltage between calibrator’s –I and Lo outputs must not exceed 10 V.<br />

Operation manual v42 111


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

Counters and oscilloscopes<br />

The calibrator can be used for basic calibration of the frequency ranges of multimeters and simple counters. The<br />

calibrator provides the following functions:<br />

• calibration of frequency functions up to 20 MHz using squarewave signal. The function is activated by<br />

pressing F direct mode button and selecting HF mode. Frequency can be set.<br />

• check of input sensitivity from 1 mV to 10 V in the frequency range up to 100 kHz. The function is<br />

activated by pressing F direct mode button and selecting PWM mode. Frequency, amplitude and duty cycle<br />

can be set.<br />

• calibration of time period using squarewave signal whose period can be set up to 10 s, with selectable duty<br />

cycle. The function is activated by pressing F direct mode button and selecting PWM mode. Frequency,<br />

amplitude and duty cycle can be set.<br />

The instrument to be calibrated connects to FREQ connector using BNC/banana cable.<br />

Connection of the frequency output<br />

Counter to be calibrated<br />

Coaxial cable<br />

The calibrator can be used for basic calibration of oscilloscopes. The calibrator provides the following functions.<br />

• check of time base up to 20 MHz using squarewave signal. The function is activated by pressing F direct<br />

mode button and selecting HF mode. Frequency can be set.<br />

• check of vertical channel input sensitivity from 1 mV to 10 V in the frequency range up to 10 kHz. The<br />

function is activated by pressing F direct mode button and selecting PWM mode. Frequency, amplitude and<br />

duty cycle can be set.<br />

• check of bandwidth using a signal up to 20 MHz with very steep rising edge (less than 5 ns). Calibration of<br />

time period using squarewave signal whose period can be set up to 10 s, with selectable duty cycle. The<br />

function is activated by pressing F direct mode button and selecting PWM mode. Frequency can be set. The<br />

delay of the signal displayed on the oscilloscope’s screen is checked.<br />

The oscilloscope to be calibrated connects to FREQ connector using a coaxial cable.<br />

Thermometers<br />

The calibrator can be used for calibration of transducers used in thermometers and heat meters, which use a<br />

thermocouple or Pt or Ni sensor. The block connected to the sensor is checked, as it is disconnected from the<br />

transducer and the calibrator’s output is connected to the transducer’s input instead. Simulation of thermal sensor<br />

is activated by pressing T direct mode button. Pt and Ni resistance temperature sensors as well as K, N, R, S, B,<br />

J, T, E thermocouples can be simulated.<br />

The calibrator can be connected to transducer’s input in one of the following ways:<br />

• without the cable adapter, directly to Hi/Lo terminals. Temperature of cold junction must be compensated by<br />

manual setting of RJ value. The connection is similar to voltmeter connection.<br />

• with cable adapter Option 140-01 to the terminals Hi/Lo. Automatic compensation of TC sensors cold<br />

junction can be used here. The cable adapter is equipped with build-in Pt1000 temperature, which measures<br />

ambient temperature. After build-in meter switching on (button INPUT ON) and with cable adapter 140-01<br />

connected, automatic compensation of TC cold junction temperature is performed. When meter is<br />

deactivated (INPUT OFF), cold junction temperature (RJ) can be set manually only. Connection of device<br />

under test is similar to the voltmeter connection.<br />

112 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

• using 140-01 cable adapter and OUTPUT terminals. Temperature of cold junction must be compensated by<br />

manual setting of RJ value.<br />

Connection of the transducer to be checked using 140-41 cable adapter<br />

DUT<br />

+TC<br />

-TC<br />

Measurement<br />

Thanks to built-in multimeter, the calibrator can be used for basic calibration of some sources of electrical<br />

signals. The table lists the type of adapter which is necessary for a particular measurement.<br />

Applications and desired options<br />

DC voltage to 12 V Opt. 40 or Opt 140-41<br />

DC current to 25 mA Opt. 40 or Opt 140-41<br />

Impulse frequency to 15 kHz Opt. 40 or Opt 140-41<br />

Temperature through external TC sensor<br />

Opt.140-41<br />

Temperature through external RTD sensor<br />

Opt. 60 or Opt.140-41<br />

Non-electric quantities with strain gauge sensors (force, pressure, Opt.140-41<br />

torque, etc.)<br />

Resistance to 2 kΩ Opt. 60 or Opt. 140-41<br />

Voltage, current and frequency<br />

10 V voltage range, 20 mA current range and frequency up to 15 kHz can be measured using Opt. 40 cable. The<br />

connection simple. The cable is connected to calibrator’s AUXILIARY connector, the other end has bananas<br />

which connect calibrator to the measured object. When making the connection, observe the polarity and connect<br />

the calibrator’s L terminal to the L (common) or grounded terminal of the measurement instrument. To activate<br />

the measurement, select the respective function mode and press INPUT to measure the input value.<br />

ATTENTION<br />

The input terminals of the built-in multimeter are floating. Maximum voltage between<br />

the input terminals and the chassis is 15 Vpk. If this value is exceeded, the multimeter<br />

can be damaged.<br />

If the measurement range is exceeded, the calibrator displays an error message; input terminals are disconnected<br />

only in the voltage and current measurement mode and remain connected in all other modes.<br />

