15.01.2015 Views

MultiRAE - Shawcity Limited

MultiRAE - Shawcity Limited

MultiRAE - Shawcity Limited

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>MultiRAE</strong><br />

MULTIPLE GAS MONITOR<br />

(PGM - 50)<br />

DIAGNOSTIC AND SERVICE MANUAL<br />

(Document No: 008-4011-000)<br />

Rev. D.<br />

RAE SYSTEMS INC.<br />

1339 Moffet Park Drive<br />

Sunnyvale, CA 94089<br />

January 2004<br />

1


I. Objective:<br />

This document provides detailed diagnostic and service procedures for service personnel<br />

in the factory and field service center. The procedure allows trained service personnel to<br />

perform a set of diagnostic tests to determine if the Unit Under Test (UUT) is functioning<br />

properly and performing within the specification. The procedure also helps to diagnose<br />

the possible causes of failures and suggest a number of corrective actions.<br />

II. Diagnostic and Service Requirements:<br />

1) In order to perform the PGM-50 diagnostic and service procedures, the service person<br />

should be very familiar with the procedure described in the PGM-50 Operation and<br />

Maintenance Manual (RAE Part Number: 008-4001).<br />

2) The following equipment is required:<br />

PC with Windows 3.1/ Windows 95 or above operating system and serial port.<br />

Computer interface cable (RAE Part Number 008-3003)<br />

ProRAE Suite software<br />

DC power supply (0-10V, 200 mA)<br />

Multi-meter (0-10VDC, 0-150 mA, 0-10 megaohm)<br />

20 ohm, 1W resistor<br />

Flat head screw driver<br />

#2 phillips screw driver<br />

Tweezers<br />

Battery connector with wires<br />

2


III. Trouble Shooting Guide:<br />

PGM-50 problems can be grouped into 4 major areas:<br />

1) Power supply - battery and charger<br />

2) Sensor and measurement<br />

3) Electronics<br />

4) Programming<br />

The following trouble shooting guide provides a quick reference for typical problems in<br />

