User <strong>Manual</strong><br />
Piccolo Range<br />
1.4 Step 1: Installation<br />
This <strong>in</strong>strument is <strong>in</strong>tended for permanent <strong>in</strong>stallation, for <strong>in</strong>door use only, and enclosed <strong>in</strong> an electrical panel<br />
Select a location which is subject <strong>to</strong> m<strong>in</strong>imum vibrations <strong>the</strong> ambient temperature is with<strong>in</strong> 0 and 55 o C (32 - 131 o F)<br />
and humidity 5 <strong>to</strong> 95% RH non condens<strong>in</strong>g.<br />
The <strong>in</strong>strument can be mounted on a panel up <strong>to</strong> 15mm thick.<br />
To ensure IP65 front protection, mount on a non-textured surface.<br />
Please read <strong>the</strong> safety <strong>in</strong><strong>format</strong>ion <strong>in</strong> section 3 before proceed<strong>in</strong>g. An EMC Booklet, part number HA025464, gives<br />
fur<strong>the</strong>r <strong>in</strong>stallation <strong>in</strong><strong>format</strong>ion and can be <strong>download</strong>ed from www.euro<strong>the</strong>rm.co.uk.<br />
1.4.1 Panel Mount<strong>in</strong>g <strong>the</strong> Controller<br />
1. Prepare a cut-out <strong>in</strong> <strong>the</strong> mount<strong>in</strong>g panel <strong>to</strong> <strong>the</strong> size shown. If a number of controllers are <strong>to</strong> be mounted <strong>in</strong> <strong>the</strong><br />
same panel observe <strong>the</strong> m<strong>in</strong>imum spac<strong>in</strong>g shown.<br />
2. Carefully remove <strong>the</strong> panel reta<strong>in</strong><strong>in</strong>g clips from <strong>the</strong> sleeve us<strong>in</strong>g figures or a small screwdriver.<br />
3. To achieve IP65 seal<strong>in</strong>g, make sure <strong>the</strong> gasket is fitted beh<strong>in</strong>d <strong>the</strong> front bezel of <strong>the</strong> controller<br />
4. Insert <strong>the</strong> controller through <strong>the</strong> cut-out<br />
5. Spr<strong>in</strong>g <strong>the</strong> panel reta<strong>in</strong><strong>in</strong>g clips back <strong>in</strong><strong>to</strong> place. Secure <strong>the</strong> controller <strong>in</strong> position by hold<strong>in</strong>g it level and<br />
push<strong>in</strong>g both reta<strong>in</strong><strong>in</strong>g clips forward.<br />
6. Peel off <strong>the</strong> protective cover from <strong>the</strong> display.<br />
7. If <strong>the</strong> panel reta<strong>in</strong><strong>in</strong>g clips subsequently need remov<strong>in</strong>g, <strong>the</strong>y can be unhooked from <strong>the</strong> side with ei<strong>the</strong>r your<br />
f<strong>in</strong>gers or a screwdriver.<br />
1.4.2 Panel Cut Out Sizes<br />
45 mm - 0.0 + 0.6<br />
1.77 <strong>in</strong> -0.00, +0.02<br />
92 mm - 0.0 + 0.8<br />
3.62 <strong>in</strong> -0.00, +0.03<br />
P116<br />
45 mm - 0.0 + 0.6<br />
1.77 <strong>in</strong> -0.00, +0.02<br />
45 mm<br />
- 0.0 + 0.6<br />
1.77 <strong>in</strong><br />
-0.00, +0.02<br />
P108<br />
92 mm<br />
- 0.0 + 0.8<br />
3.62 <strong>in</strong><br />
-0.00, +0.03<br />
92 mm<br />
- 0.0 + 0.8<br />
3.62 <strong>in</strong><br />
-0.00, +0.03<br />
P104<br />
1.4.3 Recommended m<strong>in</strong>imum spac<strong>in</strong>g of controllers<br />
Applies <strong>to</strong> all models.<br />
10mm (0.4 <strong>in</strong>)<br />
38mm (1.5 <strong>in</strong>)<br />
(Not <strong>to</strong> scale)<br />
1.4.4 To Remove <strong>the</strong> Controller from its Sleeve<br />
The controller can be unplugged from its sleeve by eas<strong>in</strong>g <strong>the</strong> latch<strong>in</strong>g ears outwards and pull<strong>in</strong>g it forward out of<br />
<strong>the</strong> sleeve. When plugg<strong>in</strong>g it back <strong>in</strong><strong>to</strong> its sleeve, ensure that <strong>the</strong> latch<strong>in</strong>g ears click back <strong>in</strong><strong>to</strong> place <strong>to</strong> ma<strong>in</strong>ta<strong>in</strong> <strong>the</strong><br />
IP65 seal<strong>in</strong>g.<br />
6 Part No HA031260 Issue 1 May -12