If 140-41 cable adapter is used, the signal to be measured is connected to U/I – COM terminals. COM is the<br />

common terminal of the multimeter.<br />

Connection of the multimeter when measuring the voltage, current and frequency using Opt. 140-41 cable<br />

adapter<br />

Operation manual v42 113


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

H<br />

L<br />

The multimeter allows the measurement of small DC voltages in the range of 0 to 2 V. 140-41 cable adapter is<br />

required. The signal to be measured connects to Hu/+mV and Lu/-mV terminals. Lu/-mV is the common<br />

terminal of the multimeter. To activate the measurement, select mVDC function mode using METER button in<br />

the function menu and press INPUT to measure the input value.<br />

Measurement of resistance or temperature using resistance temperature sensors<br />

Resistance can only be measured using four-wire connection using Opt. 60 or Opt. 140-41 cable adapter. Opt. 60<br />

cable ends with four bananas, labeled Hi, Hu, Lu, Li. Their meaning is as follows:<br />

• Hi current terminal H<br />

• Hu voltage terminal H<br />

• Lu voltage terminal L<br />

• Li current terminal L<br />

During the measurement of resistance or temperature using resistance temperature sensors, the rules applicable<br />

to the four-wire connection must be observed.<br />

Connection of Pt100 sensor using Opt. 60 cable:<br />

Pt100<br />

Hi<br />

Hu<br />

Lu<br />

Li<br />

Connection of Pt100 sensor using 140-41 cable adapter<br />

114 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

Pt100<br />

Hi<br />

Hu<br />

Lu<br />

Li<br />

Measurement of temperature using thermocouples<br />

The built-in multimeter allows the measurement of temperature using external thermocouple. 140-41 is required.<br />

The thermocouple connects to Hu/+mV and Lu/-mV terminals. Temperature of cold junction must be set<br />

manually.<br />

To activate the measurement, select T TC function mode using METER button in the function menu and press<br />

INPUT to measure the input value.<br />

Connection of thermocouple to 140-41 cable adapter:<br />

H<br />

L<br />

Strain gauge sensors for non-electrical values<br />

The calibrator can be used to measure non-electrical values using bridge-connected strain gauges. The bridge is<br />

supplied with DC voltage up to 15 V while measurement of small DC voltages between the other two points of<br />

the bridge is measured. Each sensor is characterized by calibrated sensitivity of output voltage dependency on<br />

the measured non-electrical value.<br />

Operation manual v42 115


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

140-41 cable adapter fitted with separate 7pin connector labeled SENSOR is required. The picture shows the<br />

connection of the sensor to the calibrator.<br />

SGS sensor<br />

Connection of the sensor to SENSOR connector<br />

SENSOR connector<br />

4<br />

3<br />

SGS sensor<br />

5<br />

7<br />

2<br />

6<br />

1<br />

Meaning of pins of SENSOR connector:<br />

1 Voltage signal output of the bridge<br />

2 Power supply + power to the bridge<br />

3 Not connected<br />

4 Not connected<br />

5 Voltage signal output of the bridge<br />

6 Power supply - power to the bridge<br />

7 Shielding<br />

116 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

Testing of regulation and measurement sets and evaluation units<br />

The calibrator can be used for calibration and testing of various instruments and regulators which must be fed by<br />

precise signal and whose response (electrical signal) is to be measured.<br />

Two methods of connection can be used, each of them having different function and generation and<br />

measurement ranges.<br />

Simultaneous mode can only be used if Opt 40, 60, or 140-41 cable adapter is connected to AUXILIARY<br />

terminal. When no adapter is connected, simultaneous mode cannot be activated.<br />

Use of Opt. 140-41 cable adapter<br />

This adapter allows full utilization of the features of the built-in multimeter. It can be used during the calibration<br />

of objects which require the simulation of temperature or other non-electrical sensor signals, or the generation of<br />

small DC voltages at the input side, and the measurement of standard outputs (current loop, standard voltage) at<br />

the output side.<br />

When using 140-41 cable adapter, OUTPUT 140-41 option must be set in the setup menu. This function selects<br />

the calibrator’s output terminals. If OUTPUT 140-41 option is set to PANEL, calibrator’s output is connected to<br />

front panel terminals (not to the cable adapter). If this option is set to AUX, output signals are connected to the<br />

cable adapter (not to the front panel terminals). “Multimeter/Simultaneous function” chapter defines the limits of<br />

functions and ranges.<br />

If 140-41 cable adapter is used for simultaneous calibration and OUTPUT 140-41 option is set to AUX, the<br />

following generation/measurement ranges can be set:<br />

Generator<br />

DCV<br />

DCI<br />

Resistance<br />

RTD sensor simulation<br />

TC sensor simulation<br />

Meter<br />

DCV<br />

mVDC<br />

DCI<br />

Resistance<br />

Frequency<br />

0 – 20 V<br />

0 – 20 mA<br />

0 – 50 MΩ<br />

Pt 1.385, Pt 1.392, Ni<br />

K, N, R, S, B, J, T, E<br />

0 – 12 V<br />

0 – 2 V<br />

0 – 25 mA<br />

0 – 2 kΩ<br />

1 – 15kHz<br />

Operation manual v42 117


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

If 140-41 cable adapter is used for simultaneous calibration and OUTPUT 140-41 option is set to PANEL, the<br />