each major area:<br />

1) Power Supply - Battery and Charger:<br />

Problem Possible Causes Test Procedures<br />

Can not turn on unit 1) battery charge is low<br />

2) battery pack fuse is blown<br />

3) battery is disconnected from the PCB or the<br />

battery wire is broken<br />

4) defective electronics<br />

1) Test 1.5 & 1.5A<br />

2) Test 1.1<br />

3) Test 1.1<br />

4) Test 3.1<br />

“bAt” message during<br />

operation<br />

Unit runs less than<br />

specified duration:<br />

Battery is warm during<br />

normal operation<br />

Charger light does not turn<br />

on<br />

Charger light does not<br />

change from red to green<br />

after overnight charging<br />

2) Sensor and Measurement:<br />

1) battery charge is low<br />

2) defective battery voltage measurement circuit<br />

3) excessive loading on the battery<br />

4) defective battery<br />

1) excessive loading on the battery<br />

2) memory effect of Ni-Cd battery<br />

3) old or defective Ni-Cd battery pack<br />

4) defective charger<br />

5) incomplete charging<br />

1) Test 1.5 & 1.5A<br />

2) Test 1.4<br />

3) Test 1.3<br />

4) Test 1.5 & 1.5A<br />

1) Test 1.3<br />

2) Test 1.5 & 1.5A<br />

3) Test 1.5 & 1.5A<br />

4) Test 1.6<br />

5) Test 1.7 & 1.5A<br />

1) excessive loading on the battery 1) Test 1.3<br />

1) broken wire inside PGM-50<br />

2) alkaline battery is used inside PGM-50<br />

3) defective charger<br />

4) bad LED<br />

1) defective charger<br />

2) defective charging circuit<br />

3) bad green LED<br />

4) bad battery<br />

5) ambient temperature too high or low<br />

1) Test 1.6<br />

2) Test 1.7<br />

3) Test 1.6<br />

4) Test 3.2<br />

1) Test 1.6<br />

2) Test 1.7<br />

3) Test 3.2<br />

4) Test 1.5 & 1.5A<br />

5) Test 1.8<br />

Problem Possible Causes Test Procedures<br />

Unit turns on, shows<br />

“Configuration changed”<br />

message<br />

1) sensor added or removed<br />

2) sensor is disconnected or loose from socket<br />

3) defective sensor<br />

1) Test 2.3<br />

2) Test 2.3<br />

3) Test 2.3<br />

Sensor is installed but not<br />

showing on LCD<br />

1) sensor is disconnected or loose from the socket<br />

2) defective electronics<br />

3) Sensor disabled<br />

1) Test 2.3<br />

2) Test 2.3<br />

3) Test 2.3<br />

3


No response or low<br />

response to gas input<br />

Slow response to gas input<br />

Erratic reading<br />

Non-zero background<br />

reading<br />

“CAL Err” message during<br />

calibration<br />

Flashing “CAL” message<br />

during normal operation<br />

Readings consistently<br />

wrong when exposed to gas<br />

During normal operation,<br />

“pump” message flashing<br />

“xxx installed<br />

bias error” message on<br />

LCD<br />

Calibrates OK, but LEL or<br />

VOC readings off by<br />

constant percentage.<br />

“Lamp” message during<br />

operation<br />

1) sensor input is blocked or dirty filter<br />

2) defective sensor module<br />

3) defective electronics<br />

4) weak, or leaking pump<br />

5) sensor cover plate missing, loose, or misaligned<br />

6) empty sensor position<br />

1) sensor input is blocked or dirty filter<br />

2) defective sensor module<br />

3) sensor has inherently slow response time<br />

4) weak or leaking pump<br />

5) sensor cover plate missing, loose, or misaligned<br />

6) empty sensor position<br />

1) incorrect calibration<br />

2) defective sensor module<br />

3) defective electronics<br />

4) weak or leaking pump<br />

1) dirty filter<br />

2) incorrect zero calibration or zero drift<br />

3) defective sensor module<br />

4) non-zero background<br />

5) carbon filter saturated or not installed<br />

1) incorrect calibration<br />

2) low gas pressure in gas cylinder<br />

3) defective sensor module or weak lamp<br />

4) leak in flow path<br />

5) defective electronics<br />

1) Test 2.1<br />

2) Test 2.4<br />

3) Test 2.4<br />

4) Test 2.2<br />

5) Test 2.1<br />

6) Test 2.1<br />

1) Test 2.1<br />

2) Test 2.4<br />

3) Test 2.4<br />

4) Test 2.2<br />

5) Test 2.1<br />

6) Test 2.1<br />

1).Test 2.5<br />

2) Test 2.4<br />

3) Test 2.4<br />

4) Test 2.2<br />

1) Test 2.1<br />

2) Test 2.6<br />

3) Test 2.4<br />

4) Test 2.6<br />

5) Test 2.1<br />

1) Test 2.5<br />

2) Test 2.5<br />

3) Test 2.4<br />

4) Test 2.1<br />

5) Test 2.4<br />

1) excessive negative zero drift of sensor 1) Test 2.6<br />

1) incorrect span calibration<br />

2) defective sensor module<br />

3) incorrect correction factor<br />

1) Gas inlet blocked or no connection to inlet port<br />

2) weak or inoperative pump<br />

3) pump stall threshold needs adjustment<br />

1) bias switch in wrong position for Toxic1 sensor<br />

2) defective sensor module<br />

1) Wrong calibration or measurement gas selection<br />

2) Incorrect lamp selected<br />

3) incorrect dilution ratio<br />

1) dirty PID sensor or lamp<br />

2) weak or defective lamp<br />

3) Lamp fail threshold adjustment needed<br />

1) Test 2.5<br />

2) Test 2.4<br />

3) Test 4.5<br />

1) Test 2.1<br />

2) Test 2.2<br />

3) Test 2.7<br />

1) Test 2.8<br />

2) Test 2.4<br />

1) Test 4.5<br />

2) Test 4.5<br />

3) Test 4.5<br />

1) Test 2.9<br />

2) Test 2.9<br />

3) Test 2.9<br />

3) Electronics:<br />

Problem Possible Causes Test Procedures<br />

LED, back light, buzzer or 1) defective component 1) Test 3.2<br />

vibrator alarm inoperative<br />

Missing segment on the<br />

LCD<br />

1) defective LCD<br />

2) temperature too low<br />

3) contrast set too low<br />

1) Test 3.3<br />

2) Test 3.3<br />

3).Test 3.3<br />

4


Random characters on the<br />

LCD<br />

No response to key press or<br />

can not turn off unit<br />

Back light does not turn on<br />

in low ambient light<br />

1) CPU hang-up<br />

2) defective electronics<br />

3) wrong firmware or error in firmware<br />

4) bad LCD<br />

5) temperature too high or low<br />

6) contrast too high or too low<br />

1) membrane disconnected from PCB<br />

2) defective membrane keys<br />

3) CPU hang-up<br />

1) defective electronics or component<br />

2) back light threshold needs adjustment<br />

3) unit configured for manual light turn on<br />

4) unit plugged into A/C adapter<br />

1) Test 3.3<br />

2) Test 3.3<br />

3) Test 4.1<br />

4) Test 3.3<br />

5) Test 3.3<br />

6) Test 3.3<br />

1) Test 3.1<br />

2) Test 3.1<br />

3) Test 3.1<br />

1) Test 3.3<br />

2) Test 3.3<br />

3) Test 3.3<br />

4) Test 3.3<br />

4) Programming<br />

Problem Possible Causes Test Procedures<br />

Unable to put into<br />

programming mode<br />

1) unit in display or text mode<br />

2) configured for password protection<br />

1) Test 4.2<br />

2) Test 4.3<br />

Lost password 1) unit configured for password access 1) Test 4.3<br />

“Unable to configure port” 1) port not available in this PC 1) Test 4.7<br />

message from ProRAE-50<br />

software<br />

“NAK” message from<br />

ProRAE-50 program<br />

1) Cable or connection problem<br />

2) port configuration problem<br />

3) PC hardware or operating system problem<br />

1) Test 4.7<br />

2) Test 4.7<br />

3) Test 4.7<br />

Time out message from<br />

ProRAE-50 software<br />

Can’t start data logging<br />

Shows “----“ instead of<br />

sensor name, does not<br />

display reading for sensor<br />

In data log, no readings for<br />

a sensor even though<br />

sensor and values show on<br />

display<br />

Unit beeps frequently but<br />

sensor readings are below<br />

the alarm limits<br />

No events recorded in PC<br />

after downloading data but<br />

data available in “View<br />

Datalog” menu.<br />

Unit beeps once per minute<br />

for no reason<br />

1) unit not in “Ready” mode<br />

2) cable or connection problem<br />

3) PC hardware or operating system problem<br />

1) datalogging disabled<br />

2) non-data logging unit<br />

1) Test 4.7<br />

2) Test 4.7<br />

3) Test 4.7<br />

1) Test 4.4<br />

2) Test 4.4<br />

1) sensor disabled 1) Test 4.5<br />

1) sensor not configured for data logging 1) Test 4.8<br />

1) alarm latched<br />

2) STEL or TWA limits exceeded<br />

3) low battery<br />

1) Short sample period causing data to be stored<br />

before header information written.<br />

1) Test 4.6<br />

2) Test 4.6<br />

3) Test 1.1<br />

1) Test 4.8<br />

1) comfort beep option turned on 1) Test 4.8<br />

IV.<br />

Diagnostic Mode:<br />

5


All RAE Systems gas monitors have a Diagnostic Mode for evaluating the condition of<br />

sensor, lamps, pumps, batteries, etc. This is distinct from the usual diagnostic tests the<br />

instrument performs automatically during routine start-up. All the usual functions can be<br />

performed in diagnostic mode, such as calibration and monitor set-up. In diagnostic<br />

mode all sensor outputs (and most other measured values) are displayed as raw values<br />

(a/d counts) without corrections or conversion to standard units. The following<br />

diagnostic procedures will make use of this mode when appropriate. Whenever the<br />

procedures state “place the unit in diagnostic mode” use the following steps to turn on the<br />

unit in this special mode.<br />

Entering diagnostic mode: With the unit turned off, hold down the [Y/+] key. While<br />

still holding this key down, depress and hold the [MODE] key. Hold both keys down<br />

until unit beeps and then release both keys. The <strong>MultiRAE</strong> will display “ON!” followed<br />

by the software version number, the firmware build date, and then the “Diagnostic Mode”<br />

message. If you hit the [Y/+] key when Diagnostic Mode is on the screen, the warm-up<br />

will be significantly shorter. The unit will then go through the normal start-up tests.<br />

Once the monitor is ready, the display will show raw values for each installed sensor,<br />

with the word “RAW” in the center of the bottom line of the display.<br />

In Diagnostic Mode, use [N/-] and [MODE] together; just as in normal mode, to enter the<br />

Programming Menu. CAUTION: When calibrating in Diagnostic Mode the system does<br />

not check for a difference between the span and zero calibration points. This can be<br />

useful if calibrating at a low level, (such as 10 ppm VOC). The user must manually<br />

verify, from the raw values in air and with calibration gas applied, that the sensors are<br />

working correctly. Use the [MODE] key alone to step through the special diagnostic<br />

screens in the following sequence:<br />

Version 1.xx Version 2.00 Version 3.xx<br />

Raw Sensor Values Raw Sensor Values Raw Sensor Values<br />

Sensor names Sensor names Sensor names<br />

PID Lamp Fail Threshold PID Lamp Fail Threshold RF test<br />

Battery type / Bias Battery type / Bias PID Lamp Fail Threshold<br />

PID 1X and 10X Range PID 1X and 10X Range Battery type / Bias<br />

LCD Contrast Adjustment LCD Contrast Adjustment PID 1X and 10X Range<br />

LEL Power On/Off Check LEL Power On/Off Check LCD Contrast Adjustment<br />

Clock/Voltage/Temperature Clock/Voltage/Temperature LEL Power On/Off Check<br />

Sensor expiration dates Sensor expiration dates Clock/Voltage/Temperature<br />

Pump Stall Threshold Pump Stall Threshold Sensor expiration dates<br />

Back Light On/Off Threshold Back Light On/Off Threshold Pump Stall Threshold<br />

Communicate with PC S/N and pump speed Back Light On/Off Threshold<br />

Battery deep discharge S/N and pump speed<br />

Power run down time Power run down time<br />

Communicate with PC Communicate with PC<br />

6


Software versions 1.05 and later allow access to the Programming Menu from the<br />

Diagnostic Mode even if the unit is configured in “Display” or “Text” modes. A valid<br />

password will be required if the unit has been configured with a password. This feature<br />

allows users to exit from these modes without using a PC.<br />

PROGRAMMING MODE<br />

Press [N] and [MODE] together for 3 seconds to begin.<br />

Calibrate Monitor<br />

[Y] to select, [N] to move to next menu, [MODE] to return to Display Mode.<br />

[Y] selects, [N] skips,[MODE] for next menu<br />

Fresh Air Calibration<br />

Multiple Sensor Calibration<br />

Single Sensor Calibration<br />

Modify Span Gas Value<br />

Change LEL/VOC Span Gas<br />

Change Alarm Limits<br />

Change High Alarm limit<br />

Change Low Alarm limit<br />

Change STEL alarm limit<br />

Change TWA alarm limit<br />

View or Change Datalog<br />

Clear All Data<br />

Change Datalog Period<br />

Select Data Type<br />

View Datalog<br />

Enable/Disable Datalog<br />

Change Monitor Setup<br />

Change Site ID<br />

Change User ID<br />

Change Alarm Mode<br />

Change User Mode<br />

Change Real Time Clock<br />

Change Back Light Mode<br />

Change Password<br />

Change Pump Speed<br />

Change Averaging Method<br />

Change Sensor Configuration<br />

Change LEL/VOC Gas Selection<br />

Enable / Disable Sensor<br />

Change Dilution Ratio<br />

Change PID Lamp (10.6 or 11.7)<br />

If not all of the above menus appear in Programming, go to “Change Monitor Setup” and<br />

change the User Mode to Program. Use Mode to step out of Programming and then<br />

return by pressing and holding the Mode and N/- keys.<br />

7


Once in Diagnostic Mode, Version 2.00 and above will allow switching between<br />

diagnostic and standard modes by pressing and holding both the [Y] and [MODE] keys.<br />

V. Diagnostic and Service Test Procedures:<br />

1.0 Power supply / Battery / Charger<br />

1.1 Battery Voltage Test:<br />

1.1.1 Turn off the power to the <strong>MultiRAE</strong>. Open the instrument cover by removing the 2<br />

flat head screws from the backside (see Figure 1).<br />

Figure 1<br />

8


1.1.2 Carefully remove the battery pack or alkaline battery adapter from the unit. Make<br />

sure that the battery or adapter is plugged into the jack.<br />

1.1.3 Unplug the battery from the battery jack connected to the PC board of the<br />

<strong>MultiRAE</strong>. Visually inspect the wires between the battery pack or adapter and the<br />

connector, and between the jack and the PC board, for signs of broken or frayed wires or<br />

loose connector pins.<br />

1.1.4 Use a voltmeter to measure the voltage between the 2 outside pins of the connector<br />

on the battery pack. Be careful to avoid shorting the two pins as this could blow the<br />

internal fuse and/or damage the batteries. If the voltage is less than 4.2V the unit will not<br />

turn on. If using alkaline (AAA) cells, replace them with fresh cells when the voltage is<br />

less than 4.4V. When the voltage is below 4.4V, charge the batteries, either in the unit or<br />

on the external charger (RAE part number 008-3027). After replacing or recharging the<br />

batteries, check the voltage again. Fresh alkaline batteries should read over 6V. Fully<br />

charged Li-Ion or Ni-Cad batteries should read over 4.8V.<br />

1.1.5 If the voltage reads 0, then the internal fuse is blown or there is a broken wire.<br />

Replace the battery pack or adapter.<br />

1.3 Battery loading<br />

1.3.1 Remove the battery from the unit. Set the DC power supply to 4.8V DC. Connect<br />

the negative output to pin 1 (black wire) of the spare battery connector. Connect the<br />

positive power supply output to the positive side of a current meter. Connect the negative<br />

side of the current meter to pin 3 (red wire) of the battery connector (see figure 2).<br />

