Guidant (BSC) <strong>ICDs</strong> • 1.800.227.3422<br />

NBD X-Ray Sense HV Defib<br />

Model No. Model Name Code ID Connec<strong>to</strong>r Connec<strong>to</strong>r Waveform Wrench Elective Replacement Indica<strong>to</strong>r (ERI)<br />

T177 VITALITY 2 VR EL VVEV GDT 109 V:IS-1 Bi (2) DF-1 Mono/Biphasic 2 Moni<strong>to</strong>ring Voltage 13.1 seconds.<br />

NOTE: If Moni<strong>to</strong>ring Voltage is 3.0V <strong>to</strong> 2.52V, ERI Charge<br />

Time is 18.9 seconds. Programmer Indicates ERI<br />

T180 VITALITY DR HE VVEV GDT 117 A:IS-1 Bi V:IS-1 Bi (2) DF-1 Mono/Biphasic 2 Moni<strong>to</strong>ring Voltage 14.6 seconds.<br />

NOTE: If Moni<strong>to</strong>ring Voltage is 3.0V <strong>to</strong> 2.55V, ERI Charge<br />

Time is 26.1 seconds. Programmer Indicates ERI<br />