following generation/measurement ranges can be set:<br />

Generator<br />

DC/AC V<br />

DC/ AC I<br />

Resistance<br />

RTD sensor simulation<br />

TC sensor simulation<br />

Fequency<br />

Power/Energy<br />

Meter<br />

DCV<br />

MVDC<br />

DCI<br />

Resistance<br />

Frequency<br />

0 – 1000 V<br />

0 – 20 A<br />

0 – 50 MΩ<br />

Pt 1.385, Pt 1.392, Ni<br />

K, N, R, S, B, J, T, E<br />

0.1 – 20 MHz<br />

0.2 – 240 V / 0.2 – 20 A<br />

0 – 12 V<br />

0 – 2 V<br />

0 – 25 mA<br />

0 – 2 kΩ<br />

1 – 15 kHz<br />

When 140-41 cable adapter is connected, only harmonic signals are generated (SHAPE function is inactive).<br />

Use of Option 40/60 cable adapter<br />

Opt. 40 cable supports simultaneous mode - measurement of voltage up to 12 V, current up to 25 mA and<br />

frequency up to 15 kHz. Opt. 60 cable is designed for four-wire measurement of resistance or temperature using<br />

resistance temperature sensors. The generation ranges are the same as when 140-41 cable adapter is used.<br />

Use of Option 70<br />

Adapter is aimed for four-wire generation of resistance values and for four-wire resistance temperature sensor<br />

simulation. Re-calibration of resistance function of the calibrator can be performed only with this adapter.<br />

Resistance ranges are the same as two-wire resistance ranges with output on Hi – Lo terminals on front panel.<br />

When adapter Option 60 is used, resistance output is only on output terminals on this adapter. Front panel Hi –<br />

Lo terminals are switched off.<br />

When adapter Option 60 is used, set resistance value is connected to the adapter output terminals permanently.<br />

Status of button ON/OFF doesn’t influence resistance connecting or disconnecting.<br />

Examples of tests<br />

Examples of applications<br />

Generation Measurement Application<br />

Pt 100 10 V / 20 mA / f Calibration of temperature evaluation units, setting of<br />

temperature regulators<br />

TC 10 V / 20 mA / f Calibration of temperature evaluation units, setting of<br />

temperature regulators<br />

Frequency 10 V / 20 mA Setting/calibration of energy meters<br />

Resistance 10 V / 20 mA Measurement of resistance bridges<br />

1. Calibration of industrial thermometers with Pt100 sensor and 20 mA / 10 V output:<br />

118 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

DUT<br />

INPUT OUTPUT<br />

Pt100 H L<br />

Function setup: <strong>Calibrator</strong> T RTD function<br />

OUTPUT 140-41 ON<br />

Multimeter DCV or DCI depending on the output signal type of the unit<br />

being tested<br />

Adapter Opt. 140-41<br />

2. Calibration of industrial thermometers with thermocouple sensor and frequency output:<br />

DUT<br />

INPUT OUTPUT<br />

TC H L<br />

Function setup: <strong>Calibrator</strong> T TC function<br />

OUTPUT 140-41 ON<br />

Multimeter F<br />

Adapter Opt. 140-41<br />

3. Calibration of a single phase “electrical power/current loop” transducer<br />

DUT – power transducer<br />

INPUT<br />

OUTPUT<br />

HU LU HI LI +I -I<br />

Opt. 40 adapter<br />

Function setup: <strong>Calibrator</strong> P function (voltage, current, power factor, frequency)<br />

Multimeter DCI<br />

Adapter Opt. 40<br />

Operation manual v42 119


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

Testing<br />

When the calibrator is used as a tester, the results of the tests (PASS/FAIL) can be used to sort the products etc.<br />

Contacts of the relay are connected to 140-41 cable adapter’s front panel using black terminals ∅ 4mm. Basic<br />

definition of the status of the relay is performed using SETUP MENU and is valid for all test programs.<br />

120 Operation manual v42


MEATEST<br />

Specification<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

Uncertainties include long-term stability, temperature coefficient, linearity, load and line regulation and the<br />

traceability of factory and National calibration standards. Specified accuracy is valid after one hour warm up in<br />

