30.09.2014 Views

MEASUREMENTS

MEASUREMENTS

MEASUREMENTS

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Differential Phase<br />

Select DOUBLE in the 1781R DIFF PHASE mode to make this<br />

measurement. First look at the vectorscope screen and use the<br />

phase shifter to set the signal vector to the reference 9 o'clock<br />

phase position. Neither vectorscope nor waveform monitor gain<br />

is critical in this mode (see Note 18), but setting the vector to<br />

the graticule circle is a good starting point. Now refer to the<br />

waveform monitor (right-hand) screen, and use the phase<br />

shifter to overlay the blanking level portions of the two<br />

waveforms. Press REFerence SET to set the phase readout to<br />

0.00 degree, as shown in Figure 68.<br />

The next step is to use the phase shifter to overlay the point in<br />

the R-Y waveforms that deviates most from blanking level. As<br />

shown in Figure 69, the phase readout now indicates the<br />

amount of differential phase distortion. In this example the<br />

phase error is all in one direction so peak and peak-to-peak<br />

results will be the same. If the signal has both positive and<br />

negative phase errors (the R-Y signal extends both positive and<br />

negative from blanking), repeat the process for the largest<br />

positive and largest negative signal excursions.<br />

The double trace technique is similar if you are using a 521A<br />

Vectorscope. Start by setting the "calibrated phase" dial to<br />

zero. Use theA or B0 goniometer to null the blanking level,<br />

and then use the calibrated phase shifter to null the largest<br />

excursion. The number above the calibrated phase dial will<br />

now give the amount of differential phase distortion.<br />

VM700A Automatic Measurement<br />

To make an automatic measurement of differential phase with<br />

the VM700A, select DGDP in the MEASURE mode. Both differential<br />

phase and differential gain are shown on the same<br />

display; the lower graph is differential phase. Measurement<br />

results are also available in the AUTO mode.<br />

Figure 68. The 1781R double trace DIFF PHASE<br />

display; phase readout zeroed.<br />

Figure 69. The double trace DIFF PHASE display<br />

showing results.<br />

• NOTES<br />

18. 1781R Waveform & Vector Gains<br />

In the single trace mode, the vector gain must be set so the signal<br />

vector extends to the graticule circle. The waveform gain must<br />

be in the calibrated (1 volt full scale) position. The graticule is<br />

calibrated to one degree per division only under these conditions.<br />

With the double mode display, however, you can introduce more<br />

gain for greater resolution. Additional vectorscope gain and/or<br />

waveform vertical gain can be selected without affecting the results.<br />

19. Noise Reduction Filter<br />

A digital recursive filter is available in the 1781R to facilitate differential<br />

phase and gain measurements in the presence of noise.<br />

Select the NOISE REDUCTION ON touchscreen selection in<br />

the DIFF PHASE or DIFF GAIN menu to enable this filter. The<br />

filter removes about 15 dB of noise from the signal without any<br />

loss of bandwidth or horizontal resolution. This mode is<br />

particularly useful for VTR and transmitter measurements.<br />

Figure 70. The VM700A DGDP display.<br />

51

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!