DC POWER<br />

SUPPLY<br />

+<br />

-<br />

+ -<br />

DC<br />

CURRENT<br />

METER<br />

3<br />

1<br />

Figure 2<br />

1.3.2 Plug the connector into the mating connector in the <strong>MultiRAE</strong>. Current should be<br />

about 3 ma with the unit off. Turn on the unit and wait for it to complete the start up<br />

tests. The current will depend on the configuration of the <strong>MultiRAE</strong>:<br />

Basic unit:<br />

20 mA<br />

9


PID current:<br />

Pump current:<br />

LEL sensor:<br />

Other sensors:<br />

20 mA<br />

20 mA<br />

60 mA<br />

2-5 mA<br />

1.3.3 If the current exceeds the limits in the table above, then isolate the source of the<br />

excess consumption by removing components one at a time. CAUTION: turn off the<br />

power supply before removing sensors or lamps. The pump current can be checked by<br />

stalling the pump and observing the change in current.<br />

1.3.4 Replace the component found to consume excess current. If the cause of the excess<br />

current consumption cannot be isolated, then the unit must be returned to the factory or<br />

an authorized service center.<br />

1.4 Battery voltage Measurement Circuit<br />

1.4.1 Follow the steps in 1.3.1 (the current meter is not necessary).<br />

1.4.2 Turn on the power supply. Turn on the <strong>MultiRAE</strong>. Press [MODE] to get to the<br />

battery voltage display. The display should read 4.8V<br />

1.4.3 Increase the voltage to 6V. The display should read 6V.<br />

1.4.4 Slowly decrease the power supply voltage. At 4.4V the display should begin<br />

flashing “Bat”. At 4.2V the unit should turn off.<br />

1.4.5 If any of these tests fail, the unit should be returned to the factory for repairs.<br />

1.5 Battery Condition Tests (use for NiCd battery packs only)<br />

1.5.1 Charge the batteries for 8 hours, either in the <strong>MultiRAE</strong> or in an external charger<br />

(P/N 008-3027).<br />

1.5.2 Measure the voltage across pins 1 and 3 of the battery connector (test procedure<br />

1.1.4). If the voltage is 4.2V, then the battery is good and no further testing is needed. If the voltage<br />

is


Battery capacity Discharge time<br />

900 maH 3 hours<br />

1400 maH 5 hours<br />

1700 maH 6 hours<br />

Table 1.1 Battery discharge times<br />

1.5.4 Recharge the battery for 8 hours. Or place the battery in the external charger and<br />

activate the automatic discharging function before charging. (Discharging batteries<br />

should only be done with the NiCd battery pack, PN 500-0039-000) Repeat paragraphs<br />

1.5.2<br />

1.5.3 If the battery fails again, it must be replaced.<br />

1.5A Battery Condition Test without volt meter<br />

1.5.1A Charge batteries completely inside unit charge light turn green. Unplug charging<br />

adapter and turn monitor on in diagnostic mode. Allow unit to run down and turn off due<br />

to drained battery. Recharge battery and turn unit on in diagnostic mode. Press [MODE]<br />

repeatedly until you reach Run Time screen. If run time is less than 8 hours with a NiCd,<br />

or less than 12 hours with a Li-Ion battery, than replace the battery.<br />

1.6 AC Adapter Output Test<br />

1.6.1 Plug the 12VDC adapter into an appropriate AC source. Measure the voltage at the<br />

DC connector. The inside contact is positive. If the output is not at least 10V, then the<br />

adapter is defective and must be replaced.<br />

1.6.2 Inspect the DC connector. Make sure that the pin is at least 11 mm long and is not<br />

corroded or otherwise contaminated. Plug the AC adapter into the <strong>MultiRAE</strong> and into an<br />

AC outlet. With the battery removed from the unit, measure the voltage between pins 1<br />

and 2 of the connector in the <strong>MultiRAE</strong>. If the measured voltage is less than the voltage<br />

measured in paragraph 1.6.1, then there is a loose or broken wire or a poor contact. Trace<br />

the circuit from the external connector to the PC board to the battery connector to<br />

determine the cause.<br />

1.7 Battery Charging<br />

1.7.1 Remove the cover from the <strong>MultiRAE</strong>. Make sure there is a Ni-Cad battery pack<br />

(white) or Li-Ion battery pack (black or yellow) in the unit and that the battery connector<br />

is plugged in to the mating connector. The unit will not charge or operate from the AC<br />

adapter if the alkaline battery adapter (black) is installed.<br />

11


1.7.2 Perform the Battery Condition Test if the battery is a NiCd (Section 1.5), using the<br />

<strong>MultiRAE</strong> to charge the batteries. If the batteries fail, perform the AC Adapter Output<br />

Test (section 1.6.2). If the adapter output is less than 10V, then replace the adapter. If<br />

the adapter output is good, then replace the battery pack.<br />

1.8 Ambient Temperature<br />

1.8.1 Press the [MODE] key 7 times from the main display to show the date, time, and<br />

temperature. The temperature sensor is mounted on the digital PC board.<br />

1.8.2 Compare the displayed temperature to the actual room temperature. If either<br />

temperature is >40 °C the batteries may not charge properly. Charging Ni-Cad<br />

batteries at temperatures above 40°C may permanently damage the batteries.<br />

1.8.3 With the unit off and charging (red LED on), compare the ambient temperature to<br />

the temperature displayed in the unit. If the temperature displayed in the unit is more<br />

than 20°C warmer that ambient temperature, then the batteries are overheating during<br />

charging and should be replaced.<br />

1.8.4 If the temperature is colder than 10°C the batteries will not be charged to full<br />

capacity.<br />

1.8.5 If there is a large difference between the room temperature and the displayed<br />

temperature, but the unit and batteries do not feel warmer or colder than the room, then<br />

the temperature sensor or electronics need repair.<br />

2.0 Sensor and Measurement Tests<br />

2.1 Flow Related Problems (pump versions only)<br />

2.1.1 Check the external filter and connectors for dirt, obstructions, or leaks. If the<br />

external filter is wet it will block the flow of air and gas. If the external filter is<br />

contaminated, do not attempt to clean it. It must be replaced.<br />

2.1.2 Remove the cover to the <strong>MultiRAE</strong> and remove the gas piping plate. Check the gas<br />

piping plate and gasket for obstructions or leaks. Check for the presence of a carbon<br />

impregnated foam filter in the gas piping plate. This filter should only be installed over<br />

the CO filter. Remove and discard any carbon filters installed over any other sensors.<br />

When installed over the CO filter, this filter screens the CO sensor from VOCs which<br />

may affect the CO sensor. This filter should be replaced frequently. If the <strong>MultiRAE</strong><br />

otherwise works fine, but you are unable to get a stable of accurate VOC calibration, then<br />

remove the carbon disk filter.<br />

12


2.1.3 Check the tubing between the inlet port and the pump for kinks, obstructions, or<br />

leaks. Check that the pump outlet port is not loose, broken, or leaking.<br />

2.1.4 Make sure there are no empty sensor positions. There must be either a sensor or a<br />

dummy sensor block in all five sensor positions (pump models).<br />

2.1.5 Re-install the gas piping plate. Observe that the connector nipple fits into the pump<br />

manifold (<strong>MultiRAE</strong> Plus does not have the manifold) and that the gasket fits snuggly<br />

against all the sensors.<br />

Figure 3 Sensor Positions<br />

Note: <strong>MultiRAE</strong> Plus has pump mounted to gas plate.<br />

2.2 Pump problems<br />

13


2.2.1 With the pump on, connect a flow meter to the <strong>MultiRAE</strong> inlet connector. The<br />

minimum acceptable flow rate is 200 ml/min. on the high speed and 150 ml/min. on the<br />

low speed. Rev C. boards only have one speed (high). If the flow is too low, check for<br />

obstructions and leaks as in section 2.1.<br />

2.2.2 If there are no obstructions or leaks, and the pump volume is still low, check the<br />

voltage at the pump power terminals. It should be 3.5V. If the input voltage is correct,<br />

then the pump should be rebuilt or replaced (P/N 008-3043-002). Replacement<br />

diaphragms are available through RAE Systems (P/N 081-0007-000).<br />

2.3 Sensor missing or not identified<br />

2.3.1 Always check the Programming Mode to verify that all sensors are enabled. Enter<br />

the programming menu ([N] and [MODE] together). Press [N] 4 times, or until the<br />

display reads “Change Sensor Configuration” Answer “Y”. The second option under<br />

the “Change Sensor Configuration” menu is to Enable/Disable Sensors. When this<br />

option is selected the display shows all installed sensors with a “*” next to the sensors<br />

which are enabled. Use the [MODE] key to scroll the cursor through the sensors. [Y]<br />

Enables sensors and [N] disables sensors. Hold the [MODE] key down until the display<br />

shows “save” to exit from this function. When a sensor is disabled the instantaneous<br />

display screen will be blank in that sensor location. The “Display Sensor Location”<br />

screen (press [MODE] once from the instantaneous screen) shows “---“ in the sensor<br />

positions which have been disabled.<br />

2.3.2 Verify the sensor configuration: The <strong>MultiRAE</strong> sensors have sensor type and<br />

expiration date stored internally. During the power-on tests the unit identifies all<br />

installed sensors and compares this to the configuration list. Make a note of which<br />

sensors the unit does identify by pressing [MODE] once from the instantaneous display.<br />

