Removal, disassembly, assembly, Installation & replacement Instructions<br />
(continued)<br />
7.5 Fill Connector removal, disassembly, reassembly & installation, reference Figures 28 – 33<br />
Removal<br />
7.5.1 Remove the Shroud as per section 7.4 for one (1) piece shroud or 7.4a for two (2) piece shroud.<br />
7.5.2 Using a 7/8” deep socket, hold the Fill Connector Body and use a 1” box wrench on the<br />
Compression Nut to loosen the nut.<br />
NOTE: The Fill Connector has been installed with a high torque and will require<br />
considerable force to remove.<br />
7.5.3 Unthread the Fill Connector from the Manifold Assembly and remove.<br />
NOTE: • The Compression Nut and Compression Ferrule should not come off or be<br />
removed from the Manifold Neck.<br />
• Whenever the Poppet Assembly is removed, all parts (including Neck, Compression<br />
Ferrule, Compression Nut, Poppet Assembly, Spacer and Poppet Retaining Snap<br />
Ring) should be cleaned of debris before reassembly. Debris in the Fill Connector<br />
may cause it to leak.<br />