BREAKTHROUGH IN X-RAY DIFFRACTION GENERATORS! - Rigaku
BREAKTHROUGH IN X-RAY DIFFRACTION GENERATORS! - Rigaku
BREAKTHROUGH IN X-RAY DIFFRACTION GENERATORS! - Rigaku
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Specifications<br />
X-<strong>RAY</strong> GENERATOR<br />
• Maximum Load: 18kW<br />
• Tube Voltage: 20-60kV in 1kV steps<br />
• Tube Current: 10-300mA in 1mA steps [10-450mA (40kV) in 1mA steps (Optional)]<br />
• Stability of High Voltage: +/- 0.01% for a +/-10% variation in the line voltage and +/- 5°C<br />
change in ambient temperature<br />
• Ripple: 1% rms or less (at 60kV, 300mA)<br />
ROTAT<strong>IN</strong>G ANODE ASSEMBLY<br />
• Material: Select one from Cu (standard), Cr, Fe, Co, Ni, Mo, Ag, W, and Au<br />
• Target diameter: 10cm (common to horizontal and vertical types)<br />
• Rotation speed: 6000rpm<br />
• Vacuum sealing for rotary shaft: Low torque type magnetic fluid sealing<br />
• Bearing: Ceramic bearing<br />
• Water sealing: Mechanical seal<br />
X-<strong>RAY</strong> SOURCE SIZE<br />
• Focal spot size on the target: 0.5 x 10mm 2 as standard (Option: 0.3 x 3mm 2 , 0.2 x2mm 2 ,<br />
0.1 x 1mm 2 )<br />
• Effective size at 6° take-off angle: 0.05 x 10mm 2 (line focus), 0.5 x 1mm 2 (point focus)<br />
OPERATION PANEL<br />
• Manual control: Controller from the operation panel<br />
• External control: RS-232C interface is built-in to allow control from the host computer<br />
SAFETY DESIGN<br />
• An LED is used in the warning lamp to prolong life<br />
• Shutter functions are monitored. If failure occurs, the entire system will shut down.<br />
HEAT EXCHANGER (Optional)<br />
• Water cooled type or air cooled type heat exchanger are available<br />
RADIATION ENCLOSURE (Optional)<br />
• Radiation Enclosures: 1200 x 1000 and 1800 x 1000mm (each with fail safe release function)<br />
<strong>Rigaku</strong>/MSC, Inc., 9009 New Trails Dr., The Woodlands, TX 77381