Microwave Devices <strong>Faculty</strong> <strong>of</strong> <strong>Engineering</strong> EMG<strong>2026</strong><br />

PROCEDURE:<br />

CAUTION 1: The RF power levels in the following experiments are not harmful, but a human eye<br />

may be damaged by low level <strong>of</strong> radiation. Do not look into the waveguide at any time when the<br />

equipment are on.<br />

CAUTION 2: Klystron tube get extremely hot when it is operated and must not be handled by hand.<br />

Measurement <strong>of</strong> Frequency and power with variation in repeller voltage<br />

1. Set up the equipment as shown in Figure 4. Switch on the power supply and RF output. Wait a few<br />

seconds for the Klystron tube to warm up.<br />

2. Select squarewave modulation. Set the modulation level to maximum. Adjust the repeller voltage to<br />

maximum and then decrease it slowly until the largest deflection (maximum power) is indicated on the SWR<br />

meter.<br />

3. Adjust the modulation frequency so that the SWR meter deflection is further maximised.<br />

4. Set the slotted-line probe at about 2-mm protrusion depth and adjust the tuner stub/knob for proper<br />

matching <strong>of</strong> the detector.<br />

5. Set the variable attenuator at 5-mm micrometer setting. (The reading <strong>of</strong> the SWR meter should be<br />

constant when the probe carriage is moved along the slotted-line waveguide.)<br />

6. Adjust the gain knob <strong>of</strong> the SWR meter so that the needle points at 0 dB. After this, the zeroing knob<br />

should be fixed.<br />

7. Measure the repeller voltage from the connector at the rear panel <strong>of</strong> the Klystron power supply using a<br />

multimeter and record the reading. The multimeter reading shall be multiplied by 10x to give the repeller<br />

voltage.<br />

8. Measure the frequency using the cavity wavemeter and record the result.<br />

(Starting with the cavity wavemeter dial set to maximum clockwise, unscrew the dial slowly. At first there<br />

will be little effect on the SWR meter reading until at one point the power will fall sharply. Read the<br />

frequency indicated on the scale.)<br />

You may also use the alternative method given in Appendix A for <strong>microwave</strong> frequency measurement.<br />

9. Vary the repeller voltage.<br />

10. Measure the repeller voltage and record the reading.<br />

11. Measure the frequency using the cavity wave meter and record the reading.<br />

12. Repeat steps 9-11 for the full range <strong>of</strong> repeller voltage.<br />

13. Plot relative power level in dB versus repeller voltage.<br />

14. Plot frequency versus repeller voltage.<br />