MEASUREMENTS
MEASUREMENTS
MEASUREMENTS
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Differential Phase<br />
Select DOUBLE in the 1781R DIFF PHASE mode to make this<br />
measurement. First look at the vectorscope screen and use the<br />
phase shifter to set the signal vector to the reference 9 o'clock<br />
phase position. Neither vectorscope nor waveform monitor gain<br />
is critical in this mode (see Note 18), but setting the vector to<br />
the graticule circle is a good starting point. Now refer to the<br />
waveform monitor (right-hand) screen, and use the phase<br />
shifter to overlay the blanking level portions of the two<br />
waveforms. Press REFerence SET to set the phase readout to<br />
0.00 degree, as shown in Figure 68.<br />
The next step is to use the phase shifter to overlay the point in<br />
the R-Y waveforms that deviates most from blanking level. As<br />
shown in Figure 69, the phase readout now indicates the<br />
amount of differential phase distortion. In this example the<br />
phase error is all in one direction so peak and peak-to-peak<br />
results will be the same. If the signal has both positive and<br />
negative phase errors (the R-Y signal extends both positive and<br />
negative from blanking), repeat the process for the largest<br />
positive and largest negative signal excursions.<br />
The double trace technique is similar if you are using a 521A<br />
Vectorscope. Start by setting the "calibrated phase" dial to<br />
zero. Use theA or B0 goniometer to null the blanking level,<br />
and then use the calibrated phase shifter to null the largest<br />
excursion. The number above the calibrated phase dial will<br />
now give the amount of differential phase distortion.<br />
VM700A Automatic Measurement<br />
To make an automatic measurement of differential phase with<br />
the VM700A, select DGDP in the MEASURE mode. Both differential<br />
phase and differential gain are shown on the same<br />
display; the lower graph is differential phase. Measurement<br />
results are also available in the AUTO mode.<br />
Figure 68. The 1781R double trace DIFF PHASE<br />
display; phase readout zeroed.<br />
Figure 69. The double trace DIFF PHASE display<br />
showing results.<br />
• NOTES<br />
18. 1781R Waveform & Vector Gains<br />
In the single trace mode, the vector gain must be set so the signal<br />
vector extends to the graticule circle. The waveform gain must<br />
be in the calibrated (1 volt full scale) position. The graticule is<br />
calibrated to one degree per division only under these conditions.<br />
With the double mode display, however, you can introduce more<br />
gain for greater resolution. Additional vectorscope gain and/or<br />
waveform vertical gain can be selected without affecting the results.<br />
19. Noise Reduction Filter<br />
A digital recursive filter is available in the 1781R to facilitate differential<br />
phase and gain measurements in the presence of noise.<br />
Select the NOISE REDUCTION ON touchscreen selection in<br />
the DIFF PHASE or DIFF GAIN menu to enable this filter. The<br />
filter removes about 15 dB of noise from the signal without any<br />
loss of bandwidth or horizontal resolution. This mode is<br />
particularly useful for VTR and transmitter measurements.<br />
Figure 70. The VM700A DGDP display.<br />
51