temperature range 23 ± 2 o C. Specified accuracy is one year accuracy.<br />

<strong>Calibrator</strong><br />

Voltage<br />

summary range DCV:<br />

0 µV – 1000 V<br />

summary range ACV:<br />

0.1 mV – 1000 V<br />

internal ranges:<br />

20 mV, 200 mV, 2 V, 20 V, 240 V, 1000 V<br />

voltage resolution:<br />

61/2 digit<br />

frequency resolution:<br />

6 digit, minimum step 0.001Hz<br />

frequency accuracy: 0.005 %<br />

frequency range:<br />

20 Hz to 100 kHz bellow 20 V<br />

20 Hz to 10 kHz bellow 200 V<br />

20 Hz to 1000 Hz bellow 1000 V<br />

DCV uncertainty<br />

range % value + % range max. current mA<br />

0 µV – 20.00000 mV 0.03 + 0.0 + 10 µV 5<br />

20.00001 mV – 200.0000 mV 0.01 + 0.0 + 15 µV 5<br />

0.200001 mV – 2.000000 V 0.003 + 0.0008 30<br />

2.00001 V – 20.00000 V 0.003 + 0.0005 30<br />

20.0001 V – 240.0000 V 0.003 + 0.0005 30<br />

240.001 V – 1000.000 V 0.005 + 0.005 2<br />

* 2 value of the range for uncertainty calculation is 200 V<br />

ACV uncertainty<br />

range % value + % range max. current mA % value + % range max. current mA<br />

20 Hz - 10 kHz 20 Hz - 10 kHz 10 kHz - 50 kHz 10 kHz - 50 kHz<br />

0.10000 mV – 20.00000 mV 0.2 + 0.05 + 20 µV 5 0.20 + 0.10 + 20 µV 5<br />

20.0001 mV – 200.0000 mV 0.1 + 0.03+ 20 µV 5 0.15 + 0.05 + 20 µV 5<br />

0.200001 mV – 2.000000 V 0.025 + 0.005 30 0.05 + 0.01 10<br />

2.00001 V – 20.00000 V 0.025 + 0.005 30 0.05 + 0.03 10<br />

20.0001 V – 240.0000 V * 2 0.025 + 0.010 30 - -<br />

240.001 V – 1000.000 V 0.03 + 0.02 * 1 2 - -<br />

* 1 valid for f < 1000 Hz<br />

* 2 value of the range for uncertainty calculation is 200 V, in the range 200 to 240 V is frequency limited to 1000 Hz.<br />

range % value+ % range max. current mA<br />

50 kHz - 100 kHz 50 kHz - 100 kHz<br />

0.10000 mV – 20.00000 mV 1.0 + 0.10 + 20 µV 3<br />

20.0001 mV – 200.0000 mV 0.3 + 0.05 + 20 µV 3<br />

0.200001 mV – 2.000000 V 0.2 + 0.05 5<br />

2.00001 V – 20.00000 V 0.2 + 0.05 5<br />

20.0001 V – 240.0000 V * 2 - -<br />

240.001 V – 1000.000 V - -<br />

Auxiliary parameters<br />

range 20mV 200mV 2V 20V 200V 1000V<br />

THD *3 *4 0,05% + 200 uV 0,05% + 300 uV 0,05% 0,05% 0,05% 0,2%<br />

output impedance < 10 mΩ < 10 mΩ < 10 mΩ < 10 mΩ < 100 mΩ < 100 mΩ<br />

maximal capacitance 500 pF 500 pF 500 pF 500 pF 300 pF 150 pF<br />

load<br />

*3<br />

parameter includes non-linear distortion and non-harmonic noise<br />

*4<br />

valid for frequencies to 10 kHz<br />

Operation manual v42 121


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

Function Shape<br />

voltage range:<br />

1 mV to 200 V<br />

wave form:<br />

square, positive, negative, symmetrical, ramp A, ramp B, triangle<br />

truncated sin with THD 13,45 %<br />

peak value uncertainty:<br />

0.3 % + 50 uV<br />

displayed values:<br />

peak, effective<br />

Minimum frequency for squarewave signals is 0.1 Hz, for all others 20 Hz.<br />

Current<br />

summary range DCI: 0 µA - 20 A (with current coil 140-50 to 1000A)<br />

summary range ACI: 1 µA – 20 A (with current coil 140-50 to 1000 A)<br />

internal ranges:<br />

200 µA, 2 mA, 20 mA, 200 mA, 2 A, 20 A<br />

resolution:<br />

61/2 digit<br />

frequency resolution:<br />

6 digit, minimum step 0.001Hz<br />

frequency accuracy: 0.005 %<br />

frequency range: 20 Hz to 5 kHz bellow 200 mA<br />

20 Hz to 1000 Hz bellow 20 A<br />

DCI uncertainty<br />

range % value + % range max. voltage V<br />

0.0000 µA – 200.0000 µA 0.05 + 0.0 + 20 nA 3<br />

0.200001 µA – 2.000000 mA 0.02 + 0.005 3<br />

2.00001 mA – 20.00000 mA 0.01 + 0.003 3<br />

20.0001 mA – 200.0000 mA 0.01 + 0.003 3<br />

0.200001 mA – 2.000000 A 0.015 + 0.005 3<br />

2.00001 A – 20.00000 A 0.02 + 0.010 1.5<br />

ACI uncertainty<br />

range % value + % range max voltage Vef % value + % range max voltage Vef<br />

20 Hz - 1 kHz 1 kHz – 5 kHz<br />

1.0000 µA – 200.0000 µA 0.15 + 0.0 + 20 nA 3 0.30 + 0.10 + 20 nA 3<br />

0.200001 µA – 2.000000 mA 0.07 + 0.01 3 0.20 + 0.05 3<br />

2.00001 mA – 20.00000 mA 0.05 + 0.005 3 0.20 + 0.05 3<br />

20.0001 mA – 200.0000 mA 0.05 + 0.005 3 0.20 + 0.05 3<br />

0.200001 mA – 2.000000 A 0.05 + 0.005 3 - - - -<br />

2.00001 A – 20.00000 A 0.10 + 0.03 1.5 - - - -<br />

range % value + % range max voltage<br />

Vef<br />

5 kHz – 10 kHz<br />

1.0000 µA – 200.0000 µA - - - -<br />

0.200001 µA – 2.000000 mA 0.50 + 0.07 2<br />

2.00001 mA – 20.00000 mA 0.50 + 0.07 2<br />

20.0001 mA – 200.0000 mA 0.50 + 0.07 2<br />

0.200001 mA – 2.000000 A - - - -<br />

2.00001 A – 20.00000 A - - - -<br />

When option 140-50 Current Coil is used, add uncertainty 0.3% of the set current to the value specified in above table. Output current is<br />