Turn off the unit. Remove the top cover. Disconnect the battery. Failure to disconnect<br />

the battery before removing or installing sensors could blow the battery pack<br />

internal fuse. Remove the gas piping plate. Verify the identity of each installed sensor.<br />

Remove, inspect, and re-insert each sensor making sure each is fully seated in its socket.<br />

Be sure that the shorting pin is no longer across the bottom of the sensor. The tops of the<br />

sensors should all be at nearly the same height. Pay particular attention to any sensors<br />

which were not identified. If pins are missing or the sensor or circuit board shows signs<br />

of damage do not re-insert the sensor: it must be replaced.<br />

2.3.3 Re-connect the battery and turn on power. “Configuration changed” may appear<br />

now even if all the sensors are working. Note which sensors are identified when the unit<br />

warms up and if all the installed sensors are now identified, then turn off power and reassemble<br />

the unit. A full calibration must be performed before using the unit, as<br />

calibration data may have been lost. Sensors which require a bias (NO and NH 3 ) must be<br />

installed, unit turned on and running, for 24 hours before stable readings can be obtained.<br />

14


It is always good to run the unit for at least four hours after replacing any sensor. Or use<br />

a SensorRAE to pre-Burn-in the sensors.<br />

2.3.4 If the sensor configuration is still not correct, then either the sensor is bad or there is<br />

an electronic problem. If possible, try replacing the sensor that is not recognized with a<br />

known good sensor to determine if the problem is with the sensor or the unit. Note that if<br />

NONE of the sensors can be identified; try removing sensors one at a time. A bad<br />

memory chip on one sensor may prevent the monitor from reading any of the sensor IDs.<br />

If this fails to isolate the problem, then the problem is with the boards in the unit.<br />

2.4 Sensor Response Tests<br />

2.4.1 Verify that there are no flow obstructions, the pump is operating correctly, and the<br />

gas piping plate is properly installed. Turn on the unit in Diagnostic Mode (Section IV).<br />

After the start-up tests are complete the display will show “raw” values for all installed<br />

sensors, with “RAW” in the center of the lower row. Table 2.1 has the normal raw values<br />

for the sensors in air (fresh air value) and with the standard calibration gasses applied<br />

(span value). See TN-123 for most recent changes. The most important indicator of<br />

sensor operation is the delta value, which is the difference between the span output and<br />

the zero output. Response Time is the average time required for a sensor to reach 90%<br />

output.<br />

SENSOR FRESH<br />

AIR<br />

SPAN RANGE DELTA SPAN GAS RESPONSE<br />

TIME<br />

PID 310 – 450 2000 – 8500 ≥750 100 ppm 10 sec.<br />

Isobutylene<br />

LEL 310 – 1500 1300 – 2300 ≥100 2.5% Methane 15 sec<br />

O 2 860 – 1250 330 – 500 ≥200 99.9% N 2 15 sec<br />

CO 310 – 450 450 – 900 ≥50 50 ppm CO 20 sec<br />

H 2 S 310 – 450 600– 950 ≥50 10 ppm H 2 S 30 sec<br />

NO 310 – 450 420 – 750 ≥30 25 ppm NO 20 sec<br />

NO 2 310 – 450 400 – 1400 ≥30 5 ppm NO 2 25 sec<br />

SO 2 310 – 450 500 – 1700 ≥75 5 ppm SO 2 20 sec<br />

Cl 2 310 – 450 450 – 1200 ≥50 10 ppm Cl 2 60 sec<br />

NH 3 400 – 750 700 - 1300 ≥100 50 ppm NH 3 150 sec<br />

PH 3 310 - 450 700 - 1800 ≥50 5 ppm PH 3 60 sec<br />

HCN 310 - 450 400 - 500 ≥10 10 ppm HCN 60 sec<br />

Table 2.1 Raw Sensor Values<br />

15


2.4.2 Note the raw fresh air output of the sensor. Apply the span gas for this sensor and<br />

wait until the reading is stable to within a few counts. Note the raw “span” output of the<br />

sensor. Calculate the delta (span – fresh air).<br />

Note: Most sensors should increase when Span gas has been applied; the O2 sensor with<br />

Nitrogen is the only exception. If the LEL sensor drops when gas is applied, the sensor<br />

has been poisoned and needs to be replaced. See TN-144 for further information on<br />

Sensor Poisoning.<br />

2.4.3 Fresh air values significantly below or above the values in the chart generally<br />

indicate a dead sensor but could also be due to an electronic problem, especially if all the<br />

values are low or identical. First replace the sensor, if you get the same problem then the<br />

problem is most likely in the analog or digital board, and the unit must be returned to the<br />

factory for service.<br />

2.4.4 If the CO and H2S sensor RAW readings are extremely high, they could be<br />

temporarily poisoned due to excessive Methanol exposure. Turn the unit on in<br />

Diagnostic Modes and run the monitor in a clean environment, checking to make sure<br />

that the RAW values are dropping over time. Be sure to store the Lamp Cleaning Kit<br />

solution (Methanol) in a sealed container or removed from the <strong>MultiRAE</strong> kit. If it has<br />

been previously stored in the case, allow the case to air out for 24 hours before storing the<br />

monitor inside.<br />

2.4.5 The delta is the most significant indicator of sensor condition. Fresh air values<br />

slightly below the chart values are acceptable if the delta is above the minimum value. A<br />

low delta means that the sensor is not responding to the test gas and must be replaced.<br />

2.4.6 Refer to section 2.9 for further trouble shooting of the PID sensor.<br />

2.4.7 If the sensor takes a long time (relative to the normal response time) to reach a<br />

stable output, then either the sensor is bad, the sensor is not warmed up, or there is a leak<br />

in the gas path (pump models).<br />

2.5 Calibration<br />

2.5.1 Inaccurate or improper calibrations can cause erratic readings, especially if the unit<br />

is calibrated with a gas to which the unit has a low sensitivity or with a standard gas at a<br />

very low concentration. In normal operating mode the software checks for a minimal<br />

difference in sensor output between span and zero calibration data (20 to 35 A/D counts,<br />

depending on software revision). This check is skipped if the calibration is performed in<br />

Diagnostic Mode. Software revisions before 1.05 could display erratic oxygen readings<br />

if a span calibration had not been performed (i.e. single sensor calibration at a value other<br />

than 0%). Calibration without applying the span gas can result in invalid calibration data<br />

being stored. A common error is calibrating the oxygen sensor to 0% in air. Some<br />

16


sensors require time to warm up before their zero point is stable. Calibration without<br />

sufficient warm up could result in significant zero drift as the sensor warms up. New<br />

sensors should be warmed up for at least one hour before calibrating (NH 3 and NO<br />

sensors should be installed and the unit run for 24 hours prior to calibrating in order for<br />

the bias circuit to stabilize the sensors).<br />

2.5.2 To check the system response to calibration gasses, use the raw output values in the<br />

Diagnostic Mode (section IV). Apply the calibration gasses and observe that the sensor<br />

responds sufficiently.<br />

2.5.3 If the sensor is responding well to the chosen calibration gas, allow the unit to warm<br />

up for at least 15 minutes, still in the Diagnostic Mode. Switch from diagnostic mode to<br />

standard mode by pressing and holding both the [Y] and [MODE] keys. If the sensor<br />

zero continues to drift, then the sensor should be replaced.<br />

2.5.4 If the sensor zero point is stable, then, while still in Diagnostic Mode, and proceed<br />

to the calibration menu (hold down [N] and [MODE]. Perform a fresh air calibration<br />

followed by a single gas calibration for the sensor of interest. If the sensor under test is<br />

oxygen, then note that the default is to calibrate to 0% with 99.9% nitrogen (say [Y] to<br />

the 0% question) or with another gas that has no oxygen in the mixture.<br />

2.5.5 Check the span gas values under “Modify Span Gas” to make sure that the cylinder<br />

of gas you are using matches the span values of each gas. To adjust a span concentration<br />

use the [MODE] key to scroll, the [Y] key to enable, and the [N] key to disable. Hold<br />

[MODE] until “save” appears to exit from this function.<br />

2.5.6 If calibration fails despite a good response in paragraph 2.5.2, check the level of gas<br />

in the cylinder, if it is less than 250psi, use a new bottle of gas. Otherwise there is an<br />

electronic problem and the unit should be returned to the factory for service.<br />

2.5.7 If calibration is successful, then switch from diagnostic mode to standard mode by<br />

pressing and holding both the [Y] and [MODE] keys. Apply the test gas and observe if<br />

the unit is reading correctly. If it is not then it must be returned to the factory.<br />

2.6 Non-zero Background<br />

2.6.1 “Fresh air” or “zero” calibration should be done in an area known to be free of<br />

gasses, which would affect the sensors. Zeroing in the presence of such gasses will result<br />

in inaccurate readings, especially at low concentrations, or a “NEG” alarm on one or<br />

more of your sensors.<br />

2.6.2 If you are unsure of the purity of the air, or if you are seeing non-zero background<br />

readings after a good calibration, then use a bottle of “zero” air during the fresh air<br />

