<strong>Measurement</strong> Good Practice Guide No. 39<br />
Figure 2: Airy disc intensity pattern.<br />
Figure 3: Intensity sum of two airy patterns.<br />
It will be seen later that other factors, including the coherence of the light, and the<br />
numerical aperture of the condenser must also be considered in assessing resolution.<br />
Nevertheless, expression (1) gives a reasonable guide to the resolution that can be<br />
expected from a microscope objective working under normal conditions.<br />
Equation (1) shows that the minimum resolvable object separation decreases with<br />
increasing objective numerical aperture. Resolution figures derived from this expression<br />
for typical microscope objectives are shown in Table 1 (assuming green light illumination<br />
of wavelength 550 nm).<br />
5