multiplied by factor 50.<br />

Auxiliary parameters<br />

range 200 uA 2 mA 20 mA 200 mA 2 A 10 A<br />

maximal inductive load 400 uH 400 uH 400 uH 400 uH 200 uH 100 uH<br />

THD *1 0,2% 0,2% 0,2% 0,2% 0,2% 0,3%<br />

*1<br />

parameter includes non-linear distortion and non-harmonic noise<br />

122 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

Function Shape<br />

current range:<br />

100 uA to 2 A<br />

wave form:<br />

square, positive, negative, symmetrical, ramp A, ramp B, triangle<br />

truncated sin with THD 13,45 %<br />

peak value uncertainty:<br />

0.3 % + 500 nA<br />

displayed values:<br />

peak, effective<br />

Minimum frequency for squarewave signals is 0.1 Hz, for all others 20 Hz.<br />

Resistance<br />

summary range:<br />

resolution :<br />

Resistance uncertainty<br />

0 Ω to 50 MΩ<br />

41/2 digit<br />

resistance range uncertainty of value [%] *1 *2 current range *3<br />

0.00 Ω - 100.00 Ω 0.03 + 10 mΩ 1 mA - 40 mA<br />

100.01 Ω - 400.0 Ω 0.015 400 µA - 20 mA<br />

400.1 Ω - 2.0000 kΩ 0.015 100 µA - 4 mA<br />

2.001 kΩ - 10.000 kΩ 0.015 20 µA - 1 mA<br />

10.001 kΩ - 40.00 kΩ 0.015 4 µA - 200 µA<br />

40.01 kΩ - 200.00 kΩ 0.015 1 µA - 40 µA<br />

200.1 kΩ -1.0000 MΩ 0.05 0.2 µA - 10 µA<br />

1.0001 MΩ - 4.000 MΩ 0.1 40 nA – 2 µA<br />

4.001 MΩ - 20.000 MΩ 0.2 10 nA – 500 nA<br />

20.01 MΩ - 50.00 MΩ 0.5 4 nA – 150 nA<br />

*1<br />

In range from 0 Ω to 200 kΩ uncertainty is valid for four-terminal connection with use of Option 70 or Option 140-41 cable adapter. For<br />

two-wire connection from the terminals Hi-Lo na the front panel add to the specified uncertainty next 20 mOhm.<br />

*2<br />

In range from 200 kΩ to 50 MΩ uncertainty is valid for two-terminal connection to front panel terminals Hi-Lo.<br />