17


calibration. This is the only way to positively differentiate between a faulty sensor and a<br />

real background signal. If you obtain a good calibration with “zero” air, and then recheck<br />

the zero with the same bottle of “zero” air, then you can have confidence that the<br />

instrument is working properly and the background signal you are seeing is caused by<br />

something in the environment.<br />

2.6.3 If, after a “zero” air calibration, you do not obtain a stable zero when applying the<br />

“zero” air to the instrument, then the sensor does not have a stable zero point and must be<br />

replaced.<br />

2.7 Pump Stall Adjustment.<br />

2.7.1 The Pump alarm can occur if the pump current exceeds the stall threshold value or<br />

if the pump current is 0. The pump current may increase over the threshold value as the<br />

pump ages or if the diaphragm is stiff due to cold temperatures. It may be necessary to<br />

adjust the pump stall threshold when switching from low to high pump speed or when<br />

adding additional lengths of tubing (Rev. D boards). Pump current may also increase if<br />

the inlet is blocked or the vapor trap filter is clogged. Before adjusting the pump stall<br />

threshold, make sure the flow path is not obstructed. Do not attempt to clean a clogged<br />

filter: it must be replaced.<br />

2.7.2 Turn on the unit in Diagnostic Mode (Section IV). The pump alarm is disabled in<br />

Diagnostic Mode. Press the [MODE] key repeatedly until you get to the “Pump =”<br />

screen. The top line is the measured pump current, and the bottom line is the threshold<br />

value at which the pump will stall in normal operating mode. When operating properly<br />

the stall threshold should be 6 - 10 counts higher than the operating current (typically 20 -<br />

30 counts).<br />

2.7.3 If the pump current is 0 then check for loose or broken wires. If the wires are<br />

intact, check the voltage between the red and black wires. It should be 3.5V. If the<br />

voltage is correct, then replace the pump assembly. If the voltage is not correct, then the<br />

unit will have to return to the factory for repair.<br />

2.7.4 If the pump is operating, block the inlet with your finger and observe the change in<br />

pump current:<br />

No change or only a few counts:<br />

1) Pump current < stall current means there is a leak in the flow path<br />

between your finger and the pump intake. Check the filter assembly, the<br />

inlet valve, the internal tubing, the pump inlet nipple, and the pump<br />

diaphragm. Clean or replace any parts that are dirty or leaking<br />

18


2) Pump current > stall current means that the inlet is blocked. Check the<br />

filter assembly, the inlet valve, and the internal tubing. Clean or replace<br />

any parts that are dirty or leaking<br />

Changes, but pump current still < stall current<br />

Check for leaks. If none are found, then reduce the stall threshold to be less<br />

than the current with the inlet blocked. Use the [N/-] key to reduce the stall<br />

limit.<br />

Changes, but pump current > stall current even with inlet unblocked<br />

Check for obstructions. If none are found, then raise the stall threshold.<br />

Use the [Y/+] key to raise the stall value to a point midway between the<br />

open and blocked values.<br />

2.7.4 If the pump current exceeds 50 counts with no obstructions, then the pump<br />

assembly should be replaced.<br />

2.8 Bias Switch for Toxic 1 Sensor<br />

2.8.1 The NO and NH 3 sensors require a bias to operate correctly. This bias signal is<br />

only available in the Toxic 1 position and is set by a switch or jumper on the analog<br />

board. This switch, which is located under the gas piping plate immediately behind the<br />

front panel LEDs, must be ON for these two sensors and OFF for all other sensors.<br />

2.8.2 The computer checks the bias during the start up tests and compares the sensor type<br />

to the bias condition. It will display the “xxx installed bias error” message if the bias is<br />

not correct for the type of sensor installed in the Toxic 1 position.<br />

2.8.3 Check the type of sensor installed in the Toxic 1 position and the position of the<br />

bias switch. If the switch is in the proper position for the type of sensor and the error<br />

message still appears, and then the analog board may be bad.<br />

2.8.4 Verify that the sensor type displayed in the Toxic 1 position matches the sensor<br />

type installed. If they match, then the analog board is bad and must be replaced. If they<br />

do not match, then the sensor is bad (or incorrectly programmed) and must be replaced.<br />

2.9 PID Sensor and Lamp Tests<br />

2.9.1 The “Lamp” error is an indication of a problem with the lamp current. A dirty or<br />

contaminated sensor often causes high readings of the VOC sensor, especially when<br />

exposed to moisture. A weak or inoperative lamp often causes low readings or no<br />

19


esponse to test gas. Readings that go up to 2000 ppm and then decrease may be caused<br />

by saturation of the VOC sensor by gas concentrations over 2000 ppm.<br />

2.9.2 To test the sensor and lamp, place the unit in Diagnostic Mode (Section IV).<br />

2.9.3 Press the [MODE] key to reach the Lamp Failure Threshold screen. The threshold<br />

value is set to about 6 counts above the value obtained with no lamp in the PID. Page 8-4<br />

of the <strong>MultiRAE</strong> manual contains information on setting this value.<br />

2.9.4 If the Lamp Failure threshold is OK, press [MODE] repeatedly to get back to the<br />

raw data screen. Refer to the PID values in table 2.1.<br />

2.9.5 Zero values below the normal range, or a weak response to test gas, may indicate<br />

that the lamp is not illuminating sufficiently, or that the UV light is blocked by<br />

contamination on the lamp window. Remove the cover, gas piping plate, PID static<br />

shield, and PID sensor block. Remove the lamp, clean the crystal with anhydrous<br />

methanol, and reinstall the lamp. Turn on the unit. Caution: wear safety glasses to<br />

protect your eyes from UV light. Be aware that there are high voltages across the<br />

lamp electrodes. Check for a visible violet glow from the lamp. If there is no glow,<br />

then replace the lamp.<br />

2.9.6 Examine the sensor block for contamination, condensation, or corrosion. Soak and<br />

agitate the sensor in anhydrous methanol, preferably in an ultrasonic cleaner. Dry<br />

thoroughly. If there are still visible signs of contamination or corrosion, then replace the<br />

sensor.<br />

2.9.7 Reassemble and test the raw sensor output. If the delta value is still low try<br />

replacing the lamp, then the sensor. If the delta values are still low, then there is an<br />

electronic problem.<br />

3.0 Electronic Troubleshooting<br />

3.1 Power On Problem<br />

3.1.1 If the unit will not turn on, first check the battery pack, AC adapter, and fuses. If<br />

the alkaline battery adapter is installed, then the A/C adapter will not power the unit.<br />

Both the rechargeable batteries and the alkaline adapter have internal fuses, which can<br />

not be replaced.<br />

3.1.2 Try resetting the microprocessor by unplugging the battery pack. Wait a few<br />

minutes, and then reconnect the battery.<br />

20


3.1.3 If the unit will still not turn on, then either the electronics or the membrane switch is<br />

defective, and the unit must be returned to the factory for service.<br />

3.1.4 If the unit turns off before completing the warm up tests, then the battery pack is not<br />

charged. As soon as the LEL, pump, and VOC lamp turn on the voltage drops below the<br />

shut off value. This is usually an indication that the battery internal fuses are blown. If<br />

the battery connectors appear normal, then try replacing the battery.<br />

3.2 LED, Buzzer, Back light problems<br />

3.2.1 Test the Green battery charge indicator LED by opening the unit and removing the<br />

battery pack. Then plug the AC adapter into the unit. The green LED should illuminate.<br />

If it does not, then the indicator board must be replaced by the factory.<br />

3.2.2 Test the red battery-charging indicator by plugging in the A/C adapter with a Ni-<br />

Cd/Li-Ion battery connected to the unit. The red LED should illuminate. If not, then<br />

proceed as in section 3.2.1<br />

3.3.3 The alarm buzzer should beep once during the power on self-test. The buzzer is on<br />

the analog PC board, next to the VOC and TOX1 sensor positions. If the buzzer does not<br />

beep during power on or power off, then it must be replaced. Do not attempt to replace<br />

just the buzzer; the unit will need to be sent in to the factory for service.<br />

3.3.4 Depressing the [N] key can turn on the back light LEDs for the LCD automatically<br />

in low ambient light, or manually.<br />

3.3.5 If the back light does not come on in low ambient light (in normal mode only, in<br />

diagnostic mode the backlight will not turn on), check whether the unit is in auto or<br />

manual back light mode. This information is the last selection under the Change Monitor<br />

Setup sub-menu in the programming menu. If the unit is in auto mode and the lights still<br />

do not come on in low ambient light, then check the back light threshold in Diagnostic<br />

Mode (Section IV). This menu can be used to adjust the turn on threshold and to check<br />

operation of the light sensor. The circular window at the top right of the front instrument<br />

display is for the sensor. Cover the window with a finger (with the A/C adapter<br />

unplugged) while in the back light threshold display. The value in the display should<br />

change. If it does not, then the indicator board must be replaced by the factory.<br />

3.3.6 If the backlight is in manual mode, use the [N] key to turn it on. It will turn off<br />

automatically after a few seconds to conserve battery life.<br />

3.3.7 If the back light can not be turned on in either mode, the unit must be returned for<br />

service.<br />

21


3.3.8 If the green or red battery charge indicator LEDs are on (i.e. the A/C adapter is<br />

plugged in), then auto back light mode will not work. The light from the LEDs will be<br />

detected by the photo sensor, which will keep its output above the threshold value.<br />

3.3 LCD problems<br />

3.3.1 Faint or missing characters or bleeding characters, in the LCD screen may be<br />

corrected by adjusting the display contrast. Press the [Mode] key to reach the LCD<br />

contrast screen. Use the [+] and [-] keys to increase or decrease the contrast.<br />

3.3.2 In very warm environments the LCD may show bleeding around the edge of the<br />

characters. If this is excessive, then adjust the contrast downward. In very cold<br />

environments it may be necessary to increase the contrast to see all of the characters.<br />

3.3.3 Random or mal-formed characters in the display could also be caused by a CPU<br />

hang up or by a problem with the firmware. To reset the CPU, unplug the battery pack<br />

then reconnect the batteries. If the problem persists, then reload the firmware (Section<br />

4.1).<br />

3.3.3 If these steps do not resolve the problem, then the unit must be returned to the<br />

factory to have the display replaced.<br />

4.0 Troubleshooting Software/Programming Problems<br />

4.1 Reloading<br />

4.1.1 Odd characters on the display, failure to respond to key presses with a keypad that<br />

test good or erratic operation may be caused by an error in the firmware. Re-loading the<br />

basic operating program can solve this. The ProRAE Suite diskette includes a copy of<br />

the firmware or you can download the ProRAE .Suite and the <strong>MultiRAE</strong> PGM-50<br />

firmware from our website: www.raesystems.com.<br />

4.1.2 Connect the <strong>MultiRAE</strong> to the PC serial port and press Mode repeatedly until you<br />

reach “Communicate with PC” Press Y/+, unit will say “Monitor will pause” Press<br />

Y/+ key again. Unit should read “PC Comm.”<br />

4.1.3 Open ProRAE Suite and select Option, Load Firmware<br />

4.1.4 It is very important to shut down everything on your computer that could interfere<br />

such as the Internet, screen savers, etc. If the firmware loading process is interrupted, the<br />

monitor will be turned off and will not be able to turn it on. If this happens, the unit will<br />

need to come back to the factory for a boot load.<br />

22


4.1.5 Select the most recent .AO7 file that corresponds with your monitor and select<br />

Open<br />

4.1.6 After loading the firmware, turn monitor back on in diagnostic mode and return to<br />

PC Comm.<br />

4.1.7 Select Open and File on ProRAE Suite and change Files of Type to .cfg. Select the<br />

default configuration file that corresponds to your instrument and select Open.<br />

4.1.8 You can edit the configuration file by selecting Edit, Configuration<br />

4.1.9 After making the all changes you want, select Options, Config All. This will<br />

rewrite the entire configuration and finish the rewriting of the monitor.<br />

4.2 Display or Text Only Modes<br />

4.2.1 The text mode is the simplest mode of operation. The monitor will display the<br />

sensor name after the monitor is turned on. The user can press the [MODE] key to see<br />

the instantaneous gas concentration reading, battery voltage or enter the PC<br />

communication menu. The user can get into the calibration mode from text mode to do<br />

calibrations, but cannot change most of the parameters. In the display mode, the<br />

<strong>MultiRAE</strong> PLUS Multi-gas Monitor is able to display all the information in the text<br />

mode. In addition, the following information is also available by pressing the [MODE]<br />

key: peak, minimum, STEL, TWA, run time in hours and minutes, temperature in degree<br />

C, datalog mode (or enable / disable datalogging operation in manual datalog mode),<br />

LEL/VOC gas names, printing and communication with a PC options. The user cannot<br />

get into the calibration mode from display mode, so the user cannot do calibration and<br />

cannot change any of the parameters. The third option is “Programming” mode, which<br />

gives users full access to all the menus and features.<br />

4.2.2 Modes can be changed from the PC configuration file (datalogging units only) or<br />

from the “Change Monitor Set up” sub-menu under the programming menu.<br />

4.2.3 To exit from Text or Display modes in a datalogging unit, use the PC program<br />

(ProRAE-50). Select a configuration file (.CFG file), using either the default or a custom<br />

file. Make sure the Program mode button is selected, and send the configuration to the<br />

<strong>MultiRAE</strong>.<br />

4.2.4 The procedure to exit from Display modes in non-datalog units, or if a PC is not<br />

available, depends on the firmware version number. This is the number that appears on<br />

the LCD when the unit is first turned on.<br />

23


Version 1.03 and earlier: press [MODE] to get to the “Communicate with PC”<br />

screen. Press [Y] to get the “Monitor will pause OK” screen. Press [N]. A bug<br />

in the software will drop you into the programming menu. Press [MODE] until<br />

you get to the “Change Monitor Setup” screen. Press [Y], and then press [N]<br />

until you get to the “Change User Mode” screen. Press [Y], then select<br />

Programming Mode. Press [MODE] to exit and save changes.<br />

Version 1.04: requires a PC to exit from Display mode.<br />

Version 1.05 and later: Start the unit in Diagnostic Mode (section IV). Press [N]<br />

and [MODE] together to enter the programming menu. You will be required to<br />

enter the password to proceed. The default password is 0000. Hold the MODE<br />

key down after entering the password to get to the programming menus. If you do<br />

not know the password you will not be able to proceed (see section 4.3). Once in<br />

the programming menu, proceed as above for version 1.03 to change the user<br />

mode. After making and saving the changes, turn the unit off to exit from<br />

Diagnostic Mode.<br />

4.3 Password Protection<br />

4.3.1 Units may be configured with a password from the PC configuration file. There are<br />

3 security levels that work in conjunction with the password:<br />

Security level 0:<br />

Security level 1:<br />

Security level 2:<br />

User cannot enter programming menu at all.<br />

User must enter password to enter programming menu.<br />

No password needed to enter programming menu.<br />

4.3.2 To change the password from the ProRAE software you must perform a firmware<br />

upgrade and send the configuration file to the unit using “Config All”<br />

4.4 Datalogging Enable/Disable<br />

4.4.1 Datalogging functions may be enabled in the ProRAE Suite. Refer to section 5.13<br />

of the operation manual. A three-digit code, which is unique for each serial number, is<br />

required to enable datalogging. This code is supplied when the data logging option is<br />

purchased and is on file at RAE Systems.<br />

4.4.2 To determine if datalogging is enabled, observe the software version number which<br />

is the second screen displayed when the unit is turned on. If the version number ends in<br />

“D”, then datalogging is enabled.<br />

4.4.3 Individual sensors may be selected or deselected for datalogging, either from a<br />

configuration file and the ProRAE-50 program, or from the “Enable/Disable Datalog”<br />

24


function, which is the last selection in the “View or Change Datalog” sub-menu under the<br />

programming menu. Sensors which are enabled for datalogging have a “*” next to their<br />

name in this display. As usual, use the [MODE] key to scroll, the [Y] key to enable, and<br />

the [N] key to disable. Hold [MODE] until “save” appears to exit from this function.<br />

4.5 Sensor Configuration<br />

4.5.1 The programming menu has a sub-menu for sensor configuration. Within this<br />

section users may select among the list of gasses for LEL and VOC measurement. The<br />

<strong>MultiRAE</strong> has the capability to select up to 50 gases for calibration and up to 50 different<br />

gases for measurement with both the LEL and VOC sensors. The system will<br />

automatically correct for the relative responses of any combination. The factory defaults<br />

are isobutylene for both calibration and measurement on the VOC and methane for both<br />

functions on the LEL. You can also Enable/Disable sensors (see section 2.3), set a<br />

dilution ratio (for use with dilution probe), and configure for use with either the 10.6 eV<br />

or 11.7 eV lamp. These powerful features can lead to confusion if the user has not<br />

checked the configuration before calibrating and using the unit. If the unit is tested with<br />

a known gas and the readings are always off by a constant percentage, check both the<br />

calibration gas and the measurement gas selections in the programming menu. The<br />

calibration gas is under the “Calibrate Monitor” sub-menu (“Change LEL/VOC span<br />

gas”). The measurement gas is under the “Change Sensor Configuration” menu<br />

(Change LEL/VOC gas”).<br />

4.5.2 Enter the programming menu ([N] and [MODE] together). Press [N] 4 times, or<br />

until the display reads “Change Sensor Configuration” Answer “Y”.<br />

4.5.3 The first selection is to select the LEL and VOC gas. Press [Y]. The LEL gas<br />

selection is displayed (if there is a LEL sensor installed). Press [N] to select a different<br />

gas. At this point the [+] and [-] keys scroll through the gas list. Once you have found<br />

the desired gas, press the [MODE] key and the display will say “Save”. Press [Y] to<br />

save. If the display says, “Not saved”, then the security level is set to 0 and changes are<br />

not allowed. Once a gas is selected, you can change the correction factor. Check the<br />

correction factor for accuracy by looking at TN-156. The VOC gas selection is now<br />

displayed. Follow the same steps to select a VOC gas. Once a gas is selected, you can<br />

change the correction factor. Check the correction factor for accuracy by looking at TN-<br />

106. The default gases, Methane and Isobutylene both have correction factors of 1.<br />

4.5.4 “Change Lamp Type” is used when you replace the current lamp with a new type of<br />

lamp. The instrument does not know what type of lamp it has installed until you select<br />

the correct lamp. To verify which lamp is in the unit, turn off unit. Remove the cover,<br />

gas piping plate, PID static shield, and PID sensor block. Remove the lamp and check<br />

the number etched on the side. The numbers 106 means it is a 10.6 lamp and the number<br />

117 mean that the lamp is an 11.7. Always clean the lamp before reinstalling (see section<br />

25


2.9). Turn monitor on in diagnostic mode and put unit into programming. Press [N] until<br />

you reach “Change Sensor Configuration” press [Y]. Press [N] until you reach “change<br />

Lamp Type” press [Y]. If the monitor agrees with the lamp you have installed hit [Y],<br />

other wise press [N] until the correct lamp is displayed. Press [Y] and the unit will ask to<br />

save, press [Y] again to save.<br />

4.5.5 Unless you are using the dilution ration to extend the range of the PID or to add O2<br />

to an LEL stream, the dilution ration should be 1. (See TN-167 for proper use of the<br />

dilution fitting)<br />

4.6 Alarms<br />

4.6.1 The <strong>MultiRAE</strong> has several types of alarms, which can be distinguished by the rate<br />

at which the unit beeps and by the messages on the display. See section 2.5 of the<br />

operation manual for a full listing of alarms and messages. When a sensor output<br />

exceeds an alarm setting, the unit will beep at 1, 2, or 3 beeps per second, and the sensor,<br />

which is causing the alarm, will flash on the LCD. The newest versions of the firmware<br />

also flash the name of the alarm in the sensor position (High, Low, STEL, etc.).<br />

4.6.2 If Alarm Latching is ON then the alarm will continue to beep even after the<br />

condition which caused the alarm has been cleared. Press the [Y] key to clear the alarm.<br />

For example, testing with 100 ppm isobutylene will cause a high alarm (with the factory<br />

default alarm limits) on the VOC. If the alarm is latched, then the unit will continue to<br />

beep even after the gas is removed. Alarm latching can be set or reset from the ProRAE-<br />

50 program or from the “Change Monitor Setup” sub-menu in the programming menu.<br />

The pump alarm is always a latching alarm and must be acknowledged before it will turn<br />

off.<br />

4.6.3 An alarm at 1 beep per second with a flashing sensor symbol indicates that the<br />

STEL or TWA limit has been exceeded. This alarm may continue to beep even after<br />

removal of the gas, which caused the alarm, because of the time averaged nature of these<br />

alarms. Use the [MODE] key to examine the STEL and TWA value screens to determine<br />

if this is the reason for the alarm. The only way to clear this alarm is to turn the unit off<br />

and then on again.<br />

4.6.4 If the LEL sensor exceeds 100% LEL, the LEL sensor turns off, for safety reasons.<br />

The alarm will beep at 3 flashes per second and the “LEL Off” message will flash on the<br />

screen. The unit must be powered off to restart the LEL sensor.<br />

4.7 PC Communication Problems<br />

26


4.7.1 Serial communications is a simple and reliable method of exchanging information<br />

between devices. Problems can arise from failure of the PC, failure of the unit, problems<br />

with the cable, or operator error. A serial port tester, which is nothing more than a box<br />

with a male connector on one end, a female connector on the other end, and a bank of<br />

LEDs for each line, is an extremely valuable, yet inexpensive, tool for diagnosing serial<br />

communication problems. It is connected between the two devices. The LEDs show the<br />

status of each line at a glance.<br />

4.7.2 Keep a spare serial cable for testing. Cables do break from time to time. If there is<br />

a serial communication failure, first try a different cable, or try a known good unit with<br />

the cable. If the cable tests well, then try the following tests.<br />

4.7.3 The most common PC problems are incorrect port selection, interference by a serial<br />

mouse, or failure of the serial port.<br />

4.7.4 If the port selected is not available on the PC, then Windows will provide an error<br />

message.<br />

4.7.5 If the <strong>MultiRAE</strong> is not in “Ready” mode it will not respond to serial<br />

communications. Press the [MODE] key until the “Communicate with PC” window<br />

appears. Press [Y] Answer [Y] to “Monitor will pause OK” The display then reads<br />

“Ready” and the unit is waiting to communicate. If no messages are received after a<br />

timeout period, then the unit will remain at “Ready” but will no longer communicate.<br />

Press [Mode] until you reach the communications and try again.<br />

4.7.6 If the port is available, but it is not the one the <strong>MultiRAE</strong> is connected to, there will<br />

be a time out error. This will also happen if the port is the correct port but is not<br />

transmitting properly. The port tester can quickly identify if the port is the correct port.<br />

When sending a message to the <strong>MultiRAE</strong> the LED on the transmit line from the PC<br />

should blink. If it does not, then nothing is being sent out this port.<br />

4.7.7 Similarly, the port tester can show if messages are being sent from the <strong>MultiRAE</strong>.<br />

The receive line from the PC should blink when the <strong>MultiRAE</strong> is sending data to the PC.<br />

If you see the PC transmit line blink and do not see the receive line blink in response,<br />

then the unit is not responding. It must be returned for service.<br />

4.7.8 A “NAK received” message from ProRAE-50 could have several causes:<br />

Check the cable and connections.<br />

Check for proper operation of the serial port.<br />

Try reloading the ProRAE program, as corrupted files could also cause this error.<br />

Try reloading the firmware.<br />

If all of the above fail, the unit must be returned for service.<br />

27


4.8 PC Software Problems<br />

4.8.1 The ProRAE-50 programs works reliably with Windows 3.1, Windows 95,<br />

Windows 2000 and Windows XP. Some users have reported problems with Windows<br />

NT.<br />

4.8.2 The default configuration file included on the ProRAE-50/ProRAE Suite diskette is<br />

the factory default configuration. Refer to the operation manual for details about the use<br />

of the configuration file to set-up the <strong>MultiRAE</strong>.<br />

4.8.3 If you are able to download data, but data for a particular sensor is missing, check<br />

the configuration of the unit. Each sensor must be enabled for datalogging. This is<br />

different from the “Enable/Disable Sensor” function in section 4.5.<br />

4.8.4 It is generally a good idea to clear data before beginning a datalog session. This is<br />

done in the unit via the “View or change datalog” sub-menu from the Programming<br />

Menu. If data is not cleared samples taken at different times may confuse the graphing<br />

program, especially if some events from a later clock time precede events with an earlier<br />

clock time.<br />

4.8.4 If you download data, and you see in the message line at the bottom of the screen<br />

that a number of bytes have been transferred, but there are no events listed, then check<br />

the datalog period. Very short periods may confuse the ProRAE program, because<br />

datalogging is not halted when clearing memory. If a data sample is written to memory<br />

after clearing, but before the <strong>MultiRAE</strong> writes a new event header in memory, then the<br />

ProRAE program will not find the first event header and will report no events. You can<br />

prevent this by turning off datalogging before clearing memory, or by selecting a longer<br />

datalog period temporarily. ProRAE-50 releases after 2.00 will correct this problem.<br />

4.8.5 If you receive a “…datalog.raw” error message, no data has been stored in the unit.<br />

Verify that the “L” is flashing while the unit is running, or that you are running the unit<br />

for the entire datalog period. As you lengthen the datalog period, if the unit does not run<br />

for that entire time period, no data will be stored.<br />

4.8.6 If you are able to download data, but your readings are very high, the instrument<br />

was turned on in Diagnostic mode and saved the RAW readings of the sensors. The unit<br />

must be turned on in normal operations, for correct readings.<br />

28


QUICK REFERENCE GUIDE<br />

MULTIRAE CALIBRATION WITH 4 CALIBRATION GAS MIX<br />

1. Turn on <strong>MultiRAE</strong> in a safe, clean air environment. Let unit warm up for 5 minutes.<br />

2. Press [N/-] and [MODE] together for 5 seconds. Display shows “Calibrate Monitor”<br />

3. Press [Y/+]. Display shows “Fresh Air Calibration”<br />

4. Press [Y/+]. Display shows results for each sensor followed by “Multiple sensor calibration”<br />

5. Press [Y/+]. Display names sensors selected for span calibration. All others show as “----“.<br />

6. Apply the gas mixture (RAE P/N 600-0050) to the filter fitting (pump models) or the calibration adapter<br />

(diffusion models).<br />

7. Press [Y/+]. Unit displays “Apply mixed gas now…40” and counts down for 40 seconds. The unit then shows<br />

results. After a short pause the unit displays “Calibration done! Turn off the gas” followed by “Single sensor<br />

calibration”<br />

8. Turn off the gas and disconnect the bottle. Press [MODE] twice to return to the main display, or press [Y/+]<br />

for VOC sensor calibration.<br />

9. The sensor names are displayed. Push [MODE] until VOC is selected (flashing character).<br />

10. Turn on the isobutylene gas and apply the gas to the unit.<br />

11. Press [Y/+]. Display shows “Apply VOC gas” followed by “Calibration in progress…40”. Unit counts down<br />

40 seconds then shows “VOC cal’ed. Reading=xxx”, where “xxx” is the calibration result.<br />

12. Turn off the gas and disconnect from the <strong>MultiRAE</strong>. Push [MODE] twice to return to the main display.<br />

MULTIRAE BUMP TEST PROCEDURE<br />

1. Turn on <strong>MultiRAE</strong> in a safe, well-ventilated, clean air environment. Let unit warm up for 5 minutes.<br />

2. O 2 should read 20.9%. All other sensors should read 0. If readings are not correct, perform a fresh air<br />

calibration. (See above)<br />

3. If unit zeros correctly, turn on the quick check four gas mix (RAE P/N 600-0060) and apply it to the filter fitting<br />

(pump models) or calibration adapter (diffusion models).<br />

4. After 40 seconds the unit should read as follows:<br />

CO 90 - 110 ppm<br />

H 2 S 45 - 55 ppm<br />

LEL 45 - 55%<br />

O 2 14 - 16%<br />

5. Disconnect the gas and turn off the valve.<br />

6. Turn on the 100 ppm isobutylene gas and attach it to the vapor trap or calibration adapter.<br />

7. After 40 seconds the VOC should read 90-110 ppm.<br />

8. Disconnect the gas and turn off the valve.<br />

9. Wait for the unit to return to VOC=0 and O 2 =20.9.<br />

10. If the span readings were incorrect, then recalibrate the sensors (see above).<br />

11. If the span readings were all within the specified limits the unit is ready to use.<br />

29


PGM-50 Diagnostic & Service Procedure<br />

PROGRAM MODE<br />

Instantaneous display of values<br />

[N] + [M] for 3 seconds:<br />

Security level = 0: No action<br />

Security level = 1: Password check<br />

Wrong password: Instantaneous display<br />

Correct password: Programming mode<br />

Security level = 2: Programming mode<br />

[N]: Toggle backlight<br />

[Y]: Test alarm/stop alarm<br />

[M]<br />

Display sensor positions<br />

[M]<br />

Display peak values<br />

[M]<br />

Display minimum values<br />

[M]<br />

Display STEL values<br />

[M]<br />

Display TWA values<br />

[M]<br />

Display battery voltage<br />

[M]<br />

Display clock / run time / internal temperature<br />

[M]<br />

Is unit configured for manual datalog<br />

Yes: Display Start/Stop datalog<br />

[Y]: Start or stop datalog<br />

[M]: Go to “Display LEL gas name”<br />

No: Display datalog mode<br />

[M]<br />

Display LEL gas name<br />

[M]<br />

Display VOC gas name<br />

[M]<br />

Communicate with PC<br />

[Y]: Monitor will pause OK<br />

[Y]: Ready…<br />

[N]: Return to Instantaneous display<br />

[M] or [N]<br />

Return to Instantaneous display<br />

30


PGM-50 Diagnostic & Service Procedure<br />

Firmware Releases:<br />

Version 3.12A 3.12 3.11D<br />

Date 10/14/2003 6/4/2003 5/22/2003<br />

Error<br />

"Datalog Disabled"<br />

display<br />

drive for vibration alarm to<br />

RF control fixed<br />

work in Silent Mode<br />

New<br />

RF Control Enable/Disable<br />

Version 3.11C 3.11B 3.11A<br />

Date 3/18/2003 1/13/2003 8/29/2002<br />

Error<br />

check correction factor<br />

of LEL<br />

verify Silent Mode<br />

higher than 6<br />

New<br />

Enable Run Silent Mode<br />

Version 3.11 3.10A 3.10<br />

Date 8/29/2002 11/7/2001 9/26/2001<br />

Error fix LEL sensor linearity fix Power-On Zero,<br />

Fix O2 temp<br />

compensation<br />

so Fresh Air is not required<br />

Verify LEL correction<br />

factors<br />

reset NH3 low alarm to 25ppm<br />

check LEL "OFF" alarm<br />

change time out for low battery<br />

change Lamp driver<br />

prompts Single Sensor<br />

fix first point in datalog<br />

Calibration after Power-on Zero<br />

fix alarm when sensor is<br />

disabled<br />

New<br />

add disable Fresh Air<br />

Cal<br />

Default User mode is Text<br />

Mode<br />

Calibration Date is updated<br />

during<br />

Remove "View Datalog" Default setting has On-line Single Sensor calibration<br />

add temperature units<br />

selection printing turned off add User ID<br />

"Reset Peak" moved to<br />

idle mode<br />

add wireless abilities<br />

add "Change Temperature<br />

Units"<br />

31


PGM-50 Diagnostic & Service Procedure<br />

<strong>MultiRAE</strong> Service Parts<br />

32


PGM-50 Diagnostic & Service Procedure<br />

Part Number Description Price (ea.)<br />

008-2002-005 Cover screw, PGM 50 (5) 10.00<br />

008-2003-005 Cover screw washer, PGM 50 (5) 5.00<br />

430-0070-005 Gas plate screw,4P and 5P (5) 5.00<br />

430-0058-005 Gas plate washer (5) 5.00<br />

008-2027-000 Gas inlet nipple 8.00<br />

008-2020-001 Washer for inlet nipple and pwr jack 2.00<br />

008-2009-000 Dummy sensor 5.00<br />

008-2017-000 PID static shield 5.00<br />

008-2020-001 Gas inlet power jack washer 2.00<br />

008-2027-000 Gas inlet fitting assembly. (nipple) 8.00<br />

008-3003-000 Interface cable assembly 40.00<br />

008-3006-005 Charcoal filter for CO (5) 9.00<br />

008-3006-020 Charcoal filter for CO (20) 32.00<br />

008-3007-000 PID sensor assembly 75.00<br />

008-3013-000 Calibration Adapter (Diffusion Models) 15.00<br />

008-3014-000 External Filter with quick connect adapter (1) 10.00<br />

008-3014-001 External Filter with barb fitting adapter (1) 5.00<br />

008-3016-000 Calibration adapter (pump) 10.00<br />

008-3017-000 Housing cover 25.00<br />

008-3018-000 Housing case (Bottom & Top) 75.00<br />

008-2004-000 Housing bottom 50.00<br />

027-2001-000 QRAE Housing Top 25.00<br />

027-2004-000 QRAE Housing Bottom 50.00<br />

008-3022-010 Water vapor traps (pack of 10) 38.00<br />

008-3022-100 Water vapor traps (pack of 100) 345.00<br />

008-3027-000 External battery charger for NiCd Batteries 85.00<br />

008-3029-000 I/O port rubber cover (10) 20.00<br />

008-3035-002 Gas piping plate, 5P, Aluminum/Teflon hybrid 42.00<br />

008-3036-002 Gas piping plate, 4P, Aluminum/Teflon hybrid 42.00<br />

008-3037-002 Gas piping plate, 4D, Aluminum/Teflon hybrid 42.00<br />

008-3040-000 Packing case with foam 56.00<br />

008-3043-002 Pump assembly 295.00<br />

008-3044-002 5P pump on aluminum gas plate 337.00<br />

008-3045-002 4P pump on aluminum gas plate 337.00<br />

008-4011-000 Service Manual, PGM 50 50.00<br />

050-0010-000 10.6 eV, ¼” lamp 195.00<br />

050-0011-001 11.7 eV, ¼” lamp in ampoule 345.00<br />

050-0030-000 9.8 eV, ¼” lamp 245.00<br />

081-0002-000 Lamp cleaning kit 28.00<br />

081-0007-000 Pump rebuild kit 40.00<br />

090-2001-000 Remote probe tubing adapter 9.50<br />

411-0017-024 Tygon tubing, pump connections (per foot) 5.00<br />

500-0039-000 1700 mAH Ni-Cad battery 155.00<br />

500-0037-000 Li-Ion battery pack 155.00<br />

008-2104-002 Membrane key panel 40.00<br />

027-2104-000 QRAE Membrane key panel 40.00<br />

008-2021-005 Stand-off, digital PCB (5) 5.00<br />

008-2022-005 Stand-off, analog PCB (5) 5.00<br />

500-0029-000 Alkaline Adapter for 4 AA 12.00<br />

008-1171-001 LEL Sensor Module,PGM-50 175.00<br />

008-1161-000 O2 Sensor Module,PGM-50 120.00<br />

008-1111-000 H2S Sensor Module,PGM-50 195.00<br />

008-1112-000 CO Sensor Module,PGM-50 195.00<br />

008-1113-000 SO2 Sensor Module,PGM-50 195.00<br />

33


PGM-50 Diagnostic & Service Procedure<br />

008-1114-000 NO Sensor Module,PGM-50 195.00<br />

008-1115-000 NO2 Sensor Module,PGM-50 195.00<br />

008-1116-001 Cl2 Sensor Module,PGM-50 195.00<br />

008-1117-000 HCN Sensor Module,PGM-50 395.00<br />

008-1118-000 NH3 Sensor Module,PGM-50 395.00<br />

008-1119-000 PH3 Sensor Module,PGM-50 395.00<br />

008-1120-000 ClO2 Sensor Module,PGM-50 395.00<br />

008-1121-000 EtO Sensor Module,PGM-50 395.00<br />

34

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!