Vitatron (MDT) <strong>Pacemakers</strong> • 1.800.848.2876<br />

NBG X-Ray BOL/ERI BOL/ERI BOL/ERI<br />

Model No. Model Name Code ID Connec<strong>to</strong>r Polarity Wrench Rate MAG MAG Rate Rate Elective Replacement Elective Replacement Indica<strong>to</strong>r Indica<strong>to</strong>r (ERI) (ERI)<br />

550 Legacy SR SSIR VBA IS-1 Uni/Bi 2 PR/65 (SSI) 85 (SOO)/65 (SOO) Abrupt decrease in Paced/Mag rate.<br />

Mode change <strong>to</strong> SSI from SSIR<br />

551 Legacy SR SSIR VBB 5/6 Uni 2 PR/65 (SSI) 85 (SOO)/65 (SOO) Abrupt decrease in Paced/Mag rate.<br />

Mode change <strong>to</strong> SSI from SSIR<br />

552 Legacy SR SSIR VBC IS-1 Comp Uni/Bi 2 PR/65 (SSI) 85 (SOO)/65 (SOO) Abrupt decrease in Paced/Mag rate.<br />

Mode change <strong>to</strong> SSI from SSIR<br />

562 Opal MV SSIR VBE9 3.2LP Bi 2 --- --- --<br />

600 Saphir VDDR V600 IS-1 Uni/Bi(S) 2 PR/PI+100ms 100 (DOO)/ Abrupt Increase of 100 ms in<br />

(VVI) 95 (DOO) Paced/Mag Interval<br />

601 Saphir VDDR --- IS-1 --- 2 --- --- ---<br />

611 Ceryx-6 SSI V611 5/6 Uni 2 PR/PI+100ms 100/86 Abrupt Increase of 100 ms in<br />

Paced/Mag Interval<br />

614 Ceryx-6 SSI V614 3.2LP Bi 2 PR/PI+100ms 100/86 Abrupt Increase of 100 ms in<br />

Paced/Mag Interval<br />

620 Saphir II VDDR VB IS-1 Uni/Bi 2 PR/PI+100ms 100 (VOO)/ Abrupt Increase of 100 ms in<br />

(VVI) 95 (VOO) Paced/Mag Interval<br />

621 Ceryx-6 SSI V621 5 Uni 2 PR/PI+100ms 100/86 Abrupt Increase of 100 ms in Paced/Mag Interval<br />

623 Ceryx-6 SSI V623 IS-1 Uni 2 PR/PI+100ms 100/86 Abrupt Increase of 100 ms in Paced/Mag Interval<br />

624 Ceryx-6 SSI V624 IS-1 Bi 2 PR/PI+100ms 100/86 Abrupt Increase of 100 ms in Paced/Mag Interval<br />

627 Saphir II VDDR VB IS-1 Bi 2 --- --- Abrupt decrease in magnet rate<br />

630 Vita II VDD VDD VC IS-1 A:Bi V:Uni/Bi --- PR/PI+100ms 100 (DOO)/ Abrupt Increase of 100 ms in Paced/Mag<br />

(VVI) 86 (DOO) Interval<br />

640 Saphir III VDDR VDDR VC IS-1 A:Bi V:Uni/Bi --- PR/PI+100ms 100 (DOO)/ Abrupt Increase of 100 ms in Paced/Mag<br />

(VVI) 86 (DOO) Interval<br />