*3<br />

Maximal allowed voltage on output terminals is 8 V pp.<br />

Capacitance<br />

summary range: 0.9 nF to 50 µF<br />

resolution :<br />

41/2 digit<br />

Capacitance uncertainty<br />

range uncertainty of value [%] max. frequency<br />

0.9000 nF - 2.500 nF 0.5 + 15 pF 1000 Hz<br />

2.501 nF – 10.000 nF 0.5 + 5 pF 1000 Hz<br />

10.001 nF – 50.00 nF 0.5 1000 Hz<br />

50.01 nF – 0.2500 µF 0.5 1000 Hz<br />

0.2501 µF – 1.0000 µF 0.5 500 Hz<br />

1.0001 µF – 2.500 µF 1 300 Hz<br />

2.501 µF – 5.000 µF 1 300 Hz<br />

5.001 µF – 10.000 µF 1.5 300 Hz<br />

10.001 µF – 50.00 µF 2.0 300 Hz<br />

Maximal allowed voltage on output terminals is 8 V pp.<br />

AC and DC power/energy<br />

summary voltage range:<br />

0.2 V to 240 V<br />

current capability of voltage output:<br />

depends on the voltage range<br />

summary current range:<br />

2 mA to 10 A<br />

maximal voltage on current output:<br />

depends on the current range<br />

electric power range:<br />

0.0004 to 2.4 kVA<br />

phase angle ( PF range):<br />

-180° to +180° (-1 to +1 in PF)<br />

phase angle ( PF ) resolution:<br />

0.1° (0.001 in PF)<br />

time period range:<br />

1.1 s to 1999 s<br />

time setting resolution: 0.1s<br />

time period accuracy:<br />

0.01% + 0.1 s<br />

frequency range:<br />

DC, 40 Hz to 400 Hz<br />

frequency resolution:<br />

6 digit, minimum step 0.001Hz<br />

DCV uncertainty<br />

See table of DCV uncertainty.<br />

Operation manual v42 123


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

DCI uncertainty<br />

range % value + % range max. voltage [V]<br />

2.00000 mA – 20.00000 mA 0.05 + 0.010 3<br />

20.0001 mA – 200.0000 mA 0.05 + 0.005 3<br />

0.200001 mA – 2.000000 A 0.05 + 0.005 3<br />

2.00001 A – 10.00000 A 0.05 + 0.010 1.5<br />

DC POWER uncertainty<br />

Uncertainty of DC power can be calculated from the following formula:<br />

d P = √ ( dU 2 + dI 2 + 0.01 2 ) [%]<br />

where dP is uncertainty of output power [%]<br />

dU is uncertainty of set voltage [%]<br />

dI is uncertainty of set current [%]<br />

DC ENERGY uncertainty<br />

Depends on voltage, current, time values. The best uncertainty is 0.016 %.<br />

ACV uncertainty<br />

See table of ACV uncertainty.<br />

ACI uncertainty<br />

range % value + % range max. voltage [V]<br />

2.00000 mA – 20.00000 mA 0.05 + 0.010 3<br />

20.0001 mA – 200.0000 mA 0.05 + 0.005 3<br />

0.200001 mA – 2.000000 A 0.05 + 0.005 3<br />

2.00001 A – 10.00000 A 0.05 + 0.010 1.5<br />

PHASE uncertainty<br />

frequency range [Hz] phase uncertainty dϕ [°]<br />

40.000 – 200.000 0.15<br />

200.001 – 400.000 0.25<br />

AC POWER uncertainty<br />

Uncertainty of AC power can be calculated from the following formula:<br />

for active power d P = √ ( dU 2 + dI 2 + dPF 2 + 0.03 2 ) [%]<br />

for reactive power d P = √ ( dU 2 + dI 2 + dPF* 2 + 0.03 2 ) [%]<br />

for apparent power d P = √ ( dU 2 + dI 2 + 0.03 2 ) [%]<br />

where dP is uncertainty of power [%]<br />

dU is uncertainty of set voltage [%]<br />

dI is uncertainty of set current [%]<br />

dPF is uncertainty of power factor (cosϕ) [%]<br />

For calculation of dPF is valid following formula:<br />

dPF = (1 – cos (ϕ+dϕ)/cos ϕ) * 100 [%]<br />

where ϕ is set phase shift between voltage and current outputs<br />

dϕ is uncertainty of set phase shift in table above<br />

dPF* is uncertainty of of sinϕ [%]<br />

For dPF* is valid following formula:<br />

dPF* = (1 – sin (ϕ+dϕ)/sin ϕ) * 100 [%]<br />

124 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

Example:<br />

Set parameters:<br />

U = 100 V, I = 10 A, cos ϕ = 0.5, f = 50 Hz, displayed value of active power in W<br />

Output voltage uncertainty: dU = 0.025 % value + 0.010 % range = 0.045 %<br />

Output current uncertainty: dI = 0.05 % value + 0.01 % range = 0.06 %<br />

Uncertainty due the set phase shift: PF 0.5 corresponds phase shift 60 °<br />

dPF = (1 – cos (60+0.15)/cos 60) * 100 = (1 – 0.4977/0.5) * 100 = 0.45 %<br />

Output power uncertainty: dP = √ ( 0.045 2 + 0.06 2 + 0.45 2 + 0.03 2 ) = 0.46 %<br />

POWER FACTOR (PF)<br />

range: -1.000 to +1.000<br />

PF uncertainty can be calculated for any set value of output voltage, current and PF from following formula:<br />

dPF = (1 – cos (ϕ+dϕ)/cos ϕ) * 100 [%]<br />

where ϕ is set phase shift between voltage and current<br />

dϕ is uncertainty of set phase shift from the above table<br />

AC ENERGY uncertainty<br />

It depends on set value of voltage, current, time and PF. The best uncertainty is 0.07% for apparent energy.<br />

