15.01.2015 Views

MultiRAE - Shawcity Limited

MultiRAE - Shawcity Limited

MultiRAE - Shawcity Limited

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

calibration. This is the only way to positively differentiate between a faulty sensor and a<br />

real background signal. If you obtain a good calibration with “zero” air, and then recheck<br />

the zero with the same bottle of “zero” air, then you can have confidence that the<br />

instrument is working properly and the background signal you are seeing is caused by<br />

something in the environment.<br />

2.6.3 If, after a “zero” air calibration, you do not obtain a stable zero when applying the<br />

“zero” air to the instrument, then the sensor does not have a stable zero point and must be<br />

replaced.<br />

2.7 Pump Stall Adjustment.<br />

2.7.1 The Pump alarm can occur if the pump current exceeds the stall threshold value or<br />

if the pump current is 0. The pump current may increase over the threshold value as the<br />

pump ages or if the diaphragm is stiff due to cold temperatures. It may be necessary to<br />

adjust the pump stall threshold when switching from low to high pump speed or when<br />

adding additional lengths of tubing (Rev. D boards). Pump current may also increase if<br />

the inlet is blocked or the vapor trap filter is clogged. Before adjusting the pump stall<br />

threshold, make sure the flow path is not obstructed. Do not attempt to clean a clogged<br />

filter: it must be replaced.<br />

2.7.2 Turn on the unit in Diagnostic Mode (Section IV). The pump alarm is disabled in<br />

Diagnostic Mode. Press the [MODE] key repeatedly until you get to the “Pump =”<br />

screen. The top line is the measured pump current, and the bottom line is the threshold<br />

value at which the pump will stall in normal operating mode. When operating properly<br />

the stall threshold should be 6 - 10 counts higher than the operating current (typically 20 -<br />

30 counts).<br />

2.7.3 If the pump current is 0 then check for loose or broken wires. If the wires are<br />

intact, check the voltage between the red and black wires. It should be 3.5V. If the<br />

voltage is correct, then replace the pump assembly. If the voltage is not correct, then the<br />

unit will have to return to the factory for repair.<br />

2.7.4 If the pump is operating, block the inlet with your finger and observe the change in<br />

pump current:<br />

No change or only a few counts:<br />

1) Pump current < stall current means there is a leak in the flow path<br />

between your finger and the pump intake. Check the filter assembly, the<br />

inlet valve, the internal tubing, the pump inlet nipple, and the pump<br />

diaphragm. Clean or replace any parts that are dirty or leaking<br />

18

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!