Intermedics, Inc. (BSC) <strong>Pacemakers</strong> • 1.800.227.3422<br />

NBG X-Ray BOL/ERI BOL/ERI<br />

Model No. Model Name Code ID Connec<strong>to</strong>r Polarity Wrench Rate MAG Rate Elective Replacement Indica<strong>to</strong>r (ERI)<br />

284-09E Cosmos III DDD/VVIR I*KJ IS-1 Uni/Bi SLC PR/PR** PR/90/80 (SOO) Mag Rate = 1st 5 beats @ 90 Followed by PR @<br />

BOL, or 90 @ IFI, or 80 @ ERI. 64 cycles only.<br />

284-09R Cosmos III DDD IHC 6In-Line/ Uni/Bi 2 PR/PR** PR/90/80 Mag Rate = 1st 5 beats @ 90 Followed by PR @<br />

3.2LP/IS-1<br />

BOL, or 90 @ IFI, or 80 @ ERI. 64 cycles only.<br />

291-01 Nova MR SSIR ICH IS-1 Uni SLC PR/PR (SSI) PR/90/80 Mag Rate = 1st 5 beats @ 90 Followed by PR @<br />

BOL, or 90 @ IFI, or 80 @ ERI. 64 cycles only.<br />

291-03 Dash SSIR IFE 6 Uni 2 PR/PR (SSI) PR/90/80 Mag Rate = 1st 5 beats @ 90 Followed by PR @<br />

BOL, or 90 @ IFI, or 80 @ ERI. 64 cycles only.<br />

291-09 Marathon SR SSIR IGC 6 Uni 2 PR/PR (SSI) PR/90/80 Mag Rate = 1st 5 beats @ 90 Followed by PR @<br />

BOL, or 90 @ IFI, or 80 @ ERI. 64 cycles only.<br />

292-02 Nova MR SSIR ICH IS-1 Bi SLC PR/PR (SSI) PR/90/80 Mag Rate = 1st 5 beats @ 90 Followed by PR @<br />

BOL, or 90 @ IFI, or 80 @ ERI. 64 cycles only.<br />

292-03 Dash SSIR IEK IS-1 Uni/Bi SLC PR/PR (SSI) PR/90/80 Mag Rate = 1st 5 beats @ 90 Followed by PR @<br />

BOL, or 90 @ IFI, or 80 @ ERI. 64 cycles only.<br />

292-03E Dash SSIR IEK IS-1 Uni/Bi SLC PR/PR (SSI) PR/90/80 Mag Rate = 1st 5 beats @ 90 Followed by PR @<br />

BOL, or 90 @ IFI, or 80 @ ERI. 64 cycles only.<br />

292-03R Dash SSIR IFF 6In-Line/ Uni/Bi 2 PR/PR (SSI) PR/90/80 Mag Rate = 1st 5 beats @ 90 Followed by PR @<br />

3.2LP/IS-1<br />

BOL, or 90 @ IFI, or 80 @ ERI. 64 cycles only.<br />

292-05 Dart SSIR IFJ IS-1 Uni/Bi SLC PR/PR (SSI) PR/90/80 Mag Rate = 1st 5 beats @ 90 Followed by PR @<br />

BOL, or 90 @ IFI, or 80 @ ERI. 64 cycles only.<br />

292-05E Dart SSIR I*KS IS-1 Uni/Bi SLC PR/PR (SSI) PR/90/80 Mag Rate = 1st 5 beats @ 90 Followed by PR @<br />

BOL, or 90 @ IFI, or 80 @ ERI. 64 cycles only.<br />

292-06 Unity-C VDDR IFT IS-1 A:Bi V:Uni SLC PR/PR (SSI) PR/90/80 Mag Rate = 1st 5 beats @ 90 Followed by PR @<br />

BOL, or 90 @ IFI, or 80 @ ERI. 64 cycles only.<br />

*ERI magnet rates are unique for each of these devices. Contact Bos<strong>to</strong>n <strong>Scientific</strong> CRM Technical Services <strong>to</strong> find the ERI magnet rate for a specified serial number.<br />

**ERI mode reversion <strong>to</strong> VVI is programmable ON/OFF in these devices.<br />