Frequency<br />

summary range:<br />

0.1 Hz to 20 MHz<br />

frequency uncertainty: 0.005 %<br />

frequency resolution:<br />

6 digit, minimum step 0.001Hz<br />

output:<br />

BNC connector located on the front panel<br />

modes:<br />

- PWM square wave output with calibrated duty cycle ratio, frequency and<br />

amplitude<br />

- HF square wave output with calibrated frequency and amplitude<br />

Mode PWM<br />

frequency range:<br />

0.1 Hz to 100 kHz<br />

voltage range:<br />

1 mV to 10 V<br />

duty cycle ratio range: 0.01 to 0.99<br />

wave form:<br />

square, symmetrical– positive – negative<br />

duty cycle ratio uncertainty: 0.05 %<br />

signal noise (20 Hz – 1 MHz):<br />

< 2 mV from 0.1Hz to 10 kHz<br />

< 10 mV from 10kHz to 100 kHz<br />

Amplitude uncertainty (average top/bottom square signal level)<br />

range % value + V % value + mV<br />

0.1 Hz to 10 kHz 10kHz to 100 kHz<br />

1.00000 mV – 20.00000 mV 0.2 + 0.02 mV 0.2 + 0.1 mV<br />

20.0001 mV – 200.0000 mV 0.1 + 0.02 mV 0.1 + 0.1 mV<br />

0.200001 mV – 2.000000 V 0.1 + 0.02 mV 0.1 + 0.1 mV<br />

2.00001 V – 10.00000 V 0.1 + 0.02 mV 0.1 + 0.1 mV<br />

Mode HF<br />

frequency range:<br />

0.1 Hz to 20 MHz<br />

output impedance:<br />

50 Ω<br />

wave form: square symmetrical, duty cycle ratio 1:1<br />

amplitude:<br />

4 V pk-pk<br />

output amplitude range:<br />

0, -10, -20, -30 dB +/- 1 dB<br />

amplitude uncertainty: 10 %<br />

rise/fall time:<br />

< 3 ns<br />

Operation manual v42 125


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

Temperature sensors simulation<br />

temperature scale: ITS 90, PTS 68<br />

types of sensors:<br />

RTD, TC<br />

A. RTD (resistance) sensors<br />

types:<br />

range of R0 setting:<br />

temperature range:<br />

temperature uncertainty:<br />

resolution:<br />

Ranges and uncertainties of RTD sensor simulation<br />

type range –200 – 250 o C range 250 – 850 o C<br />

Pt100 0.1 o C 0.3 o C<br />

Pt200 0.1 o C 0.2 o C<br />

Pt1000 0.2 o C 0.4 o C<br />

Pt 1.385, Pt 1.392, Ni<br />

20 Ω to 2 kΩ<br />

-200 to +850 o C<br />

0.04 o C to 0.5 o C (see table bellow)<br />

0.1 o<br />

Ni100 0.07 o C *1 - -<br />

*1<br />

Valid in range –60 to +180 o C.<br />

Uncertainties in the table are maximal uncertainties of RTD sensor simulation. Actual uncertainty for each set value of<br />

simulated temperature is determined by uncertainty of relevant resistance. Actual temperature uncertainty is displayed on the<br />

calibrator display. Actual uncertainties are always lower than those in above table.<br />

B. TC sensors:<br />

types:<br />

temperature range:<br />

temperature uncertainty:<br />

resolution:<br />

K, N, R, S, B, J, T, E<br />

-250 to +1820 o C according to the type<br />

0.4 to 4.3 o C (see table bellow)<br />

0.1 o<br />

Ranges and uncertainties of TC sensor simulation (with function AUTOCAL ON)<br />

R range [ o C] -50 - 0 0 - 400 400 – 1000 1000 – 1767<br />

uncertainty [ o C] 3.2 2.1 1.4 1.7<br />

S range [ o C] -50 - 0 0 - 250 250 – 1400 1400 – 1767<br />

uncertainty [ o C] 2.7 2.1 1.7 2.0<br />

B range [ o C] 400 – 800 800 - 1000 1000 – 1500 1500 – 1820<br />

uncertainty [ o C] 2.8 1.8 1.6 1.8<br />

J range [ o C] -210 - -100 -100 - 150 150 – 700 700 – 1200<br />

uncertainty [ o C] 0.9 0.5 0.6 0.7<br />

T range [ o C] -200 - -100 -100 - 0 0 – 100 100 – 400<br />

uncertainty [ o C] 0.9 0.5 0.4 0.4<br />

E range [ o C] -250 - -100 -100 - 280 280 – 600 600 – 1000<br />

uncertainty [ o C] 1.6 0.4 0.5 0.5<br />

K range [ o C] -200 - -100 -100 - 480 480 – 1000 1000 – 1372<br />

uncertainty [ o C] 1.0 0.6 0.7 0.8<br />

N range [ o C] -200 - -100 -100 - 0 0 – 580 580 – 1300<br />

uncertainty [ o C] 1.2 0.7 0.6 0.8<br />

Uncertainties in the table are maximal uncertainties of TC sensor simulation. Actual uncertainty for each set value of<br />

simulated temperature is determined by uncertainty of relevant resistance. Actual temperature uncertainty is displayed on the<br />

calibrator display. Actual uncertainties are always lower than those in above table.<br />

126 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

Multimeter<br />

Measuring:<br />

Ranges and uncertainties<br />

DC voltage<br />

DC current<br />

resistance, temperature<br />

strain gauge sensors<br />

function total range uncertainty (%) resolution / range<br />

DC voltage - DCV *1 0 to +/-12.0000 V 0.01 % + 300 µV 100µV / 10V<br />

DC voltage - mVDC *1 0 to +/-2.00000 V 0.02 % + 7 µV 20mV / 100nV, 200mV / 1uV, 2V / 10uV<br />

DC current *1 0 to +/-25.0000 mA 0.015 % + 300 nA 100 nA/20mA<br />

Frequency 1.000 Hz to 15 000.00 Hz 0.005 10 µHz – 0.1 Hz<br />

Resistance *2 0.000 to 2 500.00 Ω 0.02% + 10 m Ω 20Ω / 1mΩ, 200Ω / 1mΩ, 2kΩ / 10mΩ<br />

Temperature - PT sensor -200.000 to +850.000 o C *3 0.1 o C 0.1 o C<br />

Temperature – TC sensor -250.00 to +1820.00 o C *4 see Table 0.01 o C<br />

Strain gauge sensors *5 depends on sensor 0.05 % + 10 µV + uncertainty of<br />

sensor<br />

*1 Uncertainties are valid after performing ZERO correction in set mode<br />

*2 Measuring current 1 mA<br />

*3 For Pt 1000 sensor maximum temperature is 350 o C. Maximum measured resistance is 2 kΩ.<br />

