SINEAX U553<br />
SINEAX U554<br />
Transducer for alternating voltage<br />
To measure sinusoidal or distorted alternating voltages,<br />
with power supply connection.<br />
Customer benefit<br />
• RMS value measurement up to crest factor 6<br />
• Possibility of adjusting the maximum value of the measuring range on site<br />
• Standard as GL (Germanischer Lloyd), suitable for ships<br />
• Can also be used for 400 Hz systems<br />
Technical data<br />
Meas. input: Different ranges from 0…100/√3 to 0…500 V or customized<br />
0…20 V to 0…690 V, maximum 400 V nominal value of the system against earth<br />
Nominal frequency 50/60 Hz or 400 Hz<br />
Maximum value of the measuring range can be set via potentiometer<br />
(approx. ±15%)<br />
Meas. output: 0…20 mA, 4…20 mA, 0…10 V or customized<br />
0…1 to 0…20 mA or 0.2…1 to 4…20 mA or<br />
0…1 to 0…10 mA or 0.2…1 to 2…10 V<br />
Setting time 50 ms or 300 ms<br />
Accuracy: Class 0.5 at 15…30 °C<br />
Power supply: 24 –60 V AC/DC or 85–230 V AC/DC (also from measurement input) or<br />
24 V AC / 24–60 V DC, connection on the low-voltage side<br />
H x W x D: 69.1 x 70 x 112.5 mm<br />
Transducer for alternating voltage<br />
To measure sinusoidal or distorted alternating voltages, with power supply connection.<br />
Customer benefit<br />
• RMS value measurement up to crest factor 6<br />
• The measuring range which is of interest can be highlighted<br />
Technical data<br />
Meas. input: Minimum value 0 V, maximum value of the measuring range E3 = 20…690 V,<br />
Step point 0.1 · E3…0.9 · E3<br />
Nominal frequency 50/60 Hz or 400 Hz<br />
Meas. output: Maximum value A3 = 1 mA, 5 mA, 10 mA, 20 mA, 10 V or customised 1…20 mA<br />
or 1…10 V<br />
A3<br />
A2<br />
A<br />
A1<br />
E<br />
E1 E2 E3<br />
A3<br />
A2<br />
A<br />
A1<br />
E<br />
E1 E2 E3<br />
E1 = 0 E1 = 0<br />
0,1 · E3 ≤ E2 ≤ 0.9 · E3 0.1 · E3 ≤ E2 ≤ 0.9 · E3<br />
A1 = 0 A1 = 0.2 · A3<br />
A1 ≤ A2 ≤ 0.9 · A3 A1 ≤ A2 ≤ 0.9 · A3<br />
<strong>Camille</strong> <strong>Bauer</strong><br />
Voltage Transducers<br />
Accuracy: Class 0.5 at 15…30 °C<br />
Power supply: 24 –60 V AC/DC or 85–230 V AC/DC (also from measurement input)<br />
24 V, 110 V, 115 V, 120 V, 230 V, 400 V AC 50/60 Hz or<br />
24 V AC / 24–60 V DC, connection on the low-voltage side<br />
H x W x D: 69.1 x 70 x 112.5 mm<br />
9