21.11.2014 Views

Technical Service Manual PM2300 Series Liquid Oxygen Reservoir

Technical Service Manual PM2300 Series Liquid Oxygen Reservoir

Technical Service Manual PM2300 Series Liquid Oxygen Reservoir

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Removal, disassembly, assembly, Installation & replacement Instructions<br />

(continued)<br />

7.5 Fill Connector removal, disassembly, reassembly & installation, reference Figures 28 – 33<br />

Removal<br />

7.5.1 Remove the Shroud as per section 7.4 for one (1) piece shroud or 7.4a for two (2) piece shroud.<br />

7.5.2 Using a 7/8” deep socket, hold the Fill Connector Body and use a 1” box wrench on the<br />

Compression Nut to loosen the nut.<br />

NOTE: The Fill Connector has been installed with a high torque and will require<br />

considerable force to remove.<br />

7.5.3 Unthread the Fill Connector from the Manifold Assembly and remove.<br />

NOTE: • The Compression Nut and Compression Ferrule should not come off or be<br />

removed from the Manifold Neck.<br />

• Whenever the Poppet Assembly is removed, all parts (including Neck, Compression<br />

Ferrule, Compression Nut, Poppet Assembly, Spacer and Poppet Retaining Snap<br />

Ring) should be cleaned of debris before reassembly. Debris in the Fill Connector<br />

may cause it to leak.<br />

Figure 28<br />

Figure 29 Figure 30<br />

<strong>PM2300</strong> <strong>Series</strong> <strong>Liquid</strong> <strong>Oxygen</strong> <strong>Reservoir</strong><br />

<strong>Technical</strong> <strong>Service</strong> <strong>Manual</strong><br />

25

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!