*4 Maximum allowed analogue input DC voltage is 2 V.<br />

*5 Supply voltage: 2 to 10 V DC, non-symmetrical<br />

Max. current:<br />

40 mA<br />

Input resistance:<br />

min. 100 MΩ<br />

Sensitivity: setable in range 0.5 mV to 100 mV /V<br />

Displayed unit :<br />

user´s defined<br />

Ranges and uncertainties of temperature measuring with TC sensor<br />

R range [ o C] -50 - 0 0 – 400 400 – 1000 1000 – 1770<br />

uncertainty [ o C] 2.5 1.5 1.0 1.2<br />

S range [ o C] -50 - 0 0 – 250 250 – 1400 1400 – 1770<br />

uncertainty [ o C] 2.0 1.6 1.1 1.3<br />

B range [ o C] 400 - 800 800 – 1000 1000 – 1500 1500 – 1820<br />

uncertainty [ o C] 2.0 1.3 1.2 1.1<br />

J range [ o C] -210 - -100 -100 – 150 150 – 700 700 – 1200<br />

uncertainty [ o C] 0.7 0.4 0.4 0.6<br />

T range [ o C] -200 - -100 -100 - 0 0 – 100 100 – 400<br />

uncertainty [ o C] 0.8 0.5 0.4 0.4<br />

E range [ o C] -250 - -100 -100 - 280 280 – 600 600 – 1000<br />

uncertainty [ o C] 1.1 0.4 0.4 0.5<br />

K range [ o C] -200 - -100 -100 - 480 480 – 1000 1000 – 1372<br />

uncertainty [ o C] 0.8 0.4 0.6 0.8<br />

N range [ o C] -200 - -100 -100 – 0 0 – 580 580 – 1300<br />

uncertainty [ o C] 0.9 0.5 0.5 0.8<br />

Sorting function<br />

Output GO/NG:<br />

Triggering :<br />

1 x make, 1 x break contacts, 50Vpp / 100mA<br />

external, internal, manual<br />

Operation manual v42 127


M-140 <strong>Multifunction</strong> <strong>Calibrator</strong> MEATEST<br />

General data<br />

Warm up time:<br />

60 minutes<br />

Operating temperatures: 23 ± 10 º C, relative humidity < 75%<br />

Temperature coefficient<br />

Temperature coefficient for temperature outside of Tcal<br />

±2°C from +13 ° C to +33 ° C is 0.1x specified accuracy / ° C<br />

Reference temperature:<br />

23 ± 2 º C<br />

Storage temperatures:<br />

-10 º C to +55 º C<br />

Dimension:<br />

450 x 480 x 150 mm<br />

Netto weight<br />

23 kg<br />

Power line:<br />

115 - 220/230 V – 50/60 Hz<br />

Power consumption:<br />

45 VA without load<br />

Max. 150 VA with full load<br />

Safety class: I according EN 61010 ed 2<br />

Used external fuses: F4L250V 1 pcs<br />

Used internal fuses: F1.6L250V 3 pcs<br />

F200mL250V<br />

2 pcs<br />

F2.5L250V<br />

2 pcs<br />

Accessories<br />

Basic accessories (included in delivery)<br />

• Power line cord 1 pc<br />

• User´s manual 1 pc<br />

• Test report 1 pc<br />

• Spare fuse 2 pc<br />

• Test cable 1000V/20A, 1m 2 pc<br />

• Option 40 Cable adapter Canon 25 / 2 x BANANA, 1 m 1 pc<br />

• Option 60 Cable adapter Canon 25 / 4 x BANANA, 1 m 1 pc<br />

• Option 70 Adapter for four-terminal resistance generation 1 pc<br />

• Cable RS RS-232 cable 1 pc<br />

Options (extra ordered)<br />

• 140-50 Current coil 50 turn<br />

• 140-01 Cable adapter for calibration of multimeters<br />

• 140-02 Set of cables<br />

• 140-41 Cable adapter for simultaneous calibration/measuring<br />

• Option 10 Output cable 20A/1000V (black)<br />

• Option 11 Output cable 20A/1000V (red)<br />

• Option 20 Output cable BNC/BNC<br />

• Option 30 Output cable BNC/BANANA<br />

• Option 40 Output cable D-SUB25/ 2xBANANA, 1 m<br />

• Option 60 Output cable D-SUB25/ 4xBANANA, 1 m<br />

• Option 100 Adapter for ¾” output terminals spacing<br />

• Cable GPIB IEEE488/IEEE488, 2m<br />

• Cable RS Cable RS-232 for connecting to PC<br />

• WinQbase SW for calibration of instrument<br />

• MEACA Program module for multimeters<br />

• Caliber Program module for multimeters<br />

Manufacturer<br />

MEATEST, s.r.o tel: +420 – 543 250 886<br />

Zelezna 3, 619 00 Brno fax: +420 – 543 250 890<br />

Czech Republic<br />

<strong>meatest</strong>@<strong>meatest</strong>.<strong>cz</strong><br />

www.<strong>meatest</strong>.com<br />

128 Operation manual v42


MEATEST<br />

M-140 <strong>Multifunction</strong> <strong>Calibrator</strong><br />

Operation Manual Supplement<br />

Manual Title:<br />

<strong>M140</strong><br />

Issue date: 9/2012<br />

Supplement issue: 001<br />

This supplement contains information necessary to ensure the accuracy of the above manual.<br />

Change 1<br />

On page 123 the original text under the table “Resistance uncertainty” has been replaced with the following:<br />

Maximal allowed voltage on output terminals is 8 V pp . Uncertainty is valid for four-terminal connection with use<br />

of Option 70 or Option 140-41 Cable adapters. For two-wire connection from the terminals Hi-Lo na the front<br />

panel add to the specified uncertainty next 20 mOhm.<br />

Change 2<br />

On page 123 text under the table “Resistance uncertainty” has been added:<br />

In range from 200 kΩ to 50 MΩ uncertainty is valid for two-terminal connection to front panel terminals Hi-Lo.<br />

Operation manual v42 129

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!