Intermedics, Inc. (BSC) <strong>Pacemakers</strong> • 1.800.227.3422<br />

NBG X-Ray BOL/ERI BOL/ERI<br />

Model No. Model Name Code ID Connec<strong>to</strong>r Polarity Wrench Rate MAG Rate Elective Replacement Indica<strong>to</strong>r (ERI)<br />

294-03 Relay DDDR IEJ IS-1 Uni/Bi SLC PR/PR (SSI) PR/90/80 Mag Rate = 1st 5 beats @ 90 Followed by PR @<br />

BOL, or 90 @ IFI, or 80 @ ERI. 64 cycles only.<br />

294-03E Relay DDDR IHN IS-1 Uni/Bi SLC PR/PR (SSI) PR/90/80 Mag Rate = 1st 5 beats @ 90 Followed by PR @<br />

BOL, or 90 @ IFI, or 80 @ ERI. 64 cycles only.<br />

294-03R Relay DDDR I*FH 6In-Line/ Uni/Bi 2 PR/PR (SSI) PR/90/80 Mag Rate = 1st 5 beats @ 90 Followed by PR @<br />

3.2LP/IS-1<br />

BOL, or 90 @ IFI, or 80 @ ERI. 64 cycles only.<br />

294-05 Stride DDDR IFK IS-1 Uni/Bi SLC PR/PR (SSI) PR/90/80 Mag Rate = 1st 5 beats @ 90 Followed by PR @<br />

BOL, or 90 @ IFI, or 80 @ ERI. 64 cycles only.<br />

294-05E Stride DDDR I*KL IS-1 Uni/Bi SLC PR/PR (SSI) PR/90/80 Mag Rate = 1st 5 beats @ 90 Followed by PR @<br />

BOL, or 90 @ IFI, or 80 @ ERI. 64 cycles only.<br />

294-09 Marathon DR DDDR IFZ IS-1 Uni/Bi SLC PR/PR** PR/90/80 Mag Rate = 1st 5 beats @ 90 Followed by PR @<br />

BOL, or 90 @ IFI, or 80 @ ERI. 64 cycles only.<br />

294-09E Marathon DR DDDR I*JN IS-1 Uni/Bi SLC PR/PR** PR/90/80 (SOO) Mag Rate = 1st 5 beats @ 90 Followed by PR @<br />

BOL, or 90 @ IFI, or 80 @ ERI. 64 cycles only.<br />

294-09R Marathon DR DDDR IGG 6In-Line/ Uni/Bi 2 PR/PR** PR/90/80 Mag Rate = 1st 5 beats @ 90 Followed by PR @<br />

3.2LP/IS-1<br />

BOL, or 90 @ IFI, or 80 @ ERI. 64 cycles only.<br />

294-09Z Marathon DR DDDR IHJ IS-1 Uni/Bi SLC PR/PR** PR/90/80 Mag Rate = 1st 5 beats @ 90 Followed by PR @<br />

BOL, or 90 @ IFI, or 80 @ ERI. 64 cycles only.<br />

294-10 Marathon DR DDDR IGJ IS-1 Bi SLC PR/PR** PR/90/80 Mag Rate = 1st 5 beats @ 90 Followed by PR @<br />

BOL, or 90 @ IFI, or 80 @ ERI. 64 cycles only.<br />

294-23 Momentum DR DDDR IHG IS-1 Uni/Bi SLC PR/PR** PR/90/80 Mag Rate = 1st 5 beats @ 90 Followed by PR @<br />

BOL, or 90 @ IFI, or 80 @ ERI. 64 cycles only.<br />

294-23E Momentum DR DDDR I*HG IS-1 Uni/Bi SLC PR/PR** PR/90/80 Mag Rate = 1st 5 beats @ 90 Followed by PR @<br />

BOL, or 90 @ IFI, or 80 @ ERI. 64 cycles only.<br />

*ERI magnet rates are unique for each of these devices. Contact Bos<strong>to</strong>n <strong>Scientific</strong> CRM Technical Services <strong>to</strong> find the ERI magnet rate for a specified serial number.<br />

**ERI mode reversion <strong>to</strong> VVI is programmable ON/OFF